Endoplus Hand held Laparoscopic instruments are intended for grasping, cutting, clamping, biopsy, and/or cauterization of tissue. The devices are used in conjunction with a laparoscopic procedures. Instruments should be used only by personnel completely familiar with their operation. Using an instrument improperly for a task which it was not intended may result in a damaged or broken instrument.
Device Story
Endoplus Hand-held Laparoscopic Instruments are manual surgical tools used in laparoscopic procedures. The system consists of a reusable handle, shaft, and interchangeable inserts (tips/jaws) including scissors, forceps, graspers, biopsy tools, dissectors, and needle holders. The device is inserted through a cannula or trocar into the body. A mechanical linkage connects the handle to the working tip, allowing the surgeon to manipulate tissue (grasp, cut, dissect, retract, clamp, cauterize) under visual guidance from a laparoscope. The instruments feature a flush port for cleaning and sterilization. The device is operated by a physician in an operating room environment. The primary benefit is providing a modular, reusable toolset for minimally invasive surgical tissue management.
Rigid, 5mm diameter, 20-50cm length laparoscopic instruments. Pistol-grip handle with mechanical linkage for tip actuation. Monopolar electrosurgical capability. Features include shaft rotation and flush port for cleaning. Reusable construction. Sterilization via steam pre-vacuum. Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-2 standards.
Indications for Use
Indicated for grasping, cutting, dissecting, retracting, clamping, biopsy, and/or cauterization of tissue during laparoscopic procedures. For use by physicians in operating rooms.
Regulatory Classification
Identification
A unipolar endoscopic coagulator-cutter is a device designed to destroy tissue with high temperatures by directing a high frequency electrical current through the tissue between an energized probe and a grounding plate. It is used in female sterilization and in other operative procedures under endoscopic observation. This generic type of device may include the following accessories: an electrical generator, probes and electrical cables, and a patient grounding plate. This generic type of device does not include devices used to perform female sterilization under hysteroscopic observation.
Predicate Devices
Pajunks disposable inserts for handled instrument and HF Electrodes (K062072)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized design featuring three abstract human figures, possibly representing individuals or families, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 23, 2015
Endoplus % E.J. Smith Consultant Smith Associates 1468 Harwell Avenue Crofton. Marvland 21114
Re: K151567
Trade/Device Name: Endoplus Hand-held Laparoscopic Instruments Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: June 8, 2015 Received: June 10, 2015
Dear E.J. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
Page 2 - E.J. Smith
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Binita S. Ashar -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known) K151567
#### Device Name
Endopus Hand Held Laparoscopic Instruments
#### Indications for Use (Describe)
Endoplus Hand held Laparosopic instruments are intended for grasping, cutting, clamping, biopsy, and/or cauterization of tissue. The devices are used in conjunction with a laparoscopic procedures. Instruments should be used only by personnel completely familiar with their operation. Using an instrument improperly for a task which it was not intended may result in a damaged or broken instrument.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (1/14)
Page 1 of 2
SC Publishing Services (301) 443-6740
{3}------------------------------------------------
# 510(k) Summary
| SPONSOR | |
|--------------------------------|------------------------------------------------------|
| Company Name: | ENDOPLUS |
| Company Address | 750 Tower Road, Suite A<br>Mundelelin, IL 60060 |
| Telephone: | 800.236.5972 |
| Contact Person: | Matthew Gudeman |
| Summary Preparation Date: | December 23, 2015 |
| DEVICE NAME | |
| Trade Name: | ENDOPLUS Hand held Laparoscopic |
| Instruments Common/Usual Name: | Hand-held Surgical Instruments |
| Classification Name: | Electrosurgical, cutting & coagulation & accessories |
| Regulation Number: | 21 CFR 878.4400 |
| Product Code: | GEI |
| Device Class: | Class II |
### PREDICATE DEVICE
Legally Marketed Equivalent Device
| Manufacturer | Brand Name | 510(k) Number |
|--------------------------------|---------------------------------------------------------------------------|---------------|
| Pajunk GmbH Medizintechnologie | Pajunks disposable inserts for<br>handled instrument and HF<br>Electrodes | K062072 |
#### DEVICE DESCRIPTION
The devices are commonly named hand held laparoscopic instruments. Specifically, each instrument has a different tip or jaw to define the functional use, such as scissors, forceps, graspers, biopsy, dissectors or needle holders. The devices are designed to be used through a portal, the opening maintained by an introducer or cannula which allows for insertion and removal of the device without damage to surrounding soft tissue. The devices have a linkage mechanism connecting the working tips, or jaws to a handle. This allows the surgeon to cut, dissect, grasp, clamp, cauterize, palpate, or retract at the operating site in conjunction with a laparoscope.
{4}------------------------------------------------
### DEVICE INDICATIONS FOR USE
Hand held Laparoscopic instruments are intended for grasping, cutting, dissecting, retracting, clamping, biopsy, and/or cauterization of tissue. The devices are used in conjunction with a laparoscope during laparoscopic procedures. Instruments should be used only by personnel completely familiar with their operation.
| Parameters | Endoplus | Pajunk GmbH | Comments |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|
| 510(k) Number | | K062072 | |
| Indications for Use | Endoplus Hand held<br>Laparoscopic<br>instruments are<br>intended for<br>grasping, cutting,<br>dissecting,<br>retracting, clamping,<br>biopsy, and/or<br>cauterization of<br>tissue. The devices<br>are used in<br>conjunction with a<br>laparoscope during<br>laparoscopic<br>procedures.<br>Instruments should be<br>used only by personnel<br>completely familiar<br>with their operation.<br>Using an instrument<br>improperly for a task<br>which it was not<br>intended may result in<br>a damaged or broken<br>instrument. | Pajunks disposable<br>inserts are instruments<br>insulated for optional<br>monopolar coagulation<br>which enable a surgeon<br>to grasp, manipulate,<br>dissect, retrieve,<br>biopsy, cut or<br>coagulate internal<br>tissue or organs while<br>performing<br>laparoscopic<br>procedures. | Similar |
| User Environment | For use by physician<br>in the operating<br>room | For use by physician<br>in the operating<br>room | Identical |
| Monopolar | Yes | Yes | Identical |
| Length | 20cm to 50cm | 33cm to 45cm | Similar |
| Minimum Cannula<br>Size | 5.5cm | 5.5cm | Identical |
| Handle Design | Pistol Grip | Pistol Grip | Identical |
| Shaft Rotation | Yes | Yes | Identical |
| Shaft Flexibility | Rigid | Rigid | Identical |
| Parameters | Endoplus | Pajunk GmbH | Comments |
| Shaft Diameter | 5mm | 5mm | Identical |
| Articulating<br>Component | Does not articulate | Does not articulate | Identical |
| Actuation<br>Mechanism | Handle Operation<br>causes opening and<br>closing of end<br>effector | Handle Operation<br>causes opening and<br>closing of end<br>effector | Identical |
| Flush Port | Yes | No | Different |
| Insertion through<br>trocar | Yes | Yes | Identical |
| Form inserted into<br>trocar | Jaws closed | Jaws closed | Identical |
| Reusable/Disposable | Reusable Handle,<br>Shaft and Insert | Disposable Insert<br>Reusable Handle and<br>Shaft Assembly | Different |
#### SUBSTANTIALLY EQUIVALENCE COMPARISON TABLE
{5}------------------------------------------------
#### DIFFERENCES
- . Endoplus includes a flush port feature for introducing cleaning fluids during the cleaning and providing a means for steam to enter during the sterilization process.
- . Endoplus's inserts are reusable and Pajunk's inserts are disposable.
### COMPARISON OF TECHNICAL CHARACTERISTICS
The Endoplus Hand held Laparoscopic instruments has a similar Indications for Use and identical technological characteristics to the predicate device. Any material differences were addressed by biocompatibility, performance and electrical testing found below:
## PERFORMANCE DATA
### Safety Testing
### Electrical and EMC Testing
Endoplus performed the following electrical tests:
- IEC 60601-1: 2012 3rd edition Medical electrical equipment—Part 1: . General requirements for basic safety and essential performance
- . IEC 60601-1-2 Edition 3: 2007-03, medical electrical equipment - part 1-2: general requirements for basic safety and essential performance - collateral standard: electromagnetic compatibility - requirements and tests.
{6}------------------------------------------------
- IEC 60601-2-2: 2009 Medical electrical equipment Part 2-2: Particular ● requirements for the basic safety and essential performance of high frequency surgical equipment and high frequency surgical accessories
### Cleaning and Sterilization
The Reusable Hand-held Laparoscopic instruments are evaluated based on the following tests:
- Manual cleaning validation study and Laparoscopic Instrument Rack TOC analysis (AAMI TIR30:2011)
- Manual cleaning validation study and Laparoscopic Instrument Rack Protein ● analysis
- . Steam Pre-vacuum Efficacy Validation Study
- Steam Sterilization Validation Study
- Shelf Life Protocol
#### Biocompatibility
In accordance with ISO 10993-1, Hand-held Laparoscopic instruments are categorized as externally communicating devices in contact with circulating blood for less than 24 hours. The following tests are recommended by ISO 10993-1:
- · Cytotoxicity
- · Sensitization
- Irritation
- Systemic toxicity
- Intracutaneous
- Hemolysis
### Performance Testing:
- PLTW Comparative Ex-Vivo Thermal Tissue Testing Validation Study
- . Electrical Safety Validation Report (Electrical Safety Testing)
- Scissors Performance Testing
- Expected Life Testing
### CONCLUSION
The Endoplus Hand held Laparoscopic instruments has a similar Indications for Use and identical technological characteristics to the predicate device. Biocompatibility and
{7}------------------------------------------------
performance testing has been performed to verify safe use and effective use. Endoplus concludes that Endoplus Hand held Laparoscopic instruments are substantially equivalent to the predicate device and introduces no new issues of safety and effectiveness.
| Modular<br>Item #'s | diameter<br>(mm) | Type | Length | Descriptions |
|---------------------|------------------|-------------|-----------|-----------------------------------------------------------|
| 5293 | 5 | insert | standard | Delicate Maryland with Cross<br>Serrations |
| 5293L | 5 | insert | bariatric | Delicate Maryland with Cross<br>Serrations |
| 5294 | 5 | insert | standard | Johan Forcep |
| 5294L | 5 | insert | bariatric | Johan Forcep |
| 5295 | 5 | insert | standard | Medium Dolphin Nose Dissecting<br>Forsep w/ cup |
| 5295L | 5 | insert | bariatric | Medium Dolphin Nose Dissecting<br>Forsep w/ cup |
| 5296 | 5 | insert | standard | Medium Fenestrated Bowel Grasper |
| 5296L | 5 | insert | bariatric | Medium Fenestrated Bowel Grasper |
| 5297 | 5 | insert | standard | Strevy Grasper |
| 5297L | 5 | insert | bariatric | Strevy Grasper |
| 5300 | 5 | insert | standard | Allis forcep |
| 5300L | 5 | insert | bariatric | Allis forcep |
| 5301 | 5 | insert | standard | Long Allis Forcep |
| 5301L | 5 | insert | bariatric | Long Allis Forcep |
| 5302 | 5 | insert | standard | Traumatic Grasper 2x3 Teeth |
| 5302L | 5 | insert | bariatric | Traumatic Grasper 2x3 Teeth |
| 5303 | 5 | insert | standard | Endo Allis Forcep |
| 5303L | 5 | insert | bariatric | Endo Allis Forcep |
| Modular<br>Item #'s | diameter<br>(mm) | Type | Length | Descriptions |
| 5304 | 5 | insert | standard | Glassman Forcep |
| 5304L | 5 | insert | bariatric | Glassman Forcep |
| 5305 | 5 | insert | standard | Pennington Forcep |
| 5305L | 5 | insert | bariatric | Pennington Forcep |
| 5306 | 5 | insert | standard | Strong Fenestrated Grasping Forcep |
| 5306L | 5 | insert | bariatric | Strong Fenestrated Grasping Forcep |
| 5307 | 5 | insert | standard | McKernan Tri Grasping Fenestrated<br>Forcep |
| 5307L | 5 | insert | bariatric | McKernan Tri Grasping Fenestrated<br>Forcep |
| 5308 | 5 | insert | standard | Alligator Forcep |
| 5308L | 5 | insert | bariatric | Alligator Forcep |
| 5309 | 5 | insert | standard | Pointed Dissector w/ cup |
| 5309L | 5 | insert | bariatric | Pointed Dissector w/ cup |
| 5310 | 5 | insert | standard | Round Tip Allis Grasping Forcep |
| 5310L | 5 | insert | bariatric | Round Tip Allis Grasping Forcep |
| 5311 | 5 | insert | standard | Curved Dissecting Forcep |
| 5311L | 5 | insert | bariatric | Curved Dissecting Forcep |
| 5312 | 5 | insert | standard | Short Curved Dissecting Forcep |
| 5312L | 5 | insert | bariatric | Short Curved Dissecting Forcep |
| 5313 | 5 | insert | standard | Mixter Dissector |
| 5313L | 5 | insert | bariatric | Mixter Dissector |
| 5314 | 5 | insert | standard | Duckbill Dissector |
| 5314L | 5 | insert | bariatric | Duckbill Dissector |
| 5315 | 5 | insert | standard | Straight Tapered Dissector |
| Modular<br>Item #'s | diameter<br>(mm) | Type | Length | Descriptions |
| 5315L | 5 | insert | bariatric | Straight Tapered Dissector |
| 5316 | 5 | insert | standard | Straight Mini Dissector |
| 5316L | 5 | insert | bariatric | Straight Mini Dissector |
| 5317 | 5 | insert | standard | Mini Right Angle Dissector |
| 5317L | 5 | insert | bariatric | Mini Right Angle Dissector |
| 5318 | 5 | insert | standard | Atraumatic Tube Forcep |
| 5318L | 5 | insert | bariatric | Atraumatic Tube Forcep |
| 5319 | 5 | insert | standard | Delicate Cone Dissector |
| 5319L | 5 | insert | bariatric | Delicate Cone Dissector |
| 5320 | 5 | insert | standard | Micro Dolphin Dissecting Forcep |
| 5320L | 5 | insert | bariatric | Micro Dolphin Dissecting Forcep |
| 5321 | 5 | insert | standard | Claw Forcep 2x3 Teeth |
| 5321L | 5 | insert | bariatric | Claw Forcep 2x3 Teeth |
| 5322 | 5 | insert | standard | Biopsy Forcep w/ Two Teeth |
| 5322L | 5 | insert | bariatric | Biopsy Forcep w/ Two Teeth |
| 5323 | 5 | insert | standard | Dorsey Fenestrated Grasper |
| 5323L | 5 | insert | bariatric | Dorsey Fenestrated Grasper |
| 5324 | 5 | insert | standard | Endo Babcock Forcep |
| 5324L | 5 | insert | bariatric | Endo Babcock Forcep |
| 5325 | 5 | insert | standard | Debakey Tissue Forcep |
| 5325L | 5 | insert | bariatric | Debakey Tissue Forcep |
| 5326 | 5 | insert | standard | Extracting Forcep Reverse Teeth |
| 5326L | 5 | insert | bariatric | Extracting Forcep Reverse Teeth |
| 5327 | 5 | insert | standard | Grasping Forcep |
| Modular<br>Item #'s | diameter<br>(mm) | Type | Length | Descriptions |
| 5327L | 5 | insert | bariatric | Grasping Forcep |
| 5328 | 5 | insert | standard | Fenestrated Grasping Forcep |
| 5328L | 5 | insert | bariatric | Fenestrated Grasping Forcep |
| 5329 | 5 | insert | standard | Micro Dissecting Forcep |
| 5329L | 5 | insert | bariatric | Micro Dissecting Forcep |
| 5330 | 5 | insert | standard | Babcock Grasping Forcep |
| 5330L | 5 | insert | bariatric | Babcock Grasping Forcep |
| 5331 | 5 | insert | standard | Dolphin Nose Dissecting Forsep w/ cup |
| 5331L | 5 | insert | bariatric | Dolphin Nose Dissecting Forsep w/ cup |
| 5332 | 5 | insert | standard | Cobra Jaw Grasping Forcep |
| 5332L | 5 | insert | bariatric | Cobra Jaw Grasping Forcep |
| 5333 | 5 | insert | standard | Atraumatic Grasper |
| 5333L | 5 | insert | bariatric | Atraumatic Grasper |
| 5334 | 5 | insert | standard | Wave Jaw Grasping Forcep |
| 5334L | 5 | insert | bariatric | Wave Jaw Grasping Forcep |
| 5335 | 5 | insert | standard | Paddle Babcock |
| 5335L | 5 | insert | bariatric | Paddle Babcock |
| 5336 | 5 | insert | standard | Tri-Grasper |
| 5336L | 5 | insert | bariatric | Tri-Grasper |
| 5337 | 5 | insert | standard | Long Grasping Forcep |
| 5337L | 5 | insert | bariatric | Long Grasping Forcep |
| 5338 | 5 | insert | standard | Maxi Fenestrated Grasping Forcep |
| 5338L | 5 | insert | bariatric | Maxi Fenestrated Grasping Forcep |
| 5339 | 5 | insert | standard | Retrograde Grasping Forcep |
| Modular<br>Item #'s | diameter<br>(mm) | Type | Length | Descriptions |
| 5339L | 5 | insert | bariatric | Retrograde Grasping Forcep |
| 5340 | 5 | insert | standard | Kocher Grasping Forcep |
| 5340L | 5 | insert | bariatric | Kocher Grasping Forcep |
| 5341 | 5 | insert | standard | Curved Maryland Dissector |
| 5341L | 5 | insert | bariatric | Curved Maryland Dissector |
| 5342 | 5 | insert | standard | Dolphin Nose Grasping Forcep |
| 5342L | 5 | insert | bariatric | Dolphin Nose Grasping Forcep |
| 5343 | 5 | insert | standard | Micro Dolphin Dissector w/ Cup |
| 5343L | 5 | insert | bariatric | Micro Dolphin Dissector w/ Cup |
| 5344 | 5 | insert | standard | Oviduct Atraumatic Forcep |
| 5344L | 5 | insert | bariatric | Oviduct Atraumatic Forcep |
| 5345 | 5 | insert | standard | Traumatic Grasper, 2x3 Teeth S/A |
| 5345L | 5 | insert | bariatric | Traumatic Grasper, 2x3 Teeth S/A |
| 5346 | 5 | insert | standard | Atraumatic Grasper S/A |
| 5346L | 5 | insert | bariatric | Atraumatic Grasper S/A |
| 5347 | 5 | insert | standard | Grasping Forcep w/ cup |
| 5347L | 5 | insert | bariatric | Grasping Forcep w/ cup |
| 5348 | 5 | insert | standard | Ripple Claw Forcep |
| 5348L | 5 | insert | bariatric | Ripple Claw Forcep |
| 5349 | 5 | insert | standard | Mini-Micro Dissecting Forcep |
| 5349L | 5 | insert | bariatric | Mini-Micro Dissecting Forcep |
| 5350 | 5 | insert | standard | Curved Metz Scissor |
| 5350L | 5 | insert | bariatric | Curved Metz Scissor |
| 5351 | 5 | insert | standard | Straight Micro Scissor |
| Modular<br>Item #'s | diameter<br>(mm) | Type | Length | Descriptions |
| 5351L | 5 | insert | bariatric | Straight Micro Scissor |
| 5352 | 5 | insert | standard | Delicate Curved Metz Scissor |
| 5352L | 5 | insert | bariatric | Delicate Curved Metz Scissor |
| 5353 | 5 | insert | standard | Curved Micro Scissor |
| 5353L | 5 | insert | bariatric | Curved Micro Scissor |
| 5354 | 5 | insert | standard | Straight Metz Scissor |
| 5354L | 5 | insert | bariatric | Straight Metz Scissor |
| 5355 | 5 | insert | standard | Flat Dissector |
| 5355L | 5 | insert | bariatric | Flat Dissector |
| 5356 | 5 | insert | standard | Hook Scissor |
| 5356L | 5 | insert | bariatric | Hook Scissor |
| 5357 | 5 | insert | standard | Delicate S/A Scissor |
| 5357L | 5 | insert | bariatric | Delicate S/A Scissor |
| 5358 | 5 | insert | standard | Serrated Scissor |
| 5358L | 5 | insert | bariatric | Serrated Scissor |
| 5359 | 5 | insert | standard | Biopsy Scoop Forcep |
| 5359L | 5 | insert | bariatric | Biopsy Scoop Forcep |
| 5360 | 5 | insert | standard | Biopsy Forcep w/ One Tooth |
| 5360L | 5 | insert | bariatric | Biopsy Forcep w/ One Tooth |
| 5361 | 5 | insert | standard | Glassman Forcep, 20mm |
| 5361L | 5 | insert | bariatric | Glassman Forcep, 20mm |
| 5362 | 5 | insert | standard | Heavy Curved Dissector |
| 5362L | 5 | insert | bariatric | Heavy Curved Dissector |
| 5363 | 5 | insert | standard | Dorsey Grasper, 40mm |
| Modular<br>Item #'s | diameter<br>(mm) | Type | Length | Descriptions |
| 5363L | 5 | insert | bariatric | Dorsey Grasper, 40mm |
| 5364 | 5 | insert | standard | Tenaculum, 5mm |
| 5364L | 5 | insert | bariatric | Tenaculum, 5mm |
| 5365 | 5 | insert | standard | Medium Curved Dissector |
| 5365L | 5 | insert | bariatric | Medium Curved Dissector |
| 5366 | 5 | insert | standard | Bump Grasper, Single Action |
| 5366L | 5 | insert | bariatric | Bump Grasper, Single Action |
| 5367 | 5 | insert | standard | Single Tooth Grasper |
| 5367L | 5 | insert | bariatric | Single Tooth Grasper |
| 5368 | 5 | insert | standard | Glassman Forcep, 30mm |
| 5368L | 5 | insert | bariatric | Glassman Forcep, 30mm |
| 5369 | 5 | insert | standard | Fenestrated Bowel Grasper, 20mm |
| 5369L | 5 | insert | bariatric | Fenestrated Bowel Grasper, 20mm |
| 5370 | 5 | insert | standard | Single Action Needle Holder |
| 5370L | 5 | insert | bariatric | Single Action Needle Holder |
| 5371 | 5 | insert | standard | Suture Knot Pusher |
| 5371L | 5 | insert | bariatric | Suture Knot Pusher |
| 5372 | 5 | insert | standard | Credi Curved Dissector |
| 5372L | 5 | insert | bariatric | Credi Curved Dissector |
| 5373 | 5 | insert | standard | St. Joseph Dissecting Forcep |
| 5373L | 5 | insert | bariatric | St. Joseph Dissecting Forcep |
| 5374 | 5 | insert | standard | Atraumatic Wave Grasper |
| 5374L | 5 | insert | bariatric | Atraumatic Wave Grasper |
| 5375 | 5 | insert | standard | Curved Debakey Tissue Forcep |
| Modular<br>Item #'s | diameter<br>(mm) | Type | Length | Descriptions |
| 5375L | 5 | insert | bariatric | Curved Debakey Tissue Forcep |
| 5376 | 5 | insert | standard | Petrovich Curved Dissector |
| 5376L | 5 | insert | bariatric | Petrovich Curved Dissector |
| 5377 | 5 | insert | standard | Andrew Babcock Forcep |
| 5377L | 5 | insert | bariatric | Andrew Babcock Forcep |
| 5378 | 5 | insert | standard | Maryland with Cross Serrations |
| 5378L | 5 | insert | bariatric | Maryland with Cross Serrations |
| 5379 | 5 | insert | standard | Fenestrated Bowel Grasper, 40mm |
| 5379L | 5 | insert | bariatric | Fenestrated Bowel Grasper, 40mm |
| 5380 | 5 | insert | standard | Byerley Grasper |
| 5380L | 5 | insert | bariatric | Byerley Grasper |
| 5381 | 5 | insert | standard | Murray Grasper |
| 5381L | 5 | insert | bariatric | Murray Grasper |
| 5382 | 5 | insert | standard | Double Action Spoon Forcep, 5mm |
| 5382L | 5 | insert | bariatric | Double Action Spoon Forcep, 5mm |
| 5383 | 5 | insert | standard | Biopsy Forcep w/ One Tooth (on<br>bottom) |
| 5383L | 5 | insert | bariatric | Biopsy Forcep w/ One Tooth (on<br>bottom) |
| 5384 | 5 | insert | standard | 5mm Smoot Endo Cinch |
| 5384L | 5 | insert | bariatric | 5mm Smoot Endo Cinch |
| 5385 | 5 | insert | standard | Single Action Spoon Forcep |
| 5385L | 5 | insert | bariatric | Single Action Spoon Forcep |
| 5386 | 5 | insert | standard | Harvey Grasper |
| 5386L | 5 | insert | bariatric | Harvey Grasper |
| Modular<br>Item #'s | diameter<br>(mm) | Type | Length | Descriptions |
| 5387 | 5 | insert | standard | Rodriguez Grasper |
| 5387L | 5 | insert | bariatric | Rodriguez Grasper |
| 5388 | 5 | insert | standard | Forsythe Bump Grasper, Single Action |
| 5388L | 5 | insert | bariatric | Forsythe Bump Grasper, Single Action |
| 5389 | 5 | insert | standard | Angled Dissector |
| 5389L | 5 | insert | bariatric | Angled Dissector |
| 5390 | 5 | insert | standard | 5mm Meah Grasper |
| 5390L | 5 | insert | bariatric | 5mm Meah Grasper |
| 5391 | 5 | insert | standard | Kowalski Hunter Bowel Grasper |
| 5391L | 5 | insert | bariatric | Kowalski Hunter Bowel Grasper |
| 5392 | 5 | insert | standard | Devlin Microfenestrated Grasper |
| 5392L | 5 | insert | bariatric | Devlin Microfenestrated Grasper |
| 5393 | 5 | insert | standard | Holmes Atraumatic Grasper |
| 5393L | 5 | insert | bariatric | Holmes Atraumatic Grasper |
| 5394 | 5 | insert | standard | Graves Grasper |
| 5394L | 5 | insert | bariatric | Graves Grasper |
| 5395 | 5 | insert | standard | 5mm Maryland with 90 Degree Tooth |
| 5395L | 5 | insert | bariatric | 5mm Maryland with 90 Degree Tooth |
| SLRVA | 5 | handle/tube | standard | Take Apart Slide Lock with Rotation |
| SLRVL | 5 | handle/tube | bariatric | Take Apart Slide Lock with Rotation |
| 5SLR | 5 | handle/tube | standard | Take Apart Slide Lock with Rotation |
| 5SLRL | 5 | handle/tube | bariatric | Take Apart Slide Lock with Rotation |
| 5MRR | 5 | handle/tube | standard | Monopolar Handle with Rotation,<br>Ratchet on Demand |
| 5MRRL | 5 | handle/tube | bariatric | Monopolar Handle with Rotation |
| Modular<br>Item #'s | diameter<br>(mm) | Type | Length | Descriptions |
| | | | | Ratchet on Demand |
| 5MROT | 5 | handle/tube | standard | Monopolar Handle with Non-Locking<br>Rotation, NO Ratchet |
| 5MROTL | 5 | handle/tube | bariatric | Monopolar Handle with Non-Locking<br>Rotation, NO Ratchet |
Below is a complete listing of laparoscopic instruments covered by this 510(k) submission:
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
{12}------------------------------------------------
{13}------------------------------------------------
{14}------------------------------------------------
{15}------------------------------------------------
{16}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.