K151541 · GC America, Inc. · EBF · Oct 9, 2015 · Dental
Device Facts
Record ID
K151541
Device Name
G-Fix
Applicant
GC America, Inc.
Product Code
EBF · Dental
Decision Date
Oct 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Splinting mobile teeth
Device Story
G-Fix is a light-cured resin cement used by dental professionals to splint mobile teeth. The device is applied to tooth surfaces following pre-treatment with phosphoric acid. It functions via a photo-initiator system that triggers the polymerization of methacrylate ester monomers when exposed to light. The material is supplied in dental composite dispensing syringes. By providing a stable bond to the enamel, the device stabilizes mobile teeth, potentially improving patient comfort and dental function. It is intended for clinical use by dentists.
Clinical Evidence
Bench testing only. The device was evaluated against ISO 4049:2009 and ISO/TS 11405:2003 standards. Testing included sensitivity to ambient light, depth of cure (5.7mm), flexural strength (94 MPa), water sorption (35 µg/mm³), solubility (0.6 µg/mm³), color stability, and shear bond strength to enamel (29 MPa). Biocompatibility was assessed per ISO 10993-1, 10993-5, and 10993-10, confirming no cytotoxicity, irritation, or significant sensitization.
Technological Characteristics
Light-cured resin cement composed of barium glass, UDMA, Bis-MEPP, phosphoric ester monomer, silicon dioxide, photo initiator, and pigments. Operates via photo-initiated polymerization of methacrylate ester monomers. Form factor is a 2.7g dental composite dispensing syringe. Complies with ISO 4049:2009 and ISO/TS 11405:2003. No software or electronic components.
Indications for Use
Indicated for splinting mobile teeth.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized emblem. The emblem consists of three abstract human profiles facing to the right, stacked on top of each other. The text reads 'DEPARTMENT OF HEALTH & HUMAN SERVICES - USA'.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 9, 2015
GC America Inc. Mr. Mark Heiss Director, Regulatory Affairs & Academics Affairs 3737 W. 127th Street Alsip, Illinois 60803
Re: K151541
Trade/Device Name: G- Fix Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth shade resin material Regulatory Class: II Product Code: EBF Dated: July 7, 2015 Received: July 8, 2015
Dear Mr. Heiss:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Section 4 - Indications for Use Statement
### Indications for Use
510(k) Number (if known): __K151541 Device Name: G-FIX Indications for Use:
- 1. Splinting mobile teeth
Prescription Use _____________________________________________________________________________________________________________________________________________________________ Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the green text "'CC'" on a white background. The text is stylized with the letters "CC" being larger than the apostrophes. The apostrophes are also slanted. The text is in a sans-serif font.
GC AMERICA INC.
3737 WEST 127TH STREET
ALSIP, ILLINOIS 60803
TEL (708) 597-0900
FAX (708) 371-5103
- 1. Submitter Information:
GC AMERICA INC. 3737 W. 127th Street Alsip, IL 60803
| Contact Person: | Mark Heiss, D.D.S. |
|-----------------|--------------------|
| Phone: | (708) 926-3090 |
| Fax: | (708) 926-9100 |
| Date Prepared: | June 8, 2015 |
#### 2. Device Name:
| Proprietary Name: | G-Fix |
|------------------------|----------------------------|
| Classification Name: | Tooth shade resin material |
| Device Classification: | Class II, 872.3690 |
| Product Code: | EBF |
#### Predicate Devices: 3.
| Company | Device | 510(k) No. | Date Cleared |
|-----------------|---------------------------------|------------|--------------|
| GC America Inc. | GRADIA DIRECT Flo (UNIFIL FLOW) | K020342 | 03/28/2002 |
#### 4. Description of Device:
G-Fix is a light-cured resin cement for splinting mobile teeth. G-Fix is used in combination with phosphoric acid as a pre-treatment agent of tooth surfaces.
#### న్. Indications for Use:
1. Splinting mobile teeth
#### 6. Technological characteristics:
All the components of the applicant device, G-Fix, have already been used in the predicate device. The curing mechanism of the predicate is polymerization of uncured methacrylate ester monomers. This reaction is caused by photo initiator system.
#### 7. Performance Bench Tests
It is confirmed that the device conforms to the required specifications of ISO 4049:2009 and ISO/TS 11405: 2003 and is suitable for its intended use. Performance testing includes:
- Sensitivity to ambient light
- Depth of cure
- · Flexural strength
- · Water sorption
- · Solubility
- Color stability after irradiation and water sorption
- · Adhesion to tooth structure (Enamel)
{4}------------------------------------------------
#### 8. Packaging
One (1) Dental Composite Dispensing Syringe 2.7g (2.0mL), 2 dispensing tips (plastic type), 1 light protective cover
Dispensing tip package: 30 dispensing tips (plastic type), 2 light protective covers 30 dispensing tips (needle type), 2 light protective covers
- 9. Shades
Clear and TC (Tooth Color)
- 10. Shelf Life Evaluation and Storage Conditions:
- Shelf Life 2 years i
- Recommended for optimal performance, store in a cool and dark place. 4-25°C (39.2 77.0°F)
- 11. Biocompatibility
Biocompatibility studies were conducted based on ISO 10993-1, 10993-5, and 10993-10 for cytotoxicity, sensitivity, and irritation.
| Test | ISO standard | Results | Conclusion |
|--------------|--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sensitivity | 10993-10 | As defined by the scoring system of<br>Kligman, this is a Grade I reaction and<br>the test article is classified as having<br>weak allergenic potential. | Based on the criteria of the protocol, a<br>Grade I sensitization rate is not considered<br>significant and the test article meets the<br>requirements of the ISO 10993-10<br>guidelines. |
| Cytotoxicity | 10993-5 | There was no biological reactivity of the<br>cells exposed to the test article extract.<br>The response obtained from the positive<br>and negative control article extracts<br>confirmed the suitability of the test<br>system. | Based on the criteria of the protocol and the<br>ISO 10993-5 guidelines, the test article<br>meets the requirements of the test and is<br>not considered to have a cytotoxic effect. |
| Irritation | 10993-10 | The test article sites did not show a<br>significantly greater biological reaction<br>than the sites injected with the control<br>article. | Based on the criteria of the protocol, the test<br>article meets the requirements of the ISO<br>10993-10 guidelines. |
Results
In addition, according to ISO 10993-1, products currently on the market may be considered as it relates to product biocompatibility. There have been no major adverse events reported for G-Fix.
{5}------------------------------------------------
| | Property | Standards | Requirements | Test results | | Compare<br>to<br>Standard | Compare<br>to<br>Predicate<br>Device |
|---|------------------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|--------------------------------------|
| 1 | Sensitivity to<br>ambient light | ISO 4049: 2009<br>5.2.7 Sensitivity<br>to ambient light | Remain<br>physically<br>homogeneous. | G-Fix: Clear<br>Lot No.<br>1103011G<br>Conformed | GRADIA DIRECT<br>Flo (UNIFIL<br>FLOW): A1<br>Lot No. 1111211<br>Conformed | | |
| 2 | Depth of cure | ISO 4049: 2009<br>5.2.8 Depth of<br>cure | Not less than<br>1.5 mm | 5.6 mm<br>5.6 mm<br>5.7 mm<br>5.7 mm<br>5.7 mm<br>5.7(0.1) mm | 3.0 mm<br>3.0 mm<br>2.9 mm<br>3.0 mm<br>3.0 mm<br>3.0(0.0) mm | Exceeds | Exceeds |
| 3 | Flexural<br>strength | ISO 4049: 2009<br>5.2.9 Flexural<br>strength | Greater than<br>80 MPa | 100 MPa<br>89 MPa<br>99 MPa<br>93 MPa<br>90 Mpa<br>94(5) MPa | 96 MPa<br>101 MPa<br>99 MPa<br>93 MPa<br>93 Mpa<br>96(3.2) MPa | Exceeds | Equal |
| 4 | Water<br>sorption | ISO 4049: 2009<br>5.2.10 Water<br>sorption and<br>solubility | Less than 40<br>µg/mm³ | 36.1 µg/mm³<br>33.9 µg/mm³<br>34.4 µg/mm³<br>33.5 µg/mm³<br>34.6 µg/mm³<br>35(0.9) µg/mm³ | 33.1 µg/mm³,<br>35.4 µg/mm³,<br>36.7 µg/mm³,<br>34.0 µg/mm³,<br>33.9 µg/mm³<br>35(1.3) µg/mm³ | Exceeds | Equal |
| 5 | Solubility | ISO 4049: 2009<br>5.2.10 Water<br>sorption and<br>solubility | Less than 7.5<br>µg/mm³ | 0.8 µg/mm³<br>0.4 µg/mm³<br>0.6µg/mm³<br>0.7µg/mm³<br>0.7µg/mm³<br>0.6(0.1) µg/mm³ | 0.5 µg/mm³<br>0.5 µg/mm³<br>0.4 µg/mm³<br>0.6 µg/mm³<br>0.7 µg/mm³<br>0.5(0.1)µg/mm³ | Exceeds | Equal |
| 6 | Color stability<br>after<br>irradiation<br>and water<br>sorption | ISO 4049: 2009<br>5.4 Color<br>stability after<br>irradiation and<br>water sorption | No more than<br>slight change<br>in color | Conformed | Conformed | Equal | Equal |
| 7 | Adhesion to<br>tooth<br>structure<br>(Enamel) | ISO/TS 11405:<br>2003 5.2.5<br>Shear bond<br>strength | Greater than<br>5MPa | 31 MPa<br>29 MPa<br>28 MPa<br>30 MPa<br>27 Mpa<br>29(1) MPa | 34 MPa<br>31 MPa<br>29 MPa<br>28 MPa<br>32 MPa<br>31(2) MPa<br>*This result was<br>obtained by the<br>combination with<br>phosphoric acid | Equal | Equal |
{6}------------------------------------------------
| | Applicant device | Comparative device | |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name | G-Fix | GRADIA DIRECT Flo (UNIFIL FLOW) | |
| Common Name | Light-cured resin cement for splinting mobile teeth | Light-cured Flowable composite restorative | Different* |
| Product<br>Classification | Tooth Shade Resin Material | Tooth Shade Resin Material | Same |
| Company | GC Corporation | GC Corporation | Same |
| 510(k) No. | - | K020342 | |
| Indications for use | 1. Splinting mobile teeth | 1. Restoration of class I, II, III, IV, V cavities<br>2. Restoration of root surface caries<br>3. Restorations in deciduous teeth<br>4. Filling tunnel shaped cavities<br>5. Sealing hypersensitive areas<br>6. Liner / base / filling in cavity undercuts<br>7. Sealant<br>8. Fixing mobile teeth<br>9. Additions to composite restorations | Different<br>Note: #8 of<br>Comparitive<br>Device is<br>same as<br>applicant. |
| Product<br>description | G-Fix is a visible-light-cured resin cement for fixing<br>splinting mobile teeth . G-Fix is used in<br>combination with phosphoric acid as a<br>pre-treatment agent of tooth surfaces. | GRADIA DIRECT Flo (UNIFIL FLOW) is a light cured<br>highly flowable composite filled in a dental syringe.<br>The material is available in 7 shades. | Both light cured<br>Polymer based<br>material |
| Components | * Barium glass<br>* Urethane dimethacrylate (UDMA)<br>* Bisphenol A polyethoxy methacrylate<br>(Bis-MEPP)<br>* Phosphoric ester monomer<br>* Silicon dioxide<br>* Photo initiator<br>* Pigment | * Fluoro-alumino-silicate glass<br>* Urethane dimethacrylate (UDMA)<br>* Dimethacrylate<br>* Silicon dioxide<br>* Photo initiator<br>* Pigment | Components of<br>new device are<br>contained in<br>predicate and<br>device includes<br>Phosphoric<br>monomer that is<br>common in this<br>device type |
| Instructions for use | 1. Preparations<br>2. Shade Selection<br>3. Preparation in case of tooth structure (apply<br>phosphoric acid to the bonding surfaces of teeth for<br>30 seconds)<br>4. Preparation in case of restoration<br>5. Placement of G-Fix<br>6. Light Curing<br>7. Finishing and Polishing | 1. Shade Selection<br>2. Cavity Preparation<br>3. Bonding treatment<br>4. Placement of GRADIA DIRECT Flo (UNIFIL Flo)<br>5. Light Curing<br>6. Finishing and Polishing | Different in that<br>the indications<br>are narrower<br>than the<br>predicate |
| | | | Page 4 |
{7}------------------------------------------------
### Substantial equivalence:
The applicant device comples with the requirements of ISO 4049: 2009 (Dentisty - Polymer-based restorative materials) and ISO/TS 11405: 2003. The curing mechanism of the new and predicationially equivalent in principle. Therefore, the new and predicate devices are the same in function, and similar in composition and intended use. This supports that the compatibility of the applicant device are substantially equivalent to the predicate devices. Components of the new device are commonly used in other restorative materials.
The applicant device shows one indication with predicate device which is "splinting mobile teeth."
### Differences
Common Name - G-Fix description is different from predicate device even though product classification is the same.
The Indications for Use is different as clinically the product is used to splint teeth.
Comparison to performance and predicate device have been met or exceeded.
Product description is different as clinically this product is used to splint teeth (predicate is more general as a restorative).
Components - additional use of Bis-MEPP is to modify viscosity as use for splinting teeth. Phosphoric ester monomer added to allow for adhesion to enamel.
Instructions for Use is different as clinically the product is used to splint teeth (predicate is more general as a restorative).
### Technological characteristics:
All the components of the applicant device, G-Fix, have already been used in the predicate is polymerization of uncured methacrylate ester monomers. This reaction is caused by photo initiator system.
Use of 10-MDP (phosphoric ester monomer) allows for adhesion to tooth structure without the of dentin adhesive.
## Conclusion
Based on formulation, testing and meeting and/or exceeding ISO standards related to performance, as well as meeting at least equivalence versus the predicate device, we find the applicant device to be substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.