The NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube device is inserted into the stomach through the nasal cavity for the purpose of supplying a nutritional source for those patients who are unable to feed themselves.
Device Story
Enteral feeding tube for continuous or intermittent delivery of nutritional sources; inserted via nasal or oral cavity into stomach, duodenum, or small intestine. Comprises radio-opaque polyurethane tube, guide wire set, and ENFit connector. Practitioner inserts tube using guide wire; guide wire removed post-insertion. ENFit connector allows attachment of female ENFit enteral syringes. Used in clinical settings by healthcare practitioners. Provides nutritional access for patients unable to feed themselves. Design includes depth markings for insertion reference and dual distal openings for flow. Single-use, sterile device.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including biocompatibility (ISO 10993), liquid leakage, stress cracking, tensile testing, corrosion testing, and flow rate testing per EN 1618:1997 and AAMI/CN3:2014(PS).
Technological Characteristics
Polyurethane tube; radio-opaque; 6Fr to 12Fr diameter; 50cm-120cm length. Features ENFit/ISO 80369-1 compliant connector. Includes metal guide wire. Sterile (Ethylene Oxide). Single-use. Biocompatibility per ISO 10993. Mechanical testing per AAMI/CN3:2014(PS).
Indications for Use
Indicated for patients unable to feed themselves requiring nutritional support via nasogastric tube insertion.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
Predicate Devices
CORFLO Anti-IV Enteral Feeding Tube Model Number 20-XXX (K083210)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing movement or progress.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 5, 2016
Gbuk Enteral Limited % Leeanne Baker Regulatory Consultant Advena Medical Pure Offices, Plato Close Warwick, CV34 6WE United Kingdom
Re: K151503
> Trade/Device Name: NUTRICARE - ISOSAF™ Naso-Gastric Feeding Tube Regulation Number: 21 CFR 876.5980 Regulation Name: Gastrointestinal Tube and accessories Regulatory Class: Class II Product Code: PIF Dated: February 2, 2016 Received: February 12, 2016
Dear Leeanne Baker,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
**Herbert P. Lerner -S**
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151503
Device Name
Nutricare ISO-SAF Naso-Gastric Feeding Tube
Indications for Use (Describe)
The NUTRICARE ISO-SAF Naso-gastric Feeding Tube device is inserted into the stomach through the nasal cavity for the purpose of supplying a nutritional source for those patients who are unable to feed themselves.
| Type of Use (Select one or both, as applicable) | <table><tr><td><div> <span style="text-decoration: underline;">☑</span> Research Use (Per 21 CFR 201.66; Subject to Review) </div></td><td><div> ☐ Other Commercial Use (21 CFR 201.66) </div></td></tr></table> | <div> <span style="text-decoration: underline;">☑</span> Research Use (Per 21 CFR 201.66; Subject to Review) </div> | <div> ☐ Other Commercial Use (21 CFR 201.66) </div> |
|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|
| <div> <span style="text-decoration: underline;">☑</span> Research Use (Per 21 CFR 201.66; Subject to Review) </div> | <div> ☐ Other Commercial Use (21 CFR 201.66) </div> | | |
X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## for
# NUTRICARE – ISOSAF™ NASO-GASTRIC FEEDING TUBE
#### 1. Submission Sponsor
- GBUK Enteral Limited Blackwood Hall Business Park North Duffield Selby North Yorkshire, YO8 5DD United Kingdom Phone: +44 (0)1757 282945 Contact: Mark Thompson, Managing Director
#### 2. Submission Correspondent
Advena Medical Pure Offices Plato Close Warwick CV34 6WE United Kingdom Cell Phone: +44 (0)7876 688344 Office Phone +44 (0)1926 800153 Contact: Leeanne Baker, Regulatory Consultant Email: Leeanne.baker@advenamedical.com
#### 3. Date Prepared
20th May 2015
## 4. Device Identification
| Trade/Proprietary Name: | NUTRICARE – ISOSAFTM NASO-GASTRIC FEEDING TUBI |
|----------------------------|------------------------------------------------|
| Common/Usual Name: | Tube, feeding |
| Classification Name: | Gastrointestinal tube and accessories |
| Classification Regulation: | 21 CFR 876.5980 |
| Product Code: | PIF |
| Device Class: | Class II |
| Classification Panel: | Gastroenterology/ Urology |
#### 5. Legally Marketed Predicate Device(s)
K083210 – CORFLO Anti-IV Enteral Feeding Tube Model Number 20-XXX
{4}------------------------------------------------
## 6. Device Description
The NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube is an enteral feeding tube designed for continuous or intermittent use. It comprises of radio-opaque polyurethane tube with guide wire set and ENFit connector.
The guide wire set is composed of a rigid guide wire and a male threaded connector, which is inserted into the polyurethane tube. They're depth marks printed on the tube in order to provide reference about the depth of insertion to the practitioner.
The tube is to be inserted through the nasal or oral area and feed down into the alimentary canal system (i.e. stomach area, duodenum or small intestine as intended). The guide wire with male connector is removed when the insertion is done leaving the tube and ENFit connector in situ. Only a female ENFit enteral syringe can be attached to the ENFit connector when in use.
# 7. Indication for Use Statement
The NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube device is inserted into the stomach through the nasal cavity for the purpose of supplying a nutritional source for those patients who are unable to feed themselves.
# 8. Substantial Equivalence Discussion
The following table compares the NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| Table 5A - Comparison of Characteristics | | |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | GBUK ENTERAL LTD | CORPAC INC. |
| Trade Name | NUTRICARE ISO-SAF™ Naso-<br>gastric Feeding Tube | Corflo Anti IV Enteral Feeding<br>Tube |
| 510(k) Number | - | K083210 |
| Product Code | PIF | FPD |
| Regulation<br>Number | 21 CFR 876.5980 | 21 CFR 876.5980 |
| Regulation Name | Gastrointestinal tubes and<br>accessories | Gastrointestinal tubes and<br>accessories |
| Indications for<br>Use | The NUTRICARE ISO-SAF™ Naso-<br>gastric Feeding Tube device is<br>inserted into the stomach<br>through the nasal cavity for the<br>purpose of supplying a<br>nutritional source for those<br>patients who are unable to feed<br>themselves. | The CORFLO® ANTI IV Enteral<br>Feeding tube is intended for use<br>in those patients who require<br>intermittent or continuous tube<br>feedings via the nasogastric or<br>nasoenteric route |
| Material | Polyurethane | Polyurethane |
## Table 54 - Comnarison of Characteristics
{5}------------------------------------------------
| Manufacturer | GBUK ENTERAL LTD | CORPAC INC. |
|------------------------------------|---------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Trade Name | NUTRICARE ISO-SAFT™ Naso-gastric Feeding Tube | Corflo Anti IV Enteral Feeding Tube |
| Sterile | Yes - Ethylene Oxide | Rest unknown<br>Yes - Ethylene Oxide |
| Overall Design | | |
| Proximal Connector | Single connector ENFit/ISO 80369-1 compliant | Luer slip two (2) ports |
| Distal feeding port | Dual openings on opposite sides of tube at offset levels | Single opening |
| Distal Tip | Continuous with diameter of rest of tube. Extra radiopaque tip | Contoured larger than diameter of rest of tube |
| Fully Radiopaque throughout length | Yes | Yes |
| Cm Markings | Numbered markings throughout the length of the tube | Markings through the length of the tube |
| Overall length | 50cm - 120cm | 55cm - 140cm |
| Guidewire | Twisted metal wire with sealed tip ends prior to the feeding port to prevent accidental emergence | Braided metal wire<br>Tip ends prior to the feeding port to prevent accidental emergence |
| Diameter | 6Fr to 12Fr | 5Fr to 12Fr |
| Duration of Use | Short term (up to 30 days) | Short term |
| Single-Use | Yes | Yes |
| Shelf Life | 3 years | 5 years |
| Latex Free | Yes | Yes |
| Complies with ISO 10993-1 | Cytotoxicity<br>Sensitization<br>Intracutaneous Reactivity | Unknown but assumed to comply as has 510k |
| ISO 80369-1 ENFIT compliant | Yes | Current product has luer slip connector |
#### 9. Non-Clinical Performance Data
The biocompatibility of the NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube has been demonstrated through appropriate in vitro tests on the device. The device has been assessed in accordance with ISO 10993 and does not introduce any additional safety risk over the predicate device.
The following testing has been performed (per AAMI/CN3:2014(PS) where applicable) to support substantial equivalence:
- Falling drop positive pressure liquid leakage pass ●
- . Stress Cracking - pass
- . Resistance to separation from axial load – pass
- Resistance to separation from unscrewing – pass
- . Resistance to overriding – pass
- Disconnection from unscrewing pass .
{6}------------------------------------------------
- . Tensile testing of the catheter – pass
- . Tensile testing of the connectors – pass
- . Corrosion testing of guide wire – pass
- Liquid leakage - pass
A comparative study performed demonstrates that the NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube performs as intended and is safe for its intended use when compared with the predicate device.
A Risk Analysis including analysis of the risk of misconnection has been conducted in accordance with EN ISO 14971:2012
As part of demonstrating safety and effectiveness of NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube and in showing substantial equivalence to the predicate device that are subject to this 510(k) submission, GBUK Enteral Limited completed a number of test as detailed below:
- . Flow Rate Testing in accordance with EN 1618:1997
- Mechanical Testing of the connector in accordance with AAMI/CN3:2014(PS) and AAMI/CN3:2014(PS), respectively
- Dimensional analysis in accordance with AAMI/CN3:2014(PS)
The NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube passed all testing stated above as shown by the acceptable results obtained.
The NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube complies with the applicable voluntary standards for biocompatibility and sterilization. The device passed all the testing in accordance with national and international standards.
## 10. Clinical Performance Data
There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The non-clinical testing detailed in this submission supports the substantial equivalence of the device.
## 11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device. Or the device has the same intended use and different technological characteristics that can be demonstrated that the device is substantially equivalent to the predicate device, and that the new device does not raise different questions regarding its safety and effectiveness as compared to the predicate device.
It has been shown in this 510(k) submission that the difference between the NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube and the predicate devices do not raise any questions regarding its safety and effectiveness. The NUTRICARE ISO-SAF™ Naso-gastric Feeding Tube, as designed and manufactured, is determined to be substantially equivalent to the
{7}------------------------------------------------
referenced predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.