K151491 · Falck Medical, Inc. · HKY · Jan 15, 2016 · Ophthalmic
Device Facts
Record ID
K151491
Device Name
Falck Medical Multifunction Tonometer, FAT1
Applicant
Falck Medical, Inc.
Product Code
HKY · Ophthalmic
Decision Date
Jan 15, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1930
Device Class
Class 2
Indications for Use
The Falck Medical Multi-function Tonometer (FAT1) is indicated for the measurement of intraocular pressure, ocular pulsatile amplitude, force (Ophthalmodynamometry) and estimating conventional outflow facility with calibrated force application (Indentation Tonography).
Device Story
The FAT1 is a slit-lamp mounted tonometer using a disposable plastic prism. It measures intraocular pressure via the Imbert-Fick applanation method; ocular pulsatile amplitude and central retinal artery pulsation via optical capture; and conventional outflow facility via indentation tonography with calibrated force. A microprocessor controls force application and performs statistical analysis on samples taken every 7 milliseconds to ensure repeatability and accuracy. The device includes a detection system to prevent prism reuse. It is intended for use by clinicians in a clinical setting to assist in ophthalmic diagnosis and patient management.
Clinical Evidence
Clinical testing conducted under IDE (1080757) for the measurement of conventional outflow facility demonstrated the safety, repeatability, precision, and accuracy of the FAT1. Biocompatibility testing (cytotoxicity, systemic toxicity, ocular irritation) confirmed the safety of the disposable prism material.
Technological Characteristics
Slit-lamp mounted device with floating prism arm. Uses disposable plastic prisms. Sensing principle: total internal reflectance. Microprocessor-controlled force application. Sampling rate: 7ms. Connectivity: Not specified. Sterilization: Not specified.
Indications for Use
Indicated for measurement of intraocular pressure, ocular pulsatile amplitude, central retinal artery pulsation force (Ophthalmodynamometry), and estimation of conventional outflow facility (Indentation Tonography) in patients requiring ophthalmic assessment.
Regulatory Classification
Identification
A tonometer and accessories is a manual device intended to measure intraocular pressure by applying a known force on the globe of the eye and measuring the amount of indentation produced (Schiotz type) or to measure intraocular tension by applanation (applying a small flat disk to the cornea). Accessories for the device may include a tonometer calibrator or a tonograph recording system. The device is intended for use in the diagnosis of glaucoma.
Model 30 Classic Pneumatonometer with Tonography Option (K002395)
Submission Summary (Full Text)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 15, 2016
Falck Medical, Inc. Francis Y. Falck, Jr., M.D., Ph.D., M.S. Chair, Scientific Advisory Board 35 Washington Street Mystic, CT 06355
Re: K151491
Trade/Device Name: Falck Medical Multifunction Tonometer, FAT1 Regulation Number: 21 CFR 886.1930 Regulation Name: Tonometer and Accessories Regulatory Class: Class II Product Code: HKY, HPK, NJJ Dated: December 7, 2015 Received: December 14, 2015
Dear Dr. Falck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
-(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation
(21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Y. Alexander -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151491
Device Name
Falck Medical Multifunction Tonometer, FAT1
Indications for Use (Describe)
The Falck Medical Multi-function Tonometer (FAT1) is indicated for the measurement of intraocular pressure, ocular pulsatile amplitude, force (Ophthalmodynamometry) and estimating conventional outflow facility with calibrated force application (Indentation Tonography).
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Submitter Information:
| 1. Company Name/Address: | Falck Medical, Inc.<br>35 Washington Street<br>Mystic, CT 06355<br>860-536-9000 |
|----------------------------|----------------------------------------------------------------------------------|
| 2. Contact Information: | Francis Y. Falck, Jr., M.D., Ph.D., M.S.<br>CEO, Chair Scientific Advisory Board |
| 3. Manufacturing Location: | VR Industries, 333 Strawberry Field Road,<br>Warwick, RI 02886 |
# Device Information:
- 1. Device Trade Name: Falck Medical Multi-Function Tonometer (FAT1)
- 2. Classification Name: Tonometer
- 3. Classification Panel: 86 (Ophthalmic)
- 4. Classification Number: 886.1930
- 5. Product Codes: HKY, HPK, NJJ
# Substantial Equivalence:
The Falck Medical Multi-Function Tonometer, (FATI) is substantially equivalent to the FAT2 device (K071755) for the measurement of intraccular pressure, ocular pulsatile amplitude and the force required to observe pulsation of the central retinal artery (Ophthalmodynamometry). The FAT1 device is technically and functionally equivalent to the Falck Medical, Inc. FAT2 device. The FAT1 device is also substantially equivalent to the Model 30 Classic Pneumatonometer with Tonography Option (K002395) for measuring the change in intraocular pressure with force application from the outflow of aqueous humor through the conventional outflow system (indentation Tonography).
{4}------------------------------------------------
# Intended Use:
The Falck Medical Multi-function Tonometer (FAT1) is indicated for the measurement of intraocular pressure, ocular pulsatile amplitude, force (Ophthalmodynamometry) and estimating conventional outflow facility with calibrated force application (Indentation Tonography).
# Device Description:
The Falck Medical Multi-function Tonometer (FAT1) mounts on the slit lamp microscope. The device has a floating prism arm which holds a disposable plastic prism. The device has a detection system located in the prism arm head that prevents re-use of the prism.
The prism optical system uses the principle of total internal reflectance. Sampling is every 7 milliseconds and multiple samples are obtained for each measurement. Each measurement is statistically analyzed for repeatability and accuracy. Intraocular pressure is measured using the Imbert-Fick method of applanation. Force application is controlled by the microprocessor. The ocular pulsatile amplitude and central retinal artery pulsation with force application is optically captured. Indentation tonography methodology using calibrated force application is used to measure conventional outflow facility.
# Performance Testing:
The disposable prism was tested for biocompatibility. The prism and prism material in cyto-toxicity, systemic toxicity and ocular irritation testing was found to be non-toxic and safe. The FAT1 device is technically and functionally equivalent to the FAT2 device for the measurement of intraocular pressure, ocular pulsatile amplitude and the force required to observe pulsation of the central retinal artery. The accuracy and safety of the FAT2 device for the measurement of intraocular pressure, ocular pulsatile amplitude and the amount of force required to observe pulsation of the central retinal artery was demonstrated in performance testing in 510(k) K071755 and in additional correspondence with the Food and Drug Administration, Office of Device Evaluation dated 11/03/2009 and 01/10/2010 -FMI.
Bench testing and clinical testing done under FDA guidance with IDE (1080757) for the measurement of conventional outflow facility demonstrated the safety, repeatability, precision and accuracy of the FAT1.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.