K151482 · Bcr Diagnostics, Inc. · FRC · Feb 26, 2016 · General Hospital
Device Facts
Record ID
K151482
Device Name
BCR-30-min BI™ and RBIP-ST30™ Readout Accessory
Applicant
Bcr Diagnostics, Inc.
Product Code
FRC · General Hospital
Decision Date
Feb 26, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2800
Device Class
Class 2
Indications for Use
Use the BCR-30-min BI™ and the RBIP-ST30™ together for monitoring performance of gavity displacement steam sterlization cycles of 30 minutes provides final fluorescence results in 30 minutes.
Device Story
System comprises BCR-30-min BI (biological indicator) and RBIP-ST30 (readout accessory). BI consists of filter paper inoculated with Geobacillus stearothermophilus spores. After sterilization exposure, BI is placed in RBIP-ST30. Device converts spores to F-spores, then incubates at 55±2°C for 30 minutes. RBIP-ST30 uses blue LED, optical filters, and CCD array to detect fluorescence. Presence of fluorescence above a cut-off value indicates sterilization cycle failure; absence indicates success. Used in clinical settings to validate sterilization performance. Provides rapid results compared to conventional 7-day incubation, enabling timely clinical decision-making regarding instrument sterility.
Clinical Evidence
Bench testing only. Performance validated using three separate spore lots. Demonstrated compliance with ANSI/AAMI/ISO 11138-1:2006/(R) 2010 and 11138-3:2006/(R) 2010. Resistance characteristics (D-value 1.9 min, Z-value 15.5°C, survival time 7.9 min, kill time 19 min) confirmed. Rapid readout (30 min) showed >97% alignment with 7-day conventional incubation results.
Technological Characteristics
BI: Geobacillus stearothermophilus spores on glass fiber paper. Readout: 55±2°C incubation, blue LED excitation, optical filters, CCD array detection. Connectivity: 110V AC. Safety/EMC: EN 61326-1, FCC Part 15 Class A. Software: Automated fluorescence threshold detection.
Indications for Use
Indicated for monitoring gravity displacement steam sterilization cycles (30 min at 250°F/121°C) using Geobacillus stearothermophilus spores.
Regulatory Classification
Identification
Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.
Predicate Devices
3M Attest™ 1292 Rapid Readout Biological Indicator for Steam (K090569)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of three faces in profile, stacked on top of each other. The faces are stylized and appear to be moving forward.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 26, 2016
BCR Diagnostics, Inc. Boris Rotman, Ph.D. Vice President, CSO 145 S. 79th Street - Ste. 12 Chandler. AZ 85226
Re: K151482
Trade/Device Name: BCR-30-min BITM and RBIP-ST30™ Readout Accessory Regulation Number: 21 CFR 880.2800 Regulation Name: Chemical/Physical Sterilization Indicator Regulatory Class: II Product Code: FRC Dated: January 27, 2016 Received: February 1, 2016
Dear Dr. Rotman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
Page 2 - Dr. Rotman
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# A.6 Statement of Indications for Use
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
Indications for Use
| 510(k) Number (if known) | K151482 |
|--------------------------|---------|
| Device Name | |
BCR-30-min BI™ and RBIP-ST30™ Readout Accessory
Indications for Use (Describe)
lndications for Use (Desorite)
Use the BCR-30-min BI™ and the RBIP-ST30™ together for monitoring performance of gavity displacement steam
sterlization cycles of 30 minutes provides final fluorescence results in 30 minutes.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (1011) 443-6780 - EF
============================================================================================================================================================================== ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for BCR Diagnostics. The letters "BCR" are stacked on top of each other in a bold, sans-serif font. The word "Diagnostics" is written in a smaller, italicized font to the right of the letters "BCR". The logo is simple and professional.
# 510(k) Pre-market Notification Summarv K151482
#### Sponsor Information:
BCR Diagnostics, Inc. 145 S. 79th Street - Ste. 12 Chandler, AZ 85226
| Contact Person: | Boris Rotman, Ph.D., CSO. |
|------------------|---------------------------|
| Phone Number: | (480) 809-6156 |
| FAX Number: | (480) 809-6001 |
| email: | bcr@bcr.necoxmail.com |
| Date of Summary: | February 25, 2016 |
#### Device Name and Classification:
| Common or Usual Name: | Biological Indicator |
|-------------------------|-------------------------------------------------|
| | Biological Indicator Readout Accessory |
| Trade/Proprietary Name: | BCR-30-min BI™ and RBIP-ST30™ Readout Accessory |
| Classification Name: | Chemical/Physical Sterilization Indicator |
| Regulation Number: | 21 CFR 880.2800 |
| Regulatory Class: | Class II |
| Product Code: | FRC |
## Predicate Devices:
- 'Intended Use Predicate' for BCR-30-min BI™: 3M Attest™ 1292 Rapid Readout Biological Indicator for Steam, K090569.
- . 'Readout Predicate' for RBIP-ST30™: 3M Attest 490 Auto-reader, K103277.
## Description of Device:
# BCR-30-min ВГМ
The BCR-30-min BITM is designed to be used with its readout accessory, the RBIP-ST30™ robotic processor, for validating and challenging performance of gravity displacement steam sterilization cycles at 250°F (121°C).
The BCR-30-min BI™ consists of a strip (6.0 x 60 mm) of filter paper (Whatman GF/A) inoculated in the center with about 106 living spores of Geobacillus stearothermophilus ATCC No. 7953. The strip is packaged in a #30 blue glassine pouch (1 x 3 inch). The BCR-30-min BITM should be placed in the center of the load or in the most difficult to sterilize area.
{4}------------------------------------------------
#### RBIP-ST30™ Readout Accessory
In the RBIP-ST30™, the spores of the BCR-30-min BI™ are first converted into Fspores™ and then they are incubated for 30 min at 55±2°C to allow germination of living Fspores™. The device utilizes the fact that F-spores™ are not fluorescent per se but produce intense fluorescence when germinated. Accordingly, the RBIP-ST30™ detects failure of a steam sterilization cycle when significant fluorescence is present.
The RBIP-ST30™ is designed to automatically process the BCR-30-min BI™ to obtain a fluorescence reading within 30 minutes. The fluorescence results at 30 minutes meet the FDA's requirement of > 97% alignment with the result after the conventional incubation time of 7 days.
#### Indications for Use:
Use the BCR-30-min BITM and the RBIP-ST30™ together for monitoring performance of gravity displacement steam sterilization cycles of 30 minutes at 250°F (121°C). The BCR-30-min BI™ in conjunction with the RBIP-ST30TM provides final fluorescence results in 30 minutes.
# Substantial Equivalence of BCR-30-min BI™ to "Intended Use" Predicate Device 3M Attest™ 1292 Rapid Readout Biological Indicator (K090569):
Performance of the BCR-30-min BITM was determined using three (3) separate spore lots from independent spore cultures. A Summary of the Nonclinical Testing is shown in the table below.
| DEVICE CHARACTERISTICS | "INTENDED USE PREDICATE"<br>3M Attest™ 1292 BI, K090569 | NEW DEVICE<br>BCR-30-min BI™ |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use<br>• Method of sterilization | Gravity displacement steam | Gravity displacement steam |
| Indications for Use | Use the 3M Attest 1292 RRBI and the 3M Attest 490 Auto-reader together for monitoring:<br>1. Performance of gravity displacement steam sterilization cycles of 30 minutes at 250°F (121°C).<br>2. Performance of vacuum assisted steam sterilization cycles at 270°F (132°C).<br>The 3M Attest 1292 RRBI in conjunction with the 3M Attest 490 Auto-reader provide final fluorescence results in 180 minutes. | Use the BCR-30-min BI™ and the RBIP-ST30™ together for monitoring:<br>1. Performance of gravity displacement steam sterilization cycles of 30 minutes at 250°F (121°C).<br>The BCR-30-min BI™ in conjunction with the RBIP-ST30™ provide final fluorescence results in 30 minutes. |
| Organism<br>• Spores, Species, Strain | Geobacillus stearothermophilus from ATCC™ 7953 | Geobacillus stearothermophilus from ATCC™ 7953 |
| Viable spore population | About 106 spores | About 106 spores |
| Carrier Material | Polyethylene terephthalate paper | Glass fiber paper |
# Summary of a Comparative Validation of the BCR-30-min BITM
{5}------------------------------------------------
| Resistance Characteristics* | | |
|-------------------------------------------------------------------------------------|--------------------------------|-------------------------------|
| • D-Value | 1.5-2.2 min | 1.9 min |
| • Z-Value | 10-13 °C | 15.5 °C |
| • Survival Time | 7-10 min | 7.9 min |
| • Kill Time | 16-23 min | 19 min |
| Incubation Temperature | 60°C | 55±2 °C |
| Reduced Incubation Time<br>(>97% correlation to the 7-day<br>visual readout result) | Fluorescence result in 180 min | Fluorescence result in 30 min |
| Shelf-life | 2 years | Six months |
*Resistance values for the predicate were obtained from the3M Attest™ 1292 brochure.
Testing of the BCR-30-min BI™ summarized above was conducted following the FDA guidance and standards below:
- . FDA's Guidance for Industry and FDA Staff Biological Indicator (BI) Premarket Notification 510(k) Submissions; October 4, 2007.
- . ANSI/AAMI/ISO 11138-1:2006/(R) 2010 Sterilization of health care products -Biological indicators - Part 1: General Requirements.
- . ANSI/AAMI/ISO 11138-3:2006/(R) 2010 Sterilization of health care products -Biological indicators - Part 3: Biological indicators for moist heat sterilization processes
- United States Pharmacopeia, Chapter <1035> Biological Indicators for Sterilization . and Chapter <55> Biological Indicators - Resistance Performance Tests.
The results of these tests show that the BCR-30-min BI™, when used with the RBIP-ST30™, complies with ANSVAAMI/ISO 11138-1:2006/(R) 2010 and ANSI/AAMI/ISO 11138-3:2006/(R) 2010, the USP requirements for biological indicators, as well as the FDA's Guidance for Biological Indicators.
# Substantial Equivalence of RBIP-ST30™ to "Readout Predicate" Device 3M Attest 490 Auto-reader:
Since the RBIP-ST30™ is a required readout accessory of the BCR-30-min BT™, the performance of the RBIP-ST30™ was determined at each step during the validation of the BCR-30-min BI™. The table below shows the functionality and characteristics of the new device compared to the predicate.
| DEVICE CHARACTERISTICS | "READOUT PREDICATE"<br>3M Attest 490 Auto-reader<br>(K103277) | NEW DEVICE<br>RBIP-ST30TM |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | The 3M Attest 490 Auto-reader is<br>designed to incubate and<br>automatically read the 3M AttestTM<br>1492 BI for a final fluorescent result<br>at 30 min. | The RBIP-ST30TM readout<br>accessory is designed to<br>incubate and automatically read<br>the BCR-30-min BITM for a final<br>fluorescent result at 30 min. |
| Incubation Temperature | $56\pm2$ °C | $55\pm2$ °C |
{6}------------------------------------------------
| Basis of Rapid Readout | Fluorescence of BI medium | Fluorescence of F-spores™ |
|------------------------------------------------------|---------------------------------------------------|---------------------------------------------------------------|
| Method of Fluorescence<br>Detection | UV LED, optical filters, sensing by<br>photodiode | Blue LED, optical filters,<br>detection, sensing by CCD array |
| Indicator of Adequate<br>Sterilization Cycle | (-) on LCD Display. | Fluorescence below cut-off<br>value |
| Indicator of Possible<br>Sterilization Cycle Failure | (+) on LCD Display. Audible Alarm | Fluorescence above cut-off<br>value |
| Incubation wells | 10—reader/incubation wells | 10—cartridge wells |
| Voltage Range | 100-240 Volts AC | 110 Volts |
| Product Safety | UL/IEC 61010-1 | EN 61326-1 |
| EMC Compliance | FCC Part 15, Subpart B, Class A | FCC Part 15, Subpart B, Class A |
#### Statement of Substantial Equivalence:
The RBIP-ST30™ was tested for safety by a certified Testing Laboratory to demonstrate compliance to EN 61326-1. All tests were conducted using measurement procedure from CISPR 16 as appropriate.
Based on the same intended use and similar performance data, the BCR-30-min BI™ and the RBIP-ST30™ readout accessory have been demonstrated to be substantially equivalent to, and therefore, as safe and effective as the legally marketed devices: 3M Attest™ 1292, Rapid Readout Biological Indicator for Steam (K090569), and 3M Attest 490 Auto-reader (K103277), respectively.
## Conclusion
The BCR-30-min BI™ Biological Indicator and the RBIP-ST30™ readout accessory meet all applicable performance standards and are substantially equivalent to their predicate devices in terms of their intended use, physical properties and technological characteristics.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.