K151469 · Best Nomos · MUJ · Nov 25, 2015 · Radiology
Device Facts
Record ID
K151469
Device Name
CORVUS Radiation Therapy Planning System
Applicant
Best Nomos
Product Code
MUJ · Radiology
Decision Date
Nov 25, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
CORVUS is intended for use as a planning tool for conformal radiation therapy. Using operator-supplied input and patient scans, it creates a plan for treatment delivery systems and generates a set of beam weights that, when applied to a compatible system, facilitates delivery of an intensity-modulated 3D conformal radiation therapy treatment. CORVUS is intended only to suggest a delivery plan. It is the physician's responsibility to verify that the dose distributions which would result from plan implementation are appropriate for a particular patient. The CORVUS system is intended to be used as an integrated system with a modulating device for planning and delivery of conformal radiation therapy. The modulating device can be the NOMOS MIMIC, nomosSTAT MLC, or a supported MLC. CORVUS produces radiation fields which are modulated to conform to the projected tumor volume plus margins. The system tries to achieve target goals while sparing sensitive structures.
Device Story
CORVUS is a semi-automatic radiation treatment planning system; inputs include patient scans and treatment machine-specific data. Clinicians (physicists, dosimetrists, oncologists) use the system to generate conformal treatment plans for external beam photon therapy. The system uses optimization methods (simulated annealing, gradient descent) to calculate volumetric dose distributions and beam weights for intensity-modulated 3D conformal therapy. It supports integration with modulating devices like NOMOS MIMIC or nomosSTAT MLC. Output consists of a suggested delivery plan, which the physician reviews and approves. The system helps clinicians achieve target dose goals while sparing sensitive structures. It is used in clinical settings to guide radiation delivery, potentially improving treatment precision and patient outcomes.
Clinical Evidence
Bench testing only. Validation included dosimetric accuracy (film and ion chamber measurements), clinical plan quality/optimization comparisons, and system integration tests. Results showed >95% pass rate for film comparisons (4%/4mm criteria) and chamber readings within 1.5%. Cobalt-60 plans were clinically acceptable and comparable to Linac-based plans. Verification confirmed dose calculation consistency, decay calculations, and R&V system data transfer accuracy.
Technological Characteristics
Radiation therapy planning system; uses finite size pencil beam (FSPB) algorithm. Supports homogeneous and Effective Path Length (EPL) dose calculation options. Connectivity via DICOM RT and proprietary formats. Software-based optimization using simulated annealing and gradient descent. Operates on Intel-based Mac Pro hardware. Conforms to IEC 62304, BS EN 62366, and IEC 62083.
Indications for Use
Indicated for medical physicists, dosimetrists, and radiation oncologists to create conformal treatment plans using photon (x-ray, gamma ray) external beam radiation therapy. Valid only for external beam photon therapy; calculations for electrons and intracavity sources (Brachytherapy) are not supported.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 25, 2015
Best NOMOS % Vineet Gupta, Ph.D. Director of R&D One Best Drive PITTSBURGH PA 15202
Re: K151469
Trade/Device Name: CORVUS Radiation Therapy Planning System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ Dated: November 9, 2015 Received: November 10, 2015
Dear Dr. Gupta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Oolo
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use Statement Section 1 INDICATIONS FOR USE
510(k) Number (if known): _K151469
Device Name: CORVUS
Indications for Use:
The CORVUS system is a radiation treatment planning package designed to allow medical physicists, dosimetrists, and radiation oncologists to create conformal treatment plans using photon (x-ray, gamma ray) external beam radiation therapy. The treatment plans generated by CORVUS are based upon treatment machine-specific data and are intended to provide a guide to delivering external beam radiation therapy which conforms to the target volume defined by the radiation oncologist.
The CORVUS system is valid for use only with external beam photon therapy; calculations for electrons and intracavity sources (Brachytherapy) are NOT supported.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
### Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# Traditional 510(k) Summary
### A. SUBMITTERS NAME
Best NOMOS
### B. ADDRESS
One Best Drive Pittsburgh, PA 15202
### C. CONTACT
Vineet Gupta, Ph.D. Phone: (412) 312-6819 Fax: (412) 312-6701
### D. DATE PREPARED
Date: 29-May-2015
### E. DEVICE NAME
Device Trade Name: CORVUS Common/Classification Name: Radiation Therapy Treatment Planning System
### F. ESTABLISHMENT REGISTRATION NUMBER
Corporate Office Registration Number: 2434141 Manufacturing and Packaging Registration Number: 2434141
### G. DEVICE CLASS
Class II Panel: Radiology Product Code: 90-MUJ Regulation Number: 21CFR 892.5050
#### H. LEVEL OF CONCERN
#### Major Level of Concern
{4}------------------------------------------------
We believe the level of concern is Major for CORVUS 2011. CORVUS 2011 is an accessory to a medical device that has a Major Level of Concern such as delivery systems which use modulated radiation therapy for delivery and planning.
The copy of the complete decision making process for this conclusion is included in Section 13.1.
# I. STATEMENT OF INDICATIONS FOR USE
The intended use for the CORVUS Radiation Therapy Planning System has NOT changed as a result of this modification. The indications for use for the CORVUS Radiation Therapy Planning System has been modified to include usage of "gamma rays" (another form of photons) for creating conformal treatment plans. This modification does not change the fundamental principle of using photons for creating conformal treatment plans which is the same as that of its predicate device.
### Intended Use
CORVUS is intended for use as a planning tool for conformal radiation therapy. Using operator-supplied input and patient scans, it creates a plan for treatment delivery systems and generates a set of beam weights that, when applied to a compatible system, facilitates delivery of an intensity-modulated 3D conformal radiation therapy treatment. CORVUS is intended only to suggest a delivery plan. It is the physician's responsibility to verify that the dose distributions which would result from plan implementation are appropriate for a particular patient.
The CORVUS system is intended to be used as an integrated system with a modulating device for planning and delivery of conformal radiation therapy. The modulating device can be the NOMOS MIMIC, nomosSTAT MLC, or a supported MLC. CORVUS produces radiation fields which are modulated to conform to the projected tumor volume plus margins. The system tries to achieve target goals while sparing sensitive structures.
### Indications for Use
The CORVUS system is a radiation treatment planning package designed to allow medical physicists, dosimetrists, and radiation oncologists to create conformal treatment plans using photon (x-ray, gamma ray) external beam radiation therapy. The treatment plans generated by CORVUS are based upon treatment machine-specific data and are intended to provide a guide to delivering external beam radiation therapy which conforms to the target volume defined by the radiation oncologist.
The CORVUS system is valid for use only with external beam photon therapy; calculations for electrons and intracavity sources (Brachytherapy) are NOT supported.
{5}------------------------------------------------
### J. DEVICE DESCRIPTION
CORVUS is a semi-automatic planning system: rather than simply verifying a user-designed plan, the system itself suggests a plan. A clinician then reviews and approves the plan.
CORVUS is designed to generate plans for treatment delivery systems that can create multiple radiation patterns composed of pencil beams on which the intensity can be individually controlled. The treatment beams are weighted so that when they are projected into the treatment space they superimpose to give the desired dose distribution.
Each radiation field is generated using one of several optimization methods provided with the system, including simulated annealing and gradient descent.
The treatment beams are set not only to deliver the prescribed dose to the identified target volume, but also to keep the dose to other sensitive volumes below user-defined limits. Planning is done volumetrically: the beam weights for treating the entire target volume are generated simultaneously. The dose matrix is volumetric. The dose to each point is calculated to be that received from all beams and from all gantry angles. Dosage is calculated using a finite size pencil beam (FSPB) algorithm based on the beam characterization of clinically measured data. The degree to which a treatment plan is optimized is determined in part by constraints placed on the planning algorithm. The user has direct control over these constraints, which include dose goals to the target structures, dose limits to the sensitive structures, and the specification of arcs or fixed gantry positions in the treatment plan.
CORVUS treatment plans need not have the isocenter located within the target volume. An unlimited number of targets falling within the treatment volume can be planned for at the same time. Dose may be prescribed for up to 32 structures, 29 of them user-selectable, any number of which may be separate targets or radiation-sensitive structures. Each structure can have a separate dose prescription.
### K. PREDICATE DEVICE INFORMATION
The CORVUS 2011 system is substantially equivalent to its primary predicate device CORVUS 09 (K100092). The ViewRay Treatment Planning and Delivery System (K102915) for treatment planning is used only as a reference predicate device. The CORVUS 09 system was determined to be substantially equivalent to its predicate device as of February 2010.
The fundamental scientific technology for the CORVUS 2011 and CORVUS 09 systems has not changed. The intended use of the device has not changed. Based upon the performance testing results for CORVUS 2011 (as detailed in the submission), the system raises no new issues of safety or effectiveness.
{6}------------------------------------------------
### L. COMPARISON TO THE PREDICATE DEVICE
This section describes the incremental changes from CORVUS 09 to CORVUS 2011 systems and provides inputs on its equivalence to its predicate devices.
#### Hardware and Operating System
| CORVUS 09 (predicate) | | | | | |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------|--|--|--|--|
| Operating System: | MacOS X 10.5.4 | | | | |
| Workstation: | Intel-based Mac Pro hardware system (based on two 3.2GHz<br>Quad-Core Intel Xeon (8-core) CPUs). | | | | |
| WACOM Tablet: | WACOM CINTIQ 21UX Interactive Pen Display, Graphics<br>Tablet | | | | |
| Physician's Review<br>Workstation (Optional): | Intel-based iMac hardware system (based on 2.03 Ghz Intel<br>core 2 Duo CPU) | | | | |
| CORVUS 2011 | | | | | |
| Operating System: | MacOS X 10.6.7 | | | | |
| Workstation: | Intel-based Mac Pro hardware system (based on two<br>2.66GHz 6-Core Intel Xeon (12-core) CPUs | | | | |
| WACOM Tablet: | WACOM CINTIQ 21UX Interactive Pen Display, Graphics<br>Tablet | | | | |
| OR | | | | | |
| Operating System: | MacOS X 10.6.2 | | | | |
| Workstation: | Intel-based Mac Pro hardware system (upgrade of 8-core<br>Mac) | | | | |
| WACOM Tablet: | WACOM CINTIQ 21UX Interactive Pen Display, Graphics<br>Tablet | | | | |
CORVUS 2011 contains all of the features of CORVUS 09 (K100092) and adds the feature of supporting Cobalt-60 based external beam radiation treatment planning (Gamma Tomotherapy). Similar to Linac based planning, CORVUS 2011 allows users to create treatment plans for Cobalt-60 based systems. The Cobalt-60 based dose calculation, optimization, and sequencing algorithms present in CORVUS 2011 are very similar to those used in CORVUS 09 for Linacs. In addition to supporting CORVUS 09 Linac based treatment
{7}------------------------------------------------
plans and systems. CORVUS 2011 supports the following for Cobalt-60 based treatment planning solution:
- Theratron Equinox 100 Gamma Teletherapy System (Best Theratronics) (K053485) a.
- Avanza Couch/Patient Positioning Table (Best Theratronics) (K060870) b.
- C. NomosSTAT (Best Nomos) (K060895)
- Export of Cobalt-60 plans via nomosSTAT proprietary format and DICOM RT d.
In both the products (CORVUS 2011 (Linac and Cobalt-60) and CORVUS 09 (Linac)), the pencil-beam algorithm is used and dose calculation can be performed using either a homogeneous option or an Effective Path Length (EPL) option. The dose calculation based on the Lateral Disequilibrium Inclusive option (for Linacs in CORVUS 2011 and CORVUS 09) is not supported for the Cobalt-60 based planning option in CORVUS 2011. The optimization algorithms used in both cases are the same. The workflow in CORVUS 2011 for Gamma Tomotherapy when compared to its predicate device CORVUS 09 is unchanged.
To account for the larger, cylindrical source of the Cobalt-60 system and for radioactive source decay, the following modifications were made in CORVUS 2011 for Gamma Tomotherapy:
- i. During commissioning the calibration dose rate, the calibration date and the treatment start date for each patient needs to be specified. During commissioning the system performs a sanity check of the measured penumbra vs the calculated penumbra based on source size.
- ii. The pencil beam algorithm has been updated to account for source size in addition to accounting for radioactive source decay. The changes include modification to:
- a. Primary Geometric Kernel to reflect large cylindrical cobalt source. The Radiological Kernel did not require any modification.
- b. Account for large source when calculating scatter dose for small field sizes.
- c. Account for leaf curvature interplay with the large source (fluence correction for the large source in sequencing algorithm).
For the gamma ray (Cobalt-60) based treatment planning for external beam therapy the CORVUS 2011 system shares many of the technological features and characteristics as that of the predicate device ViewRay Treatment Planning and Delivery System (K102915) for treatment planning. The target population and intended use are similar to that of the predicate devices. In addition, the fundamental technical characteristics are the same as those of the predicate devices and any minor differences in features do not raise any concerns for safety, performance, or effectiveness of the device. The characteristics/features of CORVUS 2011 with respect to each of the predicate devices is described in the comparison chart in the submission. The similarities and differences are discussed in detail as part of this submission.
# M. SUMMARY OF PERFORMANCE TESTING
Activities to ensure that the modifications do not affect safety or effectiveness are provides in CORVUS 2011 Total Validation Report and CORVUS 2011 Verification Summary Report. The total validation activities included clinical workflow, treatment planning software
{8}------------------------------------------------
usability, clinical plan quality / optimization comparison, and dosimetric accuracy reports. The verification activities included system tests, anomaly verification, code review, linac dose equivalence report, and run-though integration tests.
Design verification and validation testing was performed to ensure that the device functionality for treatment planning works as per its intended use, is substantially equivalent with respect to its predicate device, all risks are mitigated, and the product conforms to the required standards. The details of the design verification and validation activities are explained in the submission (Section 15). The accuracy of treatment plans was evaluated through comparison with medical physics measurements including film and ion chamber measurements. The quality of treatment plans was evaluated based upon clinical criteria associated with conformal therapy.
CORVUS 2011, for Cobalt treatment planning, has been tested on and supports the following:
- Theratron Equinox 100 Gamma Teletherapy System (Best Theratronics) (K053485) a.
- b. Avanza Couch/Patient Positioning Table (Best Theratronics) (K060870)
- C. NomosSTAT (Best Nomos) (K060895)
- i. Plan delivery to nomosSTAT is via NOMOS proprietary format: MIMiC Delivery File Set Specification.
- Export to R&V systems via DICOM RT defined by DICOM standard (tested on d. Mosaig version 2.3)
- i. Corvus 2011 for Cobalt-60 planning uses the convention 1min = 100 MU in the treatment plan to export in DICOM RT format. This keeps the structure of the data the same as that used for exporting LINAC plans as defined by DICOM standard. However, for nomosSTAT plans, the DICOM RT plans do not contain the leaf positions as the MIMiC is a binary leaf MLC.
Summary of the tests performed, in addition to system testing, module testing, and clinical workflow testing, to ensure accuracy of treatment plans and clinical efficacy of CORVUS 2011 system include:
| Sr. No. | Test | Test Details | Results |
|---------|----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | Linac Dose Equivalence Test | To test Linac treatment | The results showed that the |
| | | plans for the same patients | Linac treatment plans |
| | | between CORVUS 09 and | recomputed (resegmented and |
| | | CORVUS 2011 | redosed) with CORVUS 2011 |
| | | | had equivalent dose and |
| | | | sequences to those created on |
| | | | CORVUS 09. |
| 2. | Clinical comparison of CORVUS | To test the clinical efficacy | The results showed that the |
| | 09 treatment plans with Cobalt | of treatment plans created | treatment plans created for |
| | treatment plans on CORVUS | for Cobalt treatment on | Cobalt based treatment were |
| | 201 1 | CORVUS 2011 with respect | clinically acceptable and were |
| | | to treatment plans created | comparable to Linac based |
| | | for Linac treatment on | treatment plans on CORVUS 09. |
| | | CORVUS 09. | |
| 3 | Dosimetric Accuracy | To determine the dosimetric accuracy of the plans delivered to phantoms and measured using EDR2 film and 0.125 cc chamber | >95% of points pass relative comparison for all film.<br>Chamber reading within 1.5% (pass/fail criteria of 4% / 4mm). |
| 4 | Comparison of calculated versus measured values | To determine if Cobalt based plans generate similar dose values as those measured using EDR2 films and CC01 ionization chamber | Multiple test results of dosimetric parameters (such as percentage dose depth and output factors for various pencil beam sizes) showed that the Cobalt plan based calculated values are similar to those measured using film and ion chambers. Output factors and percentage depth dose curves agreed within <4% (pass/fail criteria of 4% / 4mm). |
| 5 | Dose Comparisons for CORVUS 2011 using Gamma Tomotherapy | • Check for dose calculation consistency when there is a large growth region (margin) distinguishing PTV and CTV<br>• Compare EPL (Effective Path Length Heterogeneity correction) and homogeneous calculation in a water equivalent phantom<br>• Verify consistency of hybrid plans and patient plans<br>• Check the consistency of single fraction hybrid plans and multi-fraction patient plans<br>• Check the dose scaling feature<br>• Check the decay calculations | The results indicated that:<br>• The growth margin does not affect the dose distribution. Profiles of the hybrid plan and the patient plan matched exactly for EPL and homogeneous calculation.<br>• In a water equivalent phantom, EPL and homogeneous calculation matched as expected.<br>• The statistics, treatment plan, and dose profiles of the hybrid plan match the original patient plan when the original plan is applied to its own geometry (for both EPL & homogeneous calculation options).<br>• The dose to the chamber per fraction in the original multi-fraction plan and the hybrid calculation agree as expected.<br>• Scaling the dose in Display Results using the "Total delivered dose" and "Percentage of Max Delivering XX Gy" boxed |
| | | | changed the treatment time<br>and mean chamber dose<br>according as per expectations.<br>• The decay calculation agreed<br>as expected. |
| 6 | Transfer to R&V System | To test transfer to Mosaiq<br>record and verify system via<br>Dicom RT | All fields including Energy,<br>Field Size, Gantry Start Angle,<br>Gantry Stop Angle, Collimator<br>Angle, Couch Angle, and MU of<br>each arc were verified to be<br>correct. |
| 7 | Comparison of Linac plan<br>Optimization | To compare the<br>optimization of Linac plans<br>in CORVUS 2011 with<br>respect to CORVUS 09 | Comparative analysis of the<br>optimized plans (statistics,<br>isodose distributions and dose<br>volume histograms) indicated<br>that these plans were clinically<br>similar. |
| 8 | Final build – validating dose<br>accuracy | To re-assess the dosimetric<br>accuracy of CORVUS<br>2011for all cobalt validation<br>plans in the final build using<br>EDR2 film | >98% of the points pass for all<br>the plans (pass fail criteria of<br>4%/4mm) |
{9}------------------------------------------------
{10}------------------------------------------------
Detailed results of these tests are included as part of this submission. The verification and validation results demonstrate that the CORVUS 2011 system met its design requirements and specifications, is substantially equivalent to its predicate device, is safe and effective to use for its intended purpose, and conforms to the applicable sections of standards that includes IEC 63204:2006: Medical device software - Software life cycle processes, BS EN 62366:2008: Medical devices -Application of usability engineering to medical devices, and IEC 62083:2009: Requirements for the safety of radiotherapy treatment planning systems.
Refer to Section 6 for the Declaration of Conformity.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.