K151398 · Terumo Europe N.V. · FMI · Jun 25, 2015 · General Hospital
Device Facts
Record ID
K151398
Device Name
K-Pack II Needle - 29G x 5/16 Thin Wall
Applicant
Terumo Europe N.V.
Product Code
FMI · General Hospital
Decision Date
Jun 25, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 29G x 5/16" Thin Wall K-Pack II Needles being Hypodermic Single Lumen Needles are sterile medical devices for single use, intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin.
Device Story
K-Pack II Needle is a sterile, single-use hypodermic single lumen needle; consists of stainless steel cannula, polypropylene hub, epoxy glue, and silicone lubricant. Operates manually by connecting to a piston syringe via female luer connector. Used by clinicians to inject or withdraw fluids below skin surface. Device extension of existing K-Pack II line; provides shorter 5/16" length option. Benefits patient by enabling fluid delivery/aspiration via established hypodermic technique.
Clinical Evidence
No clinical data. Bench testing only. Verification activities included visual appearance, acidity/alkalinity, extractable metals, needle penetration resistance, conical fitting (ISO 594-1), fitting strength, effective length, lubrication, breakage resistance (EN ISO 9626), stiffness (EN ISO 9626), corrosion resistance, bonding strength, liquid/air leakage, flow rate, torque force, endotoxin (EP 2.6.14), toxicity (EP 2.6.9), and sterility (EP 2.6.1).
Technological Characteristics
Stainless steel cannula; polypropylene hub; epoxy glue; silicone lubricant. Dimensions: 29G x 5/16" thin wall. Manual operation. Luer connector per ISO 594-1/2. Sterilization: Ethylene oxide (EtO) to SAL 10^-6 per EN ISO 11135-1. Externally communicating, blood-contacting, limited exposure (≤ 24 hrs).
Indications for Use
Indicated for injection of fluids into or withdrawal of fluids from parts of the body below the skin surface. For single use.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the central image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 25, 2015
Terumo Europe N.V. Mrs. M. J. Aerts Regulatory Affairs Manager Interleuvenlaan 40 3001 Leuven BELGIUM
Re: K151398
Trade/Device Name: K- Pack II Needles- 29 G x 5/16" Thin Wall Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: May 21, 2015 Received: May 26, 2015
Dear Mrs. Aerts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mrs. M.J. Aerts
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indication for Use
510(k) Number (if known): K151398
Device Name: K-Pack II Needle - 29G x 5/16" Thin Wall
Indication For Use:
The 29G x 5/16" Thin Wall K-Pack II Needle being a Hypodermic Single Lumen Needle is a sterile medical device for single use, intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin.
Prescription Use X (21 CFR Part 801 Subpart D) And/Or
Over the Counter Use ______ (21 CFR Part 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
#### 510(k) SUMMARY as required by 807.92
#### Submitter information
| Prepared for: | TERUMO EUROPE N.V.<br>Interleuvenlaan 40,<br>3001 Leuven,<br>BELGIUM |
|-----------------------------|-------------------------------------------------------------------------------------------------|
| Prepared by/Contact person: | Mrs. M.J. Aerts – Manager Regulatory Affairs<br>Tel. (+32) 16 38 13 53<br>Fax (+32) 16 40 02 49 |
| Date prepared: | May 2015 |
#### II.1. Device Name
#### Proprietary Name
K-Pack II Needle - 29G x 5/16" Thin Wall
#### Classification Name
Hypodermic Single Lumen Needle
21CFR, Section 880.5570
Classification: Class II
#### Product Code
FMI
#### II.2. Predicate Devices
The following cleared devices are selected as predicate devices:
- 1. K-Pack II Needles (K984576)
- 2. 29G x ½'' Thin Wall K-Pack II Needles (K082820)
#### II.3. Reason for Submission
This 510k is being submitted to extend the cleared K-Pack II Needle (K984576) product line with the K-Pack II Needle - 29G x 5/16" Thin Wall.
The 29G Thin Wall K-Pack II Needle is the same needle as the cleared 29G Thin Wall Needle covered in K082820, the only difference is that the needle length of the 29G x 5/16" Thin Wall K-Pack II Needle is shorter than what is currently cleared under the 29G x ½2" Thin Wall Needle K082820.
The packaging of the 29G x 5/16" Thin Wall K-Pack II Needle in a short case is the same as the packaging of the cleared 29G x ½" Thin Wall K-Pack II Needle covered in K082820.
{4}------------------------------------------------
Although there are no potential issues of safety and effectiveness for a shorter length needle, this Special 510k is being submitted due to the fact that the new length is out of the range of the previously cleared needles.
This 510k will provide supporting information that the 29G x 5/16" Thin Wall K-Pack II Needles are an acceptable extension of the current K-Pack II Needle product line.
#### II.4. Intended Use
The 29G x 5/16" Thin Wall K-Pack II Needles being Hypodermic Single Lumen Needles are sterile medical devices for single use, intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin.
Note: This is the same intended use as the predicate devices, K-Pack II Needle - K984576 and 29G x ½" Thin Wall K-Pack II Needle K082820.
#### II.5. Substantial Equivalence
The 29G x 5/16" Thin Wall K-Pack II Needles are substantially equivalent in intended use, design, technology/principles of operation, materials, and performance to the following cleared devices:
- 1. K-Pack II Needles (K984576)
- 29G x 1/2" Thin Wall K-Pack II Needles (K082820) 2.
Differences between the devices do not raise any significant issues of safety and effectiveness.
| | 29G x 5/16" Thin Wall<br>K-Pack II Needle<br>(Terumo Europe, Belgium)<br>(Subject of this 510k) | 29G x 1/2" Thin Wall K-Pack II<br>Needles<br>(Terumo Europe, Belgium)<br>(K082820) | K-Pack II Needles<br>(Terumo Europe, Belgium)<br>(K984576) |
|--------------------------------|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Same as predicate | Intended to inject fluids into, or<br>withdraw fluids from, parts of<br>the body below the surface of the<br>skin. | Intended to inject fluids into, or<br>withdraw fluids from, parts of<br>the body below the surface of the<br>skin. |
| Materials | Same as predicate | Same as predicate | Same as predicate |
| Cannula | Stainless Steel | Stainless Steel | Stainless Steel |
| Hub | Polypropylene | Polypropylene | Polypropylene |
| Glue | Epoxy glue | Epoxy glue | Epoxy glue |
| Lubricant | Silicone | Silicone | Silicone |
| Description/<br>Specifications | Same as predicate | Comprised of a stainless steel<br>tube that is sharpened at one end<br>and at the other end joined to a<br>female luer connector (hub)<br>designed to be connected with a<br>male connector (nozzle) of a<br>piston syringe. | Comprised of a stainless steel<br>tube that is sharpened at one end<br>and at the other end joined to a<br>female luer connector (hub)<br>designed to be connected with a<br>male connector (nozzle) of a<br>piston syringe. |
| Needle Gauge/ length | 29G x 5/16" | 29G x 1/2" | Ranging from 18G – 27G<br>3/8" - 1 1/2" |
| Principle of Operation | Same as predicate | Manually | Manually |
The similarities and differences are summarized in the table below.
{5}------------------------------------------------
| Unit packaging | Hard pack consisting of cap and<br>short case | Hard pack consisting of cap and<br>short case | Hard pack consisting of cap and<br>long or short case |
|----------------|-----------------------------------------------|-----------------------------------------------|-------------------------------------------------------|
| Wall Thickness | Thin Wall | Thin wall | Ultra thin + thin + regular wall |
| Sterilization | EtO to SAL 10-6 | EtO to SAL 10-6 | EtO to SAL 10-6 |
| Shelf life | 5 years | 5 years | 5 years |
## II.6. Summary of Verification Activities
Design verification activities were conducted for the 29G x 5/16" Thin Wall K-Pack II Needles to demonstrate substantial equivalence to the predicate devices. This is for confirmation purpose only.
Summary of the verification activities including acceptance criteria is given in the table below:
| TEST | ACCEPTANCE CRITERIA |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. Visual appearance | Silicon amount on cannula not visible as droplets. |
| 2. Limits for acidity or alkalinity | $\Delta$ pH for K-Pack Needles extract solution is within 1 unit of the control fluid. |
| 3. Limits for extractable metals | The extract solution of the 29G K-Pack II Needles has a content of extractable metals<br>which is, when corrected for the metal content of the control fluid:<br>$\Sigma Pb, Sn, Zn, Fe \le 5 mg/l$<br>$Cd < 0.1 mg/l$ |
| 4. Needle Penetration Resistance | The penetration resistance of cannula point and drag complies with the limits specified<br>as follow: $Point value \le 0.10 N$ and $Drag Value \le 0.04 N$ |
| 5. Conical fitting | The taper of the hub meets the gauging requirements of ISO 594-1. |
| 6. Fitting strength hub/case | The force to pull axially the needle hub from the case is minimum 0.5 N and maximum<br>25 N. |
| 7. Effective needle length | The effective length = nominal length + 1 mm/-2 mm. |
| 8. Silicon amount on cannula | Needles are uniformly lubricated and the silicone is not visible as droplets on the<br>outside surface of the needle, the quantity will not exceed 0.25 mg/cm2. |
| 9. Resistance to breakage | The cannula does not break when tested in accordance with EN ISO 9626. |
| 10. Stiffness test cannula | The cannula shall show a deflection not greater than 0.37 mm in accordance with EN<br>ISO 9626. |
| 11. Corrosion Resistance<br>(cannula) | The cannula shall show no signs of corrosion caused by the immersion in a sodium<br>chloride solution 0.5 M. |
| 12.Bonding strength between hub<br>and cannula | The bonding strength between hub and cannula is $\ge$ 22N. |
| 13. Liquid leakage | There is no leak of liquid at the luer (lock) fitting when the needle hub is fixed to<br>reference control fitting in accordance with to ISO 594-1 and ISO 594-2. |
| 14. Air leakage | There is no air leakage at luer (lock) fitting when the needle hub is fixed to a reference<br>control fitting in accordance with to ISO 594-1 and ISO 594-2. |
{6}------------------------------------------------
| 15. Flow rate | Tolerance on flow rate: between 80% and 125% of nominal value. |
|---------------------------|---------------------------------------------------------------------------------------------------------------|
| 16. Torque Force Cap-Case | The torque force required to break the label and to remove the cap from the case is not<br>more than 25 N.cm. |
| 17. Endotoxin reaction | The device is free from endotoxin reaction in accordance with European Pharm. 2.6.14. |
| 18. Abnormal toxicity | The device is free from abnormal toxicity in accordance with European Pharm. 2.6.9. |
| 19. Sterility | The device is sterile in accordance with European Pharm. 2.6.1. |
### II.7. Additional Information
The sterility of the 29G x 5/16" Thin Wall K-Pack II Needles is assured by using a validated sterilization method qualified in accordance with EN ISO 11135-1: "Sterilization of Health Care Products - Ethylene oxide - Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices", to a sterility assurance level (SAL) of 10° as required by EN 556-1: "Sterilization of Medical Devices - Requirements for medical devices to be designated "STERILE" - Part - 1: Requirements for terminally sterilized medical devices".
Ethylene oxide residual levels resulting from EtO sterilization are in compliance with EN ISO 10993-7: "Biological evaluation of medical devices - Part 7: Ethylene oxide sterilization residuals".
The 29G x 5/16" Thin Wall K-Pack II Needle, like the standard K-Pack II Needle (K984576), is an Externally Communicating device, Contacting Circulating Blood, Limited Exposure (≤ 24 hrs). The device's blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993-1, "Biological Evaluation of Medical Devices Part-1: Evaluation and testing".
The expiration dating for the 29G x 5/16" Thin Wall K-Pack II Needles has been established at 5 years which is the same as the cleared K-Pack II Needles.
#### II.8. Conclusion
In summary, the 29G x 5/16" Thin Wall K-Pack II Needles are substantially equivalent in intended use, design, technology/principal of operation, materials, and performance to the following cleared devices:
- 1. K-Pack II Needles (K984576)
- 29G x 1/2" Thin Wall K-Pack II Needles (K082820) 2.
Differences between the devices do not raise any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.