K151380 · Siemens Medical Solutions USA, Inc. · MUJ · Aug 13, 2015 · Radiology
Device Facts
Record ID
K151380
Device Name
syngo.via RT Image Suite
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
MUJ · Radiology
Decision Date
Aug 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
syngo.via RT Image Suite is a 3D and 4D image visualization, multimodality manipulation and contouring tool that helps the preparation and response assessment of treatments such as, but not limited to those performed with radiation (for example, Brachytherapy, Particle Therapy, External Beam Radiation Therapy). It provides tools to efficiently view existing contours, create, edit, modify, copy contours of the body, such as but not limited to, skin outline, targets and organs-at-risk. Contours and images can subsequently be exported to a Treatment Planning System. The software combines following digital image processing and visualization tools: · Multi-modality viewing and contounical, functional, and multi-parametric images such as but not limited to CT, PET, PET/CT, MRI, and Linac Cone Beam CT (CBCT) images · Multiplanar reconstruction (MPR) thingthick, minimum intensity projection (MIP), volume rendering technique (VRT) · Freehand and semi-automatic contouring of regions-of-interest on any orientation including oblique · Creation of contours on any type of mages without prior assignment of a planning CT · Manual and semi-automatic registration using rigid and deformable registration · Supports the user in comparing, contouring, and adapting contours based on datasets acquired with different imaging modalities and at different time points Supports the user in comparing images and contours of different patients · Supports multi-modality image fusion · Visualization and contouring of moving tumors and organs
Device Story
syngo.via RT Image Suite is a post-processing software application for radiation therapy planning. It ingests multi-modality medical images (CT, PET, PET/CT, MRI, CBCT). The software provides tools for 3D/4D visualization, multiplanar reconstruction (MPR), maximum/minimum intensity projection (MIP), and volume rendering (VRT). It enables manual and semi-automatic contouring of targets and organs-at-risk, as well as rigid and deformable image registration. Used in clinical settings by medical professionals, the software assists in comparing datasets across modalities and time points. Outputs include contoured images and structure sets exported to a Treatment Planning System. The device supports clinical decision-making by facilitating accurate target definition and treatment response assessment, potentially improving radiation therapy precision.
Clinical Evidence
No clinical data. Bench testing only. Verification and validation testing, including integration and functional testing, confirmed that software specifications met acceptance criteria. Risk management and hazard analysis were performed to mitigate identified risks.
Technological Characteristics
Software-based image processing application; operates on syngo.via platform (client-server architecture). Features: multi-modality viewing, MPR, MIP, VRT, rigid/deformable registration, and 3D contouring. Standards: DICOM (PS 3.1-3.18), IEC 62304 (software lifecycle), ISO 14971 (risk management), IEC 60601-1-6 (usability), IEC 60601-1-4 (programmable electrical medical systems). Cybersecurity controls implemented for access/data protection.
Indications for Use
Indicated for use by medical professionals as a 3D/4D image visualization, manipulation, and contouring tool for radiation therapy treatment preparation and response assessment. Applicable to patients requiring radiation therapy (e.g., Brachytherapy, Particle Therapy, External Beam Radiation Therapy). No specific age or gender contraindications stated.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines extending from the bottom of the profiles.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 13, 2015
Siemens Medical Solutions USA, Inc. % Ms. Eve Davis Regulatory Affairs Specialist 51 Valley Stream Parkway MALVERN PA 19355
Re: K151380
Trade/Device Name: syngo.via RT Image Suite Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: MUJ and LLZ Dated: May 21, 2015 Received: May 22, 2015
Dear Ms. Davis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Michael D'Hara
For
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a light teal. The letters are evenly spaced and fill the majority of the frame. The background is plain and white.
DEPARTMENT OF HEALTH AND HUMAN SERVICES Form Approved: OMB No. 0910-0120 Food and Drug Administration Expiration Date: January 31, 2017 See PRA Statement below. Indications for Use 510(k) Number (if known) 51380 levice Name syngo.via RT Image Suite Indications for Use (Describe) syngo via RT Image Suite is a 3D and 4D image visualization, multimodality manipulation and contouring tool that helps the preparation and response assessment of treatments such as, but not limited to those performed with radiation (for example, Brachytherapy, Particle Therapy, External Beam Radiation Therapy). It provides tools to efficiently view existing contours, create, edit, modify, copy contours of the body, such as but not limited to, skin outline, targets and organs-at-risk. Contours and images can subsequently be exported to a Treatment Planning System. The software combines following digital image processing and visualization tools: · Multi-modality viewing and contounical, functional, and multi-parametric images such as but not limited to CT, PET, PET/CT, MRI, and Linac Cone Beam CT (CBCT) images · Multiplanar reconstruction (MPR) thingthick, minimum intensity projection (MIP), volume rendering technique (VRT) · Freehand and semi-automatic contouring of regions-of-interest on any orientation including oblique · Creation of contours on any type of mages without prior assignment of a planning CT · Manual and semi-automatic registration using rigid and deformable registration · Supports the user in comparing, contouring, and adapting contours based on datasets acquired with different imaging modalities and at different time points Supports the user in comparing images and contours of different patients · Supports multi-modality image fusion · Visualization and contouring of moving tumors and organs
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.'
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Poblishing Services (101) 443-5740 EP
{3}------------------------------------------------
# 510(K) SUMMARY FOR syngo.via RT Image Suite
Submitted by: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355
Date Prepared: May 21, 2015
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
### 1. General Information
Importer/Distributor: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355
## Establishment Registration Number: 2240869
# Manufacturing Site:
Siemens AG Medical Solutions Siemens Str. 1 D-91301 Forchheim, Germany
## Establishment Registration Number: 3004977335
### 2. Contact Person:
Eve Davis Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway, Mail Code D02 Malvern, PA 19355 Phone: (610) 219-7133 Fax: (610) 448-1787 Email: eve.davis@siemens.com
{4}------------------------------------------------
| 3. | Device Name and Classification | | | |
|----|----------------------------------------------|-----------------------------------------------|--|--|
| | Product Name: | syngo.via RT Image Suite | | |
| | Propriety Trade Name: | syngo.via RT Image Suite | | |
| | Classification Name: | system, planning, radiation therapy treatment | | |
| | Classification Panel: | Radiology | | |
| | CFR Section: | 21 CFR §892.5050 | | |
| | Device Class: | Class II | | |
| | Product Code: | MUJ | | |
| | Additional Product Code: | LLZ | | |
| | Legally Marketed Primary Predicate Device: | | | |
| | Trade Name: | syngo® Dosimetrist Workspace v2.7 | | |
| | 510(k)#: | K101119 | | |
| | Clearance Date: | June 16, 2010 | | |
| | Classification Name: | System, Planning, Radiation Therapy Treatment | | |
| | Classification Panel: | Radiology | | |
| | Classification Regulation: | 21 CFR §892.5050 | | |
| | Device Class: | II | | |
| | Product Code: | MUJ | | |
| | Recall Information: | Recall 68395 | | |
| | Legally Marketed Secondary Predicate Device: | | | |
| | Trade Name: | syngo.PET&CT Oncology | | |
| | 510(k)#: | K093621 | | |
| | Clearance Date: | February 23, 2010 | | |
| | Classification Name: | Picture Archiving and Communication Systems | | |
| | Classification Panel: | Radiology | | |
| | Classification Regulation: | 21 CFR § 892.2050 | | |
| | Device Class: | II | | |
| | Product Code: | LLZ | | |
| | Recall Information: | There have been no recalls for this device | | |
### 4. Substantial Equivalence:
The subject device, syngo.via RT Image Suite, is substantially equivalent to following medical devices in commercial distribution:
| Predicate Devices | FDA Clearance<br>Number | FDA Clearance<br>Date |
|---------------------------------------------------------------------------|-------------------------|-----------------------|
| Primary Predicate Device:<br>Siemens<br>syngo® Dosimetrist Workspace v2.7 | K101119 | June 16, 2010 |
| Secondary Predicate Device:<br>Siemens<br>syngo.PET&CT Oncology | K093621 | February 23, 2010 |
{5}------------------------------------------------
## 5. Device Description:
syngo.via RT Image Suite is an image analysis software used for viewing, manipulation, 3D and 4D visualization, comparison of medical images from multiple imaging modalities and for the segmentation of tumors and organs-atrisk, prior to dosimetric planning and response assessment in radiation therapy. syngo.via RT Image Suite combines routine and advanced digital image processing and visualization tools for easy manual and software assisted contouring of volumes of interest, identification of points of interest, registering images and exporting final results.
syngo.via RT Image Suite supports the medical professional with tools to use during different steps in radiation therapy case preparation.
The synqo.via RT Image Suite post-processing software application is embedded in the Siemens syngo.via framework (operating platform most recently cleared as VB10 K150843) of a multi-user HW/SW architecture with Client-Server support. Software information is located in Section 16.
The following modifications have been made to the previously cleared primary predicate device (syngo® Dosimetrist Workspace 2.7 K101119, clearance date June 16 2010) and the secondary predicate device (syngo.PET&CT Oncology K093621, clearance date February 23 2010).
- 1. A new software version SOMARIS/8 VB10 which supports the following:
- New post-processing software application "syngo.via RT Image . Suite" which has been designed for the new syngo.via platform (SOMARIS/8 VB10) and contains the following application modifications compared to the predicate devices:
- o Advanced Contouring: offers freehand editing and 3D Nudge tools for editing in non-original image orientations. These two features are segmentation improvements to the already cleared feature "Advanced Segmentation" described in Premarket Notification K101119. clearance date 06/16/2010.
- o Deformable Alignment: for cases where rigid alignment does not provide sufficient anatomical correspondence between secondary images and the reference image, deformable registration is an option. The deformable registration allows the user to map the corresponding anatomy captured in the two images. Rigid Alignment has been cleared within the Premarket Notification K093621, cleared date 02/23/2010.
- Contouring on 4D CT: The subject device allows the user to o display a cine loop of images acquired through gated CT. This is a new feature supported by software testing.
- 2. An extended, more descriptive Indication for Use statement to include details of device functionality
{6}------------------------------------------------
## 6. Summary of Technical Characteristics of the Subject Device as Compared with the Predicate Device:
syngo.via RT Image Suite has the same intended use and operating principle as the predicate devices. The synqo.via RT Image Suite application is designed to be operated on the syngo.via platform in a single or multi server environment.
syngo.via RT Image Suite does not have significant changes in technological characteristics when compared to the predicate devices. The subject device is the same to either one or both of the predicate devices in regards to:
- Visualization ●
- Routine Contouring .
- Annotation
- Reference Point Management ●
- Structure Set Management ●
- User Interface ●
- Archiving/Storage ●
- Hardware ●
- Image Processing and Evaluation ●
- Communication ●
The following table shows the differences in technological characteristics between the subject device and the predicate devices.
| Feature | Subject Device<br>syngo.via RT<br>Image Suite | Primary<br>Predicate Device<br>Siemens<br>syngo®<br>Dosimetrist<br>Workspace v2.7<br>(K101119) | Secondary<br>Predicate Device<br>Siemens<br>syngo.PET&CT<br>Oncology<br>(K093621) | Comparison |
|------------------------|-----------------------------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Advanced<br>Contouring | Freehand<br>editing 3D | Random Walker<br>Tools | N/A | The subject device offers<br>freehand 3D editing and 3D<br>Nudge tools for editing in<br>non-original image<br>orientations. These two<br>features are segmentation<br>improvements to the 2D<br>tools cleared in K101119. |
| | Nudge 3D | Contour Nudge<br>Tool | N/A | Contour interpolation was<br>cleared as part of both the<br>primary and secondary<br>predicate devices. |
| | Contour<br>interpolation | Contour<br>interpolation | Contour<br>interpolation | |
| Alignment | Rigid<br>alignment tools | Rigid alignment<br>tools | Rigid alignment<br>tools | The subject device contains<br>the same rigid alignment<br>tools as cleared in both the<br>primary and secondary<br>predicate devices |
# Table 1: Differences in technical characteristics
{7}------------------------------------------------
| Feature | Subject Device<br>syngo.via RT<br>Image Suite | Primary<br>Predicate Device<br>Siemens<br>syngo®<br>Dosimetrist<br>Workspace v2.7<br>(K101119) | Secondary<br>Predicate Device<br>Siemens<br>syngo.PET&CT<br>Oncology<br>(K093621) | Comparison |
|------------------------|-----------------------------------------------|------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Deformable<br>alignment tools | N/A | N/A | The Deformable alignment<br>used to map the<br>corresponding anatomy<br>captured on two registered<br>images is an optional,<br>additional feature for cases<br>where rigid alignment does<br>not provide sufficient<br>anatomical correspondence<br>between secondary images<br>and the reference image. |
| Contouring<br>on 4D CT | Present | N/A | N/A | The subject device allows<br>the user to display a cine<br>loop of images acquired<br>through gated CT. This is a<br>new feature and was not<br>available on the predicate<br>devices. |
### 7. Nonclinical Testing:
syngo.via RT Image Suite is designed to fulfill the requirements of the following safety and performance standards:
| Recognition<br>Number | Product<br>Area | Title of Standard | Reference<br>Number and<br>Date | Publication<br>Date | Standards<br>Development<br>Organization |
|-----------------------|-----------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|---------------------|------------------------------------------|
| 12-238 | Radiology | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set | PS 3.1 - 3.18 | 03/16/2012 | NEMA |
| 13-8 | Software | Medical device software -<br>Software life cycle processes | 62304 First<br>edition 2006-05 | 08/20/2012 | IEC |
| 5-40 | General | Medical devices - Application of<br>risk management to medical<br>devices | 14971 Second<br>Edition 2007-<br>03-01 | 08/20/2012 | ISO |
| 5-85 | General | Medical electrical equipment --<br>Part 1-6: General requirements<br>for basic safety and essential<br>performance -- Collateral<br>Standard: Usability | 60601-1-6<br>Edition 3.0<br>2010-01 | 1/30/2014 | IEC |
| 5-41 | General | Medical electrical equipment -<br>Part 1-4: General requirements<br>for safety -- Collateral standard:<br>Programmable electrical medical<br>systems, edition 1.1 | 60601-1-<br>4:2000, Consol.<br>Ed. 1.1 | 09/08/2009 | IEC |
{8}------------------------------------------------
# Verification and Validation
This submission contains performance data to demonstrate continued conformance with special controls for medical devices containing software. Integration and functional tests were conducted for syngo.via RT Image Suite during product development.
Software documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. The test results show that all of the software specifications have me the acceptance criteria.
The risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
Siemens conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. It is the hospital's responsibility to comply with IEC 8001-1-2010.
# Summary
Performance tests were conducted to test the functionality of the subject device, syngo.via RT Image Suite. Results of all conducted testing were found acceptable in supporting the claim of substantial equivalence.
### 8. Indications for Use:
syngo.via RT Image Suite is a 3D and 4D image visualization, multimodality manipulation and contouring tool that helps the preparation and response assessment of treatments such as, but not limited to those performed with radiation (for example, Brachytherapy, Particle Therapy, External Beam Radiation Therapy).
It provides tools to efficiently view existing contours, create, edit, modify, copy contours of regions of the body, such as but not limited to, skin outline, targets and organs-at-risk. Contours and images can subsequently be exported to a Treatment Planning System. The software combines following digital image processing and visualization tools:
- · Multi-modality viewing and contouring of anatomical, functional, and multiparametric images such as but not limited to CT, PET, PET/CT, MRI, and Linac Cone Beam CT (CBCT) images
- · Multiplanar reconstruction (MPR) thin/thick, minimum intensity projection (MIP), volume rendering technique (VRT)
{9}------------------------------------------------
- · Freehand and semi-automatic contouring of regions-of-interest on any orientation including oblique
- · Creation of contours on any type of images without prior assignment of a planning CT
- · Manual and semi-automatic registration using rigid and deformable registration
- · Supports the user in comparing, contouring, and adapting contours based on datasets acquired with different imaging modalities and at different time points. Supports the user in comparing images and contours of different patients
- · Supports multi-modality image fusion
- · Visualization and contouring of moving tumors and organs
### 9. General Safety and Effectiveness Concerns:
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing. To minimize electrical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.
# 10. Conclusion as to Substantial Equivalence
The predicate devices were cleared based on non-clinical supportive information. The subject device was tested using the same non-clinical methods. The subject device non-clinical data supports the safety of the software with verification and validation testing. Verification and Validation testing also demonstrates that syngo.via RT Image Suite performs as intended. The non-clinical test data demonstrates that syngo.via RT Image Suite device performance is comparable to the predicate devices that are currently marketed for the same intended use. In summary, Siemens is of the opinion that the syngo.via RT Image Suite does not introduce any new potential safety risk and is substantially equivalent to and performs as well as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.