Conventus PRS System

K151379 · Conventus Orthopaedics, Inc. · HRS · Aug 13, 2015 · Orthopedic

Device Facts

Record IDK151379
Device NameConventus PRS System
ApplicantConventus Orthopaedics, Inc.
Product CodeHRS · Orthopedic
Decision DateAug 13, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Conventus PRS™ System is indicated for treatment of proximal radial fractures when internal fixation is desired, and fracture fragments are not too numerous and/or too small to be stabilized with the use of the device.

Device Story

Conventus PRS™ System is an intramedullary bone fixation device for proximal radius fractures. Implant consists of self-expanding scaffold deployed into medullary canal; provides structural support for bone fragments. Fragments are secured to scaffold using fragment screws. Device used by orthopedic surgeons in clinical settings. Benefits include stabilization of fracture fragments to facilitate healing. Implant materials include titanium alloy (Ti-6Al-4V ELI) and Nitinol.

Clinical Evidence

No clinical data provided. Evidence consists of bench testing (static/cyclic axial, bend, and torsional testing; screw pullout; corrosion; wear; nickel ion release; surface analysis; Nitinol phase/transition temperature; biocompatibility; MRI testing) and animal testing.

Technological Characteristics

Intramedullary self-expanding scaffold; materials: titanium alloy (Ti-6Al-4V ELI) and Nitinol. Dimensions/form factor: designed for medullary canal deployment. Energy source: none (mechanical). Connectivity: none. Sterilization: not specified.

Indications for Use

Indicated for patients with proximal radial fractures requiring internal fixation, provided fracture fragments are of sufficient size and number for stabilization by the device.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol representing the human form, depicted as three stylized profiles facing to the right. The profiles are connected and flow together, creating a sense of unity and collaboration. August 13, 2015 Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Conventus Orthopaedics. Incorporated Mr. Kent R. Lind Vice President Quality, Regulatory, Clinical 10200 73td Avenue North, Suite 122 Maple Grove, Minnesota 55369 Re: K151379 Trade/Device Name: Conventus PRSTM System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HSB Dated: May 21, 2015 Received: May 22, 2015 Dear Mr. Lind: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing {1}------------------------------------------------ Page 2 - Mr. Kent R. Lind (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, ## Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ### Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) Device Name Conventus PRSTM System Indications for Use (Describe) The Conventus PRS™ System is indicated for treatment of proximal radial fractures when internal fixation is desired, and fracture fragments are not too numerous and/or too small to be stabilized with the use of the device. Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Company: | Conventus Orthopaedics, Inc.<br>10200 73rd Avenue North, Suite 122<br>Maple Grove, MN 55369 | |----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Trade Name: | Conventus PRSTM System | | Device Common Name: | Proximal radius fracture system | | Contact: | Kent R. Lind<br>Vice President, Quality, Regulatory, Clinical<br>Phone: (763) 515-5003<br>Fax: (763) 315-4980 | | Date Prepared: | July 27, 2015 | | Classification: | 21 CFR 888.3030, Single/multiple component metallic bone<br>fixation appliances and accessories<br>21 CFR 888.3020: Intramedullary fixation rod | | Class: | II | | Product Codes: | HRS and HSB | | Indications for Use: | The Conventus PRSTM System is indicated for treatment of<br>proximal radial fractures when internal fixation is desired, and<br>fracture fragments are not too numerous and/or too small to be<br>stabilized with the use of the device. | | Device Description: | The Conventus PRSTM is an intramedullary device intended for<br>proximal radius fractures. The PRSTM System is a self-expanding<br>implant which is deployed into the medullary canal and provides a<br>scaffold to which bone fragments are attached using fragment<br>screws. The implant is made from titanium alloy (Ti-6Al-4V ELI)<br>and Nitinol. | {4}------------------------------------------------ Substantial Equivalence: Conventus Orthopaedics has demonstrated that, for purposes of FDA's regulation of medical devices, the Conventus PRS™ System is substantially equivalent to the following devices that have been previously cleared by the FDA: - · Primary predicate: Synthes T-Plate (Pre-Amendment) - · Synthes Elastic Intramedullary Nail System (K042135) - · Conventus DRS™ System (K102689 and K131552) This finding is supported by the following pre-clinical tests that have been performed: - Static and cyclic axial/bend testing - · Static and cyclic torsional testing - · Screw pullout testing - · Corrosion testing - · Wear testing - · Nickel ion release testing - Surface analysis testing - Nitinol phase composition - Nitinol transition temperature (Af) - · Biocompatibility testing - · Animal Testing - · MRI Testing The results demonstrate that the PRS™ System is substantially equivalent to the legally marketed predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...