Inclusive Titanium Abutments compatible with: Hiossen HG Implant System
K151375 · Prismatik Dentalcraft, Inc. · NHA · Oct 7, 2015 · Dental
Device Facts
Record ID
K151375
Device Name
Inclusive Titanium Abutments compatible with: Hiossen HG Implant System
Applicant
Prismatik Dentalcraft, Inc.
Product Code
NHA · Dental
Decision Date
Oct 7, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Indications for Use
Inclusive Titanium Abutments are premanufactured prosthetic components directly connected to endosseous dental implant and are intended for use as an aid in prosthetic rehabilitation.
Device Story
Inclusive Titanium Abutments are premanufactured prosthetic components made of titanium alloy (Ti-6Al-4V). They connect directly to endosseous dental implants to support prosthetic restorations like crowns, bridges, or overdentures. The device features a hexagonal anti-rotational connection. Used in dental clinics by dental professionals, the abutments are placed in the oral cavity at the tissue level. The device provides a stable interface between the implant and the final dental prosthesis, aiding in the functional and aesthetic rehabilitation of the patient's dentition.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including reliability calculations, fatigue testing (ISO 14801), and static load failure testing of the assembled implant/abutment system.
Technological Characteristics
Material: Titanium alloy (Ti-6Al-4V ELI) per ASTM F136-12a. Design: Machined, hexagonal anti-rotational connection. Dimensions: Mini (2.1mm hex) and Standard (2.5mm hex). Sterilization: Moist heat/steam sterilization per ANSI/AAMI ST79 and ISO 17665-1. Biocompatibility: Evaluated per ISO 10993-1 via predicate comparison.
Indications for Use
Indicated for use as an aid in prosthetic rehabilitation for patients requiring endosseous dental implant-supported restorations.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
Predicate Devices
Inclusive® Titanium Abutments for Astra Tech OsseoSpeed Implants (K100993)
Inclusive® Abutment for Zimmer, 3i and Nobel Biocare Implants (K073217)
Inclusive® Titanium Abutment for Camlog Screw-Line Implants (K121391)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus-like symbol with three human profiles facing right, representing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol, indicating the department's name and national affiliation.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 7, 2015
PRISMATIK DENTALCRAFT, INC. Mr. Armin Zehtabchi Senior RA 2212 Dupont Dr., Suite P Irvine, California 92612
Re: K151375
> Trade/Device Name: Inclusive Titanium Abutments compatible with: Hiossen HG Implant System Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: May 21, 2015 Received: May 22, 2015
Dear Mr. Zehtabchi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| Form Approved: OMB No. 0910-0120 | |
|-----------------------------------|--|
| Expiration Date: January 31, 2017 | |
| See PRA Statement below. | |
| 510(k) Number (if known) | TBD |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | Inclusive® Titanium Abutments compatible with: Hiossen HG Implant System |
| Indications for Use (Describe) | Inclusive Titanium Abutments are premanufactured prosthetic components directly connected to endosseous dental implant and are intended for use as an aid in prosthetic rehabilitation. |
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <div> <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> </div> <div> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> </div> |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with different colors. To the right of the triangle is the text "PRISMATIK DENTALCRAFT, INC."
B.
## 006_510 (K) Summary-807.92(c)
This 510 (k) summary is being submitted in accordance with the requirements of SMDA of 1990 and 21 CFR 807.92.
### SUBMITTER INFORMATION A.
| Company Name: | Prismatik Dentalcraft, Inc. |
|-------------------------------|-----------------------------------------------------------------------------|
| Company Address: | 2212 Dupont Dr., Suite P,<br>Irvine, CA 92612 |
| Company Phone: | 949-225-1269 |
| Company FAX: | 949-553-0924 |
| Facility Registration Number: | 3005477956 |
| Primary Contact Person: | Armin Zehtabchi, (949) 225-1234<br>Senior RA Specialist |
| Secondary Contact Person | Marilyn Pourazar, (949) 225-1269<br>Senior Director, RA/QA |
| Date Summary Prepared: | September 25, 2015 |
| DEVICE IDENTIFICATION | |
| Trade/Proprietary Name: | Inclusive® Titanium Abutments compatible<br>with: Hiossen HG Implant System |
| 21 CFR Reference: | 21 CFR 872.3630 |
| 21 CFR Common Name: | Abutment, implant, dental, endosseous |
| Classification: | Class II, NHA |
| Panel: | Dental |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Prismatik Dentalcraft, Inc. On the left is a triangular prism with a color gradient from blue to pink. To the right of the prism is the company name, "PRISMATIK DENTALCRAFT, INC." in a sans-serif font. The text is a dark teal color.
#### C. IDENTIFICATION OF PREDICATE DEVICE
Trade/Proprietary Name:
Predicate Device: Inclusive® Titanium Abutments for Astra Tech OsseoSpeed Implants=510(k)-K100993
Reference Device: HS/HG Prosthetic System-510(k)k100245
### D. PROPOSED DEVICE DESCRIPTION
Inclusive® Titanium Abutments for Hiossen HG Implant System are made of Titanium Alloy (Ti-6Al-4V) and are premanufactured prosthetic components connected directly to the endosseous dental implant and intended as an aid in prosthetic rehabilitation. They are compatible with Hiossen HG Implant System (Mini and Standard).
The anti-rotational feature for Inclusive® Titanium Abutment is a hexagon. The numerical value which defines the across flats of the hexagon is 2.10 mm for the Mini, and 2.50 mm for Standard. Both have a tolerance range of ±0.05 mm.
#### E. INDICATIONS FOR USE
Inclusive® Titanium Abutments are premanufactured prosthetic components directly connected to endosseous dental implants and are intended for use as an aid to prosthetic rehabilitation.
#### DETERMINATION OF SUBSTANTIAL EQUIVALENCE F.
Inclusive® Titanium Abutments compatible with: Hiossen HG Implant System is substantially equivalent to the Predicate Device, Inclusive® Titanium Abutments for Astra Tech OsseoSpeed Implants-510(k)-K100993, and the Reference Device, HS/HG Prosthetic System-510(k)-K100245 identified in Section C above. They are substantially equivalent in intended use, materials, design and performance.
(See Comparison Tables below)
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. The logo consists of a colorful triangle on the left and the company name on the right. The triangle is divided into several sections, each with a different color. The company name is written in a sans-serif font, with "PRISMATIK" on the top line and "DENTALCRAFT, INC." on the bottom line.
| Attribute | Reference Device (1) | Predicate Device (1) | Proposed Device | Similarities<br>and<br>Differences<br>of Devices |
|---------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| | HS/HG Prosthetic<br>System (K100245)<br>Osstem Implant<br>Co. Ltd<br>(Subsidiary of<br>Hiossen Inc) | Inclusive® Titanium Abutment<br>for Astra Tech OsseoSpeed™<br>Implants (K100993)<br>Prismatik Dentalcraft, Inc. | Inclusive Titanium<br>Abutments compatible<br>with: Hiossen HG Implant<br>System<br>Prismatik Dentalcraft, Inc. | |
| Dimensions of<br>Abutment | Mini - 2.1 Across<br>Flats of Hex<br>Standard - 2.5<br>Across Flats of<br>Hex | 3.5/4.0 - 2.1 Across Flats of Hex<br>4.5/5.0 - 2.5 Across Flats of Hex | Mini - 2.1 Across Flats of<br>Hex<br>Standard - 2.5 Across Flats of<br>Hex | Same |
| Dimensions of<br>Abutment Screw | Mini - Length -<br>10.7mm, Screw<br>Head - 2.2mm<br>Standard - Length -<br>8.85mm, Screw<br>Head - 2.5mm | 3.5/4.0 - Length 8.25mm, Screw<br>Head 2.33mm 4.5/5.0 - Length -<br>10.3mm, Screw Head 2.33mm | Mini - Length - 10.3mm,<br>Screw Head - 2.2mm<br>Standard - Length - 8.35mm,<br>Screw Head - 2.3mm | Different<br>screws<br>dimensions |
| Indications for<br>Use | HS/HG Prosthetic<br>System is intended<br>for use with the<br>dental implant to<br>provide support for<br>prosthetic<br>restorations such as<br>crowns, bridges, or<br>overdentures. | The Inclusive Titanium<br>Abutments for Astra<br>OsseoSpeed™ Implants are<br>premanufactured<br>prosthetic components directly<br>connected to endlosseous dental<br>implants and are intended<br>for use as an aid in prosthetic<br>rehabilitation. They are<br>compatible with the Astra Tech<br>OsseoSpeed™ 3.0, 3.5, 4.0, 4.5,<br>5.0 implants. | Inclusive Titanium<br>Abutments are<br>premanufactured prosthetic<br>components directly<br>connected to endosseous<br>dental implant and are<br>intended for use as an aid in<br>prosthetic rehabilitation. | Similar<br>intended use<br>for the<br>additional use<br>with Hiossen<br>mini<br>2.1 and<br>Standard 2.5 |
| Platform<br>Diameter<br>Compatibility | Hiossen HG Mini<br>(2.1 Across Flats<br>of Hex and<br>Standard (2.5<br>Across Flats of<br>Hex | Astra Tech OsseoSpeed™<br>Implants 3.5/4.0<br>4.5/5.0 | Hiossen HG<br>Mini - 2.1 Across Flats of<br>Hex<br>Standard - 2.5 Across Flats<br>of Hex | Same as<br>Reference<br>predicate |
| Connection | Hexagonal | Hexagonal | Hexagonal | Same |
| Material | Titanium alloy<br>(Ti-6AL-4V) | Titanium alloy<br>(Ti-6AL-4V) | Titanium alloy<br>(Ti-6AL-4V) | Same |
| Design /<br>Construction | Machined | Machined | Machined | Same |
| Attribute | Reference Device (1) | Predicate Device (1) | Proposed Device | Similarities<br>and<br>Differences<br>of Devices |
| | HS/HG Prosthetic<br>System<br>(K100245)<br>Osstem Implant<br>Co. Ltd<br>(Subsidiary of<br>Hiossen Inc) | Inclusive® Titanium Abutment<br>for Astra Tech OsseoSpeed™<br>Implants (K100993)<br>Prismatik Dentalcraft, Inc. | Inclusive Titanium<br>Abutments compatible<br>with: Hiossen HG Implant<br>System<br>Prismatik Dentalcraft, Inc. | |
| Anatomical Site /<br>Placement | Oral Cavity /<br>Tissue Level | Oral Cavity / Tissue Level | Oral Cavity / Tissue Level | Same |
| Abutment Angle | 0-30° | 0-20° | 0-30° | Similar |
# Table 1 – Comparison between Predicate and Proposed Device
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. On the left is a triangular prism with different colors radiating from the center. To the right of the prism is the company name, "PRISMATIK DENTALCRAFT, INC." in blue text.
#### G. NON-CLINICAL TESTING (PERFORMANCE DATA)
Non-clinical test data was used to evaluate the device's equivalence, and determine substantial equivalence with predicate devices.
Non-clinical testing was performed in accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments" and it consisted of Reliability Calculation, Fatigue Testing and Static Load Failure Testing of finished assembled implant/abutment systems, and the testing results and evaluations demonstrate the implant to abutment compatibility between the proposed and predicate/reference devices. The detailed discussion and the testing procedure can be found in Section 018 (Performance Testing-Bench). The applicable standards that are used in this submission are listed below:
| | Applicable Standards |
|---------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM F136-12a | Standard for Wrought Titanium-6Aluminum-<br>4Vanadium ELI (Extra Low Interstitial) Alloy<br>for Surgical Implant Applications |
| AAMI/ANSI/ISO 10993-1:2009 | Biological Evaluation of Medical Devices - Part<br>1: Evaluation and Testing within a Risk<br>Management Process (Biocompatibility) |
| ISO14801:2007 | Dentistry - Implants - Dynamic fatigue test for<br>endosseous dental implants |
| AAMI/ANSI 17665:-1:2006 | Sterilization of health care products - Moist Heat<br>- Part 1: Requirements for the development,<br>validation and routine control of a sterilization<br>process for medical devices |
| AAMI/ANSI/ISO 17665-2:2009 | Sterilization of Health Care Products - Radiation<br>- Part 2: Guidance of on the application of ISO<br>17665-1 |
| ANSI/AAMI ST79:2010 &A1:2010 &<br>A2:2011 & A3:2012 & A4:2013 | (Consolidated Text) Comprehensive guide to<br>steam sterilization and sterility assurance in<br>health care facilities |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Prismatik Dentalcraft, Inc. On the left is a triangle with a gradient of colors, including blue, pink, and yellow. To the right of the triangle is the text "PRISMATIK DENTALCRAFT, INC." in a blue sans-serif font.
Inclusive® Titanium Abutments are manufactured from biocompatible titanium grade 23 (Ti-6AL-4V ELI) and it meets ASTM F-136 Standard. In accordance with FDA Guidance "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments", in lieu of performing biocompatibility testing per ISO 10993-1 for the subject device, we have identified several of Prismatik own predicate devices with identical materials, same manufacturing process, and same type/duration of patient contact:
- K073217, Inclusive® Abutment for Zimmer, 3i and Nobel Biocare Implants -
- K100993, Inclusive® Titanium Abutments for Astra OsseoSpeed Implants -
- K121391, Inclusive® Titanium Abutment for Camlog Screw-Line Implants -
In addition, sterilization validation information and a recommended sterilization method based on ANSI/AAMI ST79 and ISO 17665-1 is provided in the Information for Use.
#### CONCLUSION FROM THE NON-CLINICAL TESTING (PERFORMANCE DATA) H.
The proposed device, the Inclusive® Titanium Abutments compatible with: Hiossen HG Implant System have the same performance specifications, fundamental scientific technology and intended use as that of the Predicate Device, Inclusive® Titanium Abutments for Astra Tech OsseoSpeed Implants-510(k)-K100993, and the Reference Device, HS/HG Prosthetic System-510(k)-K100245 identified in Section C above. The testing performed demonstrated implant to abutment compatibility and supports the substantial equivalence of the subject device to the identified predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.