K151373 · Merit Medical Systems, Inc. · FCG · Aug 7, 2015 · Gastroenterology, Urology
Device Facts
Record ID
K151373
Device Name
Corvocet Coaxial Introducer
Applicant
Merit Medical Systems, Inc.
Product Code
FCG · Gastroenterology, Urology
Decision Date
Aug 7, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
The Corvocet Coaxial Introducer is intended for use as a guiding needle in obtaining core biopsy samples from soft tissues such as liver, kidney, prostate, spleen, breast, lung, lymph nodes and various soft tissue tumors.
Device Story
Corvocet Coaxial Introducer is a three-part disposable device used as a guiding needle for core biopsy procedures. It consists of a stainless steel outer cannula with an echo-enhanced tip for ultrasound visibility, a stainless steel inner trocar with a one-handed release hub, and a color-coded silicone slip ring depth stop. An optional blunt stylet is available for 17G and 19G models to navigate around vasculature or organs. The device is used by clinicians in a clinical setting to facilitate the placement of biopsy instruments. By providing a stable path and depth control, it assists in obtaining tissue samples while minimizing trauma to surrounding structures. The device is supplied sterile and is for single use only.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included mechanical testing (tensile strength, Luer performance, depth stop function, introducer visibility) and simulated use testing. Biocompatibility testing was conducted per ISO 10993-1, including cytotoxicity, sensitization, irritation, acute systemic toxicity, pyrogenicity, and chemical characterization.
Technological Characteristics
Materials: Stainless steel (cannula, trocar, stylet), silicone (depth stop). Features: Echo-enhanced tip, one-handed release trocar hub, color-coded depth stop. Standards: ISO 7864, ISO 9626, ISO 594-1/2, ISO 6009, ISO 11607-1, ISO 11135, ISO 10993 series, ASTM D4169, ASTM F1980. Sterilization: Ethylene oxide. Connectivity: None (mechanical device).
Indications for Use
Indicated for use as a guiding needle for core biopsy sampling of soft tissues (liver, kidney, prostate, spleen, breast, lung, lymph nodes, and soft tissue tumors) in patients requiring biopsy procedures.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in all caps. Inside the circle is a stylized image of an eagle with three human profiles incorporated into its design, suggesting a connection between health, humanity, and national service.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 7, 2015
Merit Medical Systems Incorporated Ms. Ileana Garcia Regulatory Affairs Specialist 1600 West Merit Parkway South Jordan, Utah 84095
Re: K151373
Trade/Device Name: Corvocet Coaxial Introducer Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: FCG Dated: May 20, 2015 Received: May 21, 2015
Dear Ms. Garcia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerelv vours.
Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices
Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K151373
Device Name Corvocet Coaxial Introducer
Indications for Use (Describe)
The Corvocet Coaxial Introducer is intended for use as a guiding needle in obtaining core biopsy samples from soft tissues such as liver, kidney, prostate, spleen, breast, lung, lymph nodes and various soft tissue tumors.
| Type of Use (Select one or both, as applicable) |
|------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Ref 21 CFR 201 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 201 Subpart G)</label> |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5.0 510(k) Summary
| General<br>Provisions | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date Prepared:<br>Registration Number: | Merit Medical Systems, Inc.<br>1600 West Merit Parkway<br>South Jordan, UT 84095<br>(801) 208-4187<br>(801) 316-4065<br>lleana Garcia, Regulatory Affairs Specialist<br>20 May 2015<br>1721504 |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject Device | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Regulatory Class:<br>Product Code:<br>21 CFR §:<br>Review Panel: | Corvocet Coaxial Introducer<br>Coaxial Introducer<br>Biopsy Needle Kit<br>FCG<br>876.1075<br>Gastroenterology/Urology |
| Predicate<br>Device | Trade Name:<br>Classification Name:<br>Premarket Notification:<br>Manufacturer: | TruGuide® Disposable Coaxial Biopsy Needle<br>Biopsy Needle Kit<br>K936194<br>Bard Peripheral Vascular, Inc.<br>This predicate has not been subject to a design-related recall. |
| Reference<br>Device | | No reference devices were used in this submission. |
| Device<br>Description | The Merit Corvocet Coaxial Introducer is used as a guiding needle for a<br>biopsy device obtaining core biopsy samples from soft tissues. It is a<br>three-part disposable device consisting of the following: a stainless<br>steel outer cannula with a female Luer lock hub and an echo-enhanced<br>tip to aid with visibility under ultrasound; a stainless steel inner trocar<br>attached to a unique trocar hub design that allows it to be released by<br>squeezing both sides of the hub; and a flexible silicone slip ring-style<br>depth stop that is color-coded for gauge identification and can be<br>manually adjusted to maintain the depth of the coaxial introducer in a<br>procedure. An optional stainless steel blunt stylet with hub can be<br>included with 17G and 19G coaxial introducers. The blunt stylet may be<br>used to manipulate through soft tissue and around vasculature or other<br>organs to minimize the risk of unintentional damage to these areas.<br>The Merit Corvocet Coaxial Introducer is supplied sterile and is single<br>use only. | |
| Indications for | | The Corvocet Coaxial Introducer is intended for use as a guiding |
| Use | needle in obtaining core biopsy samples from soft tissues such as liver,<br>kidney, prostate, spleen, breast, lung, lymph nodes and various soft<br>tissue tumors.<br><br>The Indications for Use statement for the Corvocet Coaxial Introducer is<br>not identical to the predicate device; however, the differences do not<br>alter the intended therapeutic use of the device nor do they affect the<br>safety and effectiveness of the device relative to the predicate. Both the<br>subject and predicate devices have the same intended use as a guiding<br>needle in obtaining core tissue samples from soft tissue. | |
| | The technological characteristics of the subject Corvocet Coaxial<br>Introducer are substantially equivalent to those of the predicate<br>TruGuide® Disposable Coaxial Biopsy Needle. | |
| | At a high level, the subject and predicate devices are based on the<br>following same elements: | |
| Comparison to<br>Predicate<br>Device | • Clinical use<br>• Labeling<br>• Basic design<br>• Principle of operation<br>• Performance<br>• Needle gauges | |
| | The following differences exist between the subject and predicate<br>devices: | |
| | • Soft tissues<br>• One-handed trocar removal<br>• Materials | |
| Performance<br>Data | FDA guidance and recognized performance standards have been<br>established for biopsy needle kit under Section 514 of the Food, Drug<br>and Cosmetic Act. A battery of tests was performed based on the<br>requirements of the below recognized performance standards and<br>guidance, as well as biocompatibility, sterilization, and labeling<br>standards and guidance. Conformity to these standards demonstrates<br>that the proposed Corvocet Coaxial Introducer met the standards'<br>established acceptance criteria applicable to the safety and efficacy of<br>the device. Performance testing was conducted based on the risk<br>analysis and based on the requirements of the following documents:<br>• Guidance for the Content of Premarket Notifications for Biopsy<br>Devices Used in Gastroenterology and Urology<br>• ISO 7864:1993 – Sterile hypodermic needles for single use<br>• ISO 7864-2:1993 – Sterile hypodermic syringes and needles –<br>Part 2: Specification for sterile hypodermic needles for single<br>use<br>• ISO 9626:1991 Amendment 1:2001- Stainless steel needle | |
{4}------------------------------------------------
{5}------------------------------------------------
- ISO 594-1:1986 Conical fittings with a 6% (Luer) taper for ● syringes, needles, and certain other medical equipment - Part 1: General requirements
- ISO 594-2:1998 Conical fittings with a 6% (Luer) taper for . syringes, needles, and certain other medical equipment - Part 2: Lock fittings
- . ISO 6009:1992 - Hypodermic needles for single use - Color coding for identification
- . ISO 11607-1:2006 - Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems, and packaging systems
- ASTM D4169-09 Standard Practice for Performance Testing ● of Shipping Containers and Systems
- . ISO 2233:2001 - Packaging - Complete, filled transport packages and unit loads - Conditioning for testing
- ASTM F1980-11 - Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices
- . ISO 11135:2014 - Sterilization of health care products -Ethylene oxide - Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices
- ISO 10993-1:2009 Biological evaluation of medical devices -● Part 1: Evaluation and testing within a risk management process
- ISO 10993-5:2009 - Biological evaluation of medical devices -Part 5: Tests for in vitro cytotoxicity
- ISO 10993-7:2008 Biological evaluation of medical devices -● Part 7: Ethylene oxide sterilization residuals
- . ISO 10993-10:2010 - Biological evaluation of medical devices -Part 10: Tests for irritation and skin sensitization
- ISO 10993-11:2006 - Biological evaluation of medical devices -Part 11: Tests for systemic toxicity
- United States Pharmacopoeia 37, National Formulary 32 <151> . Pyrogen Test, 2014.
The following performance data were provided in support of the substantial equivalence determination.
## Performance Testing - Bench
- Dimensions ●
- Tensile of Joints Needle-to-Hub ●
- Tensile of Joints Trocar-to-Hub ●
- Tensile of Joints Blunt Stylet-to-Hub ●
- Luer - Gauging
- Luer Liquid Leakage ●
- Luer Air Leakage ●
- Luer Separation Force ●
- Luer Unscrewing Torque ●
- Luer Ease of Assembly ●
- Luer Resistance to Overriding ●
{6}------------------------------------------------
- Luer Stress Cracking ●
- Depth Stop
- Introducer Visibility ●
- Simulated Use
## Biocompatibility testing
The biocompatibility evaluation for the Corvocet Coaxial Introducer was conducted in accordance with the FDA Blue Book Memorandum #G95-Required Biocompatibility Training and Toxicology Profiles for ー Evaluation of Medical Devices, May 1, 1995, and the International Standard ISO 10993-1:2009 Biological evaluation of medical devices -Part 1: Evaluation and testing within a risk management process as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity ●
- . Sensitization
- Irritation
- Acute Systemic Toxicity ●
- Pyrogenicity
- Chemical Characterization ●
## Summary of Substantial Equivalence
Based on the indications for use, design, safety and performance testing. the subject Corvocet Coaxial Introducer meets the requirements that are considered essential for its intended use and is substantially equivalent to the predicate device, the TruGuide® Disposable Coaxial Biopsy Needle, K936194.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.