Corporate Dental Imaging (CDI) Application

K151371 · U.S. Army Dental Command (Dencom) · LLZ · Aug 19, 2015 · Radiology

Device Facts

Record IDK151371
Device NameCorporate Dental Imaging (CDI) Application
ApplicantU.S. Army Dental Command (Dencom)
Product CodeLLZ · Radiology
Decision DateAug 19, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

Corporate Dental Imaging (CDI) Application™ is a software device for general dental diagnostic imaging. It controls capture, display, transfer, enhancement, and storage of X-ray dental images from digital imaging systems.

Device Story

CDI Application is a PACS software for dental clinics; replaces legacy DEVAA system. Inputs: X-ray images from intra-oral/extra-oral modalities, intra-oral cameras, digital cameras, and video sources. Operation: Desktop client for image capture; web-based client for viewing/diagnostics; Microsoft SQL Server backend for DICOM image storage. Features: DICOM 3.x compliant; tools for contrast/brightness enhancement, image reversal, and magnification. User: Dental clinicians. Output: Displayed images for diagnostic review. Benefits: Global image access across enterprise, migration of existing images, and standardized DICOM object creation. No automated detection or semi-automated processes for nodules/shapes.

Clinical Evidence

Bench testing only. Verification and validation testing performed on all input/output functions and operational modes. Results confirmed that all predetermined acceptance criteria were met.

Technological Characteristics

Software-based PACS; DICOM 3.x compliant. Operates on Windows 64-bit 2008 R2 Enterprise. Architecture: Desktop client for capture, web-based client for viewing, Microsoft SQL Server database. Connectivity: Ethernet/networked. No patient contact. No automated detection algorithms.

Indications for Use

Indicated for general dental diagnostic imaging, including the capture, display, transfer, enhancement, and storage of X-ray dental images from digital imaging systems.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol. The logo is simple, clean, and easily recognizable. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 19, 2015 U.S. Army Dental Command (DENCOM) % Mr. Carl Alletto Consultant OTech, Inc. 8317 Belew Drive MCKINNEY TX 75071 Re: K151371 Trade/Device Name: Corporate Dental Imaging (CDI) Application"*** Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: July 29, 2015 Received: August 4, 2015 Dear Mr. Alletto: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael O'Hara For Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) # K151371 Device Name Corporate Dental Imaging (CDI) Application™ Indications for Use (Describe) Corporate Dental Imaging (CDI) Application™ is a software device for general dental diagnostic imaging. It controls capture, display, transfer, enhancement, and storage of X-ray dental images from digital imaging systems. 2 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) # PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990. #### Date Prepared: May 5, 2015 ### Submitter's Information: 21 CFR 807.92(a)(1) Mr. William K. Porter Chief Operating Officer (COO) Information Management and Technology Division (IMTD) U.S. Army Dental Command (DENCOM) 4270 Gorges Circle Suits 101, B1070 Fort Sam Houston, Texas 78234 Tel: 210 808-2748 Fax: 210 295-3013 Email: William.k.porter.civ@mail.mil #### Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2) | Product Name: | Corporate Dental Imaging (CDI) Application™ | |----------------------|---------------------------------------------| | Common Name: | Picture, archive and communications system | | Classification Name: | System, Image Processing, Radiological | | Product Code: | LLZ | #### Predicate Device: 21 CFR 807. 92(a)(3) The Corporate Dental Imaging (CDI) Application™ system is substantially equivalent to: | Device Classification Name | system, image processing, radiological | |------------------------------|---------------------------------------------------------------------------------------------------| | 510(k) Number | K102684 | | Device Name | XELIS DENTAL | | Original Applicant | INFINITT CO., LTD.<br>12f Daerung Post Tower III<br>182-4 Guro-Dong, Guro-Gu<br>Seoul, KR 152-050 | | Regulation Number | 892.2050 | | Classification Product Code | LLZ | | Date Received | 09/17/2010 | | Decision Date | 10/01/2010 | | Regulation Medical Specialty | Radiology | | 510k Review Panel | Radiology | # Device Description: 21 CFR 807 92(a)(4) The Corporate Dental Imaging (CDI) Application™ is a software application used for storing and displaying medical dental images. The application conforms to the DICOM 3.x standard to allow interoperability with other DICOM compliant systems. The implementation of CDI interface has been tested to assure compliance with the DICOM Conformance Statement. However, the DICOM Conformance Statement for the Subject device and the DICOM standard itself does not guarantee interoperability of the system to other vendor products. The user must compare the relevant Conformance Statements and if a successful interconnection should be possible, the user is responsible to specify an appropriate test {4}------------------------------------------------ suite and to validate the required interoperability. A network environment may need additional functions out of the scope of DICOM. Note: The subject device does not include any automated or semi-automated processes for the detection of nodules or other shapes. The Corporate Dental Imaging (CDI) Application ™ system is being designed as a replacement for the legacy PACS system known as DEVAA. CDI will replace the legacy PACS technologies found in DEVAA with a more robust software and hardware solution. In addition, CDI will add additional capabilities not available in legacy capture/view tools. Additional benefits include providing a global view of images through DICOM image data query across multiple service image archives and the capability to migrate existing DICOM images to new image servers. In addition, CDI offers the capability to query and retrieve images at any location within the digital enterprise for primary, ghosted and retake images. CDI is comprised of one (1) desktop client for Modality Image Capture, one (1) web-based client for Image Viewing and Diagnostics with Microsoft SQL Server back-end database(s) for storing DICOM images. There is at least one instance of this environment. Each Military clinical post will most likely have multiple instances of this environment depending upon the volume of productivity. CDI will form the core of the acquisition and viewing operations within the enterprise architecture. CDI will capture images directly from intra-oral and extra-oral acquisition modalities, full color images from intra-oral cameras, digital cameras and other video sources. CDI is loaded onto the Acquisition PCs in the clinic where either the images will be acquired or where the images will need to be viewed. If the PC is being used to acquire images, CDI is configured to use the attached acquisition modalities. CDI is has been designed to be able to acquire an image regardless of the acquisition modality used to capture the image. CDI will use the hardware vendor's own drivers to seamlessly initialize the acquisition modality from within the CDI UI. It does not matter which type of modality is used to acquire the image, or which type of image series that might be taken, the user is always presented with the same interface. CDI will be designed to create a standard DICOM object that conforms to industry standards. The use of DICOM is an absolute necessity to ensure interoperability with current and future imaging systems. CDI will also provide easy-touse image processing tools that aid providers in analyzing images and making diagnoses. CDI will allow users to enhance the contrast and brightness of an image, reverse image values to show dental caries, and magnify specific areas of concern. Note: Modifications can be saved, but they are only saved as a copy of the original image. The original image will always remain untouched. # Indications for Use: 21 CFR 807 92(a)(5) Corporate Dental Imaging (CDI) Application™ is a software device for general dental diagnostic imaging. It controls capture, display, transfer, enhancement, and storage of X-ray dental images from digital imaging systems. # Technological Characteristics: 21 CFR 807 92(a)(6) Corporate Dental Imaging (CDI) Application™ is a software device that handles and manipulates digital medical images. The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and or printed. In general, a PACS (Picture Archiving and Communication System) is a medical imaging technology which provides storage of, and convenient access to, images from multiple modalities. Electronic images and reports are transmitted digitally via PACS; this eliminates the need to manually file, retrieve, or transport film jackets. The universal format for PACS image storage and transfer is DICOM 3.x (Digital Imaging and Communications in Medicine). Non-image data, such as scanned documents, may be incorporated using consumer industry standard formats like PDF (Portable Document Format), once encapsulated in DICOM. The subject device and predicate device are substantially equivalent in the areas of technical {5}------------------------------------------------ # 510(k) Summary of Safety characteristics, general function, application, and intended use. The new device does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices. Both systems have been developed to replace traditional film handling in radiology. The predicate and subject devices are substantially equivalent in the areas of general function, application, and intended use. Any difference between the two devices does not affect safety or efficacy. The predicate device and the new device are compared below: | Functionality | Predicate:<br>Xelis<br>DENTAL<br>(K102684) | Subject<br>Device: CDI<br>Application™ | If different, Impact on Safety and or<br>Efficacy | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Computer platform (minimum<br>requirements) | IBM<br>compatible<br>Memory:<br>1GB<br>Disk: 200MB<br>free hard<br>disk space | IBM<br>compatible<br>CPU: 1 Dual<br>Core<br>Processor<br>Memory: 8GB<br>RAM<br>HDD: C:<br>(OS)=140GB;<br>D: (Data)=2TB | Yes, there are differences.<br>Current computer platforms have these<br>added hardware features to support the<br>operating system, Microsoft Windows 64<br>bit 2008 R2 Enterprise, and they are for<br>the operating system to run efficiently.<br>The difference does not create new risks<br>and does not impact safety or efficacy as<br>explained in rationale below section 6.1.1<br>(a) below | | Computer Operating System | Microsoft<br>Windows<br>XP/Vista or<br>higher (32bit<br>& 64 bit)<br>Professional | Microsoft<br>Windows 64 bit<br>2008 R2<br>Enterprise | Yes, there are differences.<br>The Windows XP and Vista is no longer<br>supported by Microsoft. CDI uses the<br>Microsoft Windows 64 bit 2008 R2<br>Enterprise operating system. The system<br>has passed testing against pre-<br>determined criteria and the difference<br>does not create new risks and does not<br>impact safety or efficacy. | | Enterprise distribution of<br>images and data via Internet<br>or Intranet | Yes | Yes | No differences | | Networking Communications<br>Protocol - DICOM 3.x | Yes | Yes | No differences | | Image Compression: | No<br>compression<br>or Lossless | No<br>compression<br>or Lossless | No differences | | Standard Interfaces -<br>Ethernet | Yes | Yes | No differences | | Image Storage- Server & on<br>board hard disk size &<br>compression dependent. Can<br>store to Short or Long Term<br>Archives | Yes | Yes | No differences | | Multi-Planar Reformation<br>(MPR) - MPR into any classic<br>planes (axial, sagittal,<br>coronal, oblique planes) and<br>curved and free-draw planar<br>reformation. 3D image-guided<br>reformation on VR, MIP/MinIP | Yes | No | Yes, there are differences.<br>MPR is not a feature that is implemented<br>in the Medical Viewer and there are no<br>claims as such in the CDI device labeling.<br>This feature is not required to meet the<br>subject device intended use. The system<br>has been tested against pre-determined<br>criteria and passed. The difference does<br>not create new risks and does not impact<br>safety or efficacy. | | Virtual Endoscopy - Observe<br>the internal structure using<br>the virtual camera. | Yes | No | Virtual Endoscopy is not a feature that is<br>implemented in the CDI and there are no<br>claims in the CDI device labeling for this | | Functionality | Predicate:<br>Xelis<br>DENTAL<br>(K102684) | Subject<br>Device: CDI<br>Application™ | If different, Impact on Safety and or<br>Efficacy | | Reporting - Save, email,<br>DICOM send, or print to<br>standard Windows printers | Yes | No | Yes, there are differences.<br>Reporting - Save, email, DICOM send, or<br>print to standard Windows printers is not a<br>feature implemented in CDI and there are<br>no claims in the CDI device labeling for<br>this feature. This feature is not required to<br>meet the subject device intended use. The<br>system has been tested against pre-<br>determined criteria and passed. The<br>difference does not create new risks and<br>does not impact safety or efficacy. | | Move, rotate, flip and delete<br>template | Yes | Rotate and flip<br>images only | Yes there are differences.<br>The CDI application itself cannot move or<br>delete images. This feature is not required<br>to meet the subject device intended use.<br>The system has been tested against pre-<br>determined criteria and passed. The<br>difference does not create new risks and<br>does not impact safety or efficacy. | | Draw line | Yes | Yes | No differences | | Draw arrow | Yes | Yes | No differences | | Write text | Yes | Yes | No differences | | Measure distance | Yes | No | Yes, there are differences.<br>Measurement is not a feature in the CDI<br>application. This feature is not required to<br>meet the subject device intended use. The<br>system has been tested against pre-<br>determined criteria and passed. The<br>difference does not create new risks and<br>does not impact safety or efficacy. | | Measure angle | Yes | Yes | No differences | | Measure area | Yes | Yes | No differences | | Print image | Yes | Yes | No differences | | 2D Image Viewer- 2D image<br>viewer with real-time pan,<br>zoom, centering, windowing,<br>flip, rotation, free layouts, cine<br>and annotation Tools for<br>removal of obscuring<br>anatomy including ROI<br>drawing, ROI dilation and<br>erosion, paint, fence,<br>thresholding, 2D & 3D region<br>growing and balloon.<br>Analytical tools for measuring<br>distance, angle, ROI, grid,<br>table, profile, time-density<br>curve, pixel coordinate, and<br>scale bar. | Yes | 2D image<br>viewer - 2D<br>image viewer<br>has the<br>following: real-<br>time pan,<br>zoom,<br>centering,<br>windowing, flip,<br>rotation, free<br>layouts and<br>annotation<br>Tools. | Yes there are differences in the CDI<br>application which has reduced<br>functionality compared to the predicate<br>device. However, these features are not<br>required to meet the subject device<br>intended use. The system has been<br>tested against pre-determined criteria and<br>passed. The difference does not create<br>new risks and does not impact safety or<br>efficacy. | | MIP/MinIP - maximum<br>intensity projection | Yes | No | Yes there are differences in the CDI<br>application which has reduced | | Functionality | Predicate:<br>Xelis<br>DENTAL<br>(K102684) | Subject<br>Device: CDI<br>Application™ | If different, Impact on Safety and or<br>Efficacy | | | | | functionality compared to the predicate<br>device. However, this feature is not<br>required to meet the subject device<br>intended use. The system has been<br>tested against pre-determined criteria and<br>passed. The difference does not create<br>new risks and does not impact safety or<br>efficacy. | | Dental volume reformat<br>functions, such as arch/curve,<br>drawing nerve-canal, axis and<br>re-slice based on dental arch | Yes | No | Yes there are differences in the CDI<br>application which has reduced<br>functionality compared to the predicate<br>device. However, this feature is not<br>required to meet the subject device<br>intended use. The system has been<br>tested against pre-determined criteria and<br>passed. The difference does not create<br>new risks and does not impact safety or<br>efficacy. | | Dental implant planning,<br>which provides simulations of<br>implant placement, bone-<br>structure analysis and<br>location of the mandibular<br>canal. | Yes | No | Yes there is a difference. This is not a<br>feature in the CDI application. A feature<br>for simulations of implant placement,<br>bone-structure analysis and location of the<br>mandibular canal is not implemented or<br>claimed. This feature is not required to<br>meet the subject device intended use. The<br>system has been tested against pre-<br>determined criteria and passed. The<br>difference does not create new risks and<br>does not impact safety or efficacy. | | Intended Use1 | Picture<br>archiving<br>and<br>communication<br>on systems<br>(PACS) | Picture<br>archiving and<br>communication<br>systems<br>(PACS) | Yes, there are differences but both<br>systems are a picture archiving and<br>communication systems (PACS) is a<br>combination of hardware and software<br>dedicated to the short and long term<br>storage, retrieval, management,<br>distribution, and presentation of images.<br>Electronic images and reports are<br>transmitted digitally via PACS; this<br>eliminates the need to manually file,<br>retrieve, or transport film jackets. The<br>universal format for PACS image storage<br>and transfer is DICOM (Digital Imaging<br>and Communications in Medicine). There<br>are differences in the features and<br>functions used in the subject device when<br>compared to the predicate.<br>The subject device has less features.<br>However, the features and functions that<br>are no in the subject device are not | | Functionality | Predicate:<br>Xelis<br>DENTAL<br>(K102684) | Subject<br>Device: CDI<br>ApplicationTM | If different, Impact on Safety and or<br>Efficacy | | | | | required to meet the subject device<br>intended use. The system has been<br>tested against pre-determined criteria and<br>passed. The difference does not create<br>new risks and does not impact safety or<br>efficacy. | {6}------------------------------------------------ {7}------------------------------------------------ <sup>1 &</sup>quot;Intended Use" is the "Objective Intent" for the device of the persons legally responsible for labeling the device. The FDA defines "Intended Use" as: The term "intended uses" refers to the objective intent of the persons legally responsible for the labeling of the device. The intent is determined by their expressions or may be shown by the circumstances surrounding the distribution of the device. This objective intent may, for example, be shown by labeling claims, advertising matter, or oral or written statements by such representatives. It may be shown by the offering or the device, with the knowledge of such persons or their representatives, for a purpose for which it is neither labeled nor advertised. (21 CFR 801.4) {8}------------------------------------------------ #### Nonclinical Testing: The Corporate Dental Imaging (CDI) Application™ device has been assessed and tested at the factory and has passed all in-house testing criteria. The Verification & Validation Test Plan was designed to evaluate all input functions, output functions, and actions performed by the Corporate Dental Imaging (CDI) Application™ software in each operational mode and followed the process documented in the Validation Test Plan. Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met. If the device is installed by the manufacturer, integration and installations verification tests will be conducted against acceptance criteria prior to being released to the user facility. # Conclusion: 21 CFR 807 92(b)(1) The 510(k) Pre-Market Notification for the Corporate Dental Imaging (CDI) Application™ contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the predicate device. The subject device will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The subject and predicate devices are substantially equivalent in the areas of general function, application, and intended use. The subject device does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices. Nonclinical tests demonstrate that the device is safe and effective for its' intended use. Therefore, the Corporate Dental Imaging (CDI) Application™ is substantially equivalent to the predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...