The AZUR Peripheral Coil System is intended to reduce or block the rate of blood flow in vessels of the peripheral vasculature. It is intended for use in the interventional radiologic management of arteriovenous malformations, arteriovenous fistulae, aneurysms, and other lesions of the peripheral vasculature.
Device Story
The AZUR CX Peripheral Coil System (Detachable 35) is an implantable vascular embolization device used in interventional radiology. The system consists of a 3D spherical platinum alloy coil with an inner hydrogel core, attached to a delivery pusher. The device is delivered to the target lesion via a microcatheter. Once positioned, the user activates an external, separately sold AZUR Detachment Controller to release the coil from the pusher. The hydrogel core expands upon contact with blood, facilitating vessel occlusion. The device is used by physicians in clinical settings to manage peripheral vascular lesions, potentially benefiting patients by reducing or blocking blood flow to malformations, fistulae, or aneurysms.
Clinical Evidence
Bench testing only. Testing included simulated use, advance/retract performance, gel expansion, appendix strength, spring constant, and pusher sleeve retention. Extensive biocompatibility testing was performed according to ISO 10993 standards, including cytotoxicity, sensitization, irritation, hemocompatibility, systemic toxicity, genetic toxicology, and intramuscular implantation.
Technological Characteristics
Implantable 3D spherical coil made of platinum/tungsten alloy or platinum alloy-clad tantalum. Features a hydrophilic acrylic copolymer hydrogel core. Delivery system uses a polyolefin elastomer. Compatible with microcatheters (0.041"-0.047" ID). Sterile, single-use. Detachment via external electronic controller. Materials comply with ISO 10993 biocompatibility standards.
Indications for Use
Indicated for patients requiring reduction or occlusion of blood flow in peripheral vasculature lesions, including arteriovenous malformations, arteriovenous fistulae, and aneurysms.
Regulatory Classification
Identification
A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).
Predicate Devices
AZUR Peripheral HydroCoil Endovascular Embolization System - Detachable 18 (K123384)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human form.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 25, 2015
MicroVention, Inc. Cynthia Valenzuela Project Manager, Regulatory Affairs 1311 Valencia Avenue Tustin, California 92780
Re: K151358
Trade/Device Name: AZUR CX Peripheral Coil System - Detachable 35 Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: August 18, 2015 Received: August 20, 2015
Dear Cynthia Valenzuela:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151358
Device Name
AZUR CX Peripheral Coil System - Detachable 35
Indications for Use (Describe)
The AZUR Peripheral Coil System is intended to reduce or blood flow in vessels of the peripheral vasculature. It is intended for use in the interventional radiologic management of arteriovenous malformations, arteriovenous fistulae, aneurysms, and other lesions of the peripheral vasculature.
Type of Use (Select one or both, as applicable)
| <span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-------------------------------------------------------------------------------------|
| <span style="font-size:100%;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(K) Summary of Safety and Effectiveness
### Date Prepared: 11SEP2015
## 510(K) Summary
| Trade Name: | AZUR CX Peripheral Coil System - Detachable 3- |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------|
| Generic Name: | Vascular Embolization Device |
| Classification: | Class II, 21 CFR 870.3300 |
| Submitted By: | MicroVention, Inc.<br>1311 Valencia Avenue<br>Tustin, California U.S.A. |
| Contact: | Cynthia Valenzuela<br>Project Manager, Regulatory Affairs<br>MicroVention, Inc.<br>Direct: 714-247-8053<br>Cell: 949-413-0071 |
#### Predicate Device:
| Number | Description | Clearance Date |
|---------|-------------------------------------------------------------------------------------|----------------|
| K123384 | AZUR Peripheral HydroCoil<br>Endovascular Embolization System -<br>Detachable 18 | 28NOV2015 |
| K130577 | AZUR Peripheral HydroCoil<br>Endovascular Embolization System -<br>Pushable 18 & 35 | 17JUN2013 |
### Device Description
The AZUR CX Coils consist of implant coil made of platinum alloy with inner hydrogel core. The coils are designed in 3D spherical structure in various loop sizes and lengths. The AZUR CX Coil System (Detachable) consists of an implantable coil attached to a delivery pusher. The coil system is delivered to the treatment site through the microcatheter. The proximal end of the delivery pusher is inserted to the AZUR detachment controller. The detachment controller is activated by the user and this detaches the coil. The AZUR coils are designed for use with the AZUR Detachment Controller (Also known as AZUR Detachment Controller), specifically designed for coil detachment and is sold separately.
{4}------------------------------------------------
### Indication For Use
The intended use as stated in the product labeling is as follows:
The AZUR Peripheral Coil System is intended to reduce or block the rate of blood flow in vessels of the peripheral vasculature. It is intended for use in the interventional radiologic management of arteriovenous malformations, arteriovenous fistulae, aneurysms, and other lesions of the peripheral vasculature.
### Technological Comparison Chart
| Features | Cleared AZUR Pushable 18<br>& 35 (K130577) | Cleared AZUR CX (D18)<br>(K123384) | AZUR CX (D35)<br>Proposed device |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Design Attributes | | | |
| Coil Shape | Helical | 3D - Spherical | 3D - Spherical |
| Coil Filar OD (inch) | 0.00275 - 0.0050 | 0.00250 - 0.00400 | 0.0035 - 0.0050 |
| Primary Coil Wind OD<br>(inch) | 18: 0.0120<br>35: 0.0220 | 0.0140 - 0.0150 | 0.0290 |
| 3D Coil OD (mm) | 18: 2 – 10mm<br>35: 4 – 16mm | 4 - 20* | 4 - 20* |
| Restrained Coil Length<br>(cm) | 18: 2 – 20mm<br>35: 4 – 20mm | 13 - 40* | 7 - 39 |
| Pusher Length | 18: 2 - 10mm<br>35: 4 - 16mm | 175cm | 162cm |
| Materials | | | |
| Main Coil Wire | 18: Platinum (92%)/ Tungsten<br>(8%) | Platinum/Tungsten alloy<br>(92/8) | 4mm - 8mm:<br>Platinum/Tungsten alloy (92/8) |
| | 35: Platinum alloy-clad tantalum<br>[Platinum (90%)/Iridium (10%)] | | 10mm -20mm:<br>Platinum alloy-clad tantalum<br>[Platinum (90%)/Iridium (10%)] |
| | 35 (4 mm): Platinum (92%)/<br>Tungsten (8%) | | |
| Coil-to-Pusher Coupler | - | Platinum / Iridium (90/10) | No coupler |
| Adhesive | UV cure adhesive (DYMAX 1128-AM-VT) | UV cure adhesive (DYMAX<br>1128-AM-VT) | UV cure adhesive (DYMAX<br>1128-AM-VT) |
| Implant to the pusher | - | Polyolefin Elastomer | Polyolefin Elastomer |
| Stretch Resistant Filament | - | Polyolefin Elastomer | Polyolefin Elastomer |
| Hydrogel | Hydrophilic acrylic copolymer<br>(cross-linked copolymer of<br>polyethylene glycol diacrylamide<br>and acrylic acid) | Hydrophilic acrylic<br>copolymer (cross-linked<br>copolymer of polyethylene<br>glycol diacrylamide and<br>acrylic acid) | Hydrophilic acrylic copolymer<br>(cross-linked copolymer of<br>polyethylene glycol<br>diacrylamide and acrylic acid) |
| General | | | |
| Catheter Compatibility | 18: 0.021" - 0.022" ID<br>35: 0.041" - 0.047" ID | Compatible with<br>microcatheters having an<br>ID of > .019" | Compatible with catheters<br>having an ID of 0.041"-0.047" |
| Method of Supply | Sterile, single use | Sterile, single use | Sterile, single use |
| Package Configuration | Sealed in a poly (ethylene<br>terephthalate) pouch coated with<br>silica or aluminum oxide and<br>placed in a shipping carton | Dispenser coil, pouch &<br>shipping carton | Dispenser coil, pouch &<br>shipping carton |
{5}------------------------------------------------
# Verification of Test Summary
| Bench Testing | Result |
|-------------------------|--------|
| Simulated use | Passed |
| Advance/Retract | Passed |
| Gel Expansion | Passed |
| Appendix Strength | Passed |
| Spring Constant | Passed |
| Pusher Sleeve Retention | Passed |
# Biocompatibility Coil Implant Segment - Biocompatibility Summary
| Test Method | Standard |
|-----------------------------------------------|--------------|
| Cytotoxicity | |
| MEM Elution Test | ISO 10993-5 |
| ISO Cell Culture Agar Overlay | ISO 10993-5 |
| Sensitization | |
| Sensitization-Guinea Pig Maximization Test | ISO 10993-10 |
| Irritation | |
| ISO Intracutaneous Reactivity Evaluation Test | ISO 10993-10 |
{6}------------------------------------------------
# Coil Implant Segment - Biocompatibility Summary
| Test Method | Standard |
|-----------------------------------------|--------------|
| Hemocompatibility | |
| Hemolysis | ISO 10993-4 |
| Prothrombin Time Assay - ISO | ISO 10993-4 |
| Systemic Toxicity | |
| Systemic toxicity (IV injection) | ISO 10993-11 |
| Rabbit Pyrogen Test (material mediated) | ISO 10993-11 |
| Test Method | Standard |
|---------------------------------------------|-------------|
| Genetic Toxicology | |
| Bacteria Reverse Mutation Assay (Ames Test) | ISO 10993-3 |
| Intramuscular Implantation | |
| 7-day Muscle Implantation | ISO 10993-6 |
| 13-week Intramuscular Implantation Test | ISO 10993-6 |
| 26-week Intramuscular Implantation Test | ISO 10993-6 |
{7}------------------------------------------------
# Biocompatibility Summary for Delivery Pusher Segment
| Test Method | Standard |
|-----------------------------------------------|--------------|
| Cytotoxicity | |
| MEM Elution Test | ISO 10993-5 |
| ISO Cell Culture Agar Overlay | ISO 10993-5 |
| Sensitization | |
| Sensitization-Guinea Pig Maximization Test | ISO 10993-10 |
| Irritation | |
| ISO Intracutaneous Reactivity Evaluation Test | ISO 10993-10 |
| Hemocompatibility | |
| Blood compatibility Evaluation (Hemolysis) | ISO 10993-4 |
| Prothrombin Time Assay - ISO | ISO 10993-4 |
| Systemic Toxicity | |
| Systemic toxicity<br>(IV injection) | ISO 10993-11 |
| Rabbit Pyrogen Test (material mediated) | ISO 10993-11 |
{8}------------------------------------------------
### Summary of Substantial Equivalence:
The data presented in this submission demonstrates the technological similarity and equivalency of the AZUR Peripheral Coils System - CX Detachable 35 Coils when compared with the predicate devices, MicroVention, Inc. AZUR CX Detachable 18 (K1223384) and AZUR Pushable 18 & 35 (K130577).
- The device, ●
- Has the same intended use, ●
- Uses the same operating principle, ●
- Incorporates the same basic design, ●
- Uses similar construction and material, ●
- Are packaged and sterilized using the same material and processes. .
In summary, the AZUR Peripheral Coil System – CX Detachable 35 Coils described in this submission is, in our opinion equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.