K151347 · Thompson Surgical Instruments · KCT · Oct 16, 2015 · General Hospital
Device Facts
Record ID
K151347
Device Name
MIS Spine Frame Instrument Case
Applicant
Thompson Surgical Instruments
Product Code
KCT · General Hospital
Decision Date
Oct 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Thompson Surgical Instruments MIS Spine Frame Instrument Case is intended for use in healthcare facilities to store, transport, organize, and sterilize surgical instruments such as retractor blades, retractor frame components, lighting accessories, and blade or frame accessories between uses when used in combination with an FDA cleared sterilization wrap. Validated sterilization parameters include pre-vacuum steam and gravity.
Device Story
MIS Spine Frame Instrument Case (model 50000MSF) is a reusable container for storage, transport, and steam sterilization of surgical instruments (retractor blades, frame components, lighting accessories). Device consists of aluminum base, lid, and lift-out tray with silicone, aluminum, stainless steel, and nylon inserts. Features surface perforations for steam penetration; no gaskets, valves, or filters. Used in healthcare facilities; operated by clinical staff. Instruments are placed inside, the case is wrapped in an FDA-cleared sterilization wrap, and subjected to validated pre-vacuum or gravity steam sterilization cycles. Output is a sterile, organized set of instruments ready for surgical use. Benefits include protection of instruments during transport and maintenance of sterility post-processing.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing: sterilization efficacy (biological indicators showing 6-log reduction), cleaning validation, and biocompatibility (MEM elution test, non-cytotoxic).
Indicated for use in healthcare facilities to store, transport, organize, and sterilize surgical instruments, including retractor blades, frame components, and lighting accessories, when used with an FDA-cleared sterilization wrap.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
{0}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-000
October 16, 2015
Thompson Surgical Instruments, Inc. Stephanie Myers, Product Manager 10170 East Cherry Bend Road, Traverse City, MI 49684
Re: K151347
Trade/Device Name: MIS Spine Frame Instrument Case Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: II Product Code: KCT Dated: September 15, 2015 Received: September 18, 2015
Dear Ms. Myers,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/0/Picture/10 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle are three stylized human profiles facing to the right, stacked on top of each other.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151347
Device Name MIS Spine Frame Instrument Case
Indications for Use (Describe)
Thompson Surgical Instruments MIS Spine Frame Instrument Case is intended for use in healthcare facilities to store, transport, organize, and sterilize surgical instruments such as retractor frame components, lighting accessories, and blade or frame accessories between used in combination with an FDA cleared sterilization wrap. Validated sterilization parameters include pre-vacuum steam and gravity.
| CYCLE | CYCLE TEMP | EXPOSURE TIME | DRY TIME |
|-----------|---------------|---------------|------------|
| Prevacuum | 132°C (270°F) | 4 Minutes | 30 Minutes |
| Gravity | 121°C (250°F) | 30 Minutes | 30 Minutes |
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@)fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Submitter: Thompson Surgical Instruments, Inc.
Product: MIS Spine Frame Instrument Case
Image /page/3/Picture/2 description: The image shows the logo for Thompson Surgical Instruments. The logo has the company name in bold, black letters, with the words "SURGICAL INSTRUMENTS" in a smaller font size below the name. To the right of the company name is a blue symbol. Below the logo is the text "K151347".
# 510(k) Summary
| Submitter Name: | Thompson Surgical Instruments, Inc. | | | | | | | | | | | | |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|
| Submitter Address: | 10170 E Cherry Bend Rd<br>Traverse City, MI 49684 | | | | | | | | | | | | |
| Contact Person: | Stephanie Myers, Research & Development /Product Manager | | | | | | | | | | | | |
| Phone Number: | 231-922-5170 | | | | | | | | | | | | |
| Fax: | 231-922-0174 | | | | | | | | | | | | |
| Date Prepared: | September 14, 2015 | | | | | | | | | | | | |
| Device Trade Name: | MIS Spine Frame Instrument Case | | | | | | | | | | | | |
| Classification Name: | Sterilization Wrap Containers, Cassettes, and Accessories | | | | | | | | | | | | |
| Device Class: | Class II | | | | | | | | | | | | |
| Classification<br>Regulation: | 21 CFR 880.6850 | | | | | | | | | | | | |
| Product Code: | KCT | | | | | | | | | | | | |
| Predicate Device: | The instrument case described in this submission was compared and found to be<br>substantially equivalent to the K3131170 Sonicision™ Sterilization Tray<br>manufactured by Covidien. K3131170 was cleared August 28th, 2013. | | | | | | | | | | | | |
| Statement of Intended<br>Use: | Thompson Surgical Instruments MIS Spine Frame Instrument Case is intended for<br>use in healthcare facilities to store, transport, organize, and sterilize surgical<br>instruments such as retractor blades, retractor frame components, lighting<br>accessories, and blade or frame accessories between uses when used in combination<br>with an FDA cleared sterilization wrap. Validated sterilization parameters include<br>pre-vacuum steam and gravity.<br>CYCLECYCLE TEMPEXPOSURE TIMEDRY TIME Prevacuum132°C (270ºF)4 Minutes30 Minutes Gravity121ºC (250ºF)30 Minutes30 Minutes | | | | | | | | | | | | |
| Device Description: | Thompson Surgical Instruments “MIS Spine Frame Instrument Case,” model<br>50000MSF, is intended to hold surgical instruments such as retractor blades,<br>retractor frame components, lighting accessories, and blade or frame accessories<br>during storage, transportation, and sterilization, as indicated. It is composed of<br>materials that can be reused with steam sterilization methods and DCG graphics may<br>be used. It has a distribution of perforations on the outer surfaces and between case<br>levels that allow for optimal steam penetration. Vent to volume is .110. It is reusable | | | | | | | | | | | | |
{4}------------------------------------------------
Submitter: Product: Thompson Surgical Instruments, MIS Spine Frame Instrument Case Inc. and provided to the customer in a non-sterile conditions. This case can be described as an accessory for the Thompson Retractor. Technological and The MIS Spine Frame Instrument Case was found to be similar to the predicate. Performance Three fundamental similarities are: Characteristics: Basic design: Both the MIS Spine Frame Instrument Case and the predicate 1. have a basic lid/base design with latches, handles, perforations, and contoured inserts or sections for containing items for sterilization, storage, and transport. General size, shape, weight, and materials are not identical but are similar. 2. Role in sterile barrier system: The MIS Spine Frame Instrument Case and the predicate must be enclosed with a qualified FDA cleared sterilization wrap to maintain sterility. Neither contains gaskets, valves, or filters. 3. Fundamental Technology: The MIS Spine Frame Instrument Case and the predicate allow the sterilant (steam) to penetrate and render its contents sterile by relying on surface perforations.
> Primary differences are related to minor variances in size, shape, intended contents, surface perforations, and overall look. However, performance data demonstrate that the differences do not adversely affect safety and effectiveness. Please see further comparison below.
| | Proposed: MIS Spine Frame<br>Instrument Case | Predicate: Sonicision™ Sterilization Tray K131170 |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | Thompson Surgical<br>Instruments MIS Spine Frame<br>Instrument Case is intended<br>for use in healthcare facilities<br>to store, transport, organize,<br>and sterilize medical devices<br>and other instrumentation<br>between uses when used in<br>combination with a FDA<br>cleared sterilization wrap. | To encase and protect<br>reusable batteries and<br>generators of the Sonicision<br>system during sterilization<br>and storage. Compatible<br>sterilization systems are<br>indicated as follows:<br>STERRATD® 100s, STERRAD®<br>NX ®, STERRAD® 100NX®,<br>STERIS Amsco® V-PRO® 1,<br>STERIS Amsco® V-PRO® 1<br>Plus, STERIS Amsco® V-PRO®<br>max |
| Design<br>Characteristics: | | |
| -Composition: | Base, lift out tray, and lid | Base and lid |
# Comparison to Predicate Devi
{5}------------------------------------------------
Submitter:
Inc.
Thompson Surgical Instruments,
## Product: MIS Spine Frame Instrument Case
Image /page/5/Picture/2 description: The image contains the logo for Thompson Surgical Instruments. The logo features the company name in bold, sans-serif font, with "Thompson" on the top line and "SURGICAL INSTRUMENTS" on the bottom line. To the right of the name is a stylized blue symbol. Below the logo is the alphanumeric code "K151347".
| -Intended content<br>(max load): | Medical devices/instruments<br>weighing no less than 25lbs<br>total including the weight of<br>the MIS Spine Frame<br>Instrument Case. | One Sonicision generator and<br>one battery pack |
|--------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| -Inserts: | Yes | Yes |
| -Handles: | Yes | Yes |
| -Latches: | Yes | Yes |
| -Reusable: | Yes | Yes |
| Materials: | | |
| -Lid/base/lift out<br>tray not including<br>inserts: | Aluminum | Polysulfone |
| -Inserts: | Silicone, Aluminum, Stainless<br>Steel, Nylon | Silicone |
| -Latch: | Stainless steel | Polysulfone |
| Assembled<br>Dimensions: | 11" x 23.5" | 5.9" X 8.1" |
| Weight containing<br>max load: | 25 lbs | 1.58 lbs |
| Percent of surface<br>perforations on lid,<br>base, lift out tray: | Lid has 15.7%, lift out tray has<br>19.2%, and base has 13.6% | 9.7% |
# Non-Clinical Performance Studies:
The MIS Spine Frame Instrument Case was evaluated in accordance with applicable clauses/criteria specified in AAMI 5T77:2006. Test evidence supporting sterilization efficacy, cleaning, and biocompatibility are provided in the submission. A brief discussion of tests used to support the conclusion of substantial equivalence with the predicate device is provided below.
Sterilization efficacy: The MIS Spine Frame Instrument Case and the predicate containing the maximum load were inoculated with biological indicators placed in areas of the trays deemed to be most difficult for sterilant (steam) to penetrate. Following inoculation, the trays were wrapped in FDA cleared sterilization wraps.
{6}------------------------------------------------
| Submitter:<br>Thompson Surgical Instruments, Inc. | Product:<br>MIS Spine Frame Instrument Case | Thompson<br>SURGICAL INSTRUMENTS<br>K151347 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| They trays were then exposed to the sterilization cycles (Steam sterilization for the<br>MIS Spine Frame Instrument Case and STERRAD or STERIS sterilization for the<br>predicate). The results for both trays indicated that half cycles provided a six log<br>reduction of the indicator organism. | | |
| Cleaning: The MIS Spine Frame Instrument Case was tested for cleanability. The<br>results indicated a pass. Supporting evidence is included in the submission. | | |
| Biocompatibility: The MIS Spine Frame Instrument Case is considered non-cytotoxic<br>as based on MEM elution test results. Supporting evidence is found in the<br>submission. | | |
| Clinical Performance<br>Studies: | This premarket notification report does not rely on the assessment of clinical<br>performance data to demonstrate substantial equivalence. | |
| Conclusion: | The MIS Spine Frame Instrument Case is substantially equivalent to the Sonicision™<br>Sterilization Tray, K131170. None of the differences between the MIS Spine Frame<br>Instrument Case and the predicate change the intended use or raise new questions<br>of safety or effectiveness. | |
{7}------------------------------------------------
Submitter: Thompson Surgical Instruments, Inc.
Product: MIS Spine Frame Instrument Case
Image /page/7/Picture/2 description: The image shows the logo for Thompson Surgical Instruments. The logo features the word "Thompson" in bold, black letters, with the words "SURGICAL INSTRUMENTS" in smaller, thinner letters underneath. To the right of the word "Thompson" is a blue symbol that appears to be a stylized "T". Below the logo is the alphanumeric code "K151347".
5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.