K151344 · Unicare Biomedical, Inc. · NPK · Nov 19, 2015 · Dental
Device Facts
Record ID
K151344
Device Name
Cytoflex Tefguard Ti-Enforced Membrane
Applicant
Unicare Biomedical, Inc.
Product Code
NPK · Dental
Decision Date
Nov 19, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3930
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cytoflex® Tefguard® Ti-Enforced membrane is a temporarily, non-resorbable, implantable material for use as a spacemaking barrier in the treatment of periodontal defects and augmentation of alveolar ridge in accordance with guided tissue regeneration principle.
Device Story
Cytoflex Tef-Guard Ti-Enforced membrane is a passive, non-resorbable, implantable barrier; consists of titanium frame enclosed within two layers of ePTFE. Surgically placed under muco-periosteum by dental professionals; acts as space-making barrier to contain bone grafting material; prevents gingival tissue cell migration into bony defects; facilitates neovascularization and bone repair. Implanted for 30 days up to 6 months; single-use; provided sterile. Benefits include improved environment for bone regeneration and graft stability.
Clinical Evidence
Bench testing only. No clinical data required. Performance evaluated via tensile loading, bending, flexibility, and delamination testing compared to primary predicate. Biocompatibility verified per ISO 10993-5, 6, 7, 10, and 11. Sterilization validated per ISO 11135.
Technological Characteristics
Multi-layer non-resorbable membrane; ePTFE exterior layers with internal titanium frame. Thickness 0.25mm; density 3.5 g/dm². Sterile (EtO gas, SAL <10^-6). Single-use. Biocompatible per ISO 10993 standards.
Indications for Use
Indicated for patients requiring guided tissue regeneration for the treatment of periodontal defects and alveolar ridge augmentation.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
Cytoplast® Titanium reinforced high density PTFE membranes (K972278)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure in profile, composed of three overlapping silhouettes. The silhouettes are arranged in a way that suggests movement or progress. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 19,2015
Unicare Biomedical, Inc. Stan Yang Vice President 23011 Moulton Parkway, J-11 Laguna Hills, California 92653
Re: K151344
Trade/Device Name: Cytoflex Tefguard Ti-enforced Eptfe Membrane Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: Class II Product Code: NPK Dated: October 14, 2015 Received: October 19, 2015
Dear Stan Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno DDS, MA
Erin Keith Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
10(k) Number (if known) K151344
#### Device Name
Cytoflex® Tefguard® Ti-Enforced membrane
Indications for Use (Describe)
Cytoflex® Tefguard® Ti-Enforced membrane is a temporarily, non-resorbable, implantable material for use as a spacemaking barrier in the treatment of periodontal defects and augmentation of alveolar ridge in accordance with guided tissue regeneration principle.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# SECTION 6
# 510(K) SUMMARY
{4}------------------------------------------------
# 510(k) Summary (As Required by 21CFR 807.92 (c))
- 1. Submitter
| Name: | Unicare Biomedical |
|----------|-----------------------------------------------------------------|
| Address: | 23011 Moulton Parkway, J-11, Laguna Hills, California 92653 USA |
| Tel: | Tel: 949-305-1770; Fax: 949-334-1338 |
| Contact: | Stan Yang, PhD |
| Date: | June 15, 2015 |
| 2. Trade/Device Name: | Cytoflex® Tef-Guard® Ti-Enforced Membrane |
|-----------------------|-------------------------------------------|
| Common Name: | Non-resorbable barrier membrane |
| Regulation Number: | 21 CFR 872.3930 |
| Regulation Name: | Barrier, Synthetic, Intraoral |
| Regulation Class: | II |
| Product Code: | NPK |
#### 3. Predicate Devices
Primary predicate device -Cytoplast® Titanium reinforced high density PTFE membranes (K972278; Osteogenics Biomedical, Inc.) Reference predicate device -Neoss Ti Reinforced Membrane (K143327; Neoss Ltd)
- 4. Device Description
Cytoflex® Tef-Guard® Ti-Enforced membranes are a multi-layer, non-resorbable membrane intended to be surgically placed under the muco-periosteum to aid in tissue regeneration in accordance with the guided tissue regeneration principle. It is a passive, non-load bearing material with a titanium frame enclosed within two exterior layers of ePTFE material. The titanium reinforcement is intended for space and shape maintenance to contain bone grafting material and to minimize graft migration during wound healing. The membranes are designed to reduce the migration and establishment of gingival tissue derived cells into bony defects thus providing a more
{5}------------------------------------------------
favorable environment for neovascularization and bone derived cells to repopulate and repair the defect. The membranes are intended to be submerged and implanted for more than 30 days and up to 6 months. The membrane is supplied sterile and available in a variety of shapes and sizes for single use only.
| Description / Model | Dimension | Thickness |
|------------------------------------------------------|-----------|-----------|
| Cytoflex® Tef-Guard® Ti-Enforced<br>membrane / T-001 | 11x21mm | 0.25mm |
| Cytoflex® Tef-Guard® Ti-Enforced<br>membrane / T-002 | 19x26mm | 0.25mm |
| Cytoflex® Tef-Guard® Ti-Enforced<br>membrane / T-003 | 23x29mm | 0.25mm |
| Cytoflex® Tef-Guard® Ti-Enforced<br>membrane / T-004 | 32x40mm | 0.25mm |
### 5. Indication
Cytoflex® Tef-Guard® Ti-Enforced membrane is a temporarily, non-resorbable, implantable material for use as a space-making barrier in the treatment of periodontal defects and augmentation of alveolar ridge in accordance with guided tissue regeneration principle.
#### Comparison with Primary Predicate Device and Reference Predicate Device Q.
| Device Name /<br>Comments | Cytoflex<br>Tef-Guard<br>Ti-Enforced<br>Membrane | Cytoplast<br>Titanium<br>Reinforced<br>high density<br>PTFE<br>Membrane* | Neoss Ti<br>Reinforced<br>Membrane** | Comments |
|---------------------------|----------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| 510K ID | K 151344 | K 972278 | K 143327 | - |
| Indications for<br>use | Temporarily<br>implantable<br>material (non-<br>resorbable) for<br>use as a space-<br>making barrier<br>in the | Temporarily<br>implantable<br>material (non-<br>resorbable) for<br>use as a space-<br>making barrier<br>in the treatment | An implantable<br>temporary non-<br>resorbable<br>membrane for<br>use as a spacer<br>creation barrier<br>in the treatment | Substantially<br>Equivalent<br>indications<br>for use |
{6}------------------------------------------------
| | treatment of<br>periodontal<br>defects and<br>alveolar ridge<br>augmentation | of periodontal<br>defects | of local oral<br>defects in<br>conjunction<br>with tissue<br>regeneration or<br>augmentation. | |
|---------------|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|----------------------------------------------|
| Implant sites | periodontal<br>defects &<br>alveolar ridge | periodontal<br>defects | Oral bony<br>cavity | Substantially<br>equivalent<br>implant sites |
| Design | A variety of<br>configurations.<br>May be<br>trimmed and<br>shaped to fit<br>bony defect. | A variety of<br>configurations.<br>May be<br>trimmed and<br>shaped to fit<br>bony defect. | A variety of<br>configurations.<br>May be<br>trimmed and<br>shaped to fit<br>bony defect. | Substantially<br>equivalent<br>design |
| Material | ePTFE<br>membranes<br>reinforced with<br>a Titanium<br>frame | High density<br>PTFE<br>membranes<br>reinforced with<br>a Titanium<br>frame | ePTFE and<br>dense PTFE<br>membranes<br>reinforced with<br>a Titanium<br>frame | Substantially<br>Equivalent<br>material |
| Thickness | 0.25mm | 0.23mm | 0.3mm | Substantially<br>Equivalent |
| Density | 3.5 g/dm² | 4.0 g/dm² | 3.5 g/dm² | Substantially<br>Equivalent |
| Biocompatible | Yes | Yes | Yes | Substantially<br>Equivalent |
| Sterility | Sterile by EtO<br>gas, SAL <10-6 | Sterile | Sterile | Substantially<br>Equivalent |
*Primary Predicate Device; ** Reference Predicate Device
### 7. Device Characteristics
As summarized in the comparison Table above, the designs of Cytoflex® Tef-Guard® Ti-Enforced membranes are substantially equivalent to the primary predicate device, Cytoplast Titanium Reinforced high density PTFE Membrane, and reference predicate device, Neoss Ti Reinforced Membrane. The subject and the predicate devices are passive, non-load bearing membranes for use to maintain space and shape to contain bone grafting material and to minimize graft migration during wound healing. The membranes are designed to reduce the migration and establishment of gingival tissue derived cells
{7}------------------------------------------------
into bony defects thus providing a more favorable environment for neovascularization and bone derived cells to repopulate and repair the defect in accordance with the guided tissue regeneration principle to repair periodontal defects and/or augment alveolar ridge. The devices are for single use only and are provided sterile with a variety of configurations for different bony size defects.
### 8. Performance Testing
The performance of Cytoflex® Tef-Guard® Ti-Enforced membrane was evaluated against the primary predicate device for tensile loading, bending, flexibility and delaminating potential. The test results demonstrate that the subject device is substantially equivalent to the primary predicate device in functional performance.
Cytoflex® Tef-Guard® Ti-Enforced membrane is made of substantially equivalent biocompatible materials as the predicate devices. The subject device was tested according to the following ISO standards, and was found to meet the requirements of the pre-defined acceptance criteria. The results of the above testing demonstrate that the Cytoflex® Tef-Guard® Ti-Enforced membrane is substantially equivalent in safety and effectiveness, and its performance meets the requirements of its predefined acceptance criteria and intended use.
- ISO 11135 (2014) Sterilization of health-care products Ethylene oxide -. Requirements for the development, validation and routine control of a sterilization process for medical devices
- ISO 10993-5 (2009) Biological evaluation of medical devices Part 5: Tests for . in vitro Cytotoxicity
- . ISO 10993-6 (2007) - Biological evaluation of medical devices - Part 6: Tests for local effects after implantation
- ISO 10993-10 (2010) Biological evaluation of medical devices Part 10: Tests . for irritation and skin sensitization
- ISO 10993-11 (2006) Biological evaluation of medical devices Part 11: Tests . for systemic toxicity in Mice
{8}------------------------------------------------
- . ISO 10993-11 (2006) - Biological evaluation of medical devices - Part 11: Tests for systemic toxicity in Rats following subcutaneous implantation, 4 weeks
- ISO 10993-7 (2008R2012) Biological evaluation of medical devices Part 7: . Ethylene oxide sterilization residuals
- 9. Clinical Testing
Clinical data was not required to establish the substantial equivalence of the Cytoflex® Tef-Guard® Ti-Enforced Membrane.
### 10. Conclusion
Cytoflex® Tef-Guard® Ti-Enforced Membranes are substantially equivalent to the primary predicate device in that they have similar indications for use and exhibit substantially equivalent technology and device characteristics, including thickness, dimension, density, design and material composition when compared with the predicate devices. The test results demonstrate that Cytoflex® Tef-Guard® Ti-Enforced Membranes are biocompatible and perform substantially equivalent to the primary predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.