# Nexus Universal, Nexus Universal Chroma (K151332)

_Sybron Dental Specialties · EMA · Aug 17, 2015 · Dental · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151332

## Device Facts

- **Applicant:** Sybron Dental Specialties
- **Product Code:** [EMA](/productcode/EMA.md)
- **Decision Date:** Aug 17, 2015
- **Decision:** SESE
- **Submission Type:** Abbreviated
- **Regulation:** 21 CFR 872.3275
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Nexus Universal/Nexus Universal Chroma is indicated for cementation of all indirect restorations including ceramic, resin, and metal-based inlays/onlays, crowns, bridges, posts, and veneers*. Additional indications include core build-ups and cementation of crowns to implants. * Adhesive application on the preparation is required for veneer cementation using Nexus Universal Chroma.

## Device Story

Nexus Universal and Nexus Universal Chroma are self-adhesive, dual-cure resin cements used by licensed dental professionals in clinical settings. The device is supplied in dual-barrel syringes with automix tips for direct application into restorations or tooth preparations. It functions as a self-etch, self-adhesive system, eliminating separate etching/priming steps. The Chroma version includes a redox-based gel-state color indicator; the paste transitions from pink to clear to signal the optimal window for excess cement removal. The device provides permanent cementation for indirect restorations and core build-ups. By improving bond strength to dentin and compatibility with adhesives in self-cure mode, the device facilitates secure restoration placement, potentially increasing the longevity and stability of dental prosthetics.

## Clinical Evidence

No clinical data was performed. Substantial equivalence is supported by non-clinical bench testing, including bond strength to various substrates, radiopacity, film thickness, diametral strength, consistency, heat generation, water sorption, and gel-state color transition. Biocompatibility testing was conducted per ISO 10993 standards (parts 1, 3, 5, 6, 10, 11) and ISO 4049.

## Technological Characteristics

Self-adhesive, dual-cure (redox initiator for self-cure; photoinitiator for light-cure) resin cement. Paste/paste formulation. Packaged in dual-barrel syringes with automix/curved dispensing tips. Materials comply with ISO 4049 and ISO 10993 biocompatibility standards. Features include a redox-based gel-state color indicator (Chroma version only).

## Regulatory Identification

Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.

## Predicate Devices

- Maxcem 2 ([K073209](/device/K073209.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 17, 2015

Sybron Dental Specialties Ms. Courtney Clark Regulatory Affairs Manager 1717 W. Collins Avenue Orange, CA 92867

Re: K151332

Trade/Device Name: Nexus Universal, Nexus Universal Chroma Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: May 18, 2015 Received: May 19, 2015

Dear Ms. Clark:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/1/Picture/8 description: The image shows a signature with the name "Tina Kiang -S" written in a simple, sans-serif font. The first name, "Tina," is positioned above the last name, "Kiang," with a hyphen and the letter "S" following the last name. To the left of the name is a stylized, looping signature mark that appears to be a series of connected curves and lines.

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151332 Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

Device Name

Nexus Universal/ Nexus Universal Chroma

Indications for Use (Describe)

Nexus Universal/Nexus Universal Chroma is indicated for cementation of all indirect restorations including ceramic, resin, and metal-based inlays/onlays, crowns, bridges, posts, and veneers*. Additional indications include core build-ups and cementation of crowns to implants.

* Adhesive application on the preparation is required for veneer cementation using Nexus Universal Chroma.

| Type of Use (Select one or both, as applicable)                                                              |                                                             |
|--------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| <div> <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D)  </div> | <div>   Over-The-Counter Use (21 CFR 801 Subpart C)  </div> |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## SECTION 5. 510(k) SUMMARY for Nexus Universal / Nexus Universal Chroma

Image /page/3/Picture/1 description: The image shows the logo for Sybron Dental Specialties. The logo features the letters "sds" in a stylized, bold, blue font, with a curved line above the letters. Below the letters, the words "Sybron Dental Specialties" are written in a smaller, less bold font, also in blue.

### Nexus Universal /Nexus Universal Chroma

- Submitter Information: 1. Sybron Dental Specialties
1717 W. Collins Ave. Orange CA, 92687

| Contact Person:   | Courtney Clark |
|-------------------|----------------|
| Telephone Number: | 714-516-7426   |
| Fax Number:       | 714-516-7472   |

Date Prepared: May 18, 2015

- 2. Device Name:
	- Proprietary Name: Nexus Universal and Nexus Universal Chroma .
	- . Classification Name: Dental Cement w/out Zinc-Oxide Eugenol

Cement, Dental

- 872.3275 CFR Number: ●
- Device Class: II .
- Product Code: EMA .
- 3. Predicate Device:

The proposed Nexus Universal and Nexus Universal Chroma are substantially equivalent to the legally marketed predicate device. Maxcem 2 (K073209), cleared on January 29, 2008, product codes MZW, EMA.

- 4. Description of Device:
Nexus Universal / Nexus Universal Chroma is indicated for cementation of all indirect restorations including ceramic, resin, and metal-based inlays/onlays, crowns, bridges, posts, and veneers*. Additional indications include core build-ups and cementation of crowns to implants.

* Adhesive application on the preparation is required for veneer cementation using Nexus Universal and Nexus Universal Chroma.

{4}------------------------------------------------

Nexus Universal and Nexus Universal Chroma are offered in paste/paste formulations for use as self-etch, self-adhesive resin cements or as bonded resin cements. These cements are considered permanent and not temporary.

The dual-cure material is packaged in dual barrel syringes with single-use automix tips and optional curved dispensing tips to allow the user to deliver the desired volume of cement directly into the restoration and/or tooth preparation.

| Accessories Used with Nexus<br>Universal / Nexus Universal Chroma | Manufacturer of Accessory                                                 |
|-------------------------------------------------------------------|---------------------------------------------------------------------------|
| Dual barrel syringe                                               | Sulzer Mixpac AG<br>Ruetistrasse 7<br>Haag Sankt gallen, Switzerland 9469 |
| Curved dispensing tips                                            | Sulzer Mixpac AG<br>Ruetistrasse 7<br>Haag Sankt gallen, Switzerland 9469 |
| Auto-mix tips                                                     | Sulzer Mixpac AG<br>Ruetistrasse 7<br>Haag Sankt gallen, Switzerland 9469 |

Nexus Universal and Nexus Universal Chroma are offered in multiple shades. The difference between Nexus Universal and Nexus Universal Chroma is that Nexus Universal Chroma offers a gel-state color indicator, which visually displays the optimal time to remove the excess cement.

#### న. Statement of Indications for Use:

Nexus Universal/Nexus Universal Chroma is indicated for cementation of all indirect restorations including ceramic, resin, and metal-based inlays/onlays, crowns, bridges, posts, and veneers*. Additional indications include core build-ups and cementation of crowns to implants.

* Adhesive application on the preparation is required for veneer cementation using Nexus Universal and Nexus Universal Chroma.

#### 6. Description of Substantial Equivalence:

### Technological Characteristics

Nexus Universal and Nexus Universal Chroma, like their predicate Maxcem 2 (K073209), are self-adhesive paste/paste dual-cure permanent resin cements. Thev provide a simplified one-step cementation procedure by combining the etching, priming, bonding and cementing steps into a single step. This is achieved by incorporating adhesive monomers along with an acid tolerant redox initiator system into a traditional resin cement.

There are two curing mechanisms associated with the proposed Nexus Universal and Nexus Universal Chroma self-adhesive resin cements: self-curing (also referred to as dark-curing) mechanism by a redox initiator system and light-curing by a photoinitiator.

{5}------------------------------------------------

The proposed Nexus Universal and Nexus Universal Chroma offer the following changes over the predicate, Maxcem 2 (K073209):

- 1. A gel-state color indicator (Nexus Universal Chroma only) provides clinicians with a visual cue indicating the optimal timing to remove the gelled excess cement
- 2. Increased bond strength to dentin
- 3. Increased bond compatibility with adhesive even in self-cure mode (also referred to as dark-cure mode)

The proposed Nexus Universal Chroma has a gel-state indicator that is a redox color indicator that has an initial pink color for the mixed paste before gelation. The pink color will fade away once the mixed paste reaches the gel state, the ideal timing for removing excess cement. For both the Nexus Universal and Nexus Universal Chroma, the increased bond strength to dentin and the increased bond compatibility with the adhesive in self-cure mode is achieved by modifying the adhesive property of the resin matrix and wetting ability of the pastes along the acid tolerant redox initiator system used in the proposed.

## Non-Clinical Performance Data

Non-clinical performance data included testing for bond strength to various substrates, radiopacity, film thickness, diametral strength, consistency, amount of heat generated during setting, water sorption, dark-cure bond compatibility, and gel-state indicator color transition.

Gel/set time testing and biocompatibility testing were also performed. The data analyzed from the various tests substantiate that the proposed Nexus Universal and Nexus Universal Chroma are substantially equivalent to the predicate, Maxcem 2 (K073209). The following standards were utilized for the non-clinical performance testing of the proposed:

- ISO 10993-1: 2009 Biological evaluation of medical devices
- ISO 10993-3:2003 Biological Evaluation of Medical Devices- Part 3: Tests for ● Genotoxicity, Carcinogenicity, and Reproductive Toxicity
- ISO 10993-5:2009 Biological Evaluation of Medical Devices- Part 5: Tests for in ● Vitro Cytotoxicity
- ISO 10993-6:2007 Biological Evaluation of Medical Devices- Part 6: Tests for Local Effects after Implantation
- ISO 10993-10:2010 Biological Evaluation of Medical Devices- Part 10: Tests for Irritation and Skin Sensitization
- ISO 10993-11: 2006 Biological Evaluation of Medical Devices- Part 11: Tests for System Toxicity
- . ISO 4049: 2009 Dentistry - Polymer-Based Restorative Materials

Additional testing was performed as recommended by the following guidance document:

- . FDA "Guidance for Industry and FDA Staff: Dental Cements - Premarket Notification"

{6}------------------------------------------------

# Table 5.1: Predicate and Proposed Device Comparison Table

| Element                                                 | Predicate Device- Maxcem 2                                                                                                                                                                                                                                                                                                                                                                                            | Nexus Universal/Nexus<br>Universal Chroma, Dental<br>Cement                                                                                                                                                                                                                                                                                                                                                                                                  |
|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)                                                  | Maxcem 2 (K073209)                                                                                                                                                                                                                                                                                                                                                                                                    | To be assigned                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Trade Name                                              | Maxcem 2, Dental Cement                                                                                                                                                                                                                                                                                                                                                                                               | Nexus Universal and Nexus<br>Universal Chroma, Dental Cement                                                                                                                                                                                                                                                                                                                                                                                                 |
| Target Users                                            | Licensed dental professionals                                                                                                                                                                                                                                                                                                                                                                                         | Licensed dental professionals                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Indications for Use                                     | Maxcem 2 is intended for<br>cementation of indirect restorations<br>including ceramic, resin and metal-<br>based inlays, onlays, crowns, bridges,<br>posts, and veneers.* Additional<br>indications include core-buildup<br>material, pit and fissure sealant, and<br>cementation of crown restorations to<br>implants.<br>*Adhesive application on the prep is<br>required for veneer cementation<br>using Maxcem 2. | Nexus Universal/Nexus Universal<br>Chroma is indicated for<br>cementation of all indirect<br>restorations including ceramic,<br>resin, and metal-based<br>inlays/onlays, crowns, bridges,<br>posts, and veneers*. Additional<br>indications include core build-ups<br>and cementation of crowns to<br>implants.<br>*Adhesive application on the<br>preparation is required for veneer<br>cementation using Nexus<br>Universal and Nexus Universal<br>Chroma. |
| Common Name                                             | Dental Cement                                                                                                                                                                                                                                                                                                                                                                                                         | Dental Cement                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Classification Name                                     | Cement, Dental, per 21 CFR §<br>872.3275(b)                                                                                                                                                                                                                                                                                                                                                                           | Cement, Dental, per 21 CFR §<br>872.3275(b)                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Class                                                   | II                                                                                                                                                                                                                                                                                                                                                                                                                    | II                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Product Code                                            | MZW/EMA                                                                                                                                                                                                                                                                                                                                                                                                               | MZW/EMA                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Storage                                                 | Ambient Temperature                                                                                                                                                                                                                                                                                                                                                                                                   | Ambient Temperature                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Curing Mechanism                                        | Photo initiation or self-cure                                                                                                                                                                                                                                                                                                                                                                                         | Photo initiation or self-cure                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Material Compatibility                                  | Biocompatibility meets requirements                                                                                                                                                                                                                                                                                                                                                                                   | Biocompatibility meets<br>requirements                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Shelf Life                                              | 18 months based on real time data                                                                                                                                                                                                                                                                                                                                                                                     | 18 months based on accelerated<br>data                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Gel/Set Time<br>(fresh ambient/ fresh<br>oral)          | ≥ 2'00" / ≤ 3'30"                                                                                                                                                                                                                                                                                                                                                                                                     | ≥ 2'00" / ≤ 3'30"                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Bond Strength to<br>Dentin                              | 20.35 MPa                                                                                                                                                                                                                                                                                                                                                                                                             | 23.05 MPa                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Bonding strength for<br>Post Cementation                | 59.68 lbs                                                                                                                                                                                                                                                                                                                                                                                                             | 62.34 lbs                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Radiopacity, %<br>aluminum                              | 280%                                                                                                                                                                                                                                                                                                                                                                                                                  | 283%                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Film Thickness (μm)                                     | 22 μm                                                                                                                                                                                                                                                                                                                                                                                                                 | 24 μm                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Diametral Strength,<br>Mpa (Dual-Cure<br>mode)          | 54.8 MPa                                                                                                                                                                                                                                                                                                                                                                                                              | 58.1 MPa                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Consistency                                             | 3.00 cm                                                                                                                                                                                                                                                                                                                                                                                                               | 3.39 cm                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Maximum<br>Temperature During<br>Self Curing            | N/A                                                                                                                                                                                                                                                                                                                                                                                                                   | 26.89°C                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Water sorption                                          | N/A                                                                                                                                                                                                                                                                                                                                                                                                                   | 33.5 µg/mm³                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Color indicator version<br>(Nexus Universal<br>Chroma): | N/A                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Color at set time                                       | N/A                                                                                                                                                                                                                                                                                                                                                                                                                   | Pass                                                                                                                                                                                                                                                                                                                                                                                                                                                         |

{7}------------------------------------------------

Clinical Performance Data

Clinical performance testing has not been performed for the proposed devices.

Conclusion as to Substantial Equivalence

The slight modifications to the formula of the predicate, Maxcem 2 (K073209), to increase bond strength and bond compatibility with the adhesive in self-cure mode, and add the gel-state color indicator (Nexus Universal Chroma only) shall not affect the safety or efficacy of the proposed devices based on internal performance testing and biocompatibility testing. Nexus Universal and Nexus Universal Chroma are substantially equivalent to the predicate, Maxcem 2 (K073209), based on formulation, performance and testing comparisons.

---

**Source:** [https://fda-staging.innolitics.com/device/K151332](https://fda-staging.innolitics.com/device/K151332)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).