K151311 · Artsana S.P.A · FMI · Feb 10, 2016 · General Hospital
Device Facts
Record ID
K151311
Device Name
Safe Block Pen Needle
Applicant
Artsana S.P.A
Product Code
FMI · General Hospital
Decision Date
Feb 10, 2016
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Safe Block Pen Needle is intended for use with pen injector devices for the injection of fluids, including insulin. Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental sticks from the patient end of the needle. The shield also serves to conceal the needle before and after injection.
Device Story
Safe Block Pen Needle is a sterile, single-use, disposable safety needle for subcutaneous drug delivery via pen injectors. Device consists of a stainless steel cannula, plastic hub, and integrated safety shield mechanism. Operation: user attaches needle to pen; shield conceals needle before use. During injection, shield retracts against skin; upon needle withdrawal, shield automatically extends and permanently locks to cover the cannula, preventing reuse and accidental needlesticks. Used by patients or healthcare providers in clinical or home settings. Output is the delivery of medication; safety feature provides physical protection against sharps injuries. Benefits include reduced risk of accidental needlesticks and concealment of needle for patients with needle phobia.
Clinical Evidence
No clinical tests were conducted. Safety and effectiveness were demonstrated through bench testing (performance, safety, and specifications) and simulated user studies to verify equivalence to predicate devices.
Technological Characteristics
Materials: PP (hub/cover), HDPE (ring/shield/container), medical paper (seal), stainless steel (cannula/spring), UV acrylate (glue), silicone oil mixture (lubricant). Standards: ISO 11608-2 (dimensions/bond strength/lubrication), ISO 10993-1 (biocompatibility), ISO 23908 (safety mechanism). Sterilization: ETO (ISO 11135-1:2007). Form factor: 31G, 5mm/8mm lengths. Mechanical safety shield with automatic locking.
Indications for Use
Indicated for use with pen injector devices for subcutaneous injection of fluids, including insulin, in both hospital and domestic environments.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other, resembling a bird in flight.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 10, 2016
Artsana S.p.A c/o Mr. H. Carl Jenkins Official Correspondent The Wood Burditt Group, LLC 10 E. Scranton Ave., Ste. 201 Lake Bluff, Illinois 60044
Re: K151311
Trade/Device Name: Safe Block Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: January 7, 2016 Received: January 11, 2016
Dear Mr. Jenkins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151311
Device Name Safe Block Pen Needle
#### Indications for Use (Describe)
Safe Block Pen Needle is intended for use with pen injector devices for the injection of fluids, including insulin. Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental sticks from the patient end of the needle.
The shield also serves to conceal the needle before and after injection.
Type of Use (Select one or both, as applicable)
| <span style="font-family: sans-serif;"> <input type="checkbox"/> Research Use (Part 21 CFR 601 Subpart D) </span> |
|-----------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: sans-serif;"> <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 201 Subpart G) </span> |
_ | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "The Wood Burditt Group". The logo consists of the words "the WOOD BURDITT group" arranged in three lines. A teal-colored curved line wraps around the words "WOOD" and "BURDITT", visually connecting the two words.
# 510(k) Summary
Date prepared: May 12, 2015
| Submitter / Contact Person | H. Carl Jenkins<br>The Wood Burditt Group<br>10 E. Scranton Ave, Suite 201<br>Lake Bluff, IL 60044 |
|----------------------------|----------------------------------------------------------------------------------------------------|
| | (ph) 847-234-7500 x 205<br>(fax) 847-578-0728<br>(email) hcjenkins@woodburditt.com |
| Applicant | Artsana S.p.A Via Saldarini Catelli 1, 22070 Grandate<br>COMO – ITALY<br><b>Registration Number:</b> 9612764 |
|--------------|--------------------------------------------------------------------------------------------------------------|
| Manufacturer | Artsana S.p.A Unit Production N°30, Via Saldarini catelli<br>6/10, 22070 Grandate<br>COMO - ITALY |
## Device Name
| Trade Name | Safe Block Pen Needle |
|----------------------|----------------------------------------------------------------|
| Proprietary Name | Safe Block Pen Needle G31x5mm<br>Safe Block Pen Needle G31x8mm |
| Common Name | Sterile Disposable Safety Pen Needle |
| Classification Name | Hypodermic Single Lumen Needle |
| Classification Panel | 80 – General Hospital |
| Regulation | 21 CFR 880.5570 |
| Product Code | FMI |
| Classification | II |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for "The Wood Burditt Group". The logo is composed of the words "the WOOD BURDITT group" in a stacked format. A teal-colored curved line visually connects the words "WOOD" and "BURDITT".
## Reason for 510(k) Submission
The applicant device is a new device.
## Device Description
SAFETY PEN NEEDLE. Safe Block Pen Needle is designated for use with pen injector for subcutaneous injection of a desired dose of drugs approved for delivery using a pen needle. Safe Block Pen Needle are offered in one gauge size (31G) and in a various lengths (5mm and 8mm). The Safe Block Pen Needle is sterile (ETO sterilization), non toxic and non-pyrogenic. It's a disposable single use device.
Additionally the Safe Block Pen Needle is designed to reduce occurrence of accidental needle sticks from patient end of the needle by providing a shield that covers and locks the needle after use. Prior to injection the user will attach the Safe Block Pen Needle to the pen. The shield of Safe Block Pen Needle will hide the needle from the user prior to use. As the user proceeds with inserting the needle into the shield will retract. After the injection is completed and needle is removed from the skin, the shield will automatically extend to cover the needle and lock in place. The Safe Block Pen Needle should be removed from the pen and discarded.
### Indications for Use:
Safe Block Pen Needle is intended for use with pen injector devices for the injection of fluids, including insulin. Additionally, the attached safety shield automatically locks in place and reduces the occurrence of accidental sticks from the patient end of the needle. The shield also serves to conceal the needle before and after injection.
## Predicate Device Summary Table
Based on the comparison of the device features, materials, intended use and performance the Safe Block Pen Needle was shown to be substantially equivalent to the commercially available predicate devices indicated in the table below.
| Product | Applicant | 510(k) # | Clearance Date |
|----------------------------------------------------------------|------------------|----------|----------------|
| Artsana Hypodermic needles, Artsana Insupen insulin pen needle | Artsana S.p.A. | K051783 | 2005.10.04 |
| Novo Fine Auto Cover | Novo Nordisk Inc | K050106 | 2005.03.30 |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "The Wood Burditt Group". The logo is composed of the words "the WOOD BURDITT group" stacked on top of each other. To the right of the words is a teal-colored curved line that resembles a crescent shape.
## Technological characteristics and comparison
Artsana Safe Block 31Gx5 mm and 31Gx8 mm is substantially equivalent to the Novo Nordisk -Novo Fine Autocover (K050106) and the Artsana pen needles (K051783). They are all class II devices and, as explained below, the intended use is the same for all the three products. They are all designed for single use in conjunction with pen injection for subcutaneous administration of drugs approved for delivery using a pen needle, both in hospital and domestic environment. Furthermore, both NovoFine Autocover and Safe Block have safety features to prevent people from needlestick injuries. For what is concerned to the injection, the working principle is the same for all the three devices: before the injection, the protective seal is removed from the plastic primary container. Then, keeping the needle in the container, the threaded hub is screwed on the injection device (as explained in the Safe Block directions for use) and the injection is prepared according to the pen injector directions for use. After the primary container is removed, both Safe Block and Autocover have a further needle shield, that is not removable, hiding the cannula before and after the injection (this is an important feature for people that are afraid of needles).
The needle cover is pushed on the skin, retracting into the product: in this way, the cannula penetrates into the patient's skin and the drug can be delivered. Then, when the cannula is removed from the skin, the shield automatically covers the cannula again: now the shield is permanently locked, protecting the needle tip and both preventing from the pen needle reuse and reducing the risk of accidental injuries. Finally, the needle is removed from the injection device and disposed as per local safety practices and regulations on sharp devices.
| Product<br>Dimensions | Colour of<br>the hub | Colour of the<br>primary container<br>(cover) | Configuration of the tip |
|-----------------------|----------------------|-----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| G31x5mm | Purple | White | Per ISO 11608-2, section<br>4.5, visually sharp at 2.5X<br>magnification, designed to<br>minimize<br>coring and fragmentation |
| G31x8mm | Light blue | White | Per ISO 11608-2, section<br>4.5, visually sharp at 2.5X<br>magnification, designed to<br>minimize<br>coring and fragmentation |
Predicate device Autocover and applicant device Safe Block have the same working principle of the safety mechanism and they're very similar also in terms of product dimensions and shape.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "The Wood Burditt Group". The words "the" and "group" are stacked on the top and bottom right of the logo. The words "Wood" and "Burditt" are stacked on the top and bottom left of the logo. A teal-colored crescent shape is in the middle of the logo.
Component materials are presented in the table below:
| Component | Applicant Device Material | Predicate Device Material (K051783) |
|-------------------|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|
| External cover | PP (non toxic, medical grade) | Identical |
| Needle hub | PP (non toxic, medical grade) | Identical |
| Ring | HDPE (non toxic, medical grade) | Identical |
| Shield | HDPE (non toxic, medical grade) | Identical |
| Primary Container | HDPE (non toxic, medical grade) | Identical |
| Seal | Medical Paper | Identical |
| Cannula | Stainless Steel | Identical |
| Glue | UV acrylate | Identical |
| Lubricant | Silicone oil mixture (non toxic, medical grade) composed by 3 different elements (Dispersion fluid, Medical Silicone, Solvent) | Identical |
| Spring | Stainless Steel | N/A |
| Colorants | Pigments for polymer | Identical |
## Clinical test:
No clinical tests were conducted in support of this 510(k) submission.
## Non Clinical Tests performed:
Testing demonstrated the safety feature performance through laboratory testing and simulated user studies.
Bench testing related to performance, safety, effectiveness and specifications of the proposed device was also conducted in order to verify the equivalence of performances with the predicate devices.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for "The Wood Burditt Group". The words "the" and "group" are stacked on the top and bottom right of the logo. The words "Wood" and "Burditt" are stacked on the left side of the logo. A teal-colored arc is in the middle of the logo.
The Wood Burditt Group LLC
Total E. Scranton Ave., Suite 2011
10 E. Scranton Ave., Suite 2014 144
10 L. Scranton Ave., 2017 14:44
847. 2234. 17. 2234. 17. 12. 12. 12. 12. 12. www.woodburditt.com
| Table of specifications (including relevant tests) for applicant and predicate devices, below: | | | |
|------------------------------------------------------------------------------------------------|--|--|--|
| | | | |
| Feature /<br>Characteristics | Applicant Device | Predicate K051783<br>(ARTSANA PEN<br>NEEDLES,<br>ARTSANA<br>INJECTION<br>NEEDLES) | Predicate K050106<br>(NOVOFINE<br>AUTOCOVER 30G<br>x 8mm) |
|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | It is intended for use<br>with pen injector<br>devices for the<br>injection of drugs,<br>including insulin.<br>Additionally the<br>attached safety shield<br>automatically locks in<br>place and reduces the<br>occurrence of<br>accidental sticks from<br>the patient end of the<br>needle. The shield<br>also serves to hide<br>the needle before and<br>after injection | Insulin delivery | For use in<br>conjunction with<br>insulin injection<br>delivery devices for<br>subcutaneous<br>administration of<br>sterile parenteral<br>insulin products. |
| Length / Tolerance | 5 mm, 8 mm<br>(tolerance<br>as per ISO 11608-2) | 6 mm, 8 mm, 12 mm<br>(tolerance as per ISO<br>11608-2) | 8 mm (tolerance as<br>per ISO 11608-2) |
| Gauge | G31 (as per ISO<br>9626) | G31, G29 (as per ISO<br>9626) | G30 (as per ISO<br>9626) |
| Tip configuration | Per ISO 11608-2,<br>section<br>4.5, visually sharp at<br>2.5X magnification,<br>designed to minimize<br>coring and<br>fragmentation | Per ISO 11608-2,<br>section 4.5, visually<br>sharp at 2.5X<br>magnification,<br>designed to minimize<br>coring and<br>fragmentation | Per ISO 11608-2,<br>section 4.5, visually<br>sharp at 2.5X<br>magnification,<br>designed to<br>minimize coring and<br>fragmentation |
| Cover colour | White | White | Transparent / White |
| Hub/needle bond<br>strength | Per ISO 11608-2 | Per ISO 11608-2 | - |
| Biocompatibility | As per ISO 10993-1 | As per ISO 10993-1 | - |
| Compatibility test | ISO 11608-2 (torque | ISO 11608-2 (torque | - |
| (type A<br>needles) | test) | test) | |
| Lubrication | As per ISO 11608-2 | ISO 11608-2 | - |
| Safety mechanism<br>Activation | As per ISO 23908,<br>internal protocol and<br>test results | - | As per ISO 23908,<br>internal protocol and<br>test results |
| Safety<br>overriding/unlocking<br>force<br>after activation | As per ISO 23908,<br>internal protocol and<br>test results | - | As per ISO 23908,<br>internal protocol and<br>test results |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for "The Wood Burditt Group". The words "the", "WOOD", and "BURDITT" are stacked on top of each other on the left side of the logo. The word "group" is on the right side of the logo. A teal-colored curved line connects the left and right sides of the logo.
## Sterilization:
Product is EO sterilized in house by manufacturer. Sterilization process is validated using current standard ISO 11135-1:2007.
The method used to validate the sterilization cycle is HSKP method (ISO 11138-1:2006)
The shelf life of the product, considering the integrity of the packaging and the sterility and physical properties, is 5 years from the production date. The expiring date and sterilization methods are clearly indicated on the pack. The product shelf life is ensured if the product is stored and transported in compliance with the environmental condition stated on the pack and on the external carton.
## Biocompatibility:
The product has successfully undergone the biocompatibility evaluation required by ISO 10993-1 for:
- Cytotoxicity (ISO 10993-5) ,
- Emocompatibility (ISO 10993-4, ASTM F756),
- -Sensitization and skin reactivity (ISO 10993-10),
- -Acute systemic toxicity (ISO 10993-11).
# Conclusion
The applicant device is substantially equivalent in its intended use, technology / principal of operation, materials, and performance to the predicate devices identified in this 510(k) submission.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.