Schoelly Nephroscope Set, Schoelly Ultra-Mini Nephroscope Set
Applicant
Schoelly Fiberoptic GmbH
Product Code
FGA · Gastroenterology, Urology
Decision Date
Feb 9, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Schoelly Nephroscope Set is used for the disintegration and removal / extraction of kidney stones. The stones are removed, under endoscopic control, through percutaneous passages, in conjunction with intracorporeal pneumatic, ultrasound, electrohydraulic or laser lithotripters. The Schoelly Ultra-Mini Nephroscope Set is used for the disintegration and removal / extraction of kidney stones. The stones are removed, under endoscopic control, through percutaneous passages, in conjunction with laser lithotripters.
Device Story
Schoelly Nephroscope Family consists of rigid, reusable endoscopes, sheaths, obturators, and bridges for percutaneous nephrolithotomy (PCNL) and Mini PCNL. Device inputs include light from external sources; optical images transmitted via rod lens and fiber optic systems. Physician operates device under endoscopic control to visualize kidney stones; uses auxiliary instruments (lithotripters) through working channels or accessory setups to disintegrate and extract stones. Output is visual feedback on external monitor for surgeon guidance during stone removal. Used in clinical settings (OR/urology suite). Benefits include minimally invasive access for stone removal.
Clinical Evidence
Bench testing only. No clinical data. Testing included temperature measurements per IEC 60601-2-18, optical parameter verification per ISO 8600, and biocompatibility testing per ISO 10993 (cytotoxicity, sensitization, acute systemic toxicity). Reprocessing validation performed per AAMI TIR12, AAMI TIR30, ANSI/AAMI ST15883-1, ISO 14937, ANSI/AAMI ST81, ISO 17664, and ANSI/AAMI/ISO 17665-1.
Technological Characteristics
Rigid, reusable endoscopes (6Fr-19Fr diameter, 220mm length). Materials: stainless steel, glass, glass fibers, adhesive, silicone, tin/silver alloy. Optical: 0°-8° direction of view, 85° field of view. Light transmission: fiber optics. Image transmission: rigid rod lens + fiber optics. Connectivity: external light source/camera/monitor. Sterilization: steam. Compliance: IEC 60601-2-18, ISO 8600.
Indications for Use
Indicated for patients requiring percutaneous nephrolithotomy (PCNL) or Mini PCNL for the disintegration and removal of kidney stones.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus-like design with three human profiles facing right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 9, 2016
Schoelly Fiberoptic, GmbH Dr. Sandra Baumann Senior Manager Regulatory Affairs Robert-Bosch-Str. 1-3 79211 Denzlingen Germany
Re: K151308
Trade/Device Name: Schoelly Nephroscope Set, Schoelly Ultra-Mini Nephroscope Set Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Codes: FGA, FEC, FED Dated: December 23, 2015 Received: December 28, 2015
Dear Dr. Sandra Baumann,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151308
Device Name Schoelly Nephroscope Set Schoelly Ultra-Mini Nephroscope Set
#### Indications for Use (Describe)
The Schoelly Nephroscope Set is used for the disintegration and removal / extraction of kidney stones. The stones are removed, under endoscopic control, through percutaneous passages, in conjunction with intracorporeal pneumatic, ultrasound, electrohydraulic or laser lithotripters.
The Schoelly Ultra-Mini Nephroscope Set is used for the disintegration and removal / extraction of kidney stones. The stones are removed, under endoscopic control, through percutaneous passages, in conjunction with laser lithotripters.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">✖</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
02/08/2016 Schoelly Nephroscope Family
#### Section 5 – 510(k) Summary
#### I. SUBMITTER
Owner's Name: Address:
Telephone Number: Fax Number: Contact Person:
Schoelly Fiberoptic GmbH (Registration: 8043903) Robert-Bosch-Str. 1 - 3 79211 Denzlingen Germany +49-7666-980-0 +49-7666-908-380 Dr. Sandra Baumann
II. DEVICE Subject Device Name: Trade Name:
Common/Usual Name: Classification Name:
Schoelly Nephroscope Family Schoelly Nephroscope Set Schoelly Ultra-Mini Nephroscope Set Nephroscope Set FGA – Kit, Nephroscope 21 CFR 876.1500; Class II FEC - Obturator, for Endoscope 21 CFR 876.1500; Class II FED - Endoscopic Access Overtube, Gastroenterology-Urology 21 CFR 876.1500; Class II
#### III. PREDICATE/REFERENCE DEVICES
| Predicate Device Name: | Nephroscope Set |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | Nephroscope Set |
| Common/Usual Name: | Nephroscope Set |
| Classification Name: | FGA - Kit, Nephroscope<br>21 CFR 876.1500; Class II<br>FEC - Obturator, for Endoscope<br>21 CFR 876.1500; Class II<br>FED - Endoscopic Access Overtube,<br>Gastroenterology-Urology<br>21 CFR 876.1500; Class II<br>FFL - Ureteral stone dislodger<br>21 CFR 876.4680; Class II |
| Premarket Notification: | K994223, Richard Wolf Medical Instruments Corp.,<br>SE date June 5, 2000 |
| Recalls: | This predicate has not been subject to a design-<br>related recall |
| Reference Device Name: | 12 Ureteroscopes/Ureterorenoscopes |
| Trade Name: | Compact Operating Fiber Uretero-Renoscopes |
| Common/Usual Name: | Ureteroscopes |
| Classification Name: | FGB - Ureteroscope and accessories, flexible/rigid<br>21 CFR 876.1500; Class II |
| Premarket Notification: | K963855, Richard Wolf Medical Instruments Corp,<br>SE date April 3, 1997 |
| Reference Device Name: | Karl Storz Adult and Pediatric Nephroscope |
| Trade Name: | Karl Storz Adult and Pediatric Nephroscope |
| Common/Usual Name: | Nephroscope |
| Classification Name: | FGA - Kit, Nephroscope<br>21 CFR 876.1500; Class II |
| Premarket Notification: | K940594, Karl Storz Endoscopy America,<br>SE date September 09, 1994 |
| Reference Device Name: | Cook Fiber Optic Bundle and Flexor Deflecting Access Sheath |
| Trade Name: | Cook Fiber Optic Bundle and Flexor Deflecting Access Sheath |
| Common/Usual Name: | Access Sheath |
| Classification Name: | FED - Endoscopic Access Overtube, Gastroenterology-<br>Urology<br>21 CFR 876.1500; Class II |
| | FFS - Image, illumination, fiberoptic, for endoscope<br>21 CFR 876.1500; Class II |
| | FAJ - Cystoscope and accessories, flexible/rigid<br>21 CFR 876.1500; Class II |
| | FGA - Kit, Nephroscope<br>21 CFR 876.1500; Class II |
| | GCJ - Laparoscope, general & plastic surgery<br>21 CFR 876.1500; Class II |
| | FGB - Ureteroscope and accessories, flexible/rigid<br>21 CFR 876.1500; Class II |
| Premarket Notification: | K072521, Cook Urological, Inc.,<br>SE date November 20, 2007 |
| Reference Device Name: | Schoelly Sinuscope |
| Trade Name: | Schoelly Sinuscope |
| Common/Usual Name: | Sinuscope |
| Classification Name: | EOB - Nasopharyngoscope (flexible or rigid)<br>21 CFR 874.4760; Class II |
| Premarket Notification: | K142249, Schoelly Fiberoptic GmbH,<br>SE date January 27, 2015 |
In addition to the Richard Wolf predicate device, the following reference devices were used in this submission to support substantial equivalence and to further outline technological characteristics:
{4}------------------------------------------------
{5}------------------------------------------------
#### IV. DEVICE DESCRIPTION
The proposed Schoelly Nephroscope Family comprises two different application sets: One set for Percutaneous Nephrolithotomy (PCNL) with the trade name "Schoelly Nephroscope Set'' and one set for Mini PCNL with the trade name "Ultra-Mini Nephroscope Set".
Both sets include a rigid, reusable endoscope (nephroscope) that is used in conjunction with a commercially available and approved light guide, light source. video camera, monitor, and printer. Light guide, light source, video camera, monitor, and printer are not included in the scope of delivery and are further not within the scope of this application.
In case of both sets, the nephroscope is accompanied by corresponding accessories to allow for access and passage of the nephroscope and retrieval instruments and for irrigation. Those include compatible endoscopic sheaths, obturators, and bridges.
The Schoelly Nephroscope Set and the Schoelly Ultra-Mini Nephroscope set are delivered in non-sterile conditions and have already obtained CE mark.
# Technological Characteristics
#### Endoscopes (Nephroscopes):
Light that is created by an external light source is transmitted from the nephroscope light guide connector through the nephroscope itself to the tip via a fiber optic system. Images are transferred the other way back through a rigid rod lens system and a fiber optic bundle. The image can be displayed by a camera/monitor system which can be connected to the nephroscope eyepiece.
The main technical parameters that characterize the optical view of the nephroscopes of both application sets are the direction of view (0°-8°) and the field of view (85°). The insertion tube diameter of the nephroscopes is 6Fr - 19Fr and the insertion tube working length is 220mm.
The nephroscope of the Schoelly Nephroscope Set incorporates a working channel with a max. inner lumen capacity of 9Fr and is used in conjunction with an instrument bridge that is mounted to the nephroscope's proximal end (see "Bridges" Section below).
The nephroscope of the Schoelly Ultra-Mini Nephroscope Set does not have a working channel and is not equipped with an instrument bridge but is accompanied by additional accessories instead. The design of those accessories ensures that if the nephroscope is used in conjunction with them, the setup provides exactly the same functionality. Thus, if assembled appropriately, both application sets incorporate the number of inner lumens required to perform each single step of the intervention and to accommodate auxiliary instruments and irrigation.
{6}------------------------------------------------
Like other currently marketed nephroscopes, all Schoelly configurations have outer surfaces mainly made from stainless steel and further comprise fiber optics for light transmission and rigid rod lenses and fiber optics for image transmission.
# Sheaths:
The endoscopic sheaths included in this submission are rigid reusable instruments with inner lumens and mainly made from stainless steel; the sheaths serve as the most outer part of the whole device setup in PCNL and Mini PCNL procedures. The proximal end of the endoscopic sheaths is equipped with one side port for irrigation. The outer diameter of the sheaths in this submission ranges from 11Fr-24Fr to comply with the different nephroscopes and obturators of the proposed Family.
# Obturators
The obturators included in this submission are rigid reusable instruments mainly made from stainless steel with an inner lumen sized to accept the needed guide wire (not part of this submission) for correct placement of the sheath in the kidney. During insertion, the obturator fills the space inside the sheath to provide a smooth surface. The overall length of the obturators and their diameter comply with the length and the diameter of the available sheaths.
# Bridges:
This submission includes a standard instrument bridge, which is a rigid reusable instrument with an inner lumen and mainly made from stainless steel. It includes ports to allow for insertion of instruments and irrigation through the working channel of the nephroscope of the Schoelly Nephroscope Set. Channels can be sealed with a self-sealing membrane and a stop cock, respectively. It comprises an automatic locking mechanism for easy connection to the proximal end of the nephroscope.
# V. INDICATIONS FOR USE
The Schoelly Nephroscope Set is used for the disintegration and removal / extraction of kidney stones. The stones are removed, under endoscopic control, through percutaneous passages, in conjunction with intracorporeal pneumatic, ultrasound, electrohydraulic or laser lithotripters.
The Schoelly Ultra-Mini Nephroscope Set is used for the disintegration and removal / extraction of kidney stones. The stones are removed, under endoscopic control, through percutaneous passages, in conjunction with laser lithotripters.
# VI. COMPARISON OF TECHNOLOGCIAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The Indications for Use for the proposed Schoelly Nephroscope Family comprises a subset of the Indications for Use of the predicate device. Since the proposed Schoelly Nephroscope Family is not intended to be used via transurethral access as is the case for the Richard Wolf predicate device, the removal/extraction of bladder stones is not included in the Indications for Use for this submission. Remaining contents of the Indications for Use are identical for the proposed and the predicate device.
{7}------------------------------------------------
| A W Y A A A WA . | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | |
| | |
| | |
| | |
| AMAAAA | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| | |
| 1 | |
| | |
| | |
| | |
| | |
| НААН С С НАНАНИ О С | |
| | |
| | |
| | |
| | |
| | |
| | |
| 7 | |
| 1 | |
| | |
| . It also allerial and the seat the second the second the section of the section of the section of the section of the section of the section of the section of the section of | |
| | |
| of the property and any and the property of the program and<br>A A A MA A | |
| -<br>1 | |
| Technological Characteristics: Similarities and Differences between the Proposed Device and the Predicate Devices | | | |
|-------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------|--------------------------------------------------------------|
| Attribute | Proposed<br>Nephroscope Family<br>(current submission) | Endoscope (Nephroscope) | Predicate<br>Richard Wolf Nephroscope Set (K994223) |
| Light transmission | Fiber optics | Fiber optics | Fiber optics |
| Light source | External, connected via light guide to light guide connector | External, connected via light guide to light guide connector | External, connected via light guide to light guide connector |
| Image transmission | Rigid rod lenses + fiber optics | Rigid rod lenses, rigid rod lenses + fiber optics | Rigid rod lenses, rigid rod lenses + fiber optics |
| Eyepiece orientation | angled, straight | angled, angled offset | angled, angled offset |
| Direction of view | 0°-8° | 12°-25° | 12°-25° |
| Field of view | 85° | 95° (Richard Wolf Long Miniature Endoscope) | 95° (Richard Wolf Long Miniature Endoscope) |
| Image display | Camera/monitor connected via the endoscope eyepiece | Camera/monitor connected via the endoscope eyepiece | Camera/monitor connected via the endoscope eyepiece |
| Insertion tube working length | 220mm | 175-225mm | 175-225mm |
| Insertion tube outer diameter | 6Fr-19Fr | 15Fr-27Fr | 15Fr-27Fr |
| Inner lumen capacity | max.3 Fr- max. 9 Fr | max. 6Fr - max.12Fr | max. 6Fr - max.12Fr |
| Single Use / Reusable | Reusable | Reusable | Reusable |
| Reprocessing | Cleaning, sterilization (steam) | Cleaning, sterilization (steam, H2O2) | Cleaning, sterilization (steam, H2O2) |
| Patient contacting materials | Stainless steel, glass, glass fibers, adhesive | Stainless steel, glass, glass fibers, adhesive | Stainless steel, glass, glass fibers, adhesive |
| Electrical safety | IEC 60601-2-18 compliant | IEC 60601-2-18 compliant | IEC 60601-2-18 compliant |
| Instruments | | | |
| Sheath outer diameter | 11Fr-24Fr | 15Fr-27Fr | 15Fr-27Fr |
| Obturator outer diameter | Compatible with 11Fr-24Fr sheath | Compatible with 15Fr-27Fr sheath | Compatible with 15Fr-27Fr sheath |
| Single Use / Reusable | Reusable | Reusable | Reusable |
| Reprocessing | Cleaning, sterilization (steam) | Cleaning, sterilization (steam, H2O2) | Cleaning, sterilization (steam, H2O2) |
| Patient contacting materials | Stainless steel, silicone, tin/silver alloy | Stainless steel, silicone | Stainless steel, silicone |
{8}------------------------------------------------
# VII. PERFORMANCE DATA
The nephroscopes of the proposed Schoelly Nephroscope Family were subjected to temperature, optical parameter and biocompatibility testing.
Temperature testing: the devices were measured for surface temperatures at various locations as long as a steady state temperature was reached using different light sources and found to meet requirements as specified in IEC 60601-2-18.
Optical parameter testing: The devices were tested for all relevant optical parameters, e.g. field of view and direction of view accuracy and found to meet requirements as specified in ISO 8600.
Biocompatibility testing: A series of biocompatibility testing according to ISO 10993, including cytotoxicity, sensitization, and acute systemic toxicity, demonstrated that the device components that are in contact with the patient are biocompatible.
#### Reprocessing
The Schoelly Nephroscope Family was the subject of completed reprocessing validations including manual cleaning, automated cleaning and steam sterilization.
Cleaning studies have been performed in accordance with AAMI TIR12:2010 (Designing, Testing and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A Guide for Device Manufacturers), AAMI TIR30:2011 (A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable devices), and ANSI/AAMI ST15883-1: 2009 (Washer-disinfectors - Part 1 - General requirements, terms, and definitions and tests).
Sterilization studies have been performed in accordance with ISO 14937:2009 (Sterilization of health care products - General requirements for characterization of sterilizing agent and the development, validation and routine control of a sterilization process for medical devices), ANSI/AAMI ST81:2004 (Sterilization of medical devices - Information to be provided by the manufacturer for the processing of resterilizable medical devices), ISO 17664:2004 (Sterilization of medical devices -Information to be provided by the manufacturer for the processing of resterilizable medical devices), and ANSI/AAMI/ISO 17665-1:2006 (Sterilization of health care products - Moist heat - Requirements for the development, validation and routine control of sterilization process for medical devices).
# VIII. CONCLUSION
The Schoelly Nephroscope Family meets all the pre-determined testing and acceptance criteria to effectively demonstrate substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.