K151306 · Convergent Dental, Inc. · GEX · Sep 23, 2015 · General, Plastic Surgery
Device Facts
Record ID
K151306
Device Name
Solea
Applicant
Convergent Dental, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Sep 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Solea system is indicated for the following: - Ablation of hard tissue for caries removal and cavity preparation - Incision, Excision, Vaporization, Coagulation and Hemostasis of soft tissue in the oral cavity - Cutting, shaving, contouring and resection of oral osseous tissue (bone)
Device Story
Solea is a mobile, cart-based dental laser system using 9.25μm CO2 laser energy to perform hard tissue, soft tissue, and osseous tissue procedures. Operated by general practitioner dentists and specialists in a dental office, the device uses an articulating arm for beam delivery. The system functions via pulsed laser energy for ablation and continuous wave for aiming. It provides clinicians with a tool for precise tissue management, replacing traditional mechanical drills or scalpels in various oral procedures. The device benefits patients by enabling minimally invasive procedures with potential for reduced bleeding and improved healing. No software or hardware changes were made for this specific clearance; it leverages previously cleared settings for hard and soft tissue, with osseous tissue settings demonstrated to be equivalent to predicate laser systems.
Clinical Evidence
No clinical testing was performed. Substantial equivalence is supported by non-clinical bench testing, including electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), laser safety (IEC 60601-2-22, IEC 60825-1), sterilization validation (AAMI TIR12, ANSI/AAMI ST79), and usability testing (IEC 62366). Bench testing confirmed thermal effects on hard tissue are equivalent to predicate devices.
Technological Characteristics
Mobile, cart-based CO2 laser (9.25μm wavelength). Class 4 (IV) laser product. Beam delivery via articulating arm (free space). Operating modes: pulsed (ablation) and continuous (aiming). Sterilization via steam autoclave (handpieces). EMC compliance per CISPR 11 Class A.
Indications for Use
Indicated for dental patients requiring hard tissue ablation (caries removal, cavity preparation), soft tissue surgery (incision, excision, vaporization, coagulation, hemostasis), and oral osseous tissue procedures (cutting, shaving, contouring, resection).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles a stylized human figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Convergent Dental Incorporated % Ms. Carrie Hetrick Emergo 816 Congress Avenue, Suite 1400 Austin, Texas 78701
September 23, 2015
Re: K151306 Trade/Device Name: Solea Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: May 14, 2015 Received: May 18, 2015
Dear Ms. Hetrick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Joshua C. Nipper -S
For I
Binita S. Ashar, M.D., M.B.A., F.A.C.S. For Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151306
Device Name Solea
Indications for Use (Describe)
The Solea system is indicated for the following:
- Ablation of hard tissue for caries removal and cavity preparation
- Incision, Excision, Vaporization, Coagulation and Hemostasis of soft tissue in the oral cavity
- Cutting, shaving, contouring and resection of oral osseous tissue (bone)
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
for
## Solea
### 1. Submission Sponsor
Convergent Dental, Inc. 2 Vision Drive Natick, MA 01760 USA Phone: (508) 500.5656 Contact: Mike Manor, Director QA/RA
### 2. Submission Correspondent
Emergo 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: (720) 838.4113 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Carrie Hetrick, Senior Consultant, RA Email: project.management@emergogroup.com
### 3. Date Prepared
August 14, 2015
### 4. Device Identification
| Trade/Proprietary Name: | Solea |
|----------------------------|-------------------------------------------------------------------------------------|
| Common/Usual Name: | Powered laser surgical instrument |
| Classification Name: | Laser surgical instrument for use in general and plastic surgery<br>and dermatology |
| Classification Regulation: | 21 CFR §878.4810 |
| Product Code: | GEX |
| Device Class: | Class II |
| Classification Panel: | General and Plastic Surgery |
### 5. Legally Marketed Predicate Device(s)
Convergent Dental, Inc., Solea, 510(k) Numbers: K123494, K130420 Biolase Technology, Inc., Waterlase MD, 510(k) Number: K091746 Fotona d.d., LightWalker Laser System, 510(k) Number: K101817
### 6. Device Description
{4}------------------------------------------------
The Solea system is a dental laser device previously cleared by the FDA for soft tissue dental indications (K123494) and hard tissue indications (K130420). The only changes from the previously cleared device are the addition of the osseous tissue indications. There have been no software changes or graphic changes to the Solea system. The osseous tissue settings are the same as the previously cleared device (K130420). The osseous settings have substantially equivalent fluence and irradiation as the hard tissue predicate devices. There are no other hardware or software changes to the Solea system device pending herein when compared to the device cleared under K123494 and K130420.
The Solea system is a mobile, cart-based dental treatment system that uses pulsed laser energy to cut and ablate hard tissue and to cut soft tissue in the oral cavity. The Solea system utilizes advanced CO2 laser technology with a wavelength of 9.25μm to safely and effectively perform ablation, incision, excision, vaporization, coagulation and hemostasis procedures.
### 7. Indication for Use Statement
The Solea system is indicated for the following:
- । Ablation of hard tissue for caries removal and cavity preparation
- Incision, Excision, Vaporization, Coagulation and Hemostasis of soft tissue in the oral ၊ cavity
- -Cutting, shaving, contouring and resection of oral osseous tissue (bone)
### 8. Substantial Equivalence Discussion
The following table compares the Solea to the predicate devices with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
The Convergent Dental, Inc. Solea system is a dental laser previously cleared by the FDA for hard tissue indications (K130420) and soft tissue indications (123494).
The Solea system included herein is the same device that was previously cleared by the FDA in K130420 (Solea, expanded hard tissue indications) and K123494 (Solea®, indications for soft tissue indications). There are no unique applications, materials or specifications presented herein. All the presented indications for use retain the same meaning as their equivalent indications cleared by the FDA.
The subject Solea system is substantially equivalent to the Biolase Technology, Inc. Waterlase MD (K091746) and the Fotona d.d., LightWalker Laser System (K101817) for the osseous tissue indications in terms of fluence, measured comparative performance data, and all features that affect safety or effectiveness.
The following table compares the subject Convergent Dental, Inc. Solea system to the Biolase Technology Inc. Waterlase MD, Fotona d.d. LightWalker Laser System Family, and the previously cleared Solea systems, with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence for intended uses.
{5}------------------------------------------------
| | Convergent<br>Dental, Inc. | Convergent<br>Dental, Inc. | Convergent<br>Dental, Inc. | Biolase<br>Technology, Inc. | Fotona d.d. |
|---------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | Solea | Solea | Solea | Waterlase® MD | LightWalker<br>Laser System<br>Family |
| 510(k)<br>Number | K152306 | K130420 | K123494 | K091746 | K101817 |
| Common<br>name | Powered<br>Laser surgical<br>instrument | Powered<br>Laser surgical<br>instrument | Powered<br>Laser surgical<br>instrument | Powered Laser<br>surgical<br>instrument | Powered Laser<br>surgical<br>instrument |
| FDA<br>Classification<br>Names | Powered<br>Laser surgical<br>instrument | Powered<br>Laser surgical<br>instrument | Powered<br>Laser surgical<br>instrument | Powered Laser<br>surgical<br>instrument<br>Drill, bone,<br>powered<br>System, Dental,<br>Hydrokinetic,<br>Carries Removal<br>& Cavity<br>Preparation | Powered Laser<br>surgical<br>instrument |
| FDA<br>Classification<br>codes | GEX, a Class<br>II device | GEX, a Class<br>II device | GEX, a Class II<br>device | GEX, MXF, DZI, a<br>Class II device | GEX, a Class II<br>device |
| Predicates<br>cited | K130420 and<br>K123494 for<br>hard and soft<br>tissue<br>indications<br>K091746,<br>K101817 for<br>osseous<br>tissue<br>indications | K091746,<br>K101817 for<br>hard tissue<br>indications<br>K123494 for<br>soft tissue<br>indications | K091320 for<br>soft tissue<br>indications | K031140,<br>K071363,<br>K090181 for<br>hard tissue<br>indications | K093162 for<br>hard tissue<br>indications |
| Target User | General<br>practitioner<br>dentists and<br>specialists | General<br>practitioner<br>dentists and<br>specialists | General<br>practitioner<br>dentists and<br>specialists | General<br>practitioner<br>dentists and<br>specialists | General<br>practitioner<br>dentists and<br>specialists |
| Indications<br>for Use | The Solea<br>system is<br>indicated for<br>the<br>following:<br>- Ablation<br>of hard<br>tissue<br>for<br>caries<br>removal<br>and<br>cavity<br>preparation<br>- Incision, | The Solea<br>system is<br>indicated for<br>the<br>following:<br>- Ablation<br>of hard<br>tissue<br>for<br>caries<br>removal<br>and<br>cavity<br>preparation.<br>- Incision, | The Solea<br>system is<br>indicated for<br>the following:<br>- Incision,<br>- Excision,<br>- Vaporization,<br>- Coagulation<br>and<br>- Hemostas<br>is of soft<br>tissue in<br>the oral<br>cavity. | General<br>Indications*<br>Class I, II, III, IV<br>and V cavity<br>preparation<br>Caries removal<br>Hard tissue<br>surface<br>roughening or<br>etching<br>Enameloplasty,<br>excavation of<br>pits and fissures<br>for placement of<br>sealants*<br>* For use on | The LightWalker<br>Er:YAG laser,<br>and its<br>accessories, are<br>intended for<br>use in dentistry,<br>dermatology<br>and other<br>surgical areas in<br>the following<br>procedures:<br>In dentistry, for:<br>- Intra-oral<br>soft tissue<br>surgery<br>(incision, |
| Convergent | Convergent | Convergent | Biolase | Fotona d.d. | |
| Dental, Inc. | Dental, Inc. | Dental, Inc. | Technology, Inc. | | |
| Vaporiza | Vaporiza | | pediatric | ablation | |
| tion, | tion, | | patients | coagulation | |
| Coagulat | Coagulat | | Root Canal Hard | ) | |
| ion and | ion and | | Tissue | Leukoplaki | |
| Hemosta | Hemosta | | Indications | a | |
| sis of | sis of | | Tooth | - Pulpotomy | |
| soft | soft | | preparation to | as adjunct | |
| tissue in | tissue in | | obtain access to | to root | |
| the oral | the oral | | root canal | canal | |
| cavity | cavity. | | Root canal | retreatmen | |
| - Cutting, | | | preparation | t | |
| shaving, | | | including | - Pulp | |
| contouri | | | enlargement | extirpation | |
| ng and | | | Root canal | - Removal of | |
| resectio | | | debridement | fibroma | |
| n of oral | | | and cleaning | - Removal of | |
| osseous | | | Root Canal | granulated | |
| tissue | | | Disinfection | tissue | |
| (bone) | | | Laser root canal | - Caries | |
| | | | disinfection after | removal, | |
| | | | endodontic | cavity | |
| | | | instrumentation | preparatio | |
| | | | Bone Surgical | n, enamel | |
| | | | Indications | roughening | |
| | | | Cutting, shaving, | - Sulcular | |
| | | | contouring and | debrideme | |
| | | | resection of oral | nt | |
| | | | osseous tissues…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.