Airlock® osteosynthesis plate systems are single use devices intended for fixation and stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes. The system may be used in both adult and pediatric patients.
Device Story
Airlock® osteosynthesis plate system consists of titanium (Ti-6Al-4V ELI) bone plates and 3mm/3.5mm locking and non-locking screws. Intended for permanent implantation to provide fixation and stabilization of small bone fractures, revisions, joint fusions, and reconstructions in hands, feet, wrists, ankles, fingers, and toes. Used by surgeons in clinical settings. Plates feature varying curvatures, lengths, and hole configurations to accommodate anatomical requirements; drill holes are aligned to prevent screw conflict. Device provides mechanical stability to bone segments during healing.
Clinical Evidence
No clinical data or animal studies were required for this submission. Biocompatibility was evaluated per ISO 10993-1 and Blue Book Memorandum #G95-1. Substantial equivalence is supported by bench testing and mechanical performance data.
Technological Characteristics
Materials: Wrought titanium 6-aluminium 4-vanadium alloy (ISO 5832-3). Mechanical standards: ASTM F382-99 (plates), ASTM F543-13 (screws). Components: Plates of varying shapes/lengths, 3mm/3.5mm locking and non-locking screws. Single-use, permanently implanted.
Indications for Use
Indicated for fixation and stabilization of fresh fractures, revision procedures, joint fusion, and reconstruction of small bones of the hand, feet, wrist, ankles, fingers, and toes in adult and pediatric patients.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes and a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The seal is presented in black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 13, 2015
NOVASTEP Mr. Gilles Audic, QA/RA Director Espace performance Alphasis Batiment C1-C2 35769 SAINT GREGOIRE France
Re: K151277
Trade/Device Name: Airlock® osteosynthesis plate system Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS Dated: May 21, 2015 Received: May 26, 2015
Dear Mr. Audic:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{1}------------------------------------------------
Page 2 - Mr. Gilles Audic
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Section
### 4: Indications for Use Statement
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration | | Form Approved: OMB No. 0910-0120 |
|-------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| <i>Indications for Use</i> | | Expiration Date: January 31, 2017<br>See PRA Statement below. |
| 510(k) Number (if known) | K151277 | |
| Device Name | Airlock® osteosynthesis plate systems | |
| Indications for Use (Describe) | Airlock® osteosynthesis plate systems are indicated for fixation and stabilization of fresh fractures, revision procedures, joint fusion and reconstruction of small bones of the hand, feet, wrist, ankles, fingers and toes. The system may be used in both adult and pediatric patients. | |
| Type of Use (Select one or both, as applicable) | <span> <svg height="12" width="12"> <polygon points="0,0 12,0 12,12 0,12" style="fill:none;stroke:black;stroke-width:1"></polygon> <polygon points="2,2 10,2 10,10 2,10" style="fill:black"></polygon> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) <span> <svg height="12" width="12"> <polygon points="0,0 12,0 12,12 0,12" style="fill:none;stroke:black;stroke-width:1"></polygon> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) | |
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
*"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."*
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for "novastep". The logo is in a pink color. To the left of the word "novastep" is a pink design that looks like two curved lines. To the right of the word "novastep" is a registered trademark symbol.
#### "510(k) Summary" as required by section 807.92(c)
| Submitter | NOVASTEP<br>Espace performance Alphasis<br>Batiment C1-C2<br>35769 SAINT GREGOIRE<br>France<br>Phone : + 33 (0)2 99 33 86 50<br>Fax : + 33 (0)9 70 29 18 95 |
|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact person | Mister Gilles AUDIC<br>QA / RA Director<br>Cell phone: +33 (0)6 30 93 96 08<br>e-mail: gilles.audic@novastep-ortho.com |
| Preparation date | May 12th 2015 |
| Trade name | Airlock® osteosynthesis plate system |
|---------------------|---------------------------------------------------------------------------------------------------------------------|
| Common Name | Plate, Fixation, Bone |
| Classification Name | Single / Multiple component metallic bone fixation appliances and<br>accessories (21CFR 888.3030, product code HRS) |
| Regulatory class | II |
| Legally marketed<br>predicate devices | 510(k) number: K083447<br>Device name: Anchorage® Bone Plate System<br>Original applicant: MEMOMETAL TECHNOLOGIES.<br>This predicate has not been subject to a design-related recall. |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description | Airlock® osteosynthesis plate systems are single-use bone fixation<br>devices intended to be permanently implanted. Plates are designed with<br>different shapes and are made of Titanium (Alloy Ti-6Al-4V ELI).<br>The system uses either 3mm or 3,5mm locking and non-locking screws. |
vastep - Siège Social / Service Client
| n o v a s t e p - S iège Social / Service Client
Espace Performanoe Alphasis - Bâtiment C1-C2 - 35769 SAINT-GRÉGOIRE Cedex - FFRANCE
Tél. +33 (0)2 99 33 86 50 - Fa
Tél. +33 (0)2 99 33 86 50 - Fax: +33 (0)2 30 96 35 61
Code A.P.E.: 3250A - N° SIRET: 752 292 797 00020-N° TVA Intracommunautaire FR 64 752 292 797
{4}------------------------------------------------
| | The drill hole are aligned to make sure there is no risk of conflict between<br>the screws.<br>The plates vary essentially through different curvatures, lengths, number<br>of holes and shape. |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended use | Airlock® osteosynthesis plate systems are single use devices intended for<br>fixation and stabilization of fresh fractures, revision procedures, joint<br>fusion and reconstruction of small bones of the hand, feet, wrist, ankles,<br>fingers and toes. The system may be used in both adult and pediatric<br>patients. |
| Comparison of the<br>technological<br>characteristics with the<br>predicate device | The new devices Airlock® osteosynthesis plate systems plates have similar<br>technological characteristics in terms of material (ISO5832-3 Implants for<br>surgery - Metallic material - Part 3: Wrought titanium 6-aluminium 4-<br>vanadium alloy) and mechanical characteristics (ASTM F382-99 Standard<br>Specification and Test Methods for Metallic Bone plates) and thus are<br>believed to be substantially equivalent to the predicate MEMOMETAL<br>Anchorage® Bone Plate System plates (K083447). |
| | The new devices Airlock® osteosynthesis plate systems screws have<br>similar technological characteristics in terms of material (ISO5832-3<br>Implants for surgery - Metallic material - Part 3: Wrought titanium 6-<br>aluminium 4-vanadium alloy) and mechanical characteristics (ASTM F543-<br>13 Standard Specification and Test Methods for Metallic Bone screws Part<br>A1, A2 & A3) and thus are believed to be substantially equivalent to the<br>predicate MEMOMETAL Anchorage® Bone Plate System screws<br>(K083447). |
| Performance data | The biocompatibility evaluation for new devices Airlock® osteosynthesis<br>plate systems was conducted in accordance with Blue Book<br>Memorandum #G95-1 "Use of International Standard ISO-10993,<br>'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing"<br>and International Standard ISO 10993-1 "Biological Evaluation of Medical<br>Devices - Part 1: Evaluation and Testing Within a Risk Management<br>Process" as recognized by FDA. |
| | The new devices Airlock® osteosynthesis plate systems have similar<br>technological characteristics in terms of design and mechanical<br>characteristics and thus are believed to be substantially equivalent to the<br>predicate MEMOMETAL Anchorage® Bone Plate System (K083447). |
| Indication for use | The Airlock® osteosynthesis plate systems are single use devices indicated<br>for fixation and stabilization of fresh fractures, revision procedures, joint<br>fusion and reconstruction of small bones of the hand, feet, wrist, ankles,<br>fingers and toes. The system may be used in both adult and pediatric<br>patients. |
| Clinical studies | Clinical studies was not required for this submission |
| Animal Study | Animal Study was not required for this submission<br>1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 - 1 |
| Conclusion | The Airlock® osteosynthesis plate systems are substantially equivalent to<br>their predicate devices MEMOMETAL Anchorage® Bone Plate System<br>(K083447), in terms of intended use and indications for use, material,<br>design and function. Any minor differences between these two devices<br>do not raise new questions of safety and effectiveness. |
vastep - Siège Social / Service Client
| n o v a st e p - Siège Social / Service Client
| Espace Perbrimance Alphasis - Bâtiment C1-C2 - 35769 SAINT-GRÉGOIRE Cedex - FFRANCE
| Tél. +33 (0)2 99 33 86 50 - Fax
Code A.P.E.: 3250A-N° SIRET 752 292 797 00020 - N° TVA Intracommunautaire FR 64 752 292
{5}------------------------------------------------
#### Traditional 510(k) premarket notification
# step®
| n o v a s t e p - S iège Social / Service Client
| Espace Performance Alphasis - Bâtiment C1-C2 - 35769 SAINT-GRÉGOIFE Cedex - FFANCE
| Tél. +33 (0)2 99 33 86 50 - F
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.