K151275 · Nvision Medical · HIH · Nov 13, 2015 · Obstetrics/Gynecology
Device Facts
Record ID
K151275
Device Name
MAKO Device
Applicant
Nvision Medical
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Nov 13, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Indications for Use
The MAKO Device is a hysteroscope accessory, placed through the working channel of a hysteroscope to obtain samples from the proximal fallopian tube for cytological evaluation.
Device Story
MAKO Device is a sterile, single-use hysteroscope accessory for collecting proximal fallopian tube cell samples. Device consists of a handle with a drive wheel and a catheter featuring an inflatable balloon. Physician inserts device through a commercially available hysteroscope working channel under visualization; positions distal tip at the fallopian tube ostium; advances balloon into the proximal tube. Balloon inflation and subsequent deflation create wrinkles/edges that scrape and collect cells from the tissue surface. Device is used in hospital/clinic settings. Output is a cell sample for cytological evaluation, aiding clinical diagnosis. Benefits include minimally invasive access to the proximal fallopian tube for diagnostic sampling.
Clinical Evidence
Prospective, single-arm study of 40 women (80 fallopian tubes) scheduled for tubal ligation. Primary endpoints: safety, device performance, and ability to collect adequate cytological samples. Results: 75/80 (93.75%) tubes successfully navigated; 71/75 (94.67%) accessible tubes yielded adequate samples. One adverse event (fallopian tube perforation) occurred in a subject with pre-existing tubal occlusion, requiring no intervention.
Technological Characteristics
Materials: stainless steel tube, Nylon tube, Nylon 12 sheath, Polycarbonate sheath knob. Principle: balloon-based mechanical tissue scraping. Energy: manual actuation via handle drive wheel. Form factor: catheter-based accessory for hysteroscope working channel. Connectivity: none. Sterilization: ethylene oxide (EO).
Indications for Use
Indicated for women undergoing cytological evaluation of the proximal fallopian tube via hysteroscopic access.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 13, 2015
Nvision Medical % Cindy Domecus Principal Domecus Consulting Services LLC 1171 Barroilhet Drive Hillsborough, California 94010
Re: K151275
Trade/Device Name: MAKO Device Regulation Number: 21 CFR 884.1690 Regulation Name: Hysteroscope And Accessories Regulatory Class: Class II Product Code: HIH Dated: October 14, 2015 Received: October 15, 2015
Dear Cindy Domecus:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151275
Device Name MAKO Device
Indications for Use (Describe)
The MAKO Device is a hysteroscope accessory, placed through the working channel of a hysteroscope to obtain samples from the proximal fallopian tube for cytological evaluation.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size: 10px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "nVision" in black font, with a light blue circle in place of the "V". The "n" is lowercase, while the rest of the word is in uppercase. The logo appears to be for a company or product called "nVision".
# 510(k) Summary
#### SUBMITTER I.
510(k) Owner: nVision Medical 250 King St, #676 San Francisco, CA 94107 Surbhi Sarna, CEO Telephone: (408) 655-3577 surbhi@nvisionmedical.com
Contact Person: Cindy Domecus, R.A.C. (US &EU) Principal, Domecus Consulting Services LLC (650) 343-4813 (phone) (650) 343-7822 (fax) DomecusConsulting@comcast.net
Date Prepared: May 11, 2015
#### II. DEVICE
| Name of Device: | MAKO Device |
|-----------------------|-----------------------------------------|
| Common or Usual Name: | Hysteroscope Accessory |
| Classification Name: | Hysteroscope and accessories (884.1690) |
| Regulatory Class: | II |
| Product Code: | HIH |
#### III. PREDICATE DEVICE
Substantial equivalence was claimed to the following predicate device:
| 510(k) Number,<br>Company | Trade Name | Model<br>Number |
|---------------------------|-------------------------------|-----------------|
| K132384<br>EndoSee | U-Scope™ 8000 HSC+EMB Cannula | 8000 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Nvision. The logo consists of the letter 'n' in black, followed by a light blue circle with a white 'V' shape inside. The rest of the word 'vision' is written in black to the right of the circle. The font is sans-serif and appears to be bolded.
#### IV. DEVICE DESCRIPTION
The MAKO Device is a hysteroscope accessory intended to collect cell samples from the proximal fallopian tube. The device is comprised of a catheter and a handle. The catheter includes a balloon, a shaft (which is made up of a stainless steel tube and a Nylon tube), a sheath (Nylon 12), and a sheath knob (Polycarbonate). The handle includes a drive wheel and an extension tube that is attached to a luer in the handle body. The extension tube attaches to a commercially available inflation device via a commercially available 3-way stopcock.
In summary, the physician inserts the MAKO Device into the working channel of the hysteroscope until the distal tip of the catheter is positioned immediately proximal to the ostium of the fallopian tube. Then, the balloon is advanced into the proximal portion of the fallopian tube and cells are collected on its surface.
The MAKO Device is a sterile, single-use device. The device is terminally sterilized using ethylene oxide (EO).
The device is intended for use in the hospital/clinic/physician's office. The duration and type of contact is classified as follows:
- Category: External Communicating Device ●
- Contact: Mucosal membrane ●
- Duration: A Less than 24 hours ●
#### INTENDED USE/INDICATIONS FOR USE V.
The indications for use for the subject device are as follows:
The MAKO Device is a hysteroscope accessory, placed through the working channel of a hysteroscope to obtain samples from the proximal fallopian tube for cytological evaluation.
The indications for use are different than the predicate, in that the indication for use for the subject device 1) does not contain information pertaining to indications for the hysteroscope, as the device does not include a hysteroscope, but rather, is used through the working channel of a commercially available hysteroscope, and 2) refers to the collection of samples from the proximal fallopian tube instead of the uterus.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "nVision" with a stylized "V". The "n" is in black, the "V" is a light blue circle with a white "V" in the center, and the "ision" is in black. The font is sans-serif and the text is left-aligned. The image is simple and clean.
Not including information pertaining to the indications for use for the hysteroscope is not a substantive issue, as the subject device will be used through the working channel of a commercially available hysteroscope, which will include indications for use for the hysteroscope that were cleared by FDA.
Since the proximal fallopian tube represents a different anatomical location than the uterus, the FDA guidance document, "General/Specific Intended Use," (issued November 4, 1998) was used to determine whether the intended use of the subject device falls within the scope of the predicate device. Indications for use that specify a particular anatomical site, but do not imply diagnosis or therapy of a specific disease entity, typically fall within the scope of the general intended use for the purposes of determining substantial equivalence. When the decision-making criteria from the guidance are applied, the intended use comparison between the subject and predicate devices supports equivalence. In regards to risk, the specific use of the subject device does not introduce new risks not normally associated with endometrial sampling procedures. Furthermore, the proximal fallopian tube is housed within the uterine wall, so it is contiguous with the uterus. Therefore, the collection of samples from the proximal fallopian tube compared to the uterus does not represent a new intended use.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS VI.
The mechanism of action of the subject device and the predicate device is similar in that both devices extract samples from the reproductive tract by scraping against the tissue of the target anatomy. The devices are also similar in that they consist of a handle, cannula/catheter body, and, most distally, a portion that collects the sample. For all devices, this distal portion is actuated by a mechanism on the handle and a hysteroscopic view is required.
For the subject device and predicate devices, the handle is meant to 1) provide an ergonomic way of gripping the device under hysteroscopic visualization, and 2) to actuate the distal portion of the device. which then collects the sample. The cannula/catheter body of the devices is meant to provide the device stability and reach as it is placed through the working channel of the hysteroscope.
The most pertinent technological assessment is the distal end of the predicate and subject devices; specifically, the portion of the device meant to collect the sample. The primary difference between the subject and predicate devices is that the subject device uses a balloon to achieve access to the fallopian tube and to collect cells, while the U-Scope device scrapes against the uterine wall with a curette. However, the technological characteristics of the subject device do not introduce new risks when compared to the uterine sampling mechanism of the predicate device. For the subject device, wrinkles in the balloon that form as the balloon deflates create multiple edges in order to collect samples. These edges work in a manner similar to the edges of the curette of the U-Scope predicate device. Therefore, the above technological differences do not raise different questions of safety or effectiveness, as the risks associated with using the subject and predicate devices (i.e., uterine
CONFIDENTIAL
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "nVision" in black font, with the "V" in "Vision" being a light blue color. The "n" is lowercase, while the rest of the letters are uppercase. The logo is simple and modern.
perforation) are the same.
#### PERFORMANCE DATA VII.
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility testing
- Cytotoxicity
- Vaginal irritation ●
- Sensitization
### Mechanical testing
- Dimensional (sample size of n=15 was used for each test)
- Tensile strength of balloon, catheter, and shaft components (sample size of n=15 was used for each test)
- Simulated use testing in a uterine model, including compatibility with inflation device, compatibility with hysteroscope, and ability to access proximal fallopian tube (sample size of n=29 was used)
- . Balloon inflation pressure (sample size of n=29 was used)
- Device deployment (sample size of n=15 was used) ●
## Shelf Life Testing
Accelerated aging testing was conducted on the subject device to substantiate a shelf life of one year. The shelf life study was performed following sterilization in the original packaging. The following device specifications were evaluated in the accelerated aging study to demonstrate the subject device maintains its specifications throughout the duration of its shelf life:
- Seal integrity (visual assessment) ●
- Pouch leakage
- Seal strength
- Device dimensions
- Inflation pressure ●
- Inflated balloon dimensions ●
- Tensile strength (balloon, catheter, and shaft components) .
### Human factors and usability testing
- o Physician feedback on handle ergonomics, drive wheel ergonomics, hysteroscopic visibility of balloon deployment, and ability to follow IFU
- A sample size of 3 to 7 physicians per parameter was used .
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "nVision" in black font. The "n" is lowercase, and the "V" is partially covered by a light blue circle. The rest of the word is in lowercase.
# Clinical testing
A prospective, single-arm clinical study was conducted to demonstrate the safety and effectiveness of the subject device in navigating the proximal fallopian tube and obtaining cell samples. A total of 40 women (80 fallopian tubes), previously scheduled to undergo a tubal ligation, were enrolled in the study. The following performance characteristics were evaluated in the study:
- Safetv
- Device Performance
- Ability of device to access target anatomy
- Ability of the device to collect samples ●
- Ability to obtain samples that are adequate for cytological evaluation ●
The results demonstrated that the subject device was successfully able to navigate the fallopian tubes (75/80 successful) and collect adequate cell samples for cytology (adequate samples in 71/75 of accessible tubes). A single fallopian tube perforation, requiring no further medical intervention, was noted in one subject. Pre-existing proximal tubal occlusion was confirmed in this subject.
#### VIII. CONCLUSION
The bench and clinical performance data provided in support of the 510(k) confirm that the subject device is as safe and effective and substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.