SD GlucoNFC Blood Glucose Monitoring System, SD GlucoNFC Multi Blood Glucose Monitoring System
Applicant
Sd Biosensor
Product Code
NBW · Clinical Chemistry
Decision Date
Nov 24, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
SD GlucoNFC Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, forearm or upper arm. SD GlucoNFC Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. It is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The SD GlucoNFC Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes. The SD GlucoNFC Blood Glucose Monitoring System is not for use in neonates. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly). SD Gluco NFC Blood Glucose Test Strips are for use with SD GlucoNFC Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, or upper arm. SD GlucoNavii Control Solution is intended for Quality Control of the SD GlucoNFC Blood Glucose Monitoring System. The control solution helps to check that the meter and test strips are working together properly and that the test is performing correctly.. SD GlucoNFC Multi Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, forearm, or upper arm and venous whole blood. The SD GlucoNFC Multi Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system should only be used with auto-disabling, single-use landing devices. The SD GlucoNFC Multi Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes. The SD GlucoNFC Multi Blood Glucose Monitoring System is not for use in neonates. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly). SD GlucoNFC Multi Blood Glucose Test Strips are for use with SD GlucoNFC Multi Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, or upper arm and venous whole blood. SD GlucoNFC Multi Blood Glucose Test Strips are for use with SD GlucoNFC Multi Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, palm, forearm or upper arm, and venous whole blood. SD GlucoNavii Control Solution is intended for Quality Control of the SD GlucoNFC Blood Glucose Monitoring System. The control solution helps to check that the meter and test strips are working together properly and that the test is performing correctly.
Device Story
System measures glucose in capillary/venous blood using electrochemical biosensor technology; test strips contain Glucose Dehydrogenase (GDH) and potassium ferrocyanide. Blood sample triggers electric current proportional to glucose concentration; meter displays result in mg/dL. Single-patient version for home use; multi-patient version for professional settings using auto-disabling lancing devices. Includes NFC technology for data transfer to smart devices; USB connectivity also available. Users view results on LCD; system stores 7, 14, 30-day averages. Benefits include diabetes control monitoring. Healthcare providers use multi-patient version in clinics to manage patient programs. System includes check strips and control solution for quality assurance.
Clinical Evidence
Clinical study with 120 lay users evaluated accuracy of fingertip and AST (palm, forearm, upper arm) measurements against YSI reference. Results showed 100% of samples within ±15 mg/dL (for <75 mg/dL) and 100% within ±20% (for ≥75 mg/dL). Bench testing included precision (CV 2.0-3.7%), linearity (R² > 0.99), interference studies, altitude testing (up to 11,480 ft), and hematocrit range validation (10-70%). Infection control robustness validated for 10,950 cycles.
Technological Characteristics
Electrochemical biosensor using GDH enzyme. Dimensions: 48mm x 90mm x 15mm. Power: 3V CR2032 battery. Connectivity: NFC and USB. Operating range: 20-600 mg/dL. Hematocrit range: 10-70%. Standards: IEC 61010-1, IEC 61010-2-101, IEC 60601-1-2, FCC Part 15.101.
Indications for Use
Indicated for quantitative blood glucose measurement in people with diabetes. Single-patient version for home use (capillary blood); multi-patient version for professional healthcare settings (capillary and venous blood). Not for neonates, not for diagnosis/screening. Alternative site testing only during steady-state.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
K151265
B. Purpose for Submission:
New Submission
C. Measurand:
Capillary whole blood glucose from fingertip, palm, forearm, or upper arm.
D. Type of Test:
Quantitative Amperometric Assay; glucose dehydrogenase - flavin adenine dinucleotide (GDH-FAD)
E. Applicant:
SD Biosensor Inc.
F. Proprietary and Established Names:
SD GlucoNFC Blood Glucose Monitoring System
SD GlucoNFC Multi Blood Glucose Monitoring System
G. Regulatory Information:
1. Regulation section:
862.1345, Glucose Test System
862.1660, Quality control material (assayed and unassayed)
2. Classification:
Class II
Class I (reserved)
{1}
3. Product code:
NBW - System, Test, Blood Glucose, Over-the-Counter
LFR - Glucose Dehydrogenase, Glucose
JJX - Single (specified) analyte controls (assayed and unassayed)
4. Panel:
(75) Chemistry
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
SD GlucoNFC Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, forearm or upper arm. SD GlucoNFC Blood Glucose Monitoring System is intended to be used by a single person and should not be shared. It is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.
The SD GlucoNFC Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes. The SD GlucoNFC Blood Glucose Monitoring System is not for use in neonates. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
SD Gluco NFC Blood Glucose Test Strips are for use with SD GlucoNFC Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, or upper arm.
SD GlucoNavii Control Solution is intended for Quality Control of the SD GlucoNFC Blood Glucose Monitoring System. The control solution helps to check that the meter and test strips are working together properly and that the test is performing correctly..
SD GlucoNFC Multi Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood from fingertip, palm, forearm, or upper arm and venous whole blood. The SD GlucoNFC Multi Blood Glucose Monitoring System is intended for testing outside the body (in vitro diagnostic use) and is intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. This system should
{2}
only be used with auto-disabling, single-use landing devices.
The SD GlucoNFC Multi Blood Glucose Monitoring System should not be used for the diagnosis of or screening for diabetes. The SD GlucoNFC Multi Blood Glucose Monitoring System is not for use in neonates. Alternative site testing should be done only during steady-state times (when glucose is not changing rapidly).
SD GlucoNFC Multi Blood Glucose Test Strips are for use with SD GlucoNFC Multi Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, or upper arm and venous whole blood.
SD GlucoNFC Multi Blood Glucose Test Strips are for use with SD GlucoNFC Multi Blood Glucose Meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, palm, forearm or upper arm, and venous whole blood.
SD GlucoNavii Control Solution is intended for Quality Control of the SD GlucoNFC Blood Glucose Monitoring System. The control solution helps to check that the meter and test strips are working together properly and that the test is performing correctly.
3. Special conditions for use statement(s):
- For in vitro diagnostic use.
- Do not use the system to test neonates. It has not been validated for neonatal use.
- For the SD Gluco NFC BGMS: Critically ill patients should not be tested with this blood glucose meter.
- For the SD GlucoNFC Multi BGMS: This device has not been evaluated in critically ill patients.
- Not for diagnosis or screening of diabetes mellitus.
- Inaccurate results may occur in severely hypotensive individuals or patients in shock. Inaccurate results may occur for individuals experiencing a hyperglycemic hyperosmolar state, with or without ketosis.
- Severe dehydration resulting from excessive water loss may cause false low results
- Alternative site testing (AST) should only be done during steady-state times (when glucose is not changing rapidly).
- AST should not be used to calibrate continuous glucose monitors or in insulin dose calculations.
- For the SD GlucoNFC Multi BGMS only: For use with single-use auto-disabling landing devices.
4. Special instrument requirements:
SD GlucoNFC meter
SD GlucoNFC Multi meter
{3}
4
I. Device Description:
SD GlucoNFC and SD GlucoNFC Multi Blood Glucose Monitoring Systems (BGMS) are over-the-counter and prescription blood glucose monitoring systems, respectively. The SD GlucoNFC BGMS is indicated for single-patient use at home and should not be shared, while the SD GlucoNFC Multi BGMS is for multi-patient use in a professional healthcare setting, in order to help monitor the effectiveness of diabetes control. The devices contain near field communication (NFC) technology.
The BGMS comes with the SD GlucoNFC or SDGlucoNFC Multi Blood Glucose Meter, one level of SD Navii Glucose Control Solution (Level 2), and SD GlucoNFC or SD GlucoNFC Multi Blood Glucose Test Strips. The SD Navii Glucose Control Solution is used to verify the performance of the SD GlucoNFC or SD GlucoNFC Multi BGMS. A second level of control (Level 3) is available for purchase separately. The device comes with a SD Glucose check strip, which is used to check the electronic performance of the meter. Use of the check strip does not take the place of running quality control solutions.
J. Substantial Equivalence Information:
1. Predicate device name(s):
SD GlucoMentor BGMS
SD GlucoMentor Multi BGMS
SD Check Gold Control Solutions
2. Predicate 510(k) number(s):
K123517
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device
SD GlucoNFC
K151265 | Predicate Device
SD GlucoMentor BGMS
K123517 |
| Intended Use | Monitoring glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, or upper arm by people with diabetes at home to be used as an aid to monitor the effectiveness of diabetes control | Same |
| Test Time | 5 seconds | Same |
| Measuring Range | 20-600 mg/dL | Same |
| Test Principle | Electrochemical biosensor | Same |
{4}
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device SD GlucoNFC K151265 | Predicate Device SD GlucoMentor BGMS K123517 |
| Sample Type | Fresh capillary whole blood | Same |
| Sample Application | Test strip capillary draw | Same |
| Calibration | Plasma-calibrated | Same |
| Coding | none | Same |
| Power Source | 3V CR2032 Battery x1 (Replaceable) | Same |
| Monitor | LCD display | Same |
| Backlight | No | Same |
| Battery Life | Approximately 1000 Tests | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device SD GlucoNFC K151265 | Predicate Device SD GlucoMentor BGMS K123517 |
| Hematocrit range | 10-70% | 10-60% |
| Operating Temperature | 46-113°F (8-45°C) | 50-113°F (10-45°C) |
| Operating Humidity | 10-90% RH | 15-90% RH |
| Operating Altitude | Up to 11, 480 ft. | Up to 11,351 ft. |
| Sample volume | 0.5 μL | 0.3 μL |
| Test Strips | SD GlucoNFC Blood Glucose Test Strips | SD GlucoMentor Blood Glucose Test Strips |
| Test Strip Technology | Glucose Dehydrogenase (FAD) | Glucose Oxidase (GOD) |
| PC Link Feature | USB Cable or NFC Reader/Writer | USB Cable |
| Smart device link feature | Yes | No |
| Memory capacity | 300 test results | 500 test results |
| Meter Dimensions | 48 mm x 90 mm x 15 mm | 47 mm x 95 mm x 17.5 mm |
| Meter weight | 50 g with battery | 57.5 g with battery |
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device SD GlucoNFC Multi BGMS K151265 | Predicate Device SD GlucoMentor Multi NFC K123517 |
| Intended Use | Monitoring glucose in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, or upper arm and venous whole blood. Is | Same |
{5}
| Similarities | | |
| --- | --- | --- |
| Item | Candidate Device SD GlucoNFC Multi BGMS K151265 | Predicate Device SD GlucoMentor Multi NFC K123517 |
| | intended for multiple-patient use in professional healthcare settings as an aid to monitor the effectiveness of diabetes control program. | |
| Test Time | 5 seconds | Same |
| Measuring Range | 20-600 mg/dL | Same |
| Test Principle | Electrochemical biosensor | Same |
| Sample Application | Test strip capillary draw | Same |
| Sample Type | Fresh capillary and venous whole blood | Same |
| Calibration | Plasma-calibrated | Same |
| Test Strip Storage Conditions | Temperature: 2-32°C (36-90°F)
Humidity: 10-90% RH | Temperature: 2-32°C (36-90°F) |
| Power Source | 3V CR2032 Battery x1 (Replaceable) | Same |
| Monitor | LCD display | Same |
| Backlight | No | Same |
| Battery Life | Approximately 1000 Tests | Same |
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device SD GlucoNFC Multi BGMS K151265 | Predicate Device SD GlucoMentor Multi NFC K123517 |
| Hematocrit range | 10-70% | 10-60% |
| Operating Temperature | 46-113°F (8-45°C) | 50-113°F (10-45°C) |
| Operating Humidity | 10-90% RH | 15-90% RH |
| Operating Altitude | Up to 11, 480 ft. | Up to 11,351 ft. |
| Sample volume | 0.5 μL | 0.3 μL |
| Test Strips | SD GlucoNFC Multi Blood Glucose Test Strips | SD GlucoMentor Multi Blood Glucose Test Strips |
| Test Strip Technology | Glucose Dehydrogenase (FAD) | Glucose Oxidase (GOD) |
| PC Link Feature | USB Cable or NFC Reader/Writer | USB Cable |
| Smart device link feature | Yes | No |
| Memory capacity | 300 test results | 500 test results |
| Meter Dimensions | 48 mm x 90 mm x 15 mm | 47 mm x 95 mm x 17.5 mm |
{6}
| Differences | | |
| --- | --- | --- |
| Item | Candidate Device
SD GlucoNFC Multi BGMS
K151265 | Predicate Device
SD GlucoMentor Multi NFC
K123517 |
| Meter weight | 50 g with battery | 57.5 g with battery |
| Similarities and differences of the control solution | | |
| --- | --- | --- |
| | Candidate Device
SD GlucoNavii Control Solution | Predicate Device
SD Check Gold Control Solution,
k123517 |
| Matrix | Viscosity-adjusted, aqueous liquid. | Same |
| Levels | Level 2 and Level 3 | Level M and Level H |
## K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A2, Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline–Second Edition
CLSI EP6-A, Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline
CLSI EP07-A2, Interference Testing in Clinical Chemistry; Approved Guideline – Second Edition
IEC 61010-1:2001, Safety requirements for electrical equipment for measurement, control, and laboratory use – Part 1: General requirements (Second Edition)
IEC 61010-2-101:2002, Safety requirements for electrical equipment for measurement, control and laboratory use - Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment
IEC 61326-2-6:2005, Electrical equipment for measurement, control and laboratory use – EMC requirements - Part 2-6: Particular requirements - In vitro diagnostic (IVD) medical equipment
IEC 60068-2-64:2008, Environmental testing – Part 2-64: Tests – Test Fh: Vibration, broadband random and guidance
ISO 14971:2007, Medical Devices – Application of Risk Management to Medical Devices
## L. Test Principle:
The SD GlucoNFC and SD GlucoNFC Multi Blood Glucose Monitoring Systems use an amperometric biosensor technology to measure the glucose level in human blood. The reaction between glucose and the materials contained in the test strip, GDH-FAD, and potassium ferrocyanide, results in an electrical current. The glucose concentration is calculated from this current.
{7}
8
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
Performance testing was conducted on the SD GlucoNFC Blood Glucose Meter only. This is acceptable because the only differences between the SD GlucoNFC and SD GlucoNFC Multi BGMS are the name and intended use (single-patient vs. multiple-patient use).
a. Precision/Reproducibility:
Repeatability:
Venous blood was spiked with five different glucose concentrations (30-50, 51-110, 111-150, 151-250, and 251-400mg/dL) and tested on ten SD GlucoNFC meters and three lots of test strips. Ten replicates were tested per meter per glucose concentration. The results from all strip lots are summarized below:
| Glucose Level (mg/dL) | Lot | N | Mean (mg/dL) | SD (mg/dL) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| 30-50 | 1 | 100 | 48.8 | 1.0 | 2.0 |
| | 2 | 100 | 48.8 | 1.1 | 2.3 |
| | 3 | 100 | 48.9 | 1.0 | 2.1 |
| 51-110 | 1 | 100 | 82.9 | 2.0 | 2.5 |
| | 2 | 100 | 82.7 | 2.0 | 2.4 |
| | 3 | 100 | 82.5 | 1.8 | 2.2 |
| 111-150 | 1 | 100 | 143.6 | 4.8 | 3.3 |
| | 2 | 100 | 143.7 | 3.7 | 2.5 |
| | 3 | 100 | 143.6 | 3.8 | 2.6 |
| 151-250 | 1 | 100 | 223.6 | 6.6 | 3.0 |
| | 2 | 100 | 222.6 | 7.2 | 3.2 |
| | 3 | 100 | 223.9 | 7.1 | 3.2 |
| 251-400 | 1 | 100 | 333.9 | 8.2 | 2.5 |
| | 2 | 100 | 333.0 | 7.3 | 2.2 |
| | 3 | 100 | 334.4 | 7.2 | 2.2 |
Intermediate Precision:
Intermediate Precision was evaluated using three lots of test strips and ten SD GlucoNFC meters. Three levels of Glucose control solutions were used. For each level of control, ten replicates were taken each day for ten days, so that 100 individual measurements were generated per control level. The results from all strip lots are summarized below:
{8}
| Control Level | Strip Lot | N | Mean (mg/dL) | SD (mg/dL) | CV (%) |
| --- | --- | --- | --- | --- | --- |
| Level 1 | 1 | 100 | 62.4 | 1.5 | 2.4 |
| | 2 | 100 | 62.6 | 1.3 | 2.1 |
| | 3 | 100 | 62.5 | 1.5 | 2.3 |
| Level 2 | 1 | 100 | 136.1 | 3.3 | 2.4 |
| | 2 | 100 | 136.9 | 3.5 | 2.5 |
| | 3 | 100 | 136.6 | 3.4 | 2.5 |
| Level 3 | 1 | 100 | 232.6 | 8.6 | 3.7 |
| | 2 | 100 | 233.1 | 8.2 | 3.5 |
| | 3 | 100 | 233.6 | 7.0 | 3.0 |
# b. Linearity/assay reportable range:
The claimed measuring range for this device is $20 - 600\mathrm{mg / dL}$ . Linearity was evaluated using three lots of test strips, five SD GlucoNFC meters and 11 venous whole blood samples with the following glucose concentrations: 16.8, 39.3 60.4, 86.4 119.5156.8, 227.3, 299.3, 367.8, 436.5, 512.5, 573.8, 647.3, and $720.3\mathrm{mg / dL}$ , obtained by spiking with glucose solution. Each glucose level was analyzed 5 times on each meter. Linear regression analysis for each meter compared to results obtained using YSI resulted in the following:
Lot 1: $\mathrm{y} = 0.999\mathrm{x} - 0.199;\mathrm{R}2 = 0.998$
Lot 2: $\mathrm{y} = 0.994\mathrm{x} + 1.007;\mathrm{R}2 = 0.998$
Lot 3: $\mathrm{y} = 0.986\mathrm{x} + 2.394;\mathrm{R}2 = 0.997$
The results of the study support the sponsor's claimed glucose measurement range of $20 - 600\mathrm{mg / dL}$ .
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The SD GlucoNFC and SD GlucoNFC Multi systems are traceable to the NIST SRM 917c reference material. The method comparison study was performed using the YSI 2300 Glucose analyzer as the reference method (see Section M.2.a.)
# Test strip stability:
Real time closed and open vial stability for the SD GlucoNFC and SD GlucoNFC Multi blood glucose test strips were assessed in real-time studies. Study protocols and acceptance criteria were reviewed and found acceptable to support the sponsor's claimed closed vial (shelf life) stability of 24 months, and open vial stability of 6 months when stored under the recommended storage conditions of $36^{\circ}\mathrm{F}$ to $90^{\circ}\mathrm{F}$ (2- $32^{\circ}\mathrm{C}$ ) and relative humidity of $10 - 90\%$ .
{9}
10
# Control solution stability:
Real time closed and open vial stability for the control solutions were assessed in real-time testing. Protocols and acceptance criteria were reviewed and found acceptable to support the sponsor's claimed closed vial (shelf life) stability of 24 months, and open vial stability of 3 months when stored at the recommended storage temperature of 46-86°F (8-30°C).
# Value assignment of controls:
Three targeted concentrations of the controls are prepared gravimetrically and analyzed using the YSI 2300 STAT Plus. Each level of the control solution is tested 40 times. The mean along with SD and CV are used to establish the ranges for each level which are then provided on the test strip vial label.
## d. Detection limit:
The measuring range of the device is 20-600 mg/dL. This range is validated via linearity study. See section M.1.b.
## e. Analytical specificity:
The sponsor performed interference studies with spiked venous blood samples at 2 different glucose concentrations (YSI reference values of 65 and 250 mg/dL). The samples were divided into 4 groups (1 control and 3 test conditions with 3 different levels of interferent). Each sample was measured twice by the reference method and once with five GlucoNFC meters per glucose concentration. The sponsor defines no significant interference as bias < ±10% for the test compared to control samples. The following table lists the concentrations of each substance at which no significant interference was detected.
| Potential Interferent | Highest concentration tested at which no significant interference is observed (mg/dL) |
| --- | --- |
| Acetaminophen | 6 |
| Ascorbic acid | 4 |
| Bilirubin | 35 |
| Uric acid | 5 |
| Triglyceride | 1500 |
| Total Cholesterol | 500 |
| Acetyl-salicylic acid | 120 |
| Ibuprofen | 50 |
| Fructose | 15 |
| Tetracycline | 5 |
| Tolbutamide | 100 |
| Dopamine | 2.5 |
{10}
| Methyl-dopa | 2 |
| --- | --- |
| Creatinine | 30 |
| Urea | 500 |
| Tolazamide | 5 |
| Warfarin | 2 |
| Levodopa | 4 |
| Maltotetraose | 120 |
| Mannose | 5 |
| Lactose | 25 |
| Mannitol | 800 |
| Sorbitol | 10 |
| Xylitol | 25 |
| Sodium salicylate | 70 |
| Xylose | 30 |
| Maltotriose | 240 |
| Hemoglobin | 200 |
| Gentistic acid | 500 |
| Glutathione | 12.5 |
| Galactose | 60 |
| Ethanol | 400 |
| Maltose | 500 |
To address potential interference from acetaminophen $\geq 6\mathrm{mg / dL}$ , ascorbic acid $\geq 4$ $\mathrm{mg / dL}$ , uric acid $\geq 5\mathrm{mg / dL}$ , and xylose $\geq 30\mathrm{mg / dL}$ , the labeling contains the following statements:
If you are taking acetaminophen containing drugs (e.g. Tylenol), Vitamin C (ascorbic acid), you may get inaccurate results.
If you have a disease or condition in which uric acid levels in your blood may be elevated, such as gout, you may get inaccurate results with this system.
This system should not be used when undergoing xylose absorption tests.
f. Assay cut-off:
Not applicable
# 2. Comparison studies:
a. Method comparison with predicate device:
System Accuracy was assessed in a study using ten SD GlucoNFC meters and three lots of test strips. Results from the meter were compared to results from the reference method, YSI. Samples from the finger, palm, forearm, and upper arm of 115 participants were collected. The total range of samples tested was from 27.5 - 510 mg/dL. To achieve glucose concentrations less than $50\mathrm{mg / dL}$ , 5 finger samples were
{11}
allowed to glycolize. To achieve glucose concentrations greater than $400\mathrm{mg / dL}$ , 5 finger samples were spiked. Results of the meter measurements relative to YSI are summarized below:
| | Glucose concentrations <75 mg/dL | | |
| --- | --- | --- | --- |
| Site | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| Finger | 14/20 (70%) | 20/20 (100%) | 20/20 (100%) |
| Palm | 10/15 (66.7%) | 15/15 (100%) | 15/15 (100%) |
| Forearm | 10/15 (66.7%) | 15/15 (100%) | 15/15 (100%) |
| Upper arm | 10/15 (66.7%) | 15/15 (100%) | 15/15 (100%) |
| | Glucose concentrations ≥75 mg/dL | | | |
| --- | --- | --- | --- | --- |
| Site | Within ± 5 % | Within ± 10% | Within ± 15% | Within ± 20% |
| Finger | 50/100 (50%) | 90/100 (90%) | 100/100 (100%) | 100/100 (100%) |
| Palm | 46/95 (48.4%) | 86/95 (90.5%) | 95/95 (100%) | 95/95 (100%) |
| Forearm | 45/95 (47.4%) | 86/95 (90.5%) | 95/95 (100%) | 95/95 (100%) |
| Upper arm | 39/95 (41.1%) | 84/95 (88.4%) | 95/95 (100%) | 95/95 (100%) |
Linear regression results:
| | Finger | Palm | Forearm | Upper arm |
| --- | --- | --- | --- | --- |
| Slope | 0.9883 | 0.9841 | 0.9875 | 0.9853 |
| Y-intercept | 2.6706 | 1.3877 | 1.0614 | 1.7441 |
| R | 0.9953 | 0.9936 | 0.9933 | 0.9917 |
| R² | 0.9907 | 0.9873 | 0.9866 | 0.9834 |
# b. Matrix comparison:
Venous whole blood samples from 120 volunteers were collected into sodium heparin tubes. The blood glucose concentration was first measured by YSI, then the blood glucose concentration was tested using three lots of test strips on 15 SD GlucoNFC meters. The total range of samples tested was $25.9 - 513\mathrm{mg / dL}$ . To achieve concentrations of glucose $< 50\mathrm{mg / dL}$ , 5 samples were allowed to glycolize. To achieve concentrations of glucose $>400\mathrm{mg / dL}$ , 5 samples were spiked. The results relative to the reference, YSI, are summarized below:
{12}
13
| | Glucose concentrations
<75 mg/dL | | |
| --- | --- | --- | --- |
| Venous Blood | Within ± 5 mg/dL | Within ± 10 mg/dL | Within ± 15 mg/dL |
| | 18/24 (75%) | 24/24 (100%) | 24/24 (100%) |
| Glucose concentrations
≥75 mg/dL | | | |
| --- | --- | --- | --- |
| Within ± 5 % | Within ± 10% | Within ± 15% | Within ± 20% |
| 57/96 (59.4%) | 89/96 (92.7%) | 96/96 (100%) | 96/96 (100%) |
Linear Regression results:
| Slope | Y-intercept | R | R2 |
| --- | --- | --- | --- |
| 0.9922 | 1.0430 | 0.9958 | 0.9917 |
An additional study was done to support use of the device with venous whole blood when collected with additional anticoagulants. The sponsor collected venous blood samples from 100 volunteers into 4 separate tubes, containing sodium heparin, lithium heparin, EDTA-K2 and EDTA-K3, using 3 test strip lots. The total range of samples tested was 38 to 475 mg/dL. To achieve glucose concentrations < 50 mg/dL and > 400 mg/dL samples were spiked and allowed to glycolyze. The meter results relative to the reference, YSI, are summarized below:
Sodium heparin vs. reference
| Glucose concentration
<75 mg/dL | | | Glucose concentration
≥75 mg/dL | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Within ±5mg/dL | Within ±10mg/dL | Within ±15 mg/dL | Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 12/16 (75.0%) | 16/16 (100%) | 16/16 (100%) | 50/84 (59.5%) | 84/84 (100%) | 84/84 (100%) | 84/84 (100%) |
Lithium heparin
| Glucose concentration
<75 mg/dL | | | Glucose concentration
≥75 mg/dL | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Within ±5mg/dL | Within ±10mg/dL | Within ±15 mg/dL | Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| 15/16 (93.8%) | 16/16 (100%) | 16/16 (100%) | 58/84 (69.0%) | 81/84 (96.4%) | 84/84 (100%) | 84/84 (100%) |
{13}
14
EDTA-K2
| Glucose concentration <75 mg/dL | | | Glucose concentration ≥75 mg/dL | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Within ±5mg/dL | Within ±10mg/dL | Within ± 15 mg/dL | Within ± 5 % | Within ± 10% | Within ± 15% | Within ± 20% |
| 13/16 (81.3%) | 16/16 (100%) | 16/16 (100%) | 61/84 (72.6%) | 82/84 (97.6%) | 84/84 (100%) | 84/84 (100%) |
EDTA-K3
| Glucose concentration <75 mg/dL | | | Glucose concentration ≥75 mg/dL | | | |
| --- | --- | --- | --- | --- | --- | --- |
| Within ±5mg/dL | Within ±10mg/dL | Within ± 15 mg/dL | Within ± 5 % | Within ± 10% | Within ± 15% | Within ± 20% |
| 15/16 (93.8%) | 16/16 (100%) | 16/16 (100%) | 66/84 (78.6%) | 82/84 (97.6%) | 84/84 (100%) | 84/84 (100%) |
Linear regression results:
| | Na heparin | Li heparin | EDTA-K2 | EDTA-K3 |
| --- | --- | --- | --- | --- |
| slope | 0.9907 | 1.0007 | 1.0025 | 1.0012 |
| Y- intercept | -0.1967 | -0.7063 | -0.8310 | -1.3730 |
| R | 0.9946 | 0.9955 | 0.9974 | 0.9968 |
| R² | 0.9892 | 0.9910 | 0.9949 | 0.9936 |
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
A lay user study was conducted at three sites using only English speaking and reading participants. A total of 120 subjects participated in these studies. Three test strip lots were used to evaluate the accuracy between results obtained from the fingertip and the alternative sites testing (AST) using SD GlucoNFC BGMS and the YSI results. Each participant was given the Instructions for Use guide only, with no additional coaching or instruction provided.
{14}
Each subject collected fingerstick and AST (palm, forearm, upper arm) samples and tested using the NFC meter. After self-testing, the healthcare professional measured the blood glucose concentration from new fingerstick and alternative site punctures using the YSI only. After each procedure, the subjects were asked to answer a questionnaire evaluating their understanding of the procedures. The range of glucose values for the finger stick samples was $62 - 416\mathrm{mg / dL}$ as measured by the reference, YSI. The results are summarized below:
| | Glucose concentrations <75 mg/dL | | |
| --- | --- | --- | --- |
| Site | Within ± 5mg/dL | Within ± 10mg/dL | Within ± 15mg/dL |
| Finger | 9/15 (60.0%) | 15/15 (100%) | 15/15 (100%) |
| Palm | 9/15 (60.0%) | 15/15 (100%) | 15/15 (100%) |
| Forearm | 9/15 (60.0%) | 15/15 (100%) | 15/15 (100%) |
| Upper arm | 9/15 (60.0%) | 15/15 (100%) | 15/15 (100%) |
| | Glucose concentration ≥75 mg/dL | | | |
| --- | --- | --- | --- | --- |
| Site | Within ± 5 % | Within ± 10% | Within ± 15% | Within ± 20% |
| Finger | 50/105 (47.6%) | 97/105 (92.4%) | 105/105 (100%) | 105/105 (100%) |
| Palm | 53/105 (50.5%) | 93/105 (88.6%) | 105/105 (100%) | 105/105 (100%) |
| Forearm | 52/105(49.5%) | 93/105 (88.6%) | 105/105 (100%) | 105/105 (100%) |
| Upper arm | 52/105 (49.5%) | 93/105 (88.6%) | 105/105 (100%) | 105/105 (100%) |
Linear Regression results:
| Tested by lay-user | | | | |
| --- | --- | --- | --- | --- |
| Site | Fingerstick | Palm | Forearm | Upper arm |
| Slope | 0.9876 | 0.9847 | 0.9860 | 0.9842 |
| Y-intercept | 1.8072 | 2.5201 | 1.8486 | 2.8443 |
| R | 0.9971 | 0.9954 | 0.9957 | 0.9947 |
| R² | 0.9942 | 0.9919 | 0.9913 | 0.9895 |
# 4. Clinical cut-off:
Not applicable
# 5. Expected values/Reference range:
The fasting adult blood glucose range for a person without diabetes $^1$ :
- Before meals $< {100}\mathrm{\;{mg}}/\mathrm{{dL}}\left( {{5.6}\mathrm{\;{mmol}}/\mathrm{L}}\right)$
After meals: $< 140\mathrm{mg / dL}$ (7.8 mmol/L)
$^{1}$ American Diabetes Association: Clinical Practice Recommendations (2015. Diabetes Care, Vol 38, Supplement 1, p. S8-S16.
{15}
16
N. Instrument Name:
SD GlucoNFC Blood Glucose Meter
SD GlucoNFC Multi Blood Glucose Meter
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ x ☐ or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ x ☐ or No ☐
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ x ☐ or No ☐
3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
4. Specimen Sampling and Handling:
The device is intended to be used with capillary whole blood from the finger, palm, forearm, and upper arm. The SD GlucoNFC Multi BGMS may also be used with venous whole blood. The whole blood sample is applied directly to the test strip by capillary action therefore there are no special handling or storage issues.
5. Calibration:
This is a non-coding device, therefore no calibration is required by the user.
6. Quality Control:
Each SD GlucoNFC Multi BGMS comes with one level of SD GlucoNavii Control Solution (Level 2). The control solution is used to check if the meter and test strips as a
{16}
system are working normally, or if the user is performing the test correctly. Control solution tests should be performed when a new box of test strips is opened, the test strip container is left open or the test strips are damaged, if the strips are left in extreme temperature or humidity, if the patient wants to check the meter and test strips, whenever the meter is dropped, whenever a result does not agree with the level felt by the patient or the patient wants to check if they are testing correctly.
Each system includes a SD Glucose check strip, which is used to check the electronic performance of the meter. The check strip may be used before using the meter for the first time, or after a long period of disuse of the meter. The check strip test does not replace the Control Solution test.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1. Altitude study: An altitude study was performed in a barometric pressure test chamber to simulate the following 4 different altitudes: sea level, 3,280 ft, 6,561 ft, and 11,480 ft. Venous blood was spiked with glucose to achieve three glucose concentrations (45.5 mg/dL, 112 mg/dL and 348 mg/dL, as measured by YSI). Each sample was tested in duplicate with 5 meters and the results were compared to YSI. The results demonstrate acceptable bias to the reference to support the claims that altitudes up to 11,480 feet have no significant effect on blood glucose measurements with the SD GlucoNFC and SD GlucoNFC Multi Blood Glucose Monitoring Systems.
2. Hematocrit study: The effect of different hematocrit levels were evaluated using venous whole blood samples with hematocrit levels of 10%, 20%, 30%, 50%, 60%, 70%, and 75%, spiked to five glucose concentrations (42, 68, 92, 128, and 360 mg/dL, as measured by YSI). The samples were tested with 10 meters and three lots of test strips. The results were compared to YSI and the normal 40% hematocrit. The percent bias of the SD GlucoNFC meter results relative to YSI demonstrated adequate performance to support the claimed hematocrit range of 10-70%.
3. Infection control studies: The SD GlucoNFC meter is intended for single-patient use, and the SD GlucoNFC Multi meter is intended for multiple-patient use. Disinfection efficacy studies were performed on the materials comprising the meter by an outside commercial laboratory service to demonstrate complete inactivation of hepatitis B virus (HBV) with DisCide Ultra Disinfecting Towelettes (EPA Reg. No. 10492-4). Robustness studies were performed by the sponsor demonstrating that there was no change in performance or in external materials of the meter after 10,950 cleaning and disinfection cycles with DisCide Ultra Disinfecting Towelettes. The robustness studies were designed to simulate 3 years of multiple-patient use and 5 years of single-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
4. Sample volume study: A sample volume study was performed using venous whole blood samples spiked to three levels of glucose concentrations (50 mg/dL, 110 mg/dL
17
{17}
and 220 mg/dL, as measured by YSI) to evaluate the effect of different sample volumes (0.2 μL, 0.3 μL, 0.4 μL, 0.5 μL, 0.6 μL, 0.8 μL, and 1.0 μL) on the performance of the device. Results at each sample volume were compared to the corresponding YSI values. Two lots of test strips and 10 meters were used. Results from these studies support the claimed minimum sample volume of 0.5 μL. The meter displays error message E-2 when an insufficient amount of blood sample is applied. The sponsor provided adequate validation for this error message.
5. Operating Temperature and Humidity: Operating temperature and humidity conditions were evaluated using six meters and three lots of test strips with venous whole blood samples at three glucose concentrations (55 mg/dL, 110 mg/dL, and 300 mg/dL by YSI). The following temperature and humidity conditions were tested: 46°F (8°C)/10% RH, 46°F (8°C)/90% RH, 113°F (45°C)/10% RH, and 113°F (45°C)/90% RH. The results support the sponsor's claimed operating temperature from 46°F to 113°F (8°C - 45°C) and relative humidity range from 10-90%.
6. EMC testing and Electrical Safety Studies: The sponsor states that they followed requirements of IEC 61010-1:2001, IEC61010-2-101:2002, and IEC 61326-2-6:2005. The sponsor provided documentation stating that the SD GlucoNFC and SD GlucoNFC Multi meters were found to be in compliance with the test standards IEC 61010-1:2001, IEC61010-2-101:2002, and IEC 61326-2-6:2005.
7. Readability Assessment: A Flesch-Kinkaid reading level assessment was conducted demonstrating that the user manuals and test strip inserts were written at or below an 8th grade reading level.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
18
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.