K151263 · Talley Group, Ltd. · OMP · Feb 19, 2016 · General, Plastic Surgery
Device Facts
Record ID
K151263
Device Name
Venturi Gauze Wound Care Sets with Portal Drain
Applicant
Talley Group, Ltd.
Product Code
OMP · General, Plastic Surgery
Decision Date
Feb 19, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Venturi™ Gauze Wound Care Sets with Portal Drain are intended to be used with the Talley Group range of negative pressure wound therapy pumps. The Talley Group NPWT pumps are indicated use for patients with acute or chronic wounds that may be benefitted by the application of continual negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids, and infectious materials. The device is intended for use by qualified healthcare professionals in a healthcare environment.
Device Story
Dressing kit for negative pressure wound therapy (NPWT); includes antimicrobial gauze, adhesive film, and Portal Drain assembly. Portal Drain adheres to film over a cut hole, connecting to NPWT pump via tubing to transport exudate to a canister. Used by healthcare professionals in clinical environments. Facilitates fluid removal from acute/chronic wounds; aids healing via continual negative pressure. Replaces previous flat/channel drain designs with an external portal design to minimize direct patient contact.
Clinical Evidence
No clinical data was required. Substantial equivalence supported by bench testing.
Technological Characteristics
Dressing kit components include antimicrobial gauze, adhesive film, and Portal Drain assembly. Portal Drain is placed over adhesive film; connects to NPWT pump via tubing. Sterile, single-use. No software or electronic components; purely mechanical fluid transport system.
Indications for Use
Indicated for patients with acute or chronic wounds requiring negative pressure wound therapy for fluid removal. Contraindicated for patients with malignancy in the wound, untreated osteomyelitis, non-enteric/unexplored fistulas, wounds with difficult haemostasis, necrotic tissue with eschar, or exposed vasculature, bone, nerves, or organs.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
Predicate Devices
Venturi™ NPWT v.II Advanced Vacuum System (K080897)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 19, 2016
Talley Group Ltd. % Phil Bromley Consultant SGR Consulting Services 8 Hollingworth Court, Turkey Mill, Ashford Road Maidstone, Kent ME14 5PP United Kingdom
Re: K151263
> Trade/Device Name: Venturi Gauze Wound Care Sets With Portal Drain Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: January 6, 2016 Received: January 11, 2016
Dear Mr. Bromley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K151263
#### Device Name
Venturi™ Gauze Wound Care Sets with Portal Drain
#### Indications for Use (Describe)
Venturi™ Gauze Wound Care Sets with Portal Drain are intended to be used with the Talley Group range of negative pressure wound therapy pumps. The Talley Group NPWT pumps are indicated use for patients with acute or chronic wounds that may be benefitted by the application of continual negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids, and infectious materials. The device is intended for use by qualified healthcare professionals in a healthcare environment.
NPWT is contraindicated for patients with:
- 1. Malignancy in the wound
- 2. Untreated osteomyelitis
- 3. Non-enteric and unexplored fistulas
- 4. Wounds with difficult haemostasis
- 5. Necrotic tissue with eschar present
- 6. Exposed vasculature, bone, nerves, or organs
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for Talley Group. The logo features a red oval with a white "T" inside. Below the oval, the word "TALLEY" is written in a large, blue, sans-serif font. Underneath "TALLEY", the word "GROUP" is written in a smaller, red, sans-serif font, with a line on either side.
Venturi™ Gauze Wound Care Sets with Portal Drain
## 510(k) Summary
### 510(k) number K151263
This 510(k) summary is being submitted in accordance with the requirement of 21 CFR 807.92.
| Summary | |
|----------------------------------|----------------------------------------------------------------------------------------------------------------|
| Submitter's Identification: | Talley Group, Ltd,<br>Premier way,<br>Abbey park industrial estate<br>Romsey, Hampshire<br>SO51 9DQ<br>England |
| Registered Establishment Number: | 8010348 |
| Owner/Operator Number: | 8010348 |
| Contact: | Robert Macdonald<br>Quality and Regulatory Affairs Manager |
| Telephone: | +44 1794 503575 |
| Fax: | +44 1794 503539 |
| Email: | rmacdonald@talleygroup.com |
| Date of Summary: | 16th February 2016 |
| Device/Proprietary Name: | Venturi™ Gauze Wound Care Sets with Portal Drain |
| Common Name: | NPWT Dressing Kits |
| Classification: | Negative pressure wound therapy powered suction pump |
| Product Code: | OMP |
| Product Class: | II |
| Code of Federal Regulation: | 21 CFR 878.4780 |
| Panel | General & Plastic Surgery |
| Substantial Equivalence: | | | |
|----------------------------------|--------------------------------------------------|------------------------------|---------------|
| Manufacturer | Trade Name | Regulation & Product<br>Code | 510(k) Number |
| Talley Group | VenturiTM NPWT v.II<br>Advanced Vacuum<br>System | 21 CFR 878.4780;<br>OMP | K080897 |
| Mölnlycke Health Care<br>US, LLC | Avance® Foam<br>Dressing Kits | 21 CFR 878.4780;<br>OMP | K122132 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Talley Group. The logo features the word "TALLEY" in a large, bold, blue font. Above the word "TALLEY" is a red circle with a white "T" inside. Below the word "TALLEY" is the word "GROUP" in a smaller, red font with a line on either side.
## enturi™ Gauze Wound Care Sets with Portal Drain
#### Description
#### Description of Device
With the exception of the Portal Drain, there are no changes to the components in the currently marketed gauze dressing kits cleared in 510(K) K080897. The wound care sets connect to Negative Pressure Wound Therapy (NPWT) suction pumps to carry exudate from the wound.
The Venturi™ Gauze Wound Care Sets with Portal Drain consist of the following components:
Venturi™ Gauze Wound Care Set: Kerlix Antimicrobial gauze wound filler, adhesive film dressing, Portal Drain Assembly; 20ml saline solution and comes with measuring tape and instructions for use. These are supplied as sterile and single use.
The Venturi™ MINO Gauze Wound Care sets are have identical contents with the addition of a canister that attaches to the Venturi™ MINO NPWT device to contain the exudate.
Product Code Product Description 97-30-42-102 Venturi™ Gauze Wound Care Set with Portal Drain – Standard 97-30-42-109 Venturi™ Gauze Wound Care Set with Portal Drain - Large 97-30-42-114 Venturi™ MINO Gauze Wound Care Set with Portal Drain – Small
#### The wound care sets include:
The Portal Drain assembly is consist of the portal drain which is adhered to a window dressing and is connected to the drain tube which connects at the other end to the NPWT pump via the canister connector assembly. There is a 90° drain tube guide and pinch clamp at the pump end of the assembly for additional safety.
The window dressing end of the Portal Drain Assembly is self-adhesive and a small hole is cut in the adhesive film dressing directly over the gauze and the Portal Drain Assembly is placed over the adhesive window dressing ensuring the portal drain is lined up with the hole. The other end of the tubing attaches to the pump to carry exudate from the wound.
#### Indications for Use:
Venturi™ Gauze Would Care Sets with Portal Drain are intended to be used with the Talley Group range of negative pressure wound therapy pumps. The Talley Group NPWT pumps are indicated use for patients with acute or chronic wounds that may be benefitted by the application of continual negative pressure to the wound for removal of fluids, including wound exudate, irrigation fluids, and infectious materials. The device is intended for use by qualified healthcare professionals in a healthcare environment.
NPWT is contraindicated for patients with:
- 1. Malignancy in the wound
- 2. Untreated osteomyelitis
- 3. Non-enteric and unexplored fistulas
- 4. Wounds with difficult haemostasis
- 5. Necrotic tissue with eschar present
- 6. Exposed vasculature, bone, nerves, or organs.
#### Summary of the Determination of Substantial Equivalence & Performance Data
All of kits components with the exception of the Portal Drain Assembly have been cleared for use as a NPWT dressing kit under K080897. The components used in the Venturi™ Gauze Wound
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Talley Group. The logo features the word "TALLEY" in large, blue, bold letters. Above the word "TALLEY" is a red circle containing a white symbol that resembles a stylized letter "T". Below the word "TALLEY" is the word "GROUP" in smaller, red letters.
## Venturi™ Gauze Wound Care Sets with Portal Drain
Care Sets are identical to those in material, function and sterility (except the Portal Drain Assembly). The dressing kits in K080897 used a flat or channel drain that was placed underneath the adhesive film dressing and contacted the skin whereas the portal drain is placed over the adhesive film dressing and a small hole is cut above the gauze to ensure the portal drain and hole line up and there is no intended patient contact. The Portal Drain Assembly performs the same function as the flat or channel drain in the predicate (K080897).
The Venturi™ Gauze Wound Care Sets are equivalent in function to the Mölnlycke Avance® Dressing Kits incl. Transparent Film, Transfer Pad (K122132) and the Portal Drain Assembly is the equivalent of the Transfer Pad in the Avance® Dressing Kits (K122132). The Portal Drain Assembly and Transfer Pad are placed over a small cut hole on the adhesive film over the gauze and connect to the NPWT device to transport exudate to the canister in the pump.
The Venturi™ Gauze Wound Care Sets have equivalent intended uses to the predicates and are intended to contact the patient in the same way and the for same period of time and except for the Portal Drain Assembly have the same component parts as the predicate (K080897).
The component parts of the device and the predicates are all provided individually packaged, sterilised and have an equivalent shelf life. The devices have similar labelling and instructions for use.
The bench testing has demonstrated that the device does not introduce any additional risks when undertaking Negative Pressure Wound Therapy and meeting the intended use.
#### Clinical Testing
No clinical data was required.
#### Conclusion
Based on the information presented in this submission, it is concluded that the Venturi™ Gauze Wound Care Sets with Portal Drain are equivalent to the Avance® Dressing Kits (K122132) and the Venturi™ NPWT v.Il Advanced Vacuum System (K080897) with respect to intended use, design and technological characteristics.
The Venturi™ Gauze Wound Care Sets with Portal Drain is as safe and effective as the predicate devices for facilitating Negative Pressure Wound Therapy and there are no new indications for use and therefore by following the FDA 510(k) "Substantial Equivalence" decision making process, http://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/
HowtoMarketYourDevice/PremarketSubmissions/PremarketNotification510k/ucm134783.htm it is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.