← Product Code [DXN](/productcode/DXN) · K151258

# Arm automatic blood pressure monitor (K151258)

_Shenzhen Pump Medical System Co., Ltd. · DXN · Feb 18, 2016 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151258

## Device Facts

- **Applicant:** Shenzhen Pump Medical System Co., Ltd.
- **Product Code:** [DXN](/productcode/DXN.md)
- **Decision Date:** Feb 18, 2016
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.1130
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

It is a device intended to measure the systolic, diastolic blood pressure, pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff (available sizes: 22cm to 44cm (8.7in to 17.3in)) is wrapped around the single upper arm.

## Device Story

Device is an automatic, non-invasive oscillometric blood pressure monitor for home use by adults. Input: pressure signals from an inflatable cuff wrapped around the upper arm. Operation: air pump inflates cuff; pressure sensor captures arterial vibrations; sensor converts pressure to digital signal; CPU/software calculates systolic/diastolic blood pressure and pulse rate. Output: digital display of blood pressure and pulse rate. Device aids users in monitoring cardiovascular health at home. Differences from predicate include appearance and power source (battery-only vs. battery/AC).

## Clinical Evidence

No clinical testing performed. Substantial equivalence to the predicate device is supported by bench testing, including electrical, mechanical, and environmental performance requirements per FDA guidance and relevant IEC standards (IEC 60601-1, IEC 60601-1-2, IEC 80601-2-30, IEC 60601-1-11).

## Technological Characteristics

Oscillometric measurement principle; battery-powered; TN-LCD digital display; upper arm cuff (22-44cm). Standards: IEC 60601-1 (safety), IEC 60601-1-2 (EMC), IEC 80601-2-30 (automated non-invasive sphygmomanometers), IEC 60601-1-11 (home healthcare environment).

## Regulatory Identification

A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.

## Predicate Devices

- Arm automatic blood pressure monitor, Model: BF1110, BF1112, BF1113 and BF1115 ([K130325](/device/K130325.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 18, 2016

Shenzhen Pump Medical System Co., Ltd. Migo Yang Clinical Registration Specialist 2/F West, M-7 Sinosteel Building, Maqueling Estate Hi-Tech Industrial Park, Nanshan District Shenzhen, Guangdong, 518057 CN

Re: K151258

Trade/Device Name: Arm Automatic Blood Pressure Monitor, Model: BE6034, BE6134 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: January 5. 2016 Received: January 19, 2016

Dear Migo Yang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151258

Device Name

Arm automatic blood pressure monitor, Model: BE6034, BE6134, Noninvasive blood pressure measurement system.

#### Indications for Use (Describe)

It is a device intended to measure the systolic blood pressure and pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable sizes: 22cm to 44cm (8.7in to 17.3in)) is wrapped around the single upper arm.

Type of Use (Select one or both, as applicable)

- Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summarv

This summary of 510(k) safety and effectiveness information is being submitted inaccordance with the requirements of SMDA 1990 and 21 CFR §807.92.

The assigned 510(k) number is: K151258

### 1. Submitter's Identification

SHENZHEN PUMP MEDICAL SYSTEM CO., LTD.

2/F, M-7 Sinosteel Building, Maqueling Estate,Hi-Tech Industrial Park,

Nanshan District,

518057 Shenzhen,

PEOPLE'S REPUBLIC OF CHINA

Contact: Xie Qiongyu Tel:86-0755-26710795 Fax:86-0755-26498210 E-mail:xieqy@bpump.com.cn

Date Prepared: 2016-2-12

### 2. Name of the Device

Trade Name: Arm automaticblood pressure monitor Model No.: BE6034, BE6134 Regulation Description: Noninvasive blood pressure measurement system. Regulation Number: 21 CFR 870.1130 Regulation Class: II Product Code: DXN Review Panel: Cardiovascular

## 3. Information for the 510(k) Cleared Device(Predicate device)

Arm automatic blood pressure monitor, Model: BF1110, BF1112, BF1113 and BF1115, K130325, SHENZHEN PUMP MEDICAL SYSTEM CO., LTD.

# 4. Description of the Device

Arm automatic blood pressure monitor is based on pressure vibration method. Blood pressure cuff uses the air pump to inflate, and then the arteries are extruded by the cuff with pressure. Pressure sensor collects the pressure in the cuff, and then converts it to digital

{4}------------------------------------------------

signal to the CPU. Then the software calculates the systolic, diastolic blood pressure and pulse rate.

The arm automatic blood pressure monitor, Model: BE6034. BE6134 has the same basic principles, main function, performance and intended use, and they are consistent in product structure and material.

## 5. Intended Use

It is intended for measuring systolic, diastolic blood pressure and pulse rate of an adult over the counter.

## 6. Indications for Use

It is a device intended to measure the systolic, diastolic blood pressure, pulse rate of an adult individual by using a non-invasive oscillometric technique in which an inflatable cuff (available sizes: 22cm to 44cm (8.7in to 17.3in)) is wrapped around the single upper arm.

#### 7. Comparisonto the 510(k) Cleared Device (Predicate device)

The subject devicemodel BE 6034, BE6134 and the predicate devicemodel BF1110, BF1112, BF1113, BF1115 use the same oscillometric method within the software algorithm to determine the systolic, diastolic blood pressure and pulse rate. Upper arm cuff is inflated automatically by air pump and the pressure is transferred via tube into a sensor in these two units.

They differ by the appearance and the way of energy supply. The appearance of subject device is little different from predicate device; The only way of energy supply of subject device is battery while both battery and DC energy are available for predicate device. But those differences either affect the accuracy or the normal use of subject device because they use the same fundamental scientific technology based on clinical declaration of identity.

The subject devices, model BE 6034, BE6134 use the same oscillometricmethod as the predicate device BF1110, BF1112, BF1113 and BF1115 to determinate the systolic, diastolic blood pressure and pulse rate. They have the same function and intended use. Based upon the aforementioned information, these devices are substantially equivalent. Please refer to table below for comparison.

{5}------------------------------------------------

| Items                                   | Arm automatic blood pressure<br>monitor                                                                                                                                                                                                                                                                 | Arm automatic blood pressure<br>monitor  |                                                                                                                                                                                                                                                                                                      |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                            | SHENZHEN PUMP MEDICAL<br>SYSTEM CO.,LTD.                                                                                                                                                                                                                                                                | SHENZHEN PUMP MEDICAL<br>SYSTEM CO.,LTD. |                                                                                                                                                                                                                                                                                                      |
| Model<br>Specification                  | BE6034                                                                                                                                                                                                                                                                                                  | BE6134                                   | BF1112                                                                                                                                                                                                                                                                                               |
| Intended Use                            | It is intended for measuring adult<br>blood pressure and pulse rate                                                                                                                                                                                                                                     |                                          | It is intended for measuring adult blood<br>pressure and pulse rate                                                                                                                                                                                                                                  |
| Indications for<br>Use                  | It is a device intended to measure the<br>systolic, diastolic blood pressure,<br>pulse rate of an adult individual by<br>using a non invasive oscillometric<br>technique in which an inflatable cuff<br>(available sizes: 22cm to 44cm (8.7in<br>to 17.3in)) is wrapped around the<br>single upper arm. |                                          | It is a device intended to measure the<br>systolic, diastolic blood pressure, pulse<br>rate of an adult individual by using a non<br>invasive oscillometric technique in which<br>an inflatable cuff (available sizes: 22cm to<br>44cm (8.7in to 17.3in)) is wrapped around<br>the single upper arm. |
| Patient<br>population                   | Adult                                                                                                                                                                                                                                                                                                   |                                          | Adult                                                                                                                                                                                                                                                                                                |
| Environment<br>of use                   | Home                                                                                                                                                                                                                                                                                                    |                                          | Home                                                                                                                                                                                                                                                                                                 |
| Measuring<br>Principle                  | Oscillometric Method                                                                                                                                                                                                                                                                                    |                                          | Oscillometric Method                                                                                                                                                                                                                                                                                 |
| Measurement<br>Range                    | Pressure: 0mmHg~280mmHg<br>(0kPa~37.3kPa)<br>Pulse: 40bpm ~180 bpm                                                                                                                                                                                                                                      |                                          | Pressure: 0mmHg-280mmHg<br>(0kPa-37.3kPa)<br>Pulse Rate: 40bpm-80bpm                                                                                                                                                                                                                                 |
| Accuracy                                | Pressure: ±3mmHg(±0.4kPa)<br>Pulse: ±5%                                                                                                                                                                                                                                                                 |                                          | Pressure: ±3mmHg (±0.4kPa)<br>Pulse Rate: ±5%                                                                                                                                                                                                                                                        |
| Display                                 | TN-LCD Digital Display                                                                                                                                                                                                                                                                                  |                                          | TN-LCD Digital Display                                                                                                                                                                                                                                                                               |
| Screen Size                             | 99mm×86mm                                                                                                                                                                                                                                                                                               |                                          | 85mm×57mm                                                                                                                                                                                                                                                                                            |
| Memory                                  | 2 Memory sets, 60 reading each set.                                                                                                                                                                                                                                                                     |                                          | 1 Memory set, 90 reading each set.                                                                                                                                                                                                                                                                   |
| Broadcast<br>Function                   | Yes                                                                                                                                                                                                                                                                                                     | No                                       | No                                                                                                                                                                                                                                                                                                   |
| Power Source                            | 4 AA Alkaline battery                                                                                                                                                                                                                                                                                   |                                          | 4 AA Alkaline battery or<br>AC Adaptor (AC 100-240 V)                                                                                                                                                                                                                                                |
| Operating<br>Environment                | Temperature: +5℃~+40℃;<br>Humidity: 15%-93%<br>Pressure: 70.0kPa~106.0kPa<br>Altitude: ≤ 3 000 m                                                                                                                                                                                                        |                                          | Temperature: +5℃~+40℃;<br>Humidity: ≤93%RH                                                                                                                                                                                                                                                           |
| Storage and<br>Transport<br>Environment | Temperature: -25℃~+70℃;<br>Humidity: 10%~95%<br>Pressure:50.0kPa~106.0kPa                                                                                                                                                                                                                               |                                          | Temperature: -25℃~+70℃;<br>Humidity: 10%~95%                                                                                                                                                                                                                                                         |
| Weight                                  | 360g(Without batteries)                                                                                                                                                                                                                                                                                 |                                          | 294g(Without batteries)                                                                                                                                                                                                                                                                              |
| Size                                    | 140 mm×116mm×81 mm                                                                                                                                                                                                                                                                                      |                                          | 140mm×110mm×70mm                                                                                                                                                                                                                                                                                     |

The subject device is Substantially Equivalent (SE) to the predicate device which is US legal.

{6}------------------------------------------------

## 8. Discussion of Non Clinical tests Performed for Determination of Substantial

## Equivalence: are as follows:

Testing information demonstrates substantial equivalence to the predicate device and safety and effectiveness of the Arm automatic blood pressure monitor. Model:BE6034. BE6134 in the intended environment of use is supported by testing conducted in accordance with the FDA November 1993 Draft "Reviewer Guidance for Premarket Notification Submissions", which outlines Electrical, Mechanical and Environmental Performance requirements.

The following testing was conducted: a. IEC 60601-1 Test b. IEC 60601-1-2 & IEC 80601-2-30 Clause 202 Test c. IEC 60601-1-11 Test d. IEC 60601-2-30 Test

None of the testing demonstrated any design characteristics that violated the requirements of the Reviewer Guidance or resulted in any safety hazards. It was our conclusion that Arm automatic blood pressure monitor, Model: BE6034, BE6134 tested met all relevant requirements of the aforementioned tests.

## 9. Discussion of Clinical Tests Performed:

The subject device, Model BE6034, BE6134 is from the technical point of view, identical to the predicate blood pressure monitor. Model BF1112, BF1113 and BF1115. The differences between them do not relate to blood pressure measurement technology, so the clinical accuracy in terms of blood pressure detection will not be affected. Therefore repeated clinical test in accordance with the standard IEC 80601-2-30:2009 & A1:2013(The standard IEC 80601-2-30:2009 & A1:2013has substituted the clinical requirements of ANSI/AAMI SP10) is unnecessary.

## 10. Software Information:

Software validation was conducted in accordance with a moderate level of concern designation in accordance with the FDA November 2005 document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".

## 11. Conclusion

We have demonstrated that there is no significant differences between the Arm automatic blood pressure monitor (Model: BE6034, BE6134), and the predicate device: Arm automatic blood pressure monitor (Model: BF1110, BF1112, BF1113 and BF1115), in terms of safety and effectiveness based on electrical, mechanical and environmental test results.And they are substantially equivalent to the earlier identified predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K151258](https://fda-staging.innolitics.com/device/K151258)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).