← Product Code [DQK](/productcode/DQK) · K151245

# Rhythm View Workstation (non-streaming, steaming option) (K151245)

_Topera, Inc. · DQK · Sep 15, 2015 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151245

## Device Facts

- **Applicant:** Topera, Inc.
- **Product Code:** [DQK](/productcode/DQK.md)
- **Decision Date:** Sep 15, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1425
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The RhythmView Workstation is a computerized system that assists in the diagnosis of complex cardiac arrhythmias. The RhythmView Workstation is used to analyze electrogram and electrocardiogram signals and display them in a visual format.

## Device Story

RhythmView Workstation assists physicians in diagnosing complex cardiac arrhythmias during electrophysiological studies. System inputs include electrical signals from multi-polar electrophysiology catheters. Proprietary algorithms compute spatial organization of heart rhythm disorders, identifying electrical rotors or focal beat sources. Outputs consist of interactive graphic displays of electrical activity, including progressive depolarization and repolarization maps. Used in clinical settings by physicians to aid diagnosis. New features in V5.0 include real-time streaming of electrograms, a Signal Quality Indicator (SQI), multi-color Rotational Activity (RAP) display, and a spotlight feature. These tools allow clinicians to review, analyze, and replay animated representations of electrical signals, facilitating identification of arrhythmia sources.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including verification and validation of software and hardware components (video splitter, panel box, computer), electrical safety, EMC testing, and usability testing.

## Technological Characteristics

Programmable diagnostic computer; C++ software; hardware includes cart, monitor, computer, mouse, keyboard, RFID reader/writer, USB switch, and optional video splitter. Connectivity via USB. Standards: IEC 60601-1-2 (EMC/Safety), IEC 62366 (Usability).

## Regulatory Identification

A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.

## Predicate Devices

- RhythmView Workstation ([K142901](/device/K142901.md))
- RhythmView Workstation ([K130827](/device/K130827.md))
- RhythmView Workstation ([K123295](/device/K123295.md))
- RhythmView Workstation ([K110878](/device/K110878.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Topera, Inc. Dennis Pozzo Sr. Regulatory Affairs Specialist 3668 S. Geyer Road Suite 365 St. Louis, MO 63127

Re: K151245

Trade/Device Name: RhythmView Workstation SW V5.0 Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II Product Code: DQK Dated: May 8, 2015 Received: May 11, 2015

Dear Dennis Pozzo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Mitchell Stein

forBram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K151245

Device Name

Rhythm View Workstation SW V5.0

Indications for Use (Describe)

The Rhythm View Workstation is a computerized system that assists in the diagnosis of complex cardiac arrhythmias. The Rhythm View Workstation is used to analyze electrogram and electrocardiogram signals and display them in a visual format.

| Type of Use (Select one or both, as applicable) |                                             |
|-------------------------------------------------|---------------------------------------------|
| > Prescription Use (Part 21 CFR 801 Subpart D)  | Over-The-Counter Use (21 CFR 801 Subpart C) |
|                                                 |                                             |

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## Appendix 1: 510(k) Summary per 21CFR §807.92

| Submitter's<br>information        | Topera, Inc.<br>3668 S. Geyer Road, Suite 365<br>St. Louis, MO 63127<br>Contact: Dennis Pozzo<br>Phone 314-300-6580                                                                                                                                                                                                                                                                      |                                                                                  |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Device/<br>classification<br>name | Device Name: RhythmView Workstation, SW V5.0<br>Product Code: DQK<br>Classification No: 21 CFR 870 1425<br>Classification/Common name: Programmable diagnostic computer<br>The marketed device(s) to which substantial equivalence is claimed:<br>K142901, cleared Dec. 19, 2014<br>K130827, cleared Dec. 16, 2013<br>K123295, cleared April 24, 2013<br>K110878, cleared Sept. 23, 2011 |                                                                                  |
| Device description                | The RhythmView Workstation is comprised of the following components:                                                                                                                                                                                                                                                                                                                     |                                                                                  |
|                                   | Cart                                                                                                                                                                                                                                                                                                                                                                                     | Keyboard                                                                         |
|                                   | Monitor/Display                                                                                                                                                                                                                                                                                                                                                                          | Mouse                                                                            |
|                                   | Computer                                                                                                                                                                                                                                                                                                                                                                                 | Two Port USB Switch                                                              |
|                                   | Radio-Frequency Identification<br>(RFID) Reader/Writer                                                                                                                                                                                                                                                                                                                                   | Video Splitter (new component -<br>streaming option only)<br>*Panel box required |
|                                   | Solid State Hard Drive (optional<br>component)                                                                                                                                                                                                                                                                                                                                           | Software                                                                         |
|                                   | RhythmView takes electrical signals collected from multi-polar<br>electrophysiology catheters and outputs a graphic display that assists in the<br>diagnosis of cardiac arrhythmias.                                                                                                                                                                                                     |                                                                                  |
|                                   | The Rhythm View computes and displays electrical rotors or focal beat<br>sources responsible for maintaining human heart rhythm disorders including                                                                                                                                                                                                                                      |                                                                                  |

focal AT, AFL, other SVT, AF, VT and VF in a given patient. The product takes as input electrical signals recorded during the heart rhythm disorder under consideration, typically from multiple specified locations within the heart during an electrophysiological study. The RhythmView then uses proprietary patented algorithms and methods to compute spatial organization during the heart rhythm disorder. These computed elements are displayed graphically in interactive form for review to aid diagnosis by the physician during an electrophysiology study.

{4}------------------------------------------------

# Appendix 1: 510(k) Summary per 21CFR §807.92, Continued

| Indications for<br>use           | The RhythmView Workstation is a computerized system that assists in the<br>diagnosis of complex cardiac arrhythmias. The RhythmView Workstation is<br>used to analyze electrogram and electrocardiogram signals and display them<br>in a visual format.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Technological<br>characteristics | Both the proposed and predicate RhythmView Workstations allows the user<br>to:<br>Review and select a time sequence of electrical signals from various electrodes; Analyze the signals; View a graphic display (Electrical Activity) of the signal potentials showing progressive depolarization and repolarization in grayscale for the particular arrhythmia; Play/Replay the animated graphic representation of electrical signals. All display options are available for both atrial and ventricular maps.          Other than the additional display options (SQI, Multi-color RAP and the spotlight feature) associated with the proposed RhythmView workstation, the major difference between the predicate and proposed workstation is the addition of the optional streaming and associated SQI of the streaming data.<br>It is important to note that the import of the streaming electrograms vs. non-streaming electrograms is identical with the exception of a few new workflow steps. |
| Performance<br>data              | The Software and Hardware (video splitter, panel box and computer) had undergone verification and validation testing, including Electrical Safety and EMC testing to provide reasonable assurance that the proposed device conformances to the appropriate requirements for its intended use. The Workstation has also undergone usability testing. Therefore, it has been demonstrated that the RhythmView Workstation is safe and effective for its intended use.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Consensus<br>standards           | IEC 60601-1-2 Medical Electrical Equipment, General requirements for safety. Electromagnetic Compatibility and essential performance<br>IEC 62366 Medical Devices - Application of Usability Engineering to Medical Devices                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

{5}------------------------------------------------

# Appendix 1: 510(k) Summary per 21CFR §807.92, Continued

| Device Characteristic                                                  | Predicate<br>RhythmView™<br>Workstation                                                     | Proposed<br>RhythmView™<br>Workstation                                                       |
|------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
| Signal processing                                                      | Yes                                                                                         | Yes                                                                                          |
| Post-processing display                                                | Yes                                                                                         | Yes                                                                                          |
| Grid display of electrode signals                                      | Yes                                                                                         | Yes                                                                                          |
| Programming Language                                                   | C++                                                                                         | C++                                                                                          |
| Export of processed file into video<br>format                          | Yes                                                                                         | Yes                                                                                          |
| Manual tagging by user of<br>electrograms                              | No                                                                                          | No                                                                                           |
| OTS Software requirements                                              | Same                                                                                        | Same                                                                                         |
| Display options for review of<br>processed signals                     | • Electrical Activity<br>• Contours Only<br>• DContours<br>• Rotational Activity<br>Profile | • Electrical Activity<br>• Contours Only<br>• DContours<br>• Rotational Activity<br>Profile” |
| RAP display (optional)                                                 | Monochromatic only                                                                          | Multi-color with<br>monochromatic<br>option available                                        |
| RFID Reader/Writer Function                                            | Yes                                                                                         | Yes                                                                                          |
| Data transfer via Two Port Switch                                      | Yes                                                                                         | Yes                                                                                          |
| Direct data transfer via USB cable to<br>RV Workstation from EP system | Option available                                                                            | Option available                                                                             |
| Atrial Function                                                        | Yes                                                                                         | Yes                                                                                          |
| Ventricular Function                                                   | Option available                                                                            | Option available                                                                             |
| Signal Quality Indicator                                               | No                                                                                          | Yes                                                                                          |
| Spotlight Feature                                                      | No                                                                                          | Yes                                                                                          |
| Streaming real time electrograms                                       | No                                                                                          | Option available                                                                             |

---

**Source:** [https://fda-staging.innolitics.com/device/K151245](https://fda-staging.innolitics.com/device/K151245)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).