STRYKER INFRARED ILLUMINATION SYSTEM (IRIS) [AIM Light Source and IRIS Ureteral Kit]
Applicant
Stryker Endoscopy
Product Code
FCS · Gastroenterology, Urology
Decision Date
Aug 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.4020
Device Class
Class 2
Indications for Use
The Stryker AIM Light Source is intended to trans-illuminate anatomical structures during open or laparoscopic surgical procedures. The Stryker® AIM Light Source and SafeLight Cable are indicated for use to provide real-time endoscopic visible and near-infrared fluorescence imaging. The Stryker® AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging. Fluorescence imaging of biliary ducts with the Stryker® AIM Light Source and SafeLight Cable are intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization. The Stryker IRIS Ureteral Kit when used with compatible Stryker is intended to trans-illuminate the ureter during surgical procedures. The catheter accepts up to 0.038" guide wire. The kit can be used for either open surgical or laparoscopic procedures.
Device Story
System provides real-time endoscopic visible and near-infrared (NIR) fluorescence imaging; utilizes light source console and ureteral kit with light-emitting fibers. Surgeons use device during open or laparoscopic procedures to trans-illuminate ureters; enables visual assessment of vessels, blood flow, tissue perfusion, and extra-hepatic bile ducts. Input consists of light signals transmitted through fibers; output is visual image displayed for surgeon. Used in OR; operated by surgeons. Enhances anatomical identification; aids in surgical navigation; reduces risk of accidental injury to ureters or biliary structures. Complements standard white light and intraoperative cholangiography.
Clinical Evidence
No human clinical trials. Evidence includes bench performance testing, biocompatibility assessment per ISO 10993-1, electrical safety/EMC testing per IEC 60601-1/1-2, and cadaver testing to validate ureteral trans-illumination capability.
Technological Characteristics
System includes light source console and ureteral kit (catheters, fiber assembly, dispenser, luer connectors, drape clips). Sterilization via ethylene oxide per ISO 11135:2014. Connectivity includes light-emitting fibers for trans-illumination and NIR fluorescence imaging. Software developed per IEC 62304.
Indications for Use
Indicated for patients undergoing open or laparoscopic surgical procedures of the lower abdomen or pelvic areas requiring ureteral trans-illumination, or minimally invasive surgery requiring real-time endoscopic visible and near-infrared fluorescence imaging for assessment of vessels, blood flow, tissue perfusion, and major extra-hepatic bile ducts.
Regulatory Classification
Identification
A fiberoptic light ureteral catheter is a device that consists of a fiberoptic bundle that emits light throughout its length and is shaped so that it can be inserted into the ureter to enable the path of the ureter to be seen during lower abdominal or pelvic surgery.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.
Stryker® Infrared Fluorescence (IRF) Imaging System (K142310)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 20, 2015
Stryker Endoscopy Golnaz Moeini Sr. Regulatory Affairs Analyst 5900 Optical Ct San Jose, CA 95138
Re: K151243
> Trade/Device Name: Stryker Infrared Illumination System (IRIS) [AIM Light Source and Ureteral Kit] Regulation Number: 21 CFR 876.4020 Regulation Name: Fiberoptic Light Ureteral Catheter Regulatory Class: II Product Code: FCS. FCW Dated: June 26, 2015 Received: June 29, 2015
Dear Golnaz Moeini,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K151243
#### Device Name
Stryker® Infrared Illuminating System (IRIS)[AIM Light Source and IRIS Ureteral Kit]
#### Indications for Use (Describe)
The Stryker AIM Light Source is intended to trans-illuminate anatomical structures during open or laparoscopic surgical procedures.
The Stryker® AIM Light Source and SafeLight Cable are indicated for use to provide real-time endoscopic visible and near-infrared fluorescence imaging. The Stryker® AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging.
Fluorescence imaging of biliary ducts with the Stryker® AIM Light Source and SafeLight Cable are intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.
The Stryker IRIS Ureteral Kit when used with compatible Stryker is intended to trans-illuminate the ureter during surgical procedures. The catheter accepts up to 0.038" guide wire. The kit can be used for either open surgical or laparoscopic procedures.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: overline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5. 510(k) Summary
#### General Information 1.
| 510(k) Sponsor | Stryker Endoscopy |
|-------------------------|----------------------------------------------------------------------|
| Address | 5900 Optical Court<br>San Jose, CA 95138 |
| FDA Registration Number | 2936485 |
| Correspondence Person | Golnaz Moeini<br>Sr. Regulatory Affairs Analyst<br>Stryker Endoscopy |
| Contact Information | Email: golnaz.moeini@stryker.com<br>Phone: 408-754-2755 |
| Date Prepared | 6/24/2015 |
### 2. Proposed Device
Proposed Device:
| Proprietary Name | Stryker® Infrared Illuminating System (IRIS)[Consisting of AIM Light<br>Source and IRIS Ureteral Kit ] |
|---------------------|--------------------------------------------------------------------------------------------------------|
| Common Name | Light Source, Illuminator |
| Classification Name | Fiberoptic light ureteral catheter, Light Source, Fiberoptic, Routine |
| Regulation Number | 21 CFR 876.4020, 21 CFR 876.1500 |
| Product Code | FCS, FCW |
| Regulatory Class | II |
## 3. Predicate Device
### Primary Predicate Device:
| Proprietary Name | Stryker Ureteral Illumination System IV [Consisting of Infravision<br>Illuminator and Universal Ureteral Kit ] |
|------------------------|----------------------------------------------------------------------------------------------------------------|
| Premarket Notification | K061548 |
| Classification Name | Fiber Optic Light Ureteral Catheter |
| Regulation Number | 21 CFR 876.4020 |
| Product Code | FCS |
| Regulatory Class | II |
#### Secondary Predicate Device:
| Proprietary Name | Stryker® Infrared Fluorescence (IRF) Imaging System |
|------------------------|-----------------------------------------------------|
| Premarket Notification | K142310 |
| Classification Name | Confocal Optical Imaging |
| Regulation Number | 21 CFR 876.1500 |
| Product Code | OWN |
| Regulatory Class | II |
{4}------------------------------------------------
### 4. Device Description
The Stryker® Infrared Illuminating System (IRIS) is used to trans-illuminate the ureter during laparoscopic or open surgical procedures. Trans-illumination is intended to help the surgeon identify the ureter(s) during open or laparoscopic surgical procedure of lower abdomen or pelvic areas. The Stryker® Infrared Illuminating System (IRIS) consists of the following main components (herein referred to as 'proposed devices'):
- A light source console. ।
- A Ureteral Kit that consists of two catheters, one fiber assembly containing two । catheters, one fiber assembly containing two light emitting fibers, one catheter dispenser, one light emitting fiber dispenser, two luer connectors, and two drape clips.
### 5. Indications for Use
#### Indications for Use for Light Source
The Stryker AIM Light Source is intended to trans-illuminate the ureter during open or laparoscopic surgical procedures.
The Stryker® AIM Light Source and SafeLight Cable are indicated for use to provide realtime endoscopic visible and near-infrared fluorescence imaging. The Stryker® AIM Light Source and SafeLight Cable enable surgeons to perform minimally invasive surgery using standard endoscope visible light as well as visual assessment of vessels, blood flow and related tissue perfusion, and at least one of the major extra-hepatic bile ducts (cystic duct, common bile duct and common hepatic duct), using near infrared imaging.
Fluorescence imaging of biliary ducts with the Stryker® AIM Light Source and SafeLight Cable are intended for use with standard of care white light and, when indicated, intraoperative cholangiography. The devices are not intended for standalone use for biliary duct visualization.
#### Indications for Use for Ureteral Kit
The Stryker IRIS Ureteral Kit when used with compatible Stryker Infrared sources is intended to trans-illuminate the ureter during surgical procedures. The catheter accepts up to 0.038" guide wire. The kit can be used for either open surgical or laparoscopic procedures.
### 6. Comparison of Technological Characteristics with the Predicate Device
As noted above, the Stryker® Infrared Illuminating System (IRIS) is equivalent to the following cleared device in terms of its indications for use, design technology and performance specification:
With respect to the indications for use of the light source console, the proposed Stryker light source console is substantially equivalent to Stryker Ureteral Illumination System IV (K061548) and Stryker® Infrared Fluorescence (IRF) Imaging System (K142310).
{5}------------------------------------------------
With respect to the design technology and performance specifications, the Stryker proposed light source console is substantially equivalent to Stryker Ureteral Illumination System IV (K061548).
With respect to the Ureteral Kit, the proposed Stryker Ureteral Kit is substantially equivalent to Stryker Ureteral Illumination System IV (K061548) since they are both the same in design and indication for use with very minor modification to the proposed device to improve usability of the light fibers.
## 7. Performance Data
Safety and performance of the Stryker® Infrared Illuminating System (IRIS) has been evaluated and verified in accordance with design specifications and applicable performance standards through biocompatibility assessment, electrical safety and EMC testing, software validation, bench testing and animal testing. The following performance testing were conducted and are summarized in this submission:
- Biocompatibility was assessed in accordance to ISO 10993-1:2009 Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process and related collateral standards for patient contacting materials.
- The proposed Ureteral Kit is provided sterile. Ethylene oxide sterilization was verified per -ISO 11135:2014: Sterilization of health care products - Ethylene oxide -Requirements for development, validation and routine control of sterilization process for medical devices (see Section 14, Sterilization and Shelf Life).
- Electrical Safety and electromagnetic compatibility testing was performed in accordance to IEC 60601-1:2005+A1:2012 - Medical electrical equipment-Part 1: General requirements for basic safety and essential performance and IEC 60601-1-2:2007- Medical electrical equipment- Part 1-2: General requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances – Requirement and tests, respectively. Testing indicates that the proposed device conforms to the aforementioned voluntary standards.
- The software validation activities were performed in accordance with IEC 62304:2006/AC: 2008- Medical device software – Software life cycle processes as well as the FDA Guidance documents, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
- Bench performance testing was conducted to ensure that the devices functioned as intended and met design specifications and acceptance criteria. Test results obtained verified the safety and effectiveness of the devices per design specifications and applicable standards.
- Cadaver testing validated the ureteral trans-illumination capability of the Stryker Infrared Illuminating System (IRIS).
{6}------------------------------------------------
#### Conclusion 8.
Based on the information submitted in this premarket notification, and based on the indications for use, technological characteristics, performance testing and technological comparison to the predicate devices, the Stryker® Infrared Illuminating System (IRIS) raises no new questions of safety and effectiveness and is substantially equivalent to the predicate devices in terms of safety, efficacy and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.