← Product Code [PGY](/productcode/PGY) · K151238

# JUSHA-M23C LCD Monitor (K151238)

_Nanjing Jusha Display Technology Co., Ltd. · PGY · Jun 12, 2015 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151238

## Device Facts

- **Applicant:** Nanjing Jusha Display Technology Co., Ltd.
- **Product Code:** [PGY](/productcode/PGY.md)
- **Decision Date:** Jun 12, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

JUSHA-M23C LCD Monitor is intended to be used in displaying and viewing digital images by trained medical practitioners. JUSHA-M23C LCD Monitor does not support the display of mammography images for diagnosis.

## Device Story

JUSHA-M23C is a 21.3-inch monochrome TFT LCD monitor for medical imaging. It accepts digital video inputs (DVI, DisplayPort) and displays images at 1600x1200 resolution. The device features an integrated calibration sensor, ambient brightness adaptation, and presence induction to automatically adjust display parameters for optimal viewing. It utilizes a 10-DICOM and 3-GAMMA lookup table to manage grayscale output (1024 simultaneous shades). Used in clinical environments by medical practitioners to evaluate high-resolution medical images. The device provides visual output to assist clinicians in diagnostic decision-making. It does not support mammography.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified via measurement of angular luminance dependency, luminance/chromaticity non-uniformity (per TG18 guidelines), small-spot contrast ratio, temporal response, and luminance stability. Electrical safety and EMC verified per IEC 60601-1 and IEC 60601-1-2.

## Technological Characteristics

21.3-inch monochrome TFT LCD panel; 1600x1200 resolution; 1000 cd/m² luminance; 1400:1 contrast ratio; LED backlight; 10-bit DICOM LUT; 3-GAMMA LUT; ambient brightness/presence sensors; DVI/DisplayPort connectivity; IEC 60601-1/60601-1-2 compliant.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- RADIFORCE GX240 ([K120407](/device/K120407.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a stylized caduceus symbol. The text reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". The caduceus is depicted with a modern, abstract design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 12, 2015

Nanjing Jusha Display Technology Co., Ltd. % Mr. Ma Jing Certifical Engineer 301 Hanzhongmen Street, 8F Block A, No. 1 Nanjing International Service Outsourcing Mansion, Nanjing, 210036 CHINA

Re: K151238

Trade/Device Name: JUSHA-M23C LCD Monitor Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: March 16, 2015 Received: May 11, 2015

Dear Ma Jing:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K151238

Device Name JUSHA-M23C LCD monitor

#### Indications for Use (Describe)

JUSHA-M23C LCD Monitor is intended to be used in displaying digital images by trained medical practitioners. JUSHA-M23C LCD Monitor does not support the display of mammography images for diagnosis.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

# 510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided:

| Date:                                        | May 6, 2015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |
|----------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Submitter:                                   | Nanjing Jusha Display Technology Co., Ltd<br>Add: 8A, Block 1. Nanjing International Service Outsourcing Mansion, No. 301, Hanzhongmen street, Nanjing, China.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |
| Contact Person:                              | Ma Jing<br>Certification Engineer<br>Nanjing Jusha Display Technology Co., Ltd<br>Tel: +86-25- 83305050<br>Fax: +86-25- 58783271                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |  |
| Device Trade Name:                           | JUSHA-M23C LCD Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |  |
| Common/Usual Name:                           | 2MP Monochrome LCD Monitor                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |
| Classification Name: Product Code:           | System, image processing 21CFR 892.2050<br>PGY                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |
| Predicate Device(s):                         | RADIFORCE GX240; K120407                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |  |
| Device Description:                          | JUSHA-M23C LCD Monitor is the display system with the high resolution (1600*1200), high luminance (1000cd/m²), and 1024 simultaneous shades of gray out of a palette of 4096, 10 DICOM look up table and 3 GAMMA look up table inside. JUSHA-M23C has ambient brightness adapting and presence induction system, with these this display can automatic adjustment according to different requirements in order to achieve the best results.<br>The product is consisted of the following components:<br>- 21.3inches, Mono-TFT LCD Panel<br>- JUSHA- SMS_19inch motherboard/FR-4/REV:0.1<br>- JUSHA-M23C LCD Monitor software<br>- Power Adapter<br>- Data Cable.                                                                                                                                     |  |
|                                              | The LCD Monitor is designed, tested, and will be manufactured in<br>accordance with both mandatory and voluntary standards:<br><br>1. IEC 60601-1Medical equipment medical electrical<br>equipment - Part 1: General requirements for basic safety<br>and essential performance<br>2005+CORR.1(2006)+CORR.2(2007)<br><br>2. IEC 60601-1-2 Edition 3:2007, Medical electrical<br>equipment - Part 1-2: General requirements for basic<br>safety and essential performance - Collateral standard:<br>Electromagnetic compatibility - Requirements and tests.                                                                                                                                                                                                                                            |  |
| Intended Use:                                | JUSHA-M23C LCD Monitor is intended to be used in displaying and<br>viewing digital images by trained medical practitioners.<br>JUSHA-M23C LCD Monitor does not support the display of<br>mammography images for diagnosis.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |
| Technology:                                  | JUSHA-M23C LCD Monitor is the display system with the high<br>resolution monitor (2 megapixels) with electronic capabilities for<br>evaluation of high resolution medical images, high luminance (1000<br>cd/m²) and 1024 simultaneous shades of gray out of a palette of 4096,<br>10 DICOM look up table and 3 GAMMA look up table inside.<br>JUSHA-M23C has ambient brightness adapting and presence<br>induction system, with these this display can automatic adjustment<br>according to different requirements in order to achieve the best<br>results.                                                                                                                                                                                                                                          |  |
| Determination of Substantial<br>Equivalence: | Summary of Non-Clinical Tests:<br>The LCD Monitor complies with voluntary standards as following:<br><br>1 IEC 60601-1Medical equipment medical electrical equipment -<br>Part 1: General requirements for basic safety and essential<br>performance 2005+CORR.1(2006)+CORR.2(2007)<br><br>2 IEC 60601-1-2 Edition 3:2007, Medical electrical equipment -<br>Part 1-2: General requirements for basic safety and essential<br>performance - Collateral standard: Electromagnetic compatibility<br>- Requirements and tests.<br><br>JUSHA-M23C is substantially equivalent to RADIFORCE GX240.<br>JUSHA-M23C employs the maximum resolution values same as that<br>of RADIFORCE GX240. Comparison table of the principal<br>characteristics of 2 devices is shown in the Attachment 1.<br>Attachment 1 |  |
|                                              | development of the system:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |
|                                              | • Risk Analysis                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |  |
|                                              | • Requirements Reviews                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |  |
|                                              | • Design Reviews                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |  |
|                                              | • Raw materials verification                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |  |
|                                              | • Testing on unit level (Module verification)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |  |
|                                              | • Integration testing (System verification)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |  |
|                                              | • Final acceptance testing (Validation)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |  |
|                                              | • Performance testing (Verification)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |  |
|                                              | • Safety testing (Verification)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |  |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |  |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |  |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |  |
|                                              | Summary of Clinical Tests:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |  |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |  |
|                                              | The subject of this premarket submission, LCD Monitor, did not                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |  |
|                                              | require clinical studies to support substantial equivalence.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |  |
|                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |  |
|                                              | The proposed device is Substantially Equivalent (SE) to the predicate                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |  |
|                                              | device which is US legally market device. Therefore, the subject                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |  |
|                                              | device is determined as safe and effectiveness.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |  |
| Conclusion:                                  | Nanjing Jusha Display Technology Co., Ltd Considers the                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |  |
|                                              | JUSHA-M23C LCD Monitor to be as safe, as effective, and                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |  |
|                                              | performance is substantially equivalent to the predicate device(s).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |  |

{4}------------------------------------------------

{5}------------------------------------------------

{6}------------------------------------------------

This comparison identifies the similarities and differences of the proposed JUSHA-M23C LCD Monitor device to the legally marketed predicate EIZO RADIFORCE GX240 LCD Monitor device to which substantial equivalency is claimed.

| Attributes                             | Predicate Device                                                                                                                                                                                                                                | Proposed Device                                                                                                                                                                                                                                 |
|----------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product                                | RADIFORCE GX240                                                                                                                                                                                                                                 | JUSHA-M23C LCD Monitor                                                                                                                                                                                                                          |
| 510(k) Number                          | K120407                                                                                                                                                                                                                                         | /                                                                                                                                                                                                                                               |
| Display Performance/Specifications     |                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                 |
| Screen technology                      | 21.3" Monochrome TFT<br>LCD Panel                                                                                                                                                                                                               | 21.3" Monochrome TFT<br>LCD Panel                                                                                                                                                                                                               |
| Viewing angle (H, V)                   | Horizontal 176 °,Vertical<br>176 °                                                                                                                                                                                                              | Horizontal 176 °,Vertical<br>176 °                                                                                                                                                                                                              |
| Resolution                             | 1600 x 1200/1200 x 1600                                                                                                                                                                                                                         | 1600 x 1200/1200 x 1600                                                                                                                                                                                                                         |
| Display area                           | 432.0 (H) x 324.0(V) mm                                                                                                                                                                                                                         | 432.0 (H) x 324.0(V) mm                                                                                                                                                                                                                         |
| Contrast Ratio                         | 1400:1                                                                                                                                                                                                                                          | 1400:1                                                                                                                                                                                                                                          |
| Scanning frequency (H; V)              | 31~100 kHz;59~61Hz                                                                                                                                                                                                                              | 52~76 kHz;59~61Hz                                                                                                                                                                                                                               |
| Recommended Luminance                  | 500cd/m²                                                                                                                                                                                                                                        | 500cd/m²                                                                                                                                                                                                                                        |
| Pixel Pitch                            | 0.27x0.27 mm                                                                                                                                                                                                                                    | 0.27x0.27 mm                                                                                                                                                                                                                                    |
| Backlight                              | LED                                                                                                                                                                                                                                             | LED                                                                                                                                                                                                                                             |
| DICOM LUT                              | 10-bit (Display Port) : 1024<br>8-bit: 256                                                                                                                                                                                                      | 12-bit:4096                                                                                                                                                                                                                                     |
| Luminance calibration                  | Built in calibration sensor provided                                                                                                                                                                                                            | Built in calibration sensor provided                                                                                                                                                                                                            |
| Input signals                          | DVI standard 1.0,<br>DisplayPort 1.1a                                                                                                                                                                                                           | DVI standard 1.0,<br>DisplayPort 1.1a                                                                                                                                                                                                           |
| Input terminational                    | DVI-D (dual link) x 1,<br>DisplayPort x 1                                                                                                                                                                                                       | DVI-D (dual link) x 1,<br>DisplayPort x 1                                                                                                                                                                                                       |
| Display controller                     | Off the shelf                                                                                                                                                                                                                                   | Off the shelf                                                                                                                                                                                                                                   |
| Power Requirement                      | AC 100~240V 50~60Hz                                                                                                                                                                                                                             | AC 100~240V 50~60Hz                                                                                                                                                                                                                             |
| Attributes                             | Predicate Device                                                                                                                                                                                                                                | Proposed Device                                                                                                                                                                                                                                 |
| Product                                | RADIFORCE GX240                                                                                                                                                                                                                                 | JUSHA-M23C LCD Monitor                                                                                                                                                                                                                          |
| 510(k) Number                          | K120407                                                                                                                                                                                                                                         | /                                                                                                                                                                                                                                               |
| Power<br>Consumption/Save Mode         | 36W/less than1.6W                                                                                                                                                                                                                               | 45W/less than 3W                                                                                                                                                                                                                                |
| Power<br>Management                    | DVI DMPM<br>DisplayPort 1.1a                                                                                                                                                                                                                    | DVI DMPM<br>DisplayPort 1.1a                                                                                                                                                                                                                    |
| USB<br>Ports/standard                  | 1 upstream (endpoint),<br>2 downstream/ Rev. 2.0                                                                                                                                                                                                | 1 upstream (endpoint),<br>2 downstream/ Rev. 2.0                                                                                                                                                                                                |
| Dimensions w/o<br>stand<br>(W x H x D) | Without stand:<br>376mmx505mmx98mm<br>With stand:<br>376mmx599mmx245.5 mm                                                                                                                                                                       | Without stand:<br>382mm x490mm x75mm<br>With stand:<br>382mm x533mm x238mm                                                                                                                                                                      |
| Indication for<br>use                  | RADIFORCE GX240 is<br>intended to be used in<br>displaying and viewing<br>digital images by trained<br>medical practitioners.<br>RADIFORCE GX240<br>Monitor does not support the<br>display of mammography<br>images for diagnosis.             | JUSHA-M23C LCD Monitor<br>is intended to be used in<br>displaying and viewing<br>digital images by trained<br>medical practitioners.<br>JUSHA-M23C LCD Monitor<br>does not support the display<br>of mammography images for<br>diagnosis.       |
| Attributes                             | Predicate Device                                                                                                                                                                                                                                | Proposed Device                                                                                                                                                                                                                                 |
| Product                                | RADIFORCE GX240                                                                                                                                                                                                                                 | JUSHA-M23C LCD Monitor                                                                                                                                                                                                                          |
| 510(k) Number                          | K120407                                                                                                                                                                                                                                         | /                                                                                                                                                                                                                                               |
| Applicable<br>standard                 | 1 IEC 60601-1Medical<br>equipment medical electrical<br>equipment - Part 1: General<br>requirements for basic safety<br>and essential performance<br>2005+CORR.1(2006)+COR<br>R.2(2007)                                                         | 1 IEC 60601-1Medical<br>equipment medical electrical<br>equipment - Part 1: General<br>requirements for basic safety<br>and essential performance<br>2005+CORR.1(2006)+COR<br>R.2(2007)                                                         |
|                                        | 2 IEC 60601-1-2 Edition<br>3:2007, Medical electrical<br>equipment - Part 1-2:<br>General requirements for<br>basic safety and essential<br>performance - Collateral<br>standard: Electromagnetic<br>compatibility - Requirements<br>and tests. | 2 IEC 60601-1-2 Edition<br>3:2007, Medical electrical<br>equipment - Part 1-2:<br>General requirements for<br>basic safety and essential<br>performance - Collateral<br>standard: Electromagnetic<br>compatibility - Requirements<br>and tests. |

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

### PERFORMANCE DATA:

The following performance data were provided in support of the substantial equivalence determination.

## Bench testing:

Bench testing was conducted to demonstrate the JUSHA-M23C meets all performance standards as follows:

- Measurement of the angular dependency of luminance response in horizontal, vertical and diagonal directions
- Measurement of the luminance non-uniformity characteristics of the display screen as specified in TGI18 guideline.
- Measurement of the chromaticity non-uniformity characteristics of the display screen as specified in TG18 guideline.
- Measurement of small-spot contrast ratio.
- Measurement of temporal response
- . Performance data on luminance stability

## Electrical safety and electromagnetic compatibility (EMC)

Electrical safety and EMC testing were conducted on the JUSHA-M23C. The device complies with the IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC.

## Animal and clinical study

The subject of this premarket submission, JUSHA-M23C, does not require animal or clinical studies to support substantial equivalence.

# CONCLUSIONS

JUSHA-M23C Medical Display is substantially equivalent to the predicate device with respect to technical characteristics, performance, application and intended use. The non-clinical data support the safety of the device should perform as intended in the specified use conditions. Nanjing Jusha Display Technology Co., Ltd considers the JUSHA-M23C Medical Display does not raise any new issues of safety or effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K151238](https://fda-staging.innolitics.com/device/K151238)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).