K151229 · Conmed Corporation · GEI · Jun 2, 2016 · General, Plastic Surgery
Device Facts
Record ID
K151229
Device Name
ABC D-FLEX PROBE
Applicant
Conmed Corporation
Product Code
GEI · General, Plastic Surgery
Decision Date
Jun 2, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ABC® D-Flex Probes are intended for use with a CONMED ABC® Generator for delivery of argon gas and electrosurgical current for non-contact ablation and hemostasis at the operative site. The ABC® D-Flex Probes are indicated for use during general, urological, and gynecological laparoscopic surgical procedures when manipulated by CONMED manual laparoscopic graspers or similar manual instruments.
Device Story
Monopolar electrosurgical device; delivers argon gas and high-frequency current for non-contact tissue ablation and hemostasis. Input: argon gas and electrical energy from compatible CONMED ABC generator. Output: ionized argon beam for coagulation. Used in laparoscopic surgery; operated by surgeon via footswitch; probe manipulated by manual laparoscopic graspers. Benefits: provides controlled hemostasis and ablation in minimally invasive settings. Device components: catheter tube, ceramic tip, D-Flex attachment, insufflation seal, shrink tube. Single-use, sterile.
Clinical Evidence
Bench testing only. Verification testing included grasping, release, manipulation, axial pull force, and argon flow. Functionality testing included beam initiation, non-contact ablation, and reliability. Ex-vivo comparison testing assessed thermal spread depth at comparable power settings. Materials comply with ANSI/AAMI/ISO 10993-1:2009.
Indicated for general, urological, and gynecological laparoscopic surgical procedures requiring non-contact ablation and hemostasis. Used with CONMED ABC generators and manipulated via manual laparoscopic graspers.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Omniguide Laser System with FlexGuide Ultra (K140378)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 2, 2016
CONMED Corporation Ms. Lisa Anderson Regulatory Affairs Specialist 525 French Road Utica, New York 11502
Re: K151229
Trade/Device Name: ABC D-Flex Probes Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: December 22, 2015 Received: December 24, 2015
Dear Ms. Anderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151229
Device Name
ABC D-Flex Probes
#### Indications for Use (Describe)
The CONMED ABC D-Flex Probes are intended for use with a CONMED ABC Generator for delivery of argon gas and electrosurgical current for non-contact ablation and hemostasis at the ABC D-Flex Probes are indicated for use during general, urological and gynecological laparoscopic surgical procedures when manipulated by CONMED manual laparoscopic graspers or similar manual instruments.
Type of Use (Select one or both, as applicable)
| <div> <span>☑</span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span>☐</span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
# CONMED ABC® D-Flex Probes™
This 510(k) summary is being submitted in accordance with the requirements of 21 CFR 807.92. CONMED Corporation is hereby submitting the 510(k) Summary for 510(k) number K151229 as of June 1, 2016.
## A. Submitter
CONMED Corporation 525 French Road Utica, N.Y. 13502 Registration Number: 1320894
## B. Company Contact
Lisa Anderson Manager, Regulatory Affairs CONMED Corporation 525 French Road Utica, N.Y. 13502 Phone: 315-624-3371 Fax: 315-624-3225 Email: lisaanderson@conmed.com
# C. Device Name
Trade Name: CONMED ABC® D-Flex™ Probes Common Name: ABC® Flexible Laparoscopic Probes Classification Name: Electrosurgical cutting and coagulation device and accessories
# D. Predicate Device
Laparoscopic 5mm ABC Probe, K925903 Birtcher Medical Systems acquired by CONMED Corporation Product Code: GEI Requlation Number: 878.440
Reference Device: CONMED ABC Flex Probes, K050161 Omniguide Laser System with FlexGuide Ultra, K140378
# E. Device Description
CONMED ABC® D-Flex Probes™ are monopolar, electrosurgical devices indicated for non-contact ablation and hemostasis of the operative site in laparoscopic surgery. The CONMED ABC® D-Flex Probes™ are composed of a catheter tube, ceramic tip, the D-
{4}------------------------------------------------
Flex attachment, insufflation seal, and a shrink tube. The CONMED ABC® D-Flex Probes™ are provided sterile for single use only and are controlled by the appropriate ABC® footswitch used in conjunction with compatible CONMED ABC generators.
The ABC® D-Flex Probes™ are intended for use with a CONMED ABC® Generator for delivery of argon gas and electrosurgical current for non-contact ablation and hemostasis at the operative site. The probes should be used with a maximum output of 6.5kV at a power setting of 80W.
The Laparoscopic 5mm Argon Beam Coagulation (ABC) Probe is intended to be used through a 5 millimeter (mm) laparoscopic trocar sleeve for coagulation and tissue desiccation only: this device will not cut or dissect tissue. The Laparoscopic ABC Probe is indicated for use only in situations where you would normally use monopolar energy.
Both the proposed and predicate devices are equivalent in their intended use for noncontact ablation and hemostasis, and utilize the same method of access via a cannula to the surgical site.
#### F. Intended Use of Device / Indications for Use
The ABC® D-Flex Probes are intended for use with a CONMED ABC® Generator for delivery of argon gas and electrosurgical current for non-contact ablation and hemostasis at the operative site. The ABC® D-Flex Probes are indicated for use during general, urological, and gynecological laparoscopic surgical procedures when manipulated by CONMED manual laparoscopic graspers or similar manual instruments.
#### H. Technological Characteristics
The proposed devices are equivalent to the predicate devices in their technological characteristics and performance. Both are designed for use with electrosurgical generators with ABC and monopolar high frequency current.
Both devices are designed to be activated using a foot control and to be used thru a laparoscopic cannula. However, the ABC D-Flex Probes are designed to be held by manual laparoscopic instruments compared to the Laparoscopic 5mm ABC probes which are hand-held devices.
#### l. Performance Data
Both devices are designed to comply with AAMI / ANSI ES60601-1:2005/(R)2012 Medical electrical equipment - Part 1: General requirements for basic safety and essential performance, IEC 60601-2-2:2009, and ANSI/AAMI ST67:2011. Materials analysis demonstrates the ABC® D-Flex Probe™ materials comply with the requirements of ANSI/AAMI/ISO 10993-1:2009 Biological Evaluation of Medical Devices Evaluation and Testing within a Risk Management Process. Benchtop
{5}------------------------------------------------
verification testing included grasping, release, manipulation, axial pull force, and argon flow. Functionality testing included beam initiation, non-contact ablation, and reliability. Ex-vivo comparison testing included an assessment of thermal spread depth at comparable power settings. The test results demonstrate the ABC® D-Flex Probes™ are safe and effective for their intended use and are substantially equivalent to the predicate devices.
### J. Substantial Equivalence
The differences between the predicate device and the proposed device do not raise any new issues of safety or efficacy. Supporting documentation per this premarket submission confirms that ABC® D-Flex Probes™ are substantially equivalent in design, manufacturing materials, intended use, principals of operation, and technical characteristics to the Laparoscopic 5mm ABC Probes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.