K151221 · Kub Technologies, Inc. · MQB · Jun 23, 2015 · Radiology
Device Facts
Record ID
K151221
Device Name
Kubtec DIGIVIEW 395
Applicant
Kub Technologies, Inc.
Product Code
MQB · Radiology
Decision Date
Jun 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The DIGIVIEW 395, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.
Device Story
DIGIVIEW 395 is a flat panel X-ray detector; utilizes amorphous silicon panel with directly deposited CsI:TI scintillator. Integrates into fixed room X-ray systems to enable digital radiography. Captures X-ray signals; transforms into digital images via 100μm pixel pitch detector. Operates via Gigabit Ethernet or 802.11n WiFi. Used in clinical settings by qualified service personnel/technicians. Includes auto exposure detection (AED) for integration with OEM systems lacking direct Prep/Expose signal access. Output sent to hardcopy printers, softcopy diagnostic workstations, or archives. Healthcare providers use images for diagnostic procedures; facilitates clinical decision-making by providing digital radiographic visualization of human anatomy.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence was demonstrated through bench testing to recognized consensus standards, including IEC 62220-1, IEC 62220-1-3, IEC 62494-1, IEC PAS 61910-1, NEMA PS 3.1-3.20, IEC 60601-2-43, and NEMA XR 24-2008.
Technological Characteristics
Flat panel detector; amorphous silicon TFT/diode array; CsI:TI scintillator. Active area 43.2 cm x 35.5 cm; 100μm pixel pitch. Housing: aluminum with carbon-fiber. Energy range: 20-150kVp. Connectivity: Gigabit Ethernet, 802.11n WiFi. Power: external supply or battery. Standards: IEC 62220-1, IEC 62494-1, DICOM (NEMA PS 3.1-3.20).
Indications for Use
Indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures in patients requiring conventional screen-film, digital radiography, or computed radiography. Not intended for mammographic use.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
Perkinelmer, XRpad 4336 MED Flat Panel Detector (K140551)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of unity and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular fashion around the emblem.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 23, 2015
KUB Technologies, Inc. % Vikram Butani President 270 Rowe Avenue, Unit E MILFORD CT 06461
Re: K151221
Trade/Device Name: Kubtec DIGIVIEW 395 Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MOB Dated: May 5, 2015 Received: May 7, 2015
Dear Vikram Butani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
For
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
# K151221
Device Name Kubtec DIGIVIEW 395
### Indications for Use (Describe)
The DIGIVIEW 395, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic X-ray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.
Type of Use (Select one or both, as applicable):
| <div> <span style="font-size:12pt"> <span style="font-family:Symbol"> ☑ </span> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:12pt"> <span style="font-family:Symbol"> ☐ </span> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for Kubtec. The logo features a stack of three squares in different shades of gray and blue on the left side. To the right of the squares is the word "kubtec" in red, with the tagline "GOING BEYOND THE SURFACE" in gray underneath.
# DIGIVIEW 395 510(K) Summary
## Identification of Device:
| Proprietary/Trade Name: | DIGIVIEW 395 (DV395) |
|-------------------------|--------------------------|
| Classification name: | SYSTEM, STATIONARY X-RAY |
| Classification: | Class II |
| CFR Section: | 21 CFR 892.1680 |
| Product Codes: | MQB |
| Common Name: | Stationary X-ray system |
## Predicate Device:
| Device Name: | Perkinelmer, XRpad 4336 MED Flat Panel Detector |
|--------------------|-------------------------------------------------|
| 510(k) Number: | K140551 |
| Classification | Class: II |
| Regulation Number: | 21 CFR 892.1680 |
| Product Code: | MQB |
| Regulation Name: | Stationary x-ray system |
#### I. INDICATIONS FOR USE
The DIGIVIEW 395, when used with a radiographic imaging system, is indicated for use in generating radiographic images of human anatomy for diagnostic Xray procedures, wherever conventional screen-film (SF), digital radiography (DR), or computed radiography (CR) systems may be used. It is not intended for mammographic use.
#### II. DEVICE DESCRIPTION
The DIGIVIEW 395 is a flat panel X-ray detector consisting of an amorphous silicon panel with a directly deposited CsI:TI scintillator.
The DIGIVIEW 395 detector has an active area of 43.2 cm x 35.5 cm at a pixel pitch of 100μm. Data and control communication is accomplished via a Gigabit Ethernet interface or 802.11n WiFi.
The detector can be integrated into a fixed room X-ray system to enable digital radiography. The following accessories are available for the DIGIVIEW 395
- DV395 LBP (Lithium Battery Pack) ●
- DV395 LBC (Lithium Battery Charger) ●
- DV395 IPU (Interface Power Unit) ●
- DV395 LPT Detector Cable, 3m/100ft ●
- DV395 Protective Insert ●
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for Kubtec. To the left of the logo are three overlapping squares in different shades of gray and blue. To the right of the squares is the word "kubtec" in red font, with the words "GOING BEYOND THE SURFACE" in gray font underneath.
- . DV395 Connector Cover Set
- Trigger cable (5m or 20m) ●
- . GigE interface cable (7.6m, 15.25m, or 30.5m)
The DIGIVIEW 395 is designed to work with X-ray systems (consisting of an Xray source, generator, collimator, and positioner) intended for use in generating radiographic images of human anatomy for diagnostic X-ray procedures that the following parameters may fulfil. Applicable detector parameters, such as dynamic range, exposure time range, energy range, image size, resolution, detective quantum efficiency, etc. are designed to support the necessary compatibility.
Energy range for compatibility is 20-150kVp.
The DIGIVIEW 395's auto exposure detection (AED) mode allows integration with OEM systems that there is no access to intercept the Prep/Expose signal.
Any modifications for hard wire integration to a previously installed OEM x-ray system that was utilized for analog or Computed Radiography are limited to integration between the hand-switch and it's connector to the x-ray console.
All integration work must be conducted by qualified service personnel and proper operation verified before releasing system for use on patients.
#### SUMMARY OF STUDIES AND SAFETY III.
Kubtec's DIGIVIEW 395 successfully completed internal safety testing requirements to recognized consensus standards listed below. No clinical studies were conducted in support of the DIGIVIEW 395 as agreed upon during Pre-Submission discussions with the Agency for the predicate device (K140551). The conduct of a clinical concurrence study was deemed unnecessary to demonstrate substantial equivalence.
Recognized Consensus Standards
IEC 62220-1 Edition 1.0 (2003-10) Medical electrical equipment -. Characteristics of digital X-ray imaging devices - Part 1: Determination of the detective quantum efficiency
IEC 62220-1-3 Edition 1.0 (2008-06) Medical electrical equipment -● Characteristics of digital X-ray imaging devices - Part 1-3: Determination of the detective quantum efficiency - Detectors used in dynamic imaging
IEC 62494-1 Edition 1.0 (2008-08) Medical electrical equipment -. Exposure index of digital X-ray imaging systems - Part 1: Definitions and requirements for general radiography
IEC PAS 61910-1 First Edition 2007-07 Medical electrical equipment -. Radiation dose documentation - Part 1: Equipment for radiography and radioscopy
. NEMA PS 3.1 - 3.20 (2011) Digital Imaging and Communications in Medicine (DICOM) Set
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Kubtec. The logo features a series of overlapping squares in shades of gray and blue on the left. To the right of the squares is the word "kubtec" in red, with the tagline "GOING BEYOND THE SURFACE" in smaller, dark blue letters underneath.
. IEC 60601-2-43 - Ed. 2.0 2010-03 Medical electrical equipment - Part 2 43: Particular requirements for the safety and essential performance of X-ray equipment for interventional procedures
NEMA XR 24-2008 (R2014) Primary user controls for interventional . angiography x-ray equipment
#### IV. BASIS FOR THE SUBMISSION
This summary of 510(k) safety and effectiveness information is supplied in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 to show Substantial Equivalence of the proposed device and the predicate device.
- PerkinElmer XRpad 4336 MED (K140551)
There is no change for the indication for use from the predicate 510(k) approved device.
| Design and Use of the Device | | | |
|------------------------------------------------------------------------|--|-----|----|
| Question | | YES | NO |
| Is the device intended for prescription use (21 CFR 801 Subpart D) ? | | X | |
| Is the device intended for over-the-counter use (21 CFR807 Subpart C)? | | | X |
| Is the device provided sterile? | | | X |
| Is the device intended for single use? | | | X |
| Is the device a reprocessed single use device? | | | X |
| If yes, does this device type require reprocessed validation data? | | | |
| Does the device contain a drug? | | | X |
| Does the device contain a biologic? | | | X |
| Does the device use software? | | X | |
| Does the submission include clinical information? | | | X |
| Is the device implanted? | | | X |
#### V. COMPARISON CHART
| Characteristics | Proposed device Kubtec<br>DIGIVIEW 395 | Predicate device<br>(K140551)<br>PerkinElmer XRpad<br>4336 MED |
|-----------------|------------------------------------------|----------------------------------------------------------------|
|-----------------|------------------------------------------|----------------------------------------------------------------|
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Kubtec logo. On the left side of the logo are three overlapping squares that are different shades of gray and blue. To the right of the squares is the word "kubtec" in red font. Below the word "kubtec" is the phrase "GOING BEYOND THE SURFACE" in a smaller, gray font.
| Intended Use/ Indications<br>for Use | The DIGIVIEW 395, when used with a<br>radiographic imaging system, is indicated<br>for use in generating radiographic images<br>of human anatomy for diagnostic X-ray<br>procedures, wherever conventional<br>screen-film (SF), digital radiography (DR),<br>or computed radiography (CR) systems<br>may be used. It is not intended for<br>mammographic use. | The XRPad 4336 MED,<br>when used with a<br>radiographic imaging<br>system, is indicated for<br>use in generating<br>radiographic images of<br>human anatomy for<br>diagnostic X-ray<br>procedures, wherever<br>conventional screen-<br>film (SF), digital<br>radiography (DR), or<br>computed radiography<br>(CR) systems may be<br>used. It is not intended<br>for mammographic<br>use. |
|--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Panel | Single substrate amorphous silicon active<br>TFT/diode array | Same |
| Scintillator | Direct deposition Csi:TI | Same |
| Pixel matrix | 3556 x 4320 pixels | Same |
| Pixel Pitch | 100 μm | Same |
| Active area | 355 mm x 432 mm | Same |
| External dimensions<br>(w x l x h) | 384 mm x 460 mm x 15 mm | Same |
| Weight | Approximately 4 kg | Same |
| Housing material | Aluminum with carbon-fiber | Same |
| Communication interface | Gb Ethernet or 802.11n WiFi | Same |
| Power | External power supply or battery | Same |
#### SUBSTANTIAL EQUIVALENCE VI.
The proposed device and predicate device DIGIVIEW 395 is substantially equivalent to the following currently cleared predicate device:
- PerkinElmer XRpad 4336 MED (K140551) a.
The proposed and predicate devices utilize the same technology and materials, comparable safety and effectiveness features, and is similar in design and construction. The Indications for Use and labeling are the same and our labeling contain the required Cautions, Warnings and Contraindications consistent to those required for similar cleared devices. Both systems produce digital images which can be sent to hardcopy printers, softcopy diagnostic workstations and/or stored in archive. The only difference is utilization of Kubtec's DIGICOM
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Kubtec. The logo consists of a series of overlapping squares in shades of gray and blue, followed by the word "kubtec" in red, and the tagline "GOING BEYOND THE SURFACE" in a smaller, sans-serif font. The logo is clean and modern, with a focus on the company name and tagline.
software to manage operation of detector which had previously received 510(K with the Kubtec KUB250 (K141539) and the Kubtec Digiview 250 (K103348)
#### VII. CONCLUSION
The DIGIVIEW 395 is as safe and effective as the predicate device, has no technological differences in the hardware, and has no new indications for use, thus rendering it substantially equivalent to the predicate devices and conforms to applicable medical device safety standards. The only difference is utilization of Kubtec's DIGICOM software to manage operation of detector which had previously received 510(K) with the Kubtec KUB250 (K141539) and the Kubtec Digiview 250 (K103348)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.