← Product Code [LLZ](/productcode/LLZ) · K151212

# Ziostation2 (K151212)

_Qi Imaging, LLC · LLZ · Nov 4, 2015 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151212

## Device Facts

- **Applicant:** Qi Imaging, LLC
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Nov 4, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

Ziostation2 is an image processing application software available for installation onto customer owned hardware. This application software can be networked to provide for sharing of resources. This application software receives medical images from modalities (mage scanning devices such as CT) or image archives such as PACS through network or media and provides for the viewing, quantification, manipulation, communication, printing, and management of medical images. This application software is intended for use by trained medical professionals to supplement generally accepted methods of interpreting radiological images. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using a monitor whose characteristics are approved by the regulatory agency governing the market within which Ziostation2 is being offered.

## Device Story

Ziostation2 is a DICOM-based image processing software installed on customer-owned hardware. It ingests medical images from modalities (CT, MRI, Ultrasound, etc.) or PACS. The software provides tools for viewing, quantification, manipulation, and management of these images. It is operated by trained medical professionals in clinical settings to assist in the analysis of anatomy, physiology, and pathology. The software includes specialized modules for cardiac analysis (coronary, RV, TAVR), perfusion analysis (CT/SPECT), lesion analysis, liver analysis, and MR tractography. Clinicians use the output to supplement standard radiological interpretation; the clinician retains ultimate responsibility for diagnosis. The device benefits patients by providing advanced visualization and quantitative tools to aid in clinical decision-making and monitoring of disease progression or treatment response.

## Clinical Evidence

Bench testing only. The device underwent integration testing, verification testing, regression testing, and Beta validation. Risk management processes were applied to address potential hazards. No clinical data was required or provided.

## Technological Characteristics

Software-only device; no physical materials. Operates as a DICOM image management and processing system. Features include image fusion, noise reduction, motion coherence, automated vessel labeling, HU range color coding, ROI measurement, polar map generation (stress/rest/reversibility), and auto-segmentation. Connectivity via network or media for PACS/modality integration.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Ziostation ([K070205](/device/K070205.md))
- Ziostation with Cardiac Function Analysis and Calcium Scoring Software ([K083446](/device/K083446.md))
- CARDIQ Express v2.0 ([K073138](/device/K073138.md))
- EBW NM 2.0 ([K111336](/device/K111336.md))
- PET VCAR ([K063324](/device/K063324.md))
- CT Perfusion 4 ([K052839](/device/K052839.md))
- Vitrea CT Myocardial Analysis ([K112531](/device/K112531.md))
- Syngo Volume Perfusion, CT Body ([K092013](/device/K092013.md))
- AW Volume Share 5 ([K110834](/device/K110834.md))
- Synapse 3D Liver Analysis ([K110186](/device/K110186.md))
- Signa HDxT HD ([K121676](/device/K121676.md))
- Achieva R2 1.5T/3.0T ([K110151](/device/K110151.md))
- SureSubtraction Ortho CSSO-001A ([K130960](/device/K130960.md))
- CardEP ([K031261](/device/K031261.md))
- Vitrea CT Transcatheter Aortic Valve Replacement Planning ([K122578](/device/K122578.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 4, 2015

Qi Imaging, LLC % Mr. Richard Ball Director, Regulatory & Quality Affairs 1301 Shoreway Road. Suite 325 BELMONT CA 94002

Re: K151212 Trade/Device Name: Ziostation2 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 8, 2015 Received: October 14, 2015

Dear Mr. Ball:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Ochs

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K151212

Device Name Ziostation2

Indications for Use (Describe)

Ziostation2 is an image processing application software available for installation onto customer owned hardware. This application software can be networked to provide for sharing of resources.

This application software receives medical images from modalities (mage scanning devices such as CT) or image archives such as PACS through network or media and provides for the viewing, quantification, manipulation, communication, printing, and management of medical images.

This application software is intended for use by trained medical professionals to supplement generally accepted methods of interpreting radiological images.

Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using a monitor whose characteristics are approved by the regulatory agency governing the market within which Ziostation2 is being offered.

| Type of Use (Select one or both, as applicable)                                      |                                               |
|--------------------------------------------------------------------------------------|-----------------------------------------------|
| <span style="font-size: 12px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary Ziostation2 Qi Imaging, LLC

This 510(k) summary is prepared in accordance with 21 CFR 807.92.

#### General Information

| Trade Name                         | Ziostation2                                                                                                            |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------|
| Common Name<br>Classification Name | Picture Archiving and Communications System (PACS)<br>System, Image Processing, Radiological (21 CFR § 892.2050 - LLZ) |
| Applicant:                         | Qi Imaging, LLC<br>1301 Shoreway Road, Suite 325<br>Belmont, CA 94002<br>Tel 650-413-1364<br>Fax 650-596-7319          |
| Contact                            | Richard Ball<br>Sr. Director, Regulatory and Quality Affairs                                                           |

#### Intended Use

- . Ziostation (hereinafter "this application software") is an image processing application software available for installation onto customer-owned hardware. This application software can be networked to provide for sharing of resources.
- This application software receives medical images from modalities (medical image scanning devices such as CT) or image archives such as PACS through network or media and provides for the viewing, quantification, manipulation, communication, printing, and management of medical images.
- . This application software is intended for use by trained medical professionals to supplement generally accepted methods of interpreting radiological images.
- . Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretations. Mammographic images may only be interpreted using a monitor whose characteristics are approved by the regulatory agency governing the market within which Ziostation2 is being offered.

Note: The clinician retains the ultimate responsibility for making the proper diagnosis based on standard radiological practices and visual comparison of the separate, unprocessed images. Ziostation2 is a tool to be used in support of those standard practices and visual comparisons.

{4}------------------------------------------------

## Predicate Devices

Predicate devices to Ziostation2 Software Tool

| Device Name                                                               | 510(k)<br>Number |
|---------------------------------------------------------------------------|------------------|
| Ziostation<br>(Multiple Protocol Enhancements)                            | K070205          |
| Ziostation with<br>Cardiac Function Analysis and Calcium Scoring Software | K083446          |
| CARDIQ Express v2.0                                                       | K073138          |
| EBW NM 2.0                                                                | K111336          |
| PET VCAR                                                                  | K063324          |
| CT Perfusion 4                                                            | K052839          |
| Vitrea CT Myocardial Analysis                                             | K112531          |
| Syngo Volume Perfusion, CT Body                                           | K092013          |
| AW Volume Share 5                                                         | K110834          |
| Synapse 3D Liver Analysis                                                 | K110186          |
| Signa HDxT HD                                                             | K121676          |
| Achieva R2 1.5T/3.0T                                                      | K110151          |
| SureSubtraction Ortho CSSO-001A                                           | K130960          |
| CardEP                                                                    | K031261          |
| Vitrea CT Transcatheter Aortic Valve Replacement Planning                 | K122578          |

#### Device Description

ZIOSTATION2 is a basic DICOM image management system to further aid clinicians in their analysis of anatomy, physiology and pathology. Universal functions such as data retrieval, storage, management, querying and listing, and output are handled by the basic Ziostation2 software. Various imaging tools and techniques can be invoked to process images from the following image types: CT, MRI, Ultrasound, Digital X-ray X-ray Angiography, PET, SPECT, NM, SC, Mammography, X-ray Radiofluoroscopic image, RT Image.

#### New Features – Substantial Equivalence Discussion

|                                       | Image Viewing/Tool Enhancements – Ref. Ziostation K070205 and Ziostation w/ Cardiac Function                                                                                                                          |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Analysis K083446 as predicate devices |                                                                                                                                                                                                                       |
| Capability                            | Substantial Equivalence                                                                                                                                                                                               |
| Fusion Viewer                         | K070205 Fusion Viewer is a workflow enhancement and is substantially equivalent to the current Ziostation ability to fuse images.                                                                                     |
| PhyZiodynamics                        | K070205 PhyZiodynamics reconstructs data with the addition of noise reduction and motion coherence to improve image quality and is substantially equivalent to the current Ziostation data reconstruction capability. |

{5}------------------------------------------------

| CT Coronary<br>Analysis    | K070205 (Secondary K073138) Improvements to CT Coronary include<br>automatic vessel labeling and color coding of HU ranges of specified areas<br>for easier visualization of suspected areas of interest. This tool with its<br>expanded capabilities is substantially equivalent to the current Ziostation<br>coronary analysis tool.                    |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CT Colon Analysis          | K070205 (Secondary K110834) The added capability to automatically<br>measure and report on user-specified ROI's is a workflow enhancement.<br>This capability is substantially equivalent to the current Ziostation CT<br>Colon tool.                                                                                                                     |
| CT/SPECT Cardiac<br>Fusion | K070205 (Secondary K111336) The functionality being added to the<br>already cleared CT/SPECT Cardiac Fusion ability of the current Ziostation<br>device is to provide data and polar maps of stress, rest, reversibility and<br>washout. This workflow enhancement is substantially equivalent to the<br>current Ziostation CT/SPECT Cardiac Fusion tool. |
| SUV Measurement            | K070205 (Secondary K063324) Uses current multi-exam comparison<br>capability of PET images to visualize and monitor disease progression or<br>response to treatment or therapy. This workflow enhancement is<br>substantially equivalent to the current Ziostation PET imaging tool.                                                                      |
| CT Subtraction             | K070205 (Secondary K130960) CT Subtraction improves the current<br>capability of contrast enhancement visualization. This improvement is<br>substantially equivalent to the current Ziostation contrast enhancement<br>capability.                                                                                                                        |
| EP Planning                | K070205 (Secondary K031261) Using currently available display and<br>measurement tools, EP Planning allows the physician to concentrate on<br>certain specific areas of the heart. This tool is substantially equivalent to<br>the current Ziostation display and measurement capability.                                                                 |
| TAVR                       | K070205 (Secondary K122578) Using currently available display and<br>measurement tools, TAVR allows the physician to concentrate on certain<br>specific areas of the heart, including aortic valve and surrounding<br>structures. This tool is substantially equivalent to the current Ziostation<br>display and measurement capability.                  |
| CT Cardiac RV              | K083446 (Secondary K110834) Using currently available display and<br>measurement tools, CT Cardiac RV adds the capability to calculate and<br>display various functional parameters of the right ventricle. This added<br>capability is substantially equivalent to the current Ziostation capability to<br>provide functional ventricular information.   |

## Additional Image Processing Functionalities

| CT Perfusion Analysis    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CT Perfusion<br>Analysis | K052839 Both the Qi Imaging feature and the GE predicate feature are<br>image analysis software applications that allow evaluation of dynamic CT<br>data following the injection of a compact bolus of contrast material,<br>generating information with regard to changes in image intensity over<br>time. Both features allow assessment of the type and extent of various<br>perfusion related parameters. The Qi Imaging CT Perfusion feature is<br>substantially equivalent to GE CT Perfusion 4. |
| CT Volume<br>Perfusion   | K052839 Both the Qi Imaging feature and the GE predicate feature are<br>image analysis software applications that allow evaluation of dynamic CT<br>data following the injection of a compact bolus of contrast material,<br>generating information with regard to changes in image intensity over                                                                                                                                                                                                     |

{6}------------------------------------------------

|                                    | time. Both features allow assessment of the type and extent of various<br>perfusion related parameters in user-specified planes of orientation. The Qi<br>Imaging CT Volume Perfusion feature is substantially equivalent to GE<br>CT Perfusion 4.                                                                                                                                                                   |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CT Myocardial<br>Perfusion         | K112531 Both the Qi Imaging feature and the Vital Images feature<br>provide color-coded polar maps of the contrasted myocardium based on<br>the calculation of transmural perfusion ratios. The Qi Imaging CT<br>Myocardial Perfusion feature is substantially equivalent to Vital Images<br>Vitrea CT Myocardial Analysis                                                                                           |
| CT Dynamic<br>Myocardial Perfusion | K092013 Both the Qi Imaging feature and the Siemens feature provide 3-<br>dimensional calculation of Blood Flow, Blood Volume and other<br>functional parameters by displaying color coded maps of individual<br>parameters on user-specified regions of the myocardium. The Qi Imaging<br>CT Dynamic Myocardial Perfusion feature is substantially equivalent to<br>Siemens syngo Volume Perfusion – CT Body        |
| CT Lesion Analysis                 | K110834 Both the Qi Imaging feature and the GE feature allow extraction<br>and measurement of lesions, enabling observation of changes over time.<br>The Qi Imaging CT Lesion Analysis feature is substantially equivalent to<br>GE AW Volume Share 5                                                                                                                                                                |
| CT Liver Analysis                  | K110186 Both the Qi Imaging CT Liver feature and the Fuji Film feature<br>allow auto-segmentation/calculation of liver features such as liver<br>structure, volumes, tumors and manual definition of separation plane<br>proposals. The Qi Imaging CT Liver Analysis feature is substantially<br>equivalent to Fuji Film Synapse 3D Liver Analysis                                                                   |
| MR Tractography                    | K121676 and K110151 Qi Imaging MR Tractography feature, the GE<br>device and the Philips device all present DTI Tractography data showing<br>colored tracts of various anatomical fibers of interest overlaid onto the<br>surrounding anatomy. The Qi Imaging MR Tractography feature is<br>substantially equivalent to the comparable software features of both GE<br>Signa HDXT and Philips Achieva R41.5T & 3.0T. |

# Materials

This software tool consists entirely of software. No materials are contained in this product.

# Testing Summary

The ZIOSTATION2 software package successfully completed integration testing/verification testing prior to Beta validation. Regression testing was also performed on all functionality present on Ziostation. Software Beta testing/validation was successfully completed prior to final testing and release. In addition, potential hazards have been addressed by the Qi Imaging Risk Management process.

# Summary of Substantial Equivalence

ZIOSTATION2 is substantially equivalent in intended use and function to the composite of predicate devices identified above.

---

**Source:** [https://fda-staging.innolitics.com/device/K151212](https://fda-staging.innolitics.com/device/K151212)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
