The Klockner Dental Implant Abutments are intended to be placed into dental implants to provide support for prosthetic reconstructions such as crowns, bridges. The Klockner Dental Implant Abutments include protective cap, temporary abutment, angled abutments, straight abutments, cast abutments are intended to be used with the Klockner Dental Implant Systems, models: Essential EC, Essential EC 1.5.
Device Story
Prosthetic components for dental implants; includes protective caps, temporary abutments, angled/straight abutments, and cast abutments. Used by dental clinicians to support crowns and bridges. Protective caps/temporary abutments used provisionally; angled/straight abutments used for definitive screw-retained restorations; cast abutments used for cement- or screw-retained restorations via casting technique. Components connect to Klockner Essential EC/EC 1.5 implants. Clinicians select components based on anatomical needs (transmucosal height, angulation) to facilitate prosthetic fitting. Benefits include anatomical characterization of gum margin and restoration of dental function. Provided non-sterile; requires clinician-performed steam sterilization before use.
Clinical Evidence
No clinical data. Bench testing only, including fatigue testing per ISO 14801, biocompatibility evaluation per ISO 10993-1/ISO 7405, and corrosion resistance testing (ISO 10993-15, ASTM F746, ASTM G71).
Technological Characteristics
Materials: Titanium alloy (ASTM F136, ISO 5832-3), Gold alloy 6019, Co-Cr-Mo alloy (ASTM F1537, ASTM F799, ISO 5832-12). Connection: Internal octagonal cone. Sterilization: Steam (ISO 17665-1).
Indications for Use
Indicated for patients requiring prosthetic reconstruction (crowns, bridges) supported by Klockner Dental Implant Systems (Essential EC, Essential EC 1.5).
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features an abstract design of three human profiles facing right, stacked on top of each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 29, 2016
SOADCO, S.L. Ms. Maria Mitjaneta Ouality Manager Avgda Fiter I Rosell., 4 Bis-local 2 Escaldes-Engordany, AD-700 ANDORRA
Re: K151194
> Trade/Device Name: Klockner Dental Implant Abutments (II) Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: II Product Code: NHA Dated: December 16, 2015 Received: December 28, 2015
Dear Ms. Mitianeta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for SOAD for KLOCKNER. The letters are large and gray, with the word "SOAD" in black. There is a globe image in the middle of the logo. The words "for KLOCKNER" are in a smaller font below the globe.
# INDICATIONS FOR USE
510(k) Number (if known):
K151194
Device Name:
Klockner Dental Implant Abutments (II)
#### Indications for Use:
The Klockner Dental Implant Abutments are intended to be placed into dental implants to provide support for prosthetic reconstructions such as crowns, bridges. The Klockner Dental Implant Abutments include protective cap, temporary abutment, angled abutments, straight abutments, cast abutments are intended to be used with the Klockner Dental Implant Systems, models: Essential EC, Essential EC 1.5.
V Prescription Use (Part 21 CFR 801 Subpart D) AND / OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for SOAD for KLOCKNER. The logo is in black and white and features the letters "SOAD" in a stylized font. The letters are arranged in a slightly overlapping manner. The words "for KLOCKNER" are written in a smaller font below the letters "SOAD". There is a globe image in the middle of the logo.
## 510(k) SUMMARY K151194
#### Date of Preparation of Summary: January 25, 2016
| Submitter/owner name: | SOADCO, S.L. | | |
|--------------------------|----------------------------------------|--|--|
| Submitter/owner address: | Avgda. Fiter i Rossell, 4bis - Local 2 | | |
| | ESCALDES - ENGORDANY | | |
| | AD-700 (ANDORRA) | | |
| Contact person: | Maria Mitjaneta | | |
| Phone: | +376 800 590 | | |
| Fax: | +376 800 594 | | |
| e-mail: | calidad@soadco.com | | |
| Device Trade Name: | Klockner Dental Implant Abutments (II) | | |
| Common Name: | Dental Abutments | | |
| Classification Name: | Endosseous Dental Implant Abutments | | |
| Regulation Number: | 21 CFR 872.3630 | | |
| Class: | Class II | | |
| Panel: | Dental | | |
| Product code: | NHA | | |
#### Legally Marketed (Predicate) Device:
| PREDICATE DEVICE<br>510(k) Number | Device Trade Name | Manufacturer |
|-----------------------------------|-----------------------------------|--------------|
| K122988 | Klockner Dental Implant Abutments | SOADCO S.L. |
#### Device Description:
Klockner Dental Implant Abutments (II) consist of a group of prosthetic components to be used with dental implants as an aid in prosthetic rehabilitation. According to their function, they are classified in:
- -Protective cap, made of titanium alloy, intended to be provisionally connected to dental abutment as a protection when choosing direct oral impression on the abutment.
- ı Temporary abutment, made of titanium alloy, intended to be provisionally connected to dental abutment to retain a provisional screw-retained multiple-unit restoration.
- Angled abutments, made of titanium alloy, intended to be definitively connected to dental implant to enable screw-retained multiple-unit restorations. These abutments, used when it is necessary to correct prosthesis axis with respect to implant axis, are available with different angulation and different transmucosal height, enabling the anatomic characterization of the gum margin and facilitating the prosthetic fitting.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo for SOAD for KLOCKNER. The logo is in black and white and features the letters "SOAD" in a stylized font. The letter "O" is replaced by a globe. Below the letters "SOAD" is the text "for KLOCKNER" in a smaller font. The logo is likely for a company or organization called SOAD that is associated with KLOCKNER.
- Straight abutments, made of titanium alloy, intended to be definitively connected to dental implant to enable screw-retained multiple-unit restorations, with not correction of implant angulation. These abutments are available with different transmucosal height, enabling the anatomic characterization of the qum margin and facilitating the prosthetic fitting.
- Cast abutments, made of gold alloy and Co-Cr-Mo alloy, intended for cement-retained single- unit and multiple-unit restorations and for screw-retained single-unit restorations on dental implants using the casting technique.
All these abutments are designed for Klockner Dental Implant Systems, models Essential EC (K080224) and Essential EC 1.5 (K082200), with internal octagonal cone connection.
#### Indications for Use:
The Klockner Dental Implant Abutments are intended to be placed into dental implants to provide support for prosthetic reconstructions such as crowns, bridges. The Klockner Dental Implant Abutments include protective cap, temporary abutments, straight abutments, cast abutments. All abutments are intended to be used with the Klockner Dental Implant Systems, models: Essential EC, Essential EC 1.5.
### Summary of Comparison with Predicate Device:
In the establishment of substantial equivalence, Klockner Dental Implant Abutments (II) are compared to a legally marketed device with the same intended use.
Proposed device intended use, indications for use and technological characteristics have been compared with those of predicate device (see next pages), following the guidelines set out in guidance document "Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments".
Features of proposed device not shared with predicate device have been substantiated using two reference devices. These differences are not considered to raise different questions of safety and effectiveness than the predicate.
All issues affecting performance have been reviewed and discussed. Any differences have been addressed in the bench testing performed on the proposed device.
| REFERENCE DEVICE<br>510(k) Number | Device Trade Name | Manufacturer |
|-----------------------------------|------------------------------------------|--------------------------|
| K080224 | Klockner Essential Dental Implant System | SOADCO S.L. |
| K121843 | NP-Cast Abutment System | OSSTEM Implant Co., Ltd. |
- NOTE: Numbers noted as 'ref. xx xx xx' (where x means a number) in the table below refer to specific catalog numbers of devices in the Klockner Implant System.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for SOADC for Klockner. The logo is in gray and features the letters "SOADC" in a stylized font. The "O" in "SOADC" contains a globe graphic. Below the "C" is the text "for KLOCKNER" in a smaller font.
| Features | PROPOSED DEVICE:<br>Klockner Dental Implant Abutments (II) | PREDICATE DEVICE K122988<br>(Klockner Dental Implant Abutments) | REFERENCE DEVICE K080224<br>(Essential Implant System) | REFERENCE DEVICE<br>K121843<br>(NP-Cast Abutment System) |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for use | The Klockner Dental Implant Abutments are<br>intended to be placed into dental implants to<br>provide support for prosthetic reconstructions<br>such as crowns, bridges. The Klockner Dental<br>Implant Abutments include protective cap,<br>temporary abutment, angled abutments, straight<br>abutments, cast abutments. All abutments are<br>intended to be used with the Klockner Dental<br>Implant Systems, models: Essential EC, Essential<br>EC 1.5. | The Klockner Dental Implant Abutments are<br>intended to be placed into dental implants to<br>provide support for prosthetic reconstructions<br>such as crowns, bridges or overdentures. The<br>Klockner Dental Implant Abutments include<br>healing caps, protective caps, temporary<br>abutments, angled abutments, straight<br>abutments, and overdentures. All abutments<br>are intended to be used with the Klockner<br>Dental Implant Systems, models: Essential<br>EC, Essential ES, Essential ECK, Essential<br>EC 1.5, SK2 and NK2. | The Klockner Essential implant system is<br>especially designed for surgical insertion into<br>the bone using additional material to replace<br>the root of the teeth, acting as the support for<br>the dental implants formed by implant<br>accessories.<br>The Essential Solid implants are fitted with an<br>internal octagonal conical connection<br>combined with an external octagonal<br>connection measuring 1.2 mm in height.<br>The Essential Cone implants are fitted with an<br>internal octagonal conical connection.<br>The Essential ECK implants are fitted with a<br>larger internal octagonal conical connection<br>than that of the Essential Cone implants.<br>Immediate loading is appropriate for the<br>Essential Solid, Essential Cone and Essential<br>ECK implants when good primary stability is<br>achieved with appropriate occlusal loading.<br>Abutments can be used in single tooth<br>replacements and multiple tooth restorations.<br>8mm Implants are not indicated for use as<br>unitary implants and for immediate load. | NP-Cast Abutment System is<br>intended for use with a dental<br>implant to provide support for<br>prosthetic restorations such as<br>crowns, bridges, or overdentures. |
| Design -<br>Protective<br>сар | EC multi-cone protective cap (ref. 10 05 16)<br>Protective cap intended to be provisionally<br>connected (contact ≤ 6 months) to 30º and 17º<br>angular Octacone® multi-cone abutments (ref. 10<br>10 37, 10 10 38, 10 10 43, 10 10 44, 10 10 45)<br>included in this submission, and also to 30º<br>anqular abutment multi-cone [occlusal screw] (ref<br>10 10 36) cleared in K122988, in order to cover<br>and protect their inner configuration when<br>choosing direct oral impression on these<br>abutments. | Proposed device has the same intended use<br>and is functionally equivalent to these<br>predicate devices:<br>EC 12º protective cap (ref. 10 05 14)<br>EC 250 protective cap (ref. 10 05 15)<br>These devices are also connected to dental<br>abutments in order to cover and protect their<br>inner configuration when performing direct<br>oral impression on them. | N/A | N/A |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for SOAD for KLOCKNER. The logo is in black and white, and the letters are large and bold. The letter "O" in SOAD has a globe in the middle of it. The words "for KLOCKNER" are in a smaller font below the letter "O".
| Features | PROPOSED DEVICE:<br>Klockner Dental Implant Abutments (II) | PREDICATE DEVICE K122988<br>(Klockner Dental Implant Abutments) | REFERENCE DEVICE K080224<br>(Essential Implant System) | REFERENCE DEVICE<br>K121843<br>(NP-Cast Abutment System) |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Design -<br>Temporary<br>abutment | Circular Ti fitting - EC multi-cone (ref. 10 13 16)<br>Abutment intended to be provisionally connected<br>(contact ≤ 6 months) to 30° and 17° angular<br>Octacone® multi-cone abutments (ref. 10 10 37,<br>10 10 38, 10 10 43, 10 10 44, 10 10 45) included<br>in this submission, and also to 30° angular<br>abutment multi-cone [occlusal screw] (ref. 10 10<br>36) cleared in K122988, for retaining a provisional<br>screw-retained multiple-unit restoration while<br>definitive prosthesis is manufactured. | Proposed device has the same intended use<br>and is functionally equivalent to this predicate<br>device:<br>Circular Ti Fitting - EC 25° (ref. 10 13 15)<br>This device is also provisionally connected to<br>dental abutment to retain a provisional screw-<br>retained multiple-unit restoration while<br>definitive prosthesis is manufactured. | N/A | N/A |
| Design -<br>Cast<br>abutments | Abutments intended for cement-retained single-<br>unit and multiple-unit restorations and for screw-<br>retained single-unit restorations on Essential<br>Cone dental implants, using casting technique.<br>The design consists on a metal-based part with a<br>casting part over-injected in plastic (burned-out in<br>casting process):<br>- G Gold Octacone® abutment (ref. 10 10 13 G):<br>gold cast abutment intended to be connected to<br>wide platform Essential Cone implants.<br>- Co-Cr Octacone® abutment (ref. 10 10 16): Co-<br>Cr cast abutment intended to be connected to<br>regular platform Essential Cone implants.<br>- G Co-Cr Octacone® abutment (ref. 10 10 16 G):<br>Co-Cr cast abutment intended to be connected<br>to wide platform Essential Cone Implants. | N/A | Proposed devices have the same intended<br>use and are functionally equivalent to<br>reference device:<br>- Gold Octacone® Abutment (ref. 10 10 13):<br>connected to regular platform Essential<br>Cone implants.<br>This abutment is also intended for cement-<br>retained single-unit and multiple-unit<br>restorations and for screw-retained single-unit<br>restorations on Essential Cone dental<br>implants, using casting technique.<br>The design also consists on a metal-based<br>part with a casting part over-injected in plastic<br>(burned-out in casting process). | N/A |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the logo for SOADC for Klockner. The logo is in black and white, and the letters are large and bold. The letter "C" in SOADC has a globe inside of it. The words "for Klockner" are in a smaller font size and are located below the "C".
| Features | PROPOSED DEVICE:<br>Klockner Dental Implant Abutments (II) | PREDICATE DEVICE K122988<br>(Klockner Dental Implant Abutments) | REFERENCE DEVICE K080224<br>(Essential Implant System) | REFERENCE DEVICE<br>K121843<br>(NP-Cast Abutment System) |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------|----------------------------------------------------------|
| Design -<br>Straight<br>abutments | Straight abutments intended to be connected to<br>Essential Cone dental implants (regular platform)<br>for screw-retained multiple-unit restorations, with<br>not correction of implant angulation.<br>Designed with top conical shape (25° respect to<br>the implant axis) to make easier the placement of<br>the multiple-unit restoration, allowing the<br>correction of possible divergences among the axis<br>of the implants involved.<br>Available with different height from implant<br>platform and different transmucosal height (TH)<br>for enabling the anatomic characterization of the<br>gum margin and facilitating the prosthetic fitting:<br>- 25° Pillar abutment - [2.5 MM] - [TH 1 MM] (ref.<br>10 10 40)<br>- 25° Pillar abutment - [4 MM] - [TH 2.5 MM] (ref.<br>10 10 41)<br>- 25° Pillar abutment - [5.5 MM] - [TH 4 MM] (ref.<br>10 10 42) | Proposed devices have the same intended<br>use and are functionally equivalent to<br>predicate device:<br>- 25° pillar abutment - 1mm (ref. 10 10 15)<br>This abutment is also intended to be<br>connected to Essential Cone dental implants<br>for screw-retained multiple-unit restorations,<br>with not correction of implant angulation.<br>It is designed with the same top conical shape<br>(25° respect to the implant axis) as that of<br>proposed devices, with 1 mm of height from<br>implant platform and without transmucosal<br>height (TH = 0). | N/A | N/A |
| Design -<br>Angled<br>abutments | Abutments with angulation that modifies the<br>prosthesis axis respect to the implant axis.<br>Intended to be connected to Essential Cone<br>dental implants (regular platform) for screw-<br>retained multiple-unit restorations.<br>Available with different angulation (30° and 17°)<br>and different transmucosal height (TH)1 for<br>enabling the anatomic characterization of the gum<br>margin and facilitating the prosthetic fitting:<br>- 30° angular Octacone® multi-cone abutment<br>[TH 4 mm] (ref. 10 10 37)<br>- 30° angular Octacone® multi-cone abutment<br>[TH 5 mm] (ref. 10 10 38)<br>- 17° angular Octacone® multi-cone abutment<br>[TH 2 mm] (ref. 10 10 43)<br>- 17° angular Octacone® multi-cone abutment<br>[TH 3 mm] (ref. 10 10 44)<br>- 17° angular Octacone® multi-cone abutment<br>[TH 4 mm] (ref. 10 10 45) | Proposed devices have the same intended<br>use and are functionally equivalent to<br>predicate device:<br>- 30° angular abutment multi-cone [occlusal<br>screw] (ref. 10 10 36)<br>This abutment is also intended to be<br>connected to Essential Cone dental implants<br>and be used for screw-retained multiple-unit<br>restorations.<br>It is designed with 30° angulation that modifies<br>the prosthesis axis respect to the implant axis,<br>and with transmucosal height (TH)1 of 3 mm. | N/A | N/A |
<sup>්</sup> Transmucosal height (TH) is the distance between implant-abutment connection] and prosthetic platform [abutment-prosthesis connection].
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image is a logo for SOADC for KLOCKNER. The logo is in black and white and features the letters "SOADC" in a stylized font. The "C" in SOADC is a globe. Below the letters is the text "for KLOCKNER" in a smaller font.
| Features | PROPOSED DEVICE:<br>Klockner Dental Implant Abutments (II) | PREDICATE DEVICE K122988<br>(Klockner Dental Implant Abutments) | REFERENCE DEVICE K080224<br>(Essential Implant System) | REFERENCE DEVICE<br>K121843<br>(NP-Cast Abutment System) |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| Material /<br>Composition | - Titanium alloy (ASTM F136, ISO 5832-3): used<br>in protective cap, temporary abutment, straight<br>abutments and angled abutments.<br>- Gold alloy 6019: used in gold cast abutment.<br>- Co-Cr-Mo alloy (ASTM F1537 alloy 1, ASTM<br>F799, ISO 5832-12): used in Co-Cr cast<br>abutments.<br>- POM C: used in casting part of cast abutments<br>(material not intended to be in contact with the<br>patient, since it is burned-out during the casting<br>process of the abutments). | - Titanium alloy (ASTM F136, ISO 5832-3):<br>used in protective caps, temporary abutment<br>and straight abutment. | - Gold alloy 6019: used in Gold Octacone®<br>Abutment.<br>- POM C: used in casting part of Gold<br>Octacone® Abutment. | - Co-Cr-Mo alloy: used in NP-<br>Cast Abutment System. |
| Sterility | All devices (protective cap, temporary abutment,<br>cast abutments, straight abutments and angled<br>abutments) are provided non-sterile.<br>These devices are individually packed and<br>provided non-sterile, but intended to be sterilized<br>before placement in the mouth. Instructions for<br>use contain the recommended sterilization<br>process validated as per standard ISO 17665-1. | Predicate devices (protective caps, temporary<br>abutment, straight abutment and angled<br>abutment) are also provided non-sterile.<br>As proposed devices, these predicates are<br>individually packed and provided non-sterile,<br>but intended to be sterilized before placement<br>in the mouth. The same sterilization process<br>(validated as per ISO 17665-1) is<br>recommended. | Reference device (cast abutment) is also<br>provided non-sterile.<br>As proposed devices, this reference device is<br>individually packed and provided non-sterile,<br>but intended to be sterilized before placement<br>in the mouth. The same sterilization process<br>(validated as per ISO 17665-1) is<br>recommended. | N/A |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for SOADC for KLOCKNER. The letters SOAD are in black, and the letters C and D are in white with a black outline. The words "for KLOCKNER" are in a smaller font size and are located below the letters C and D. The background of the logo is gray.
#### Summary Discussion of Non-Clinical Data:
- Biocompatibility .
All materials used in the manufacture of Klockner Dental Implant Abutments (II) have been subject to biological evaluation taking account the intended use of the devices and the nature and duration of contact with the patient, according to the requirements of applicable recognized standards ISO 10993-1:2009 and ISO 7405:2008. In the case of Co-Cr cast abutments, the potential degradation of the material in biological environments has also been contemplated; corrosion resistance and galvanic corrosion assays have been carried out according to ISO 10993-15, ASTM F746 and ASTM G71, and test results confirm the high corrosion resistance of Co-Cr-Mo alloy and show an insignificant galvanic corrosion of this material with respect to unalloyed titanium (Ti c.p.) and titanium alloy (Ti-6Al-4V). The materials and manufacturing methods used for the subject device are identical to those
used in the previously cleared devices, and so additional biocompatibility testing was not considered to be necessary to support the substantial equivalence of the subject device.
- Sterilization .
Klockner Dental Implant Abutments (II) have been subject to bioburden and sterility testing in accordance with ISO 11737-1 and ISO 11737-2, respectively. Steam sterilization validation according to ISO 17665-1 has been carried out to confirm the sterilization parameters.
- . Bench testing
Fatigue testing, conducted according to ISO 14801 and the Guidance for Industry and FDA Staff: Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Implant Abutments, has been performed considering worst-case conditions. Results obtained show that devices performance is acceptable as per their intended use.
#### Summary Discussion of Clinical Data:
No clinical data are included in this submission.
#### Conclusions:
We consider that the intended use, the operation principle and the technological characteristics of Klockner Dental Implant Abutments (II) are equivalent to those of predicate devices, although there is a slight difference in the indications for use statement of the subject device compared to the predicate with respect to implant compatibility. There are minor technological differences between the subject and predicate devices. The results of bench testing support the conclusion that risks associated with these technological changes have been adequately mitigated.
Therefore, substantial equivalence of Klockner Dental Implant Abutments (II) to the predicate devices may be established.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.