← Product Code [MQP](/productcode/MQP) · K151167

# NuVasive PEEK Corpectomy System (K151167)

_Nu Vasive, Incorporated · MQP · Sep 3, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151167

## Device Facts

- **Applicant:** Nu Vasive, Incorporated
- **Product Code:** [MQP](/productcode/MQP.md)
- **Decision Date:** Sep 3, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Nuvasive® PEEK Corpectomy System is a partial or total vertebral body replacement device indicated for use in the thoracolumbar spine (T1 to L5) to replace a diseased or damaged vertebral body caused by tumor or fracture, to restore height of a collapsed vertebral body, and to achieve decompression of the spinal cord and neural tissues. The PEEK Corpectomy System is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft or autograft may be used at the surgeon's discretion.

## Device Story

Vertebral body replacement system; replaces diseased/damaged vertebral bodies in thoracolumbar spine (T1-L5). Composed of PEEK core and modular endcaps; includes titanium or tantalum pins for radiopacity. Used by surgeons in clinical/OR settings to restore height and decompress neural tissues. Requires supplemental internal spinal fixation. Allograft/autograft optional. Non-sterile; user-sterilized before use.

## Clinical Evidence

Bench testing only. No clinical studies conducted. Performance demonstrated via static and dynamic axial compression, static and dynamic torsion (ASTM F2077), push-out testing, and subsidence analysis.

## Technological Characteristics

Materials: PEEK Optima LT-1 (ASTM F2026), titanium alloy (ASTM F136/ISO 5832-3), tantalum (ASTM F560/ISO 13782). Modular core and endcap design. Non-sterile; user-sterilized. No software or electrical components.

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- NuVasive Mesh ([K032476](/device/K032476.md))
- NuVasive X-Core® Expandable VBR System ([K142205](/device/K142205.md))
- Osteotech VBR ([K012254](/device/K012254.md))
- Medtronic Verte-Stack® Spinal System ([K070173](/device/K070173.md))
- Novel VBR Spinal System ([K050553](/device/K050553.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized graphic. The graphic depicts three overlapping human profiles facing to the right, symbolizing health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 3, 2015

NuVasive, Incorporated Ms. Olga Lewis Lead Specialist, Regulatory Affairs 7475 Lusk Boulevard San Diego, California 92121

Re: K151167

Trade/Device Name: NuVasive® PEEK Corpectomy System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: MQP Dated: August 4, 2015 Received: August 7, 2015

Dear Ms. Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Ms. Olga Lewis

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known)

K151167

### Device Name NuVasive® PEEK Corpectomy System

### Indications for Use (Describe)

The Nuvasive® PEEK Corpectomy System is a partial or total vertebral body replacement device indicated for use in the thoracolumbar spine (T1 to L5) to replace a diseased or damaged vertebral body caused by tumor or fracture, to restore height of a collapsed vertebral body, and to achieve decompression of the spinal cord and neural tissues. The PEEK Corpectomy System is intended to be used with supplemental internal spinal fixation systems that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft or autograft may be used at the surgeon's discretion.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the Nuvasive logo. The logo consists of a purple and gray abstract shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in a smaller font, with a gradient effect.

## 510(k) Summary

In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:

A. Submitted by:

> Olga Lewis Lead Specialist, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-3360

Date Prepared: August 4, 2015

#### B. Device Name

| Trade or Proprietary Name: | NuVasive® PEEK Corpectomy System             |
|----------------------------|----------------------------------------------|
| Common or Usual Name:      | Spinal Vertebral Body Replacement Device     |
| Classification Name:       | Spinal Intervertebral Body Fixation Orthosis |

| Device Class:   | Class II          |
|-----------------|-------------------|
| Classification: | 21 CFR § 888.3060 |
| Product Code:   | MQP               |

#### C. Predicate Devices

The subject PEEK Corpectomy System is substantially equivalent to a primary predicate device, NuVasive Mesh (K032476), and additional predicate devices, NuVasive X-Core® Expandable VBR System (K142205), Osteotech VBR (K012254), Medtronic Verte-Stack® Spinal System (K070173) and Novel VBR Spinal System (K050553).

#### D. Device Description

The NuVasive PEEK Corpectomy System is a vertebral body replacement system manufactured from Polyetheretherketone (PEEK) Optima LT-1 conforming to ASTM F2026. The PEEK Corpectomy System is made up of two primary components: a core and a set of endcaps. The core is offered in a variety of shapes and sizes to suit individual pathology and anatomical conditions of the patient. The modular endcaps are offered in multiple footprints and lordosis options.

The device contains pins made of either titanium alloy (Ti-6Al-4V) conforming to ASTM F136 or ISO 5832-3 or tantalum conforming to ASTM F560 or ISO 13782 to serve as radiopaque markers. The System is provided non-sterile, and is designed to be sterilized by the user before each use.

#### E. Indications For Use

The Nuvasive® PEEK Corpectomy System is a partial or total vertebral body replacement device indicated for use in the thoracolumbar spine (T1 to L5) to replace a diseased or damaged vertebral body caused by tumor or fracture, to restore height of a collapsed vertebral body, and to achieve decompression of the spinal cord and neural tissues. The PEEK Corpectomy System is intended to be used with supplemental internal spinal fixation

{4}------------------------------------------------

systems that are cleared by the FDA for use in the thoracic and lumbar spine. Allograft or autograft material may be used at the surgeon's discretion.

#### F. Technological Characteristics

The subject PEEK Corpectomy System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to have equivalent technological characteristics to its predicate devices through comparison in areas including design, intended use, material composition, and function. This device does not contain software or electrical equipment.

#### G. Performance Data

Nonclinical testing was performed to demonstrate that the subject PEEK Corpectomy System is substantially equivalent to other predicate devices. The following testing and analysis was performed:

- Static and dynamic axial compression per ASTM F2077
- Static and dynamic torsion per ASTM F2077 ●
- Push-out testing ●
- Subsidance analysis

The results demonstrate that the subject PEEK Corpectomy System is substantially equivalent to predicate devices. No clinical studies were conducted.

#### H. Conclusions

Based on the indications for use, technological characteristics, mechanical testing, and comparison to predicate devices, the subject PEEK Corpectomy System has been shown to be substantially equivalent to legally marketed predicate devices for its intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K151167](https://fda-staging.innolitics.com/device/K151167)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
