K151154 · Prosidyan, Inc. · MQV · Nov 10, 2015 · Orthopedic
Device Facts
Record ID
K151154
Device Name
FIBERGRAFT BG Morsels
Applicant
Prosidyan, Inc.
Product Code
MQV · Orthopedic
Decision Date
Nov 10, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
FIBERGRAFT™ BG Morsels - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ BG Morsels is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ BG Morsels must be used with autogenous bone marrow aspirate and autograft in the posterolateral spine. FIBERGRAFT™ BG Morsels is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization.
Device Story
Osteoconductive, resorbable bone graft substitute; composed of crystalline 45S5 bioactive glass fibers and microspheres. Granular form; ultra-porous scaffold; flash-sintered porous shell. Used in surgical settings by orthopedic surgeons; packed into non-load-bearing bony voids or gaps (spine, extremities, pelvis). Functions as scaffold for bone regeneration; resorbs over time; replaced by natural bone. Requires combination with autogenous bone marrow aspirate and autograft for posterolateral spine applications. Provides radiopacity and improved handling; facilitates hydroxyapatite layer formation in aqueous environments to bond with host bone. Does not provide structural support; requires rigid internal/external stabilization.
Clinical Evidence
No human clinical data. Bench testing included physical property evaluation, packaging integrity, and ISO 10993 biocompatibility. In vivo performance supported by two animal studies: 1) Ovine model (n=58) comparing device to NovaBone Putty and sham control in distal femur defects over 24 weeks; 2) Rabbit posterolateral spine fusion model (n=41) comparing device to predicate and autograft control over 26 weeks. Endpoints included radiographic, histological, histomorphometric, and biomechanical data. Results demonstrated new bone formation and substantial equivalence to predicates.
Technological Characteristics
Material: Crystalline 45S5 bioactive glass. Form: Granular, spherical, ultra-porous scaffold. Principle: Osteoconductive, forms hydroxyapatite surface layer in aqueous phosphate solutions. Sterilization: Not specified. Connectivity: None. Software: None.
Indications for Use
Indicated for patients requiring bone void filling in non-load-bearing skeletal defects (posterolateral spine, extremities, pelvis) resulting from surgery or trauma. Must be used with autogenous bone marrow aspirate and autograft in the posterolateral spine. Contraindicated for load-bearing applications.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
FIBERGRAFT™ BG Morsels Bone Graft Substitute (K141956)
FIBERGRAFT™ BG Morsels Bone Graft Substitute (K132805)
Reference Devices
NovaBone Resorbable Bone Graft Substitute (K052494)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing right, arranged in a way that suggests unity and connection. The profiles are black and are set against a white background. Encircling the profiles are the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 10, 2015
Prosidyan Incorporated % Ms. Janice Hogan Regulatory Counsel Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, Pennsylvania 19103
Re: K151154
Trade/Device Name: FIBERGRAFT™ BG Morsels - Bone Graft Substitute Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler device Regulatory Class: Class II Product Code: MQV Dated: September 18, 2015 Received: September 18, 2015
Dear Ms. Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
510(k) Number (if known) K151154
#### Device Name
FIBERGRAFT™ BG Morsels - Bone Graft Substitute
Indications for Use (Describe)
FIBERGRAFT™ BG Morsels - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ BG Morsels is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ BG Morsels must be used with autogenous bone marrow aspirate and autograft in the posterolateral spine.
FIBERGRAFT™ BG Morsels is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D) Subpart C)
_ Over-The-Counter Use (21 CFR 801
Subpart C
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995. *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number.
{3}------------------------------------------------
## 510(k) SUMMARY
## Prosidyan's FIBERGRAFT™ BG Morsels
## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Prosidyan, Inc. 30 Technology Drive Warren, NJ 07059 Phone: (908) 517-3666 Facsimile: (908) 325-0058 Contact Person: Charanpreet S. Bagga
Date Prepared: October 28, 2015
### Name of Device and Name/Address of Sponsor
FIBERGRAFT™ BG Morsels Bone Graft Substitute
### Common or Usual Name
Bone Void Filler
#### Classification Name
Resorbable Calcium Salt Bone Void Filler, 21 CFR 888.3045, product code MQV
### Predicate and Reference Devices
Predicate Devices: NovaBone Products, LLC's NovaBone Putty (K112773) (Primary Predicate); Prosidyan Inc., FIBERGRAFT™ BG Morsels Bone Graft Substitute (K141956, K132805) Reference Device: NovaBone Resorbable Bone Graft Substitute (K052494)
#### Intended Use / Indications for Use
FIBERGRAFT™ BG Morsels - Bone Graft Substitute is indicated only for bony voids or gaps that are not intrinsic to the stability of the bony structure. FIBERGRAFT™ BG Morsels is indicated to be gently packed into bony voids or gaps of the skeletal system (i.e., posterolateral spine, extremities and pelvis). These defects may be surgically created osseous defects or osseous defects created from traumatic injury to the bone. The product provides a bone void filler that resorbs and is replaced with bone during the healing process. FIBERGRAFT™ BG Morsels must be used with autogenous bone marrow aspirate and autograft in the posterolateral spine.
FIBERGRAFT™ BG Morsels is not indicated for use in load-bearing applications; therefore, standard internal or external stabilization techniques must be followed to obtain rigid stabilization.
{4}------------------------------------------------
# Technological Characteristics
The technological characteristics of the FIBERGRAFT™BG Morsels have not changed since the prior clearances of the product. The FIBERGRAFT™BG Morsels provide an osteoconductive, resorbable, biocompatible bone graft substitute that is to be gently packed into defect sites. The FIBERGRAFT™BG Morsels are made from crystalline 45S5 bioactive glass. Each granule of BG Morsels is created from a matrix of bioactive glass fibers and microspheres. Bioactive glass is defined as a group of glasses that has a compositional range that allows the formation of hydroxyapatite (HA) as a surface layer when exposed to an aqueous phosphate-containing solution such as simulated body fluid. The HA layer that forms in an aqueous phosphate-containing solution plays a significant role in forming a strong bond with natural bone. The granules provide an ultra-porous scaffold for desired biological response and improved handling characteristics, while optimizing radiopacity and resorption. BG Morsels are generally spherical in appearance and provided in granular form. The matrix is flash sintered to form a porous shell at its surface, which creates the generally spherical structure of the granules, while maintaining a level of porosity within each granule.
# Performance Data
Physical property evaluations, functional animal studies, and biocompatibility tests were performed for the FIBERGRAFT™BG Morsels device. Specifically, simulated distribution, whole package integrity and seal strength were tested on the defined packaging configuration. Accelerated and real time aging tests were also performed with passing results. The biocompatibility of the BG Morsels is demonstrated by ISO 10993 testing and the long history of clinical use of the bioactive glass material for the same intended use. In addition, the BG Morsels is composed of the same bioactive glass material with the same chemical composition and the same type and duration of patient contact as the NovaBone predicates. The physical functions and bioactivity of the BG Morsels were also evaluated. and results met the testing requirements. The device is considered bioactive based on in vitro studies that show apatite layer formation on the surface of the implant following immersion in simulated body fluid (SBF). These results have not been correlated to clinical performance.
Animal testing demonstrated evidence of new bone formation in critical size defects, consistent with FDA's recommendations for Class II synthetic bone graft substitutes. Two separate animal studies were conducted. The Prosidyan FIBERGRAFT™ BG Morsels product was evaluated and compared to a positive control, the predicate device (NovaBone Putty), and to a negative sham control (untreated defect) in an ovine model. The animal study evaluated the device performance in critical sized cancellous bone defects in the lateral distal femurs from 58 skeletally mature sheep, including radiographic, histological, histomorphometric, and biomechanical data. Testing of the BG Morsels in the ovine model is representative of the indications for use and range of anatomical sites proposed for the subject device. In addition, the study was conducted for a duration of 24 weeks with several interim evaluation points (i.e., 4, 8, 12, and 24 weeks), including a minimum of 3 animals per time point per treatment group. The results of the study demonstrated that the FIBERGRAFT™BG Morsels device performs as safely and as effectively as the predicate device, and any differences between the results of the device groups do not raise new types of safety or effectiveness concerns.
In addition, the FIBERGRAFT BG Morsels product was also evaluated in a rabbit study to further support device performance for its indications for use. The FIBERGRAFT™ BG Morsels product was compared to its predicate device as well as autograft (positive) control. The animal study evaluated device performance in a rabbit posterolateral spine fusion
{5}------------------------------------------------
model with 41 skeletally mature rabbits, including radiographic, histological, histomorphometric, and biomechanical data. Testing of the FIBERGRAFT ™ BG Morsels in the rabbit model is representative of the indications for use and range of anatomical sites proposed for the subject device. The results of the study through 26 week follow up demonstrated that the FIBERGRAFT™ BG Morsels device performs substantially equivalently to the predicate device, and any minor technological differences between the device groups do not raise new types of safety or effectiveness concerns. The device was compared to autograft in animal performance testing in order to demonstrate substantial equivalence to the predicate. ISO 10993-6 testing was also conducted, and results supported substantial equivalence of the FIBERGRAFT BG Morsels compared to its predicate. Fluid absorbability testing further confirmed the use of BMA and the of the product.
Therefore, performance testing demonstrated that the FIBERGRAFT™ BG Morsels device functions as intended and meets the requirements of class II bone void fillers as compared to the predicate devices.
# Substantial Equivalence
As demonstrated in performance testing, the FIBERGRAFT™ BG Morsels product is substantially equivalent to its predicate devices. FIBERGRAFT™ BG Morsels has the same intended use and similar indications, technological characteristics, and principles of operation as its predicate devices. Thus, FIBERGRAFT™ BG Morsels is substantially equivalent.
# Conclusions
FIBERGRAFT™ BG Morsels is an osteoconductive, resorbable, biocompatible bone graft substitute composed of bioactive glass. The product is substantially equivalent to its predicate devices for its intended use as a synthetic bone void filler and bone graft extender. Performance testing, including bench and in vivo data, demonstrated that the device functions as intended without raising new safety or effectiveness questions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.