Fabrication of temporary crowns, bridges, inlays, onlays, and veneers Fabrication of long-lasting temporary restorations Lining material for prefabricated temporary crowns made of composite (e.g., Protemp™ Crown) and metal (e.g., Iso-Form crowns)
Device Story
SuPro 100 is a two-component composite material based on multifunctional methacrylate esters used for fabricating temporary dental restorations. It is delivered via the Garant™ mixing and dispensing system. The device is used in a dental clinic setting by a dentist or dental professional. The material is mixed and applied to create temporary crowns, bridges, inlays, onlays, and veneers, or to line prefabricated crowns. The clinician uses the material to provide patients with functional, long-lasting temporary restorations during the interim period before permanent restoration placement. The increased filler content (approx. 50% by weight) compared to the predicate improves mechanical properties like flexural strength, e-Modulus, and tensile strength, potentially benefiting the patient through increased durability of the temporary restoration.
Clinical Evidence
Bench testing only. Physical and mechanical properties compared to predicate included flexural strength, e-Modulus, compressive strength, tensile strength, impact strength, surface hardness, and abrasion behavior. Biocompatibility evaluated per ISO 7405 and ISO 10993 series. Results indicate SuPro 100 has higher flexural strength, e-Modulus, and tensile strength, and comparable impact strength and abrasion behavior to the predicate.
Technological Characteristics
Two-component composite material based on multifunctional methacrylate esters. Delivery via Garant™ mixing and dispensing system. Filler content approx. 50% by weight. Standards referenced: ISO 179-1, ISO 527-1, ISO 4049, ISO 9917, ISO 7405, ISO 10993 series, and ISO 14971 for risk management.
Indications for Use
Indicated for dental patients requiring temporary restorations including crowns, bridges, inlays, onlays, veneers, and lining for prefabricated temporary crowns.
Regulatory Classification
Identification
A temporary crown and bridge resin is a device composed of a material, such as polymethylmethacrylate, intended to make a temporary prosthesis, such as a crown or bridge, for use until a permanent restoration is fabricated.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 14, 2015
3M Deutschland GmbH Dr. Desi W. Soegiarto Group Leader Regulatory Affairs ESPE Platz Seefeld, Bavaria 82234
Re: K151144
Trade/Device Name: SuPro 100 Regulation Number: 21 CFR 872.3770 Regulation Name: Temporary crown and bridge resin Regulatory Class: II Product Code: EBG Dated: August 05, 2015 Received: July 31, 2015
Dear Dr. Soegiarto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
510(k) Number (if known)
K151144
Device Name SuPro 100
Indications for Use (Describe)
· Fabrication of temporary crowns, bridges, inlays, onlays, and veneers
· Fabrication of long-lasting temporary restorations
· Lining material for prefabricated temporary crowns made of composite (e.g., Protemp™ Crown) and metal (e.g., Iso-Form crowns)
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
FORM FDA 3881 (1/14)
{3}------------------------------------------------
# K151144: 510(k) SUMMARY
## Submitter
| Company: | 3M Deutschland GmbH |
|-----------------------------------|---------------------------------------------------|
| Street: | ESPE Platz |
| ZIP-Code, City: | D-82229 Seefeld |
| Federal State: | Bavaria |
| Country: | Germany |
| Establishment Registration Number | 9611385 |
| Official Correspondent & Contact: | Dr. Desi W. Soegiarto,<br>Regulatory Affairs |
| Phone: | +49-8152-700 1169 |
| Fax: | +49-8152-700 1869 |
| E-mail: | desi.soegiarto@mmm.com |
| Date: | April 27, 2015<br>(revised on September 10, 2015) |
## Name of Device
| Product Code: | EBG |
|--------------------|--------------------------------------------------------|
| Common Name: | Composite based temporary crown and<br>brigde material |
| Classification: | Crown and Bridge, Temporary, Resin |
| Regulation Number: | 21 CFR 872.3770 |
| Proprietary Name: | SuPro 100 |
## Predicate Device
| TempXN28 | K073296 |
|----------|---------------------------------|
| | by 3M Deutschland GmbH, Germany |
{4}------------------------------------------------
#### Description for the Premarket Notification
SuPro 100 is classified as temporary crown and bridge resin (21 C.F.R. § 872.3770, Class II device).
As its predicate device TempXN28 (K073296, by 3M Deutschland GmbH - Germany), SuPro 100 is a composite material based on multifunctional methacrylates esters, therefore, both TempXN28 and SuPro 100 as two-component system have filling material-like properties. As TempXN28, Supro 100 will be available in Garant™ mixing and dispensing system (by 3M Deutschland GmbH, Germany).
Investigations have been carried out to characterize the performance of SuPro 100 to be used for fabrication of temporary restorations.
Comparison for indications for use, performance, and chemistry shows that SuPro 100 is substantially equivalent to the predicate device.
Physical and mechanical properties have been compared to properties of TempXN28. Comparison showed that physical and mechanical properties of SuPro 100 are substantially equivalent to TempXN28.
Biocompatibility evaluations have been performed for SuPro 100 in consideration of FDA & internationally recognized guidelines. The conclusion of the assessments is that SuPro 100 is biocompatible for its intended use.
### Indications for Use
- Fabrication of temporary crowns, bridges, inlays, onlays, and veneers
- Fabrication of long-lasting temporary restorations .
- . Lining material for prefabricated temporary crowns made of composite (e.g., ProtempTM Crown) and metal (e.g., Iso-Form crowns)
{5}------------------------------------------------
| | SuPro 100 | TempXN28 (K073296) |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | Fabrication of temporary<br>crowns, bridges, inlays,<br>onlays, and veneers | Fabrication of temporary<br>crowns, bridges, inlays,<br>onlays, and veneers |
| | Fabrication of long-<br>lasting temporary<br>restorations | Fabrication of long-<br>lasting temporary<br>restorations |
| | Lining material for<br>prefabricated temporary<br>crowns made of<br>composite (e.g.,<br>Protemp™ Crown) and<br>metal (e.g., Iso-Form<br>crowns) | Lining material for<br>prefabricated temporary<br>crowns made of<br>composite (e.g.,<br>Protemp™ Crown) and<br>metal (e.g., Iso-Form<br>crowns) |
| | | |
| | SuPro 100 | TempXN28 (K073296) |
| Technological<br>Characteristic | Delivery: Garant™<br>mixing and dispensing<br>system | Delivery: Garant™<br>mixing and dispensing<br>system |
| | Composite material<br>based on multifunctional<br>methacrylate esters | Composite material<br>based on multifunctional<br>methacrylate esters |
| | Two component system:<br>Base paste and catalyst<br>paste | Two component system:<br>Base paste and catalyst<br>paste |
# Tab. Substantial equivalence: Comparisons
{6}------------------------------------------------
| Test | Results |
|------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------|
| Flexural strength [MPa] | Flexural strength of SuPro 100 was higher than that<br>of TempXN28. |
| e-Modulus [GPa] | e-Modulus of SuPro 100 was higher than that of<br>TempXN28. |
| Deflection [mm] | Deflection values of SuPro 100 was comparable to<br>that of TempXN28. |
| Compression strength [MPa] | SuPro 100 shows lower values for compressive<br>strength in compare to that of TempXN28. |
| Compression at break [%] | SuPro100 has lower compression percentage at<br>break compared to TempXN28. |
| Tensile strength [MPa] | Tensile strength of SuPro 100 was higher than that<br>of TempXN28. |
| Elongation at break [%] | Value for the elongation till break of SuPro 100 was<br>comparable to that of TempXN28. |
| Impact strength [kJ/mm²] | Impact strength of SuPro 100 was comparable to<br>that of TempXN28. |
| Surface hardness (Vickers scale): 1d,<br>36°C in water & 4d, 36°C in water | SuPro 100 had a significantly higher surface<br>hardness compared to TempXN28. |
| Temperature peak (1cm³ Volume)<br>[°C] | The maximum temperatures resulting from the<br>polymerization reaction were comparable for SuPro<br>100 and TempXN28. |
| Setting characteristics (Physica;<br>23°C). Working time (tA) [min],<br>Final setting (tE) [min], Setting<br>transition (dt) [min] | Compared to TempXN28, SuPro 100 generally has a<br>faster setting. Working time, transition time and<br>final setting time are shorter for SuPro 100. |
| ACTA abrasion (3 body wear) | SuPro 100 had a similar abrasion behavior in the<br>ACTA 3 body wear test in compare to TempXN28. |
| Chewing simulation volume loss<br>[µm³] | SuPro 100 had lower rate of volume loss than<br>TempXN28. |
| Elcometer Abrasion depths samples<br>[µm] | Abrasion of SuPro 100 is lower than of TempXN28. |
| Elcometer Volume loss steatite<br>antagonist [mm] | SuPro 100 showed lower rate of wear on the steatite<br>antagonist in compare to TempXN28. |
{7}------------------------------------------------
Following Standards have been referenced for both SuPro 100 and the predicate device TempXN28 (K073296):
There are no standards specifically for temporary crown and bridge materials. Therefore, for some material testing standards were referenced: ISO 179-1, ISO 527-1, ISO 4049, ISO 9917.
Biocompatibility: ISO 7405, ISO 10993-1, ISO 10993-3, ISO 10993-5, ISO 10993-10, ISO 10993-12
Risk Management: ISO 14971
Main difference between SuPro 100 and TempXN28 is amount of fillers. To improve mechanical properties e.g. flexural strength, e-Modulus and tensile strength, amount of fillers of SuPro 100 is increased in compare to TempXN28 (SuPro 100: approx. 50% of its weight, TempXN28: approx. 33% of its weight).
SuPro 100 and its predicate device TempXN28 (K073296) are very similar in material class, composition, and technology. SuPro 100 and TempXN28 have same indications for use and procedure as bis-acrylic composite dental temporary materials. Comparisons showed that physical and mechanical properties of SuPro 100 are comparable to those of TempXN28. Both SuPro 100 and TempXN28 are biocompatible for its intended use. In summary, it can be concluded that SuPro 100 is substantially equivalent to the predicate device TempXN28 (K073296).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.