PROBEAT-V

K151132 · Hitachi, Ltd., Health Care Company · LHN · Jun 1, 2015 · Radiology

Device Facts

Record IDK151132
Device NamePROBEAT-V
ApplicantHitachi, Ltd., Health Care Company
Product CodeLHN · Radiology
Decision DateJun 1, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Device Story

PROBEAT-V is a proton beam therapy system; generates and delivers therapeutic proton beams to localized tumors. System comprises beam production (LINAC, synchrotron, beam transport) and beam delivery subsystems (rotating gantry or fixed beam room, scanning nozzle, patient positioning system, X-ray imaging). Operated by clinical staff in a hospital/clinic setting. Device delivers prescribed dose and dose distribution to target site. X-ray imaging assists in patient positioning. System performance verified through mechanical, beam dose, and safety interlock testing. Benefits patients by providing targeted radiation therapy for tumors.

Clinical Evidence

Bench testing only. Performance evaluation included mechanical testing of gantry and patient couch, beam dose shape and distribution accuracy, and safety interface testing (beam stop, dose monitoring, interlocks). Electrical safety and EMC testing performed per IEC 60601-1 and IEC 60601-1-2.

Technological Characteristics

Proton beam irradiation system. Components: LINAC, synchrotron, beam transport, rotating gantry, scanning nozzle, patient couch, X-ray imaging. Energy source: charged-particle radiation. Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC).

Indications for Use

Indicated for patients with localized tumors and other conditions susceptible to radiation therapy.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 June 1, 2015 Hitachi, Ltd., Health Care Company % Mr. Jonathan Kahan Regulatory Counsel Hogan Lovells US LLP 555 13th Street NW WASHINGTON DC 20016 Re: K151132 Trade/Device Name: PROBEAT-V Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: LHN Dated: April 28, 2015 Received: April 28, 2015 Dear Mr. Kahan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael O'Hara For Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known) K151132 Device Name #### PROBEAT-V Indications for Use (Describe) The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov {3}------------------------------------------------ # 510(k) SUMMARY # Hitachi PROBEAT-V # Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Hitachi, Ltd., Health Care Company 4-14-1 Sotokanda, Chiyodaku, Tokyo, 101-8010, Japan Telephone: +81 (3) 45643563 Facsimile: +81 (3) 45642882 Contact Person: Naoya Nishimura Date Prepared: April 28, 2015 # Name of Device and Name/Address of Sponsor PROBEAT-V Proton Beam Therapy System Hitachi, Ltd. Power Systems Company 3-1-1 Saiwai-cho, Hitachi-shi, Ibaraki-ken, 317-8511 Japan #### Common or Usual Name Proton beam therapy system # Classification Name Medical Charged-Particle Radiation Therapy System, 21 CFR 892.5050, Product Code LHN #### Predicate Devices Hitachi Ltd. PROBEAT with Discrete Spot Scanning System (K073059) # Intended Use / Indications for Use The PROBEAT-V is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. # Technological Characteristics The PROBEAT-V is a proton beam irradiation system, which provides a therapeutic proton beam for clinical treatment. It is designed to deliver a proton beam with the prescribed dose, dose distribution and directed to the prescribed patient site. The equipment to perform the above work is composed of two main components: (1) a beam delivery system {4}------------------------------------------------ properly delivered and (2) equipment necessary to generate the proton beam and direct it to the beam delivery system for patient treatment. The beam delivery system is composed of the following components" - o Gantry Room - · Rotating Gantry - . Scanning Nozzle - . Patient Couch (or Patient Positioning System) - . X-ray Imaging System - o Fixed Beam Room - Scanning Nozzle . - . Patient Couch (or Patient Positioning System) - . X-ray Imaging System The beam production system is composed of the following components" - Accelerator system (LINAC, Synchrotron) O - Beam transport system (Low/High Energy Beam Transport systems) O # Performance Data The company performed testing to evaluate, - The mechanical performance of the rotating gantry and patient couch o - o Beam performance testing to evaluate beam dose shape and beam dose - Safety interface testing to evaluate beam stop control, dose monitor, area safety, o and mechanical interlocks. Further, electrical safety and electromagnetic compatibility testing was also performed in accordance with IEC 60601-1 and IEC 60601-1-2. In all instances, the PROBEAT-V functioned as intended and met its specifications. # Substantial Equivalence The PROBEAT-V has the same intended use and indications for use, as well as substantially similar principles of operation and technological characteristics, as compared to Hitachi's cleared PROBEAT with DSSS (K073059). Both systems comprise the same components of the beam production and beam delivery subsystems. The minor differences between the PROBEAT-V and the cleared PROBEAT do not raise different questions of safety or effectiveness. Thus, the PROBEAT with DSSS is substantially equivalent to its predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...