K151110 · Medtronic, Inc. · DTN · May 27, 2015 · Cardiovascular
Device Facts
Record ID
K151110
Device Name
Intersept Filtered Cardiotomy Reservoir
Applicant
Medtronic, Inc.
Product Code
DTN · Cardiovascular
Decision Date
May 27, 2015
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Intersept filtered cardiotomy reservoir is indicated for use in the cardiopulmonary bypass circuit during surgery for: · an air/fluid separation chamber; - · a temporary storage reservoir for priming solution and blood; - · the filtration of particulates from bank blood and filtration of blood recovered from the field by suction: - · the addition of medications or other fluids. The Intersept filtered cardiotomy reservoir is also indicated for use after open heart surgery for: • the post-operative collection of autologous blood from the chest and the aseptic return of blood to the patient for volume replacement.
Device Story
Single-use cardiotomy reservoir; 2,600 ml capacity; 2 LPM max flow. Functions as air/fluid separation chamber, storage for priming solution/blood, and filtration unit for bank/field-recovered blood. Features eight access ports (suction, pre/post-defoamer, vent). Contains open-cell polyurethane defoamer and 20-micron microaggregate filter with polyester sleeve. Used in OR during cardiopulmonary bypass or post-operatively for autotransfusion. Operated by perfusionists or clinical staff. Filters particulates; allows medication/fluid addition; enables aseptic return of autologous blood to patient for volume replacement. Benefits patient by reducing particulate load and facilitating blood salvage.
Clinical Evidence
Bench testing only. Testing included filtration efficiency, blood trauma, and biocompatibility assessments. All tests passed.
Technological Characteristics
Single-use reservoir; 2,600 ml capacity; 2 LPM max flow. Materials: Polypropylene frame, open-cell polyurethane defoamer, 20-micron microaggregate filter, polyester sleeve. Eight luer/0.6 cm ID access ports. Sterile, non-pyrogenic fluid path. Mechanical filtration and air/fluid separation principle.
Indications for Use
Indicated for patients undergoing cardiopulmonary bypass surgery or post-operative autologous blood collection. Contraindicated in cases of lung air leaks, chest wall perforations, infection, malignancy, gross contamination (bone cement, etc.), aspiration of non-parenteral agents, open chest with vacuum, post-protamine administration, return to surgery, or use of unregulated vented chest tubes.
Regulatory Classification
Identification
A cardiopulmonary bypass blood reservoir is a device used in conjunction with short-term extracorporeal circulation devices to hold a reserve supply of blood in the bypass circulation.
Special Controls
*Classification.* Class II (special controls), except that a reservoir that contains a defoamer or filter is classified into the same class as the defoamer or filter. The device, when it is a cardiopulmonary bypass blood reservoir that does not contain defoamers or blood filters, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
Intersept® Cardiotomy Reservoir with Filter
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol consists of three stylized human profiles facing right, stacked on top of each other, with a ribbon-like element flowing beneath them.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 27, 2015
Medtronic, Inc. Chelsea Pioske Regulatory Affairs Specialist 7611 Northland Drive Minneapolis, Minnesota 55428
Re: K151110
Trade/Device Name: Intersept Filtered Cardiotomy Reservoir Regulation Number: 21 CFR 870.4400 Regulation Name: Cardiopulmonary Bypass Blood Reservoir Regulatory Class: Class II Product Code: DTN. DTP Dated: April 24, 2015 Received: April 27, 2015
Dear Chelsea Pioske:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Hilleman
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K151110
Device Name
Intersept™ Filtered Cardiotomy Reservoir
Indications for Use (Describe)
The Intersept filtered cardiotomy reservoir is indicated for use in the cardiopulmonary bypass circuit during surgery for: · an air/fluid separation chamber;
- · a temporary storage reservoir for priming solution and blood;
- · the filtration of particulates from bank blood and filtration of blood recovered from the field by suction:
- · the addition of medications or other fluids.
The Intersept filtered cardiotomy reservoir is also indicated for use after open heart surgery for:
• the post-operative collection of autologous blood from the chest and the aseptic return of blood to the patient for volume replacement.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Date Prepared: | April 23, 2015 |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Medtronic, Inc.<br>Medtronic Perfusion Systems<br>7611 Northland Drive<br>Minneapolis, MN 55428<br>Establishment Registration Number: 2184009 |
| Contact Person: | Chelsea Pioske<br>Regulatory Affairs Specialist<br>Phone: (763) 514-9838<br>Fax: (763) 367-8361<br>E-mail: chelsea.pioske@medtronic.com |
| Alternate Contact: | Susan Fidler<br>Senior Regulatory Affairs Manager<br>Medtronic Perfusion Systems<br>Phone: 763.514.9839<br>Fax: 763.367.8360<br>Email: susan.c.fidler@medtronic.com |
| Trade Name: | Intersept™ Filtered Cardiotomy Reservoir |
| Classification Name: | Reservoir, Blood, Cardiopulmonary Bypass; Defoamer,<br>Cardiopulmonary Bypass |
| Regulation Number: | 21 CFR 870.4400, 21 CFR 870.4230 |
| Product Classification: | Class II |
| Product Code: | DTN, DTP |
Name of Predicate Device: Intersept® Cardiotomy Reservoir with Filter
#### Device Description
The Intersept™ Filtered Cardiotomy Reservoir is a single-use, device with a sterile, nonpyrogenic fluid path. The maximum capacity of each reservoir is 2,600 ml. The maximum recommended flow rate is 2 liters per minute (LPM).
Each reservoir has eight luer or 1/4" (0.6 cm) ID access ports;
- four suction access ports .
- two pre-defoamer and two post-defoamer access ports. One of the post-defoamer access . ports is intended for use as a vent.
{4}------------------------------------------------
Each Intersept™ Filtered Cardiotomy Reservoir contains an open cell polyurethane defoamer with a 20 micron microaggregate filter covered with a polyester sleeve.
## Indications for Use
The Intersept filtered cardiotomy reservoir is indicated for use in the cardiopulmonary bypass circuit during surgery for:
- an air/fluid separation chamber; .
- a temporary storage reservoir for priming solution and blood;
- the filtration of particulates from bank blood and the storage and filtration of blood ● recovered from the field by suction;
- . the addition of medications or other fluids.
The Intersept filtered cardiotomy reservoir is also indicated for use after open heart surgery for:
- the post-operative collection of autologous blood from the chest and the aseptic return of ● blood to the patient for volume replacement.
## Contraindications
This product used for any other purposes than for the indicated use is the responsibility of the user.
The cardiotomy reservoir is contraindicated for use in post-operative chest drainage/autotransfusion procedures when:
- . there is an air leak in the lung or gross perforations to the chest wall exist;
- pericardial, mediastinal, pulmonary or systemic infection or malignancy is present;
- gross contamination (e.g., bone cement, lymphatic failure, perforated intestine) is present ● or suspected:
- aspirating blood from a site containing topical hemostatic agents, bactericidal wound irritants or antibiotics not intended for parenteral administration;
- the chest is open and vacuum is applied;
- protamine has been administered prior to the reservoir being removed from the bypass circuit;
- the patient is returned to surgery for any reason;
- vented chest tubes not incorporating vent flow regulation such as stopcock are used.
# Comparison to Predicate Devices
A comparison of the Intersept™ Filtered Cardiotomy Reservoir to the predicate device (the Intersept® Cardiotomy Reservoir with Filter) indicates the following similarities:
- Intended Use: The intended use is the same as predicate device. ●
- Design: The overall design is the same as the predicate.
- Materials: The material types used are the same as the predicate.
- Principles of Operation and Technology: The principles of operation are the same as the predicate device.
- . Performance: The performance is substantially equivalent to the predicate and/or reference device.
{5}------------------------------------------------
#### Summary of Testing
Testing has demonstrated that the Intersept™ Filtered Cardiotomy Reservoir is substantially equivalent to the predicate.
The following tests were conducted:
| Component | Base Material Changes | Verification/Validation | Results |
|-----------|-------------------------|-------------------------|---------|
| Frame | Current: Polypropylene | Filtration Efficiency | Pass |
| Frame | | Blood Trauma | Pass |
| Frame | Proposed: Polypropylene | Biocompatibility | Pass |
## Conclusion
The data included in this submission is sufficient to demonstrate substantial equivalence of the Intersept™ Filtered Cardiotomy in terms of safety and effectiveness as compared to the predicate device, the Intersept® Cardiotomy Reservoir with Filter.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.