The Circumplast single use circumcision device is intended to be used to perform circumcisions on infant males. This device consists of two components: Circumplast Circumcision Device and Ligature. The Ligature was tested for use with Circumplast Circumcision Device only.
Device Story
Circumplast is a single-use, sterile circumcision device consisting of a transparent plastic conic tube and a linen ligature. Used by healthcare practitioners in a clinical setting. The inner ring is placed over the penis at the coronal sulcus; the foreskin is fixed over the ring; the ligature is tied tightly around the groove of the bell to perform circumferential crushing of the foreskin. Excess skin is trimmed. The device remains on the patient for 3-7 days before falling off. The device facilitates the circumcision procedure by providing a mechanical clamp and ligature system to achieve hemostasis and tissue separation, potentially simplifying the procedure and improving patient outcomes.
Clinical Evidence
Two clinical studies conducted in England and Australia demonstrated the safety and effectiveness of the device. No specific performance metrics (e.g., complication rates) were provided in the summary.
Technological Characteristics
Device consists of a plastic conic tube (bell) and a linen ligature. Available in four sizes (0.95, 1.1, 1.2, 1.3 cm). Operates via mechanical circumferential crushing. Provided sterile for single use. Complies with ISO 10993 for biocompatibility.
Indications for Use
Indicated for infant males undergoing circumcision.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 18, 2015
Novadien Healthcare % Diane Sudduth Sr. Regulatory Consultant Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701
Re: K151095
> Trade/Device Name: Circumplast Circumcision Device Regulation Number: 21 CFR 884.4530 Regulation Name: Obstetric-gynecologic specialized manual instrument Regulatory Class: II Product Code: HFX Dated: August 11, 2015 Received: August 19, 2015
Dear Diane Sudduth,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/8 description: The image shows the name "Herbert P. Lerner" in a bold, sans-serif font. Below the name, there is a "-S". In the background, there is a faint, stylized FDA logo, which is partially obscured by the text.
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
{3}------------------------------------------------
## 1. Submission Sponsor
Novadien Healthcare Postal Address: PO BOX 1318, Burwood NSW 1805, Australia Physical Address: 14/60 Belmore Street, Burwood NSW 2134, Australia Phone: +61 416 178 617 Fax: +61 297 067 888 Contact: Milad Melhem, Chief Executive Officer (CEO) Email: milad.melhem@novadien.com
# 2. Submission Correspondent
Emergo Group 816 Congress Avenue, Suite 1400 Austin, TX 78701 Cell Phone: (561) 305.5075 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Diane Sudduth, Senior Consultant, QA Email: project.management@emergogroup.com
#### 3. Date Prepared
September 18, 2015
#### 4. Device Identification
| Trade/Proprietary Name: | Circumplast circumcision device |
|----------------------------|-----------------------------------------------------|
| Common/Usual Name: | Circumcision clamp |
| Classification Name: | Obstetric-gynecologic specialized manual instrument |
| Classification Regulation: | 21 CFR 884.4530 |
| Product Code: | HFX |
| Device Class: | Class II |
| Classification Panel: | Obstetrics/Gynecology Panel |
#### 5. Legally Marketed Predicate Device(s)
SmartKlamp (Circumvent BV, K032091)
#### 6. Device Description
The Circumplast is a device composed of two components, a transparent plastic conic tube, and a linen ligature. The device is sold sterile for single use, packaged in a sterile pouch.
{4}------------------------------------------------
The Circumplast is available in four (4) sizes, to accommodate size of the infant's penis.
The Circumplast is designed to perform a circumferential crushing of the foreskin (prepuce) between the inner ring and outer ring. The inner ring is placed over the penis, at the level of the coronal sulcus. After local anesthesia, the healthcare Practitioner fixes the patient's foreskin over the inner ring and then the prepuce is tied all around, over the groove of the bell, with a tight ligature. Skin forward of the crush line is trimmed away.
After the procedure, the Circumplast device remains on the patient for 7 days. The Circumplast falls off after 3 - 7 days.
#### 7. Indication for Use
The Circumplast single use circumcision device is intended to be used to perform circumcisions on infant males. This device consists of two components: Circumplast Circumcision Device and Ligature. The Ligature was tested for use with Circumplast Circumcision Device only.
# 8. Substantial Equivalence Discussion
The following table compares the Circumplast to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| Manufacturer | Novadien Healthcare | Circumvent BV | Significant<br>Differences |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | New Device<br>Circumplast | Predicate<br>SmartKlamp | |
| 510(k) Number | Not assigned | K032091 | N/A |
| Product Code | HFX | HFX | N/A |
| Regulation Number | 884.4530 | 884.4530 | N/A |
| Regulation Name | Obstetric-gynecologic<br>specialized manual<br>instrument | Obstetric-gynecologic<br>specialized manual<br>instrument | N/A |
| Intend Use/<br>Indications for Use | The Circumplast® single use<br>circumcision device is<br>intended to be used to<br>perform circumcisions on<br>infant males. | Circumcision of newborns<br>and older males, defined<br>as circumferential<br>excision of the foreskin or<br>prepuce at or near the<br>level of coronal sulcus,<br>with minimal amount of<br>preputial skin remaining. | Similar; subject device is<br>indicated for use on<br>infant males while<br>predicate is for use on<br>both newborns and older<br>males. |
| Patient population | Infant males | Newborn and older males | Similar; subject device<br>patient population is a<br>subset to the predicate<br>device. |
| Mechanism of<br>Action | Circumferential crushing of<br>the foreskin (i.e. crush the<br>foreskin between the inner<br>tube and outer clamp) | Circumferential crushing<br>of the foreskin (i.e. crush<br>the foreskin between the<br>inner tube and outer<br>clamp) | None |
Table 5A - Comparison of Characteristics
{5}------------------------------------------------
| Manufacturer<br>Trade Name | Novadien Healthcare | Circumvent BV | Significant<br>Differences |
|-----------------------------------------------------|----------------------------------|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| | New Device<br>Circumplast | Predicate<br>SmartKlamp | |
| Sizes | 4 sizes (0.95, 1.1, 1.2, 1.3 cm) | 4 sizes (1.0,1.3,1.6,2.1 cm) | Similar; subject device<br>sizes are a subset of the<br>sizes available for the<br>predicate device. |
| Material(s) | Bell: plastic<br>Ligature: linen | Cylinder: transparent<br>polycarbonate<br>Outer clamp: nylon | Similar; both devices are<br>composed of plastic,<br>however, subject device<br>uses ligature for clamping<br>while predicate uses<br>nylon outer clamp. |
| Need for suture | No | No | None |
| Additional<br>instruments<br>needed | Clamp, scissors | Clamp, scissors | None |
| Device remains on<br>the patient after<br>procedure | Yes, falls off in 3 - 7 days | Yes, removed after 5 days | Similar; devices remain on<br>the patient within a<br>variance of +/- 2 days. |
| Single Use | Yes | Yes | None |
| Provided Sterile | Yes | Yes | None |
| Complies with ISO<br>10993 | Yes | Yes | None |
## 9. Non-Clinical Performance Data
As part of demonstrating safety and effectiveness of Circumplast and in showing substantial equivalence to the predicate devices that are subject to this 510(k) submission, Novadien completed a number of tests.
The following testing has been performed to support substantial equivalence:
- . Compression Testing - Pass
- Tensile Strength - Pass
- Torsion Testing - Pass
- Sterilization Testing Pass ●
Finished, sterilized devices were tested for biocompatibility according to ISO 10993-1 for a surface device in prolonged contact with intact skin. The results confirm that the Circumplast device and ligature are biocompatible for its intended use.
The Circumplast device meets all the requirements for overall design, sterilization and biocompatibility confirms that the output meets the design inputs and specifications. The Circumplast passed all testing stated above as shown by the acceptable results obtained.
{6}------------------------------------------------
# 10. Clinical Performance Data
Two clinical studies conducted in England and Australia, respectively, demonstrated that the subject device is safe and effective.
# 11. Statement of Substantial Equivalence
The differences between the Circumplast and the predicate devices cited do not raise any different questions regarding safety and effectiveness. The differences in the technological characteristics are minimal, and the associated procedures are nearly identical. The indications for use are identical to the indications of one of the previously cleared predicate devices. The device, as designed and manufactured, is as safe and effective as the predicate devices, and the device is substantially equivalent to the referenced predicate devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.