# BR-ABVS Viewer 1.0 (K151075)

_Taihao Medical, Inc. · LLZ · Jan 15, 2016 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151075

## Device Facts

- **Applicant:** Taihao Medical, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jan 15, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

BR-ABVS Viewer 1.0 is intended as a standalone software device installed on a standalone windows-based computer to assist the physician to visualize any orientation of three-dimensional (3-D) breast ultrasound images generated by Siemens ACUSON S2000 Automated Breast Volume Scanner, ABVS (cleared in K081148). The software device is indicated for use to assist the physicians in their review and analysis of the 3-D breast ultrasound images generated by ABVS. Caution: Federal law restricts this device to sale by or on the order of a physician.

## Device Story

BR-ABVS Viewer 1.0 is a standalone software application for Windows-based computers; it processes 3-D breast ultrasound image volumes (DICOM format) acquired from the Siemens ACUSON S2000 ABVS. The device enables physicians to visualize images in any orientation (axial, sagittal, coronal) based on anatomical coordinate systems and DICOM tag spacing. It includes functionality to annotate markers and automatically generate reports containing sub-image and location information. Used in clinical settings by physicians, the device serves as an adjunct to mammography, aiding in the review and analysis of breast ultrasound data to support clinical decision-making.

## Clinical Evidence

Bench testing only. No clinical trials performed. Validation consisted of a performance study comparing the subject device to the predicate (ABVS Workplace) using clinical 3-D breast ultrasound images acquired from a Siemens ACUSON S2000 ABVS. The study validated 3-D image loading, visualization across axial, sagittal, and coronal planes, and adherence to anatomical coordinate systems defined by DICOM tags.

## Technological Characteristics

Standalone software for Windows-based PCs. Processes DICOM 3-D ultrasound volumes. Features include multi-planar reconstruction (axial, sagittal, coronal), marker annotation, and automated report generation. No specific hardware requirements beyond standard PC/review station. Software development follows documented design, verification, and validation processes with risk assessment controls.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- ABVS Workplace ([K092067](/device/K092067.md))

## Reference Devices

- Siemens ACUSON S2000 Automated Breast Volume Scanner ([K081148](/device/K081148.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the logo is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 15, 2016

TaiHao Medical, Inc. % Chiu S. Lin, Ph.D. President LIN & ASSOCIATES, LLC 9223 Cambridge Manor Court POTOMAC MD 20854

Re: K151075

Trade/Device Name: BR-ABVS Viewer 1.0 Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 15, 2015 Received: December 15, 2015

Dear Dr. Lin:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

For

Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151075

Device Name BR-ABVS Viewer 1.0

#### Indications for Use (Describe)

BR-ABVS Viewer 1.0 is intended as a standalone software device installed on a standalone windows-based computer to assist the physician to visualize any orientation of three-dimensional (3-D) breast ultrasound images generated by Siemens ACUSON S2000 Automated Breast Volume Scanner, ABVS (cleared in K081148). The software device is indicated for use to assist the physicians in their review and analysis of the 3-D breast ultrasound images generated by ABVS. Caution: Federal law restricts this device to sale by or on the order of a physician.

Type of Use (Select one or both, as applicable)

| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 5. 510(k) Summary of Safety and Effectiveness Information

This 510(k) summary of safety and effectiveness information is submitted as part of the Premarket Notification in compliance with requirements of CFR Part 807, Subpart E and Section 807.92

## 5.1. Identification of Submitter:

Submitter: TaiHao Medical Inc.

Address: 7F., No.410, Sec. 5, Zhongxiao E. Rd., Xinyi Dist., Taipei City 110, Taiwan (R.O.C.)

Phone: 886-2-2858-2357

Contact: HSIN HUNG (Simon) LAI

Title: President

Phone: 886-2-2858-2357

Email: simonlai@taihaomed.com

Manufacturer: TaiHao Medical Inc.

US Agent and Contact: Chiu S. Lin, Ph.D.

LIN & ASSOCIATES, LLC

Address: 9223 Cambridge Manor Court

Potomac, MD 20854 U.S.A.

Phone: (0) 301-591-3895

Email: cslin@lin-associates.com

Date prepared: April 8, 2015

## 5.2. Identification of Product

Device Trade Name: BR-ABVS Viewer 1.0

Common and Usual Name: Ultrasound Image Display Device

Device Classification Name: Picture Archiving and Communication System

Regulation Number: 21 CFR 892.2050

{4}------------------------------------------------

### Classification Product Code: LLZ

Classification: Class II Radiology Devices Classification Panel:

Manufacturer: TaiHao Medical Inc.

### 5.3. Predicate Device

This subject software medical device is substantially equivalent to the devices listed below:

Model: ABVS Workplace

Manufacturer: Siemens Medical Solutions

510(k) Number: K092067, cleared on September 18, 2009

### 5.4. Device Description

BR-ABVS Viewer 1.0 is intended as a standalone software device installed on a standalone windows-based computer to assist the physician to visualize any orientation of three-dimensional (3-D) breast ultrasound images generated by Siemens ACUSON S2000 Automated Breast Volume Scanner, ABVS (cleared in K081148). The software also automatically generates reports to provide the sub-image and location information of markers annotated during the image review.

#### 5.5. Indications for Use

BR-ABVS Viewer 1.0 is intended as a standalone software device installed on a standalone windows-based computer to assist the physician to visualize any orientation of three-dimensional (3-D) breast ultrasound images generated by Siemens ACUSON S2000 Automated Breast Volume Scanner, ABVS (cleared in K081148). The device is indicated for use to assist the physicians in their review and analysis of the 3-D breast ultrasound images generated by ABVS.

#### 5.6. Comparison with Predicate Devices

BR-ABVS Viewer 1.0 is substantially equivalent to ABVS Workplace with a general intended use for viewing and analyzing ultrasound image data to physicians. Minor technological characteristics differences do not raise any new questions of safety and effectiveness.

The comparison table between our device and the predicate devices is provided below:

|  |  | <b>BR-ABVS Viewer 1.0</b> | <b>ABVS Workplace</b> |
|--|--|---------------------------|-----------------------|
|--|--|---------------------------|-----------------------|

{5}------------------------------------------------

| Manufacturer                                       | TaiHao Medical Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Siemens Medical Solutions                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
|----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                                      | K151075                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | K092067                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Device Common Name                                 | Picture archiving and<br>communications system                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Picture archiving and<br>communications system                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Regulation Number                                  | 21 CFR 892.2050                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | 21 CFR 892.2050                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Regulation Name                                    | Picture archiving and<br>communications system                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Picture archiving and<br>communications system                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Product Code                                       | LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Indications for Use                                | BR-ABVS Viewer 1.0 is<br>intended as a standalone<br>software device installed on a<br>standalone windows-based<br>computer to assist the<br>physician to visualize any<br>orientation of three-<br>dimensional (3-D) breast<br>ultrasound images generated<br>by Siemens ACUSON S2000<br>Automated Breast Volume<br>Scanner, ABVS (cleared in<br>K081148). The device is<br>indicated for use to assist the<br>physicians in their review and<br>analysis of the 3-D breast<br>ultrasound images generated<br>by ABVS. | ABVS Workplace is intended<br>to display ultrasound images<br>of the breast acquired from B-<br>mode imaging using an<br>automatic or handheld<br>scanning linear transducer.<br>The images may be reviewed<br>and analyzed by the<br>physician. The ABVS<br>Workplace is indicated for use<br>as an adjunct to<br>mammography. The ABVS<br>Workplace is not intended to<br>be used as a replacement for<br>screening.                                                         |
| Functional Capability of<br>Image Processing       | BR-ABVS Viewer 1.0<br>provides visualization of any<br>orientation of 3-D image.                                                                                                                                                                                                                                                                                                                                                                                                                                        | The ABVS Workplace applies<br>post-processing algorithms<br>based on the nipple location.<br>A reverberation removal<br>algorithm determines tissue<br>contact areas. This suppresses<br>reverberation artifacts from<br>the non-contact area. A<br>proprietary adaptive nipple<br>shadow reduction tool<br>analyzes the data volume and<br>enhances structures in the<br>retroareolar area to improve<br>visualization of this typically<br>challenging anatomical<br>region. |
| Software Design                                    | Image processing techniques                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Platform                                           | Window-based                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Operating System                                   | Standard PC or review station                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Own workstation, which<br>includes IT hardware and pre-<br>installed software                                                                                                                                                                                                                                                                                                                                                                                                  |
| Clinical Application                               | As an adjunct to<br>mammography screening                                                                                                                                                                                                                                                                                                                                                                                                                                                                               | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Image Type to Be Processed<br>by The Device        | Automated breast ultrasound<br>images generated by Siemens<br>ACUSON S2000 Automated<br>Breast Volume Scanner,<br>ABVS (cleared in K081148).                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Image Format                                       | DICOM images acquired on<br>Siemens ACUSON S2000<br>Automated Breast Volume                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
|                                                    | Scanner, ABVS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Automatically Generating<br>Report                 | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Performance Testing to<br>Support SE Determination | Results from Software<br>Validation Report of BR-<br>ABVS Viewer 1.0                                                                                                                                                                                                                                                                                                                                                                                                                                                    | From the 510(k) Summary<br>that is available on the FDA<br>database, it appears that no<br>data from performance testing<br>were submitted.                                                                                                                                                                                                                                                                                                                                    |

{6}------------------------------------------------

{7}------------------------------------------------

#### 5.7. Performance Standards

No applicable FDA performance standards have been issued.

#### 5.8. Software

Software development for BR-ABVS Viewer 1.0 follows documented processes for software design, verification and validation testing. A risk assessment has been completed to identify potential design hazards. Appropriate steps have been taken to control all identified risks for this type of image viewing and quantification device.

#### 5.9. Summary of Performance Data to Support Substantial Equivalence

BR-ABVS Viewer 1.0 displays the 3-D image volume by axial, sagittal, and coronal plane according to the anatomical coordinate system to provide an overall observation. The actual image size is obtained by considering the spacings of three axes specified in the standard DICOM tags. TaiHao Medical Inc. has conducted a performance study to validate and assess the performance of BR-ABVS Viewer 1.0 for its-intended use. An actual clinical image generated by a Siemens ACUSON S2000 Automated Breast Volume Scanner in 2014 was used to do the comparison testing between BR-ABVS Viewer 1.0 and the predicate (ABVS Workplace, K092067) in terms of substantial equivalence in 3-D image loading.

#### 5.10. Conclusions

The intended use, technological characteristics, and major functionality of BR-ABVS Viewer 1.0 are similar to the predicate device and no new issues of safety or effectiveness are introduced by using this device. The performance data generated, as described, demonstrates that our software device is as

{8}------------------------------------------------

safe and effective, as compared to the predicate. Therefore we believe BR-ABVS Viewer 1.0 is Substantially Equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K151075](https://fda-staging.innolitics.com/device/K151075)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).