The Magnetic Resonance Imaging System produces transversal, sagittal and oblique cross-section images of the limbs and joints. It's intended for imaging portions of the arm, including the hand, wrist, forearm, elbow, and imaging portion of the leg, including the foot, ankle, calf and knee, but excluding the thigh. Magspin's magnetic resonance imaging system is a fixed system, it produces images correspond to the spatial distribution of protons (hydrogen nuclei) that embodies magnetic resonance properties of human body and is weighted by the MR parameters such as spinlattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density and "chemical shif". When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.
Device Story
Fixed MRI system for limb/joint imaging; utilizes permanent magnet, magnetic gradient system, RF coils, spectrometer, and RF transmitter/receiver. Inputs: proton spatial distribution (hydrogen nuclei) within human body; transforms inputs via MR parameters (T1, T2, nuclei density, chemical shift) into cross-sectional images. Operated by trained medical experts in clinical settings. Output: transversal, sagittal, coronal, and oblique images displayed on computer. Healthcare providers interpret images to obtain diagnostically useful information for clinical decision-making. Benefits: non-invasive visualization of internal limb/joint structures.
Clinical Evidence
No clinical data. Bench testing only. Compliance demonstrated via IEC 60601-2-33 (safety/performance), IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and NEMA MS series (MS-1, MS-2, MS-3, MS-4, MS-5, MS-8, MS-12) for image quality metrics including SNR, slice thickness, acoustic noise, geometric distortion, uniformity, and SAR characterization.
Indicated for imaging limbs and joints, specifically arm (hand, wrist, forearm, elbow) and leg (foot, ankle, calf, knee), excluding thigh. For use by medical experts trained in MR equipment.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
ADVANCED IMAGING LABORATORY, MRI SYSTEM, MODEL MAGFINDER II / AI 3200 (K072850)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 22, 2015
Jiangsu Magspin Instrument Co., Ltd. % Mr. Mike Gu Regulatory Affairs Manager Guangzhou Osmunda Medical Device Technology Service Co. Ltd. 7th Floor, Jingui Business Building, 982 Congyun Road, Baiyun District Guangzhou. Guangdong 510420 CHINA
Re: K151074
Trade/Device Name: Magnetic Reasonance Imaging System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: November 20, 2015 Received: November 25, 2015
Dear Mr. Gu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Oolo
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
# 510(k) Number (if known)
Device Name Magnetic Resonance Imaging System
### Indications for Use (Describe)
The Magnetic Resonance Imaging System produces transversal, sagittal and oblique cross-section images of the limbs and joints. It's intended for imaging portions of the arm, including the hand, wrist, forearm, elbow, and imaging portion of the leg, including the foot, ankle, calf and knee, but excluding the thigh.
Magspin's magnetic resonance imaging system is a fixed system, it produces images correspond to the spatial distribution of protons (hydrogen nuclei) that embodies magnetic resonance properties of human body and is weighted by the MR parameters such as spinlattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density and "chemical shif". When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MAGSPIN. The logo is in blue and red. The word "MAGSPIN" is written in blue, with a red swoosh underneath the letters "MA".
### 510(k) Summary
In accordance with 21 CFR 807.92 the following summary of information is provided: SUBMITTER l.
Jiangsu Magspin Instrument Co., Ltd.
Building 9, No. 299, Yuyang Road, Kunshan, Jiangsu, China
Phone: (+86) 512-36690072
Fax: (+86) 512-55110481
Primary Contact Person: Mike Gu
Regulatory Affairs Manager
OSMUNDA Medical Device Consulting Co., Ltd
Tel: (+86) 20-6232 1333
Fax: (+86) 20-8633 0253
Secondary Contact Person: Leo Li
Quality Manager
ll
LNH
Jiangsu Magspin Instrument Co., Ltd
Date Prepared:
II. DEVICE
Name of Device: Magnetic Resonance Imaging System
Common/Usual Name: System, Nuclear Magnetic Resonance Imaging
March 27, 2015
Classification Names: Magnetic resonance diagnostic device (21 CFR 892.1000)
Regulation Class:
- Product Code:
- == PREDICATE DEVICE
{5}------------------------------------------------
# MAGSPIN
Esaote S.p. A.'s O-Scan, K092469
ADVANCED IMAGING LABORATORY, MRI SYSTEM, MODEL MAGFINDER II / AI 3200, K072850
This predicate has not been subject to a design-related recall.
#### IV. DEVICE DESCRIPTION
Magnetic Resonance Imaging System produces images of the internal structures of the patient's limbs and joints.
The MRI system consists of the following components:
- Permanent Magnetic
- Magnetic Gradient system
- Dedicated RF coils
- Spectrometer
- Radiofrequency Transmitter
- Receiver coil
- Computer equipped with imaging software
- Patient chair
#### V. INDICATIONS FOR USE
The Magnetic Resonance Imaging System produces transversal, sagittal and coronal and oblique cross-section images of the limbs and joints. It's intended for imaging portions of the arm, including the hand, wrist, forearm, elbow, and imaging of the shoulder, and imaging portion of the leg, including the foot, ankle, calf and knee, but excluding the thigh.
Magspin's magnetic resonance imaging system is a fixed system, it produces images correspond to the spatial distribution of protons (hydrogen nuclei) that embodies magnetic resonance properties of human body and is weighted by the MR parameters such as spinlattice relaxation time (T1), spin-spin relaxation time (T2), nuclei density, flow velocity and "chemical shift". When interpreted by a medical expert trained in the use of MR equipment, the images can provide diagnostically useful information.
{6}------------------------------------------------
MAGSPIN
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The proposed Magnetic Resonance Imaging System employs the same technology as its primary predicate devices K092469 and reference device K072850.
| Item | The Proposed | Primary Predicate K092469 | Reference K072850 |
|------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| Operating<br>Modes<br>IEC 60601-2-33<br>(2010-03) | Normal operating mode | Normal operating mode | Normal operating mode |
| Safety<br>Parameter<br>Display | SAR dB/dt | SAR dB/dt | SAR dB/dt |
| Max SAR | Normal operating mode specified in IEC 60601-2-33 (2010) | Normal operating mode specified in IEC 60601-2-33 (2010) | Normal operating mode specified in IEC 60601-2-33 (2010) |
| Max dB/dt | <Normal operating mode specified in IEC 60601-2-33 (2010) | <Normal operating mode specified in IEC 60601-2-33 (2010) | <Normal operating mode specified in IEC 60601-2-33 (2010) |
| Potential<br>emergency<br>condition and<br>means provided<br>for<br>shutdown | Shut down by Emergency Ramp Down Unit to prevent abnormal movement of patient chair | Shut down by Emergency Ramp Down Unit to prevent abnormal movement of patient chair | Shut down by Emergency Ramp Down Unit to prevent abnormal movement of patient chair |
#### VII. NON CLINICAL TESTING PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
### Bench testing:
Bench testing was conducted to demonstrate the Magnetic Resonance Imaging System meets all performance standards as follows:
- . IEC 60601-2-33 Edition 3.0 2010-03 Medical electrical equipment - Part 2-33: Particular requirements for the basic safety and essential performance of magnetic resonance equipment for medical diagnostic
- . NEMA MS-1-2008 Determination of Signal-to-Noise Ratio (SNR) in Diagnostic
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the word "MAGSPIN" in a sans-serif font. The letters are a light blue color. There is a red curved line underneath the letters "MAG".
Magnetic Resonance Imaging
- . NEMA MS 5-2010 Determination of Slice Thickness in Diagnostic Magnetic Resonance Imaging
- . NEMA MS 4-2010 Acoustic Noise Measurement Procedure for Diagnosing Magnetic Resonance Imaging Devices
- . NEMA MS 12-2010 Quantification and Mapping of Geometric Distortion for Special Applications
- NEMA MS 3-2008 Determination of Image Uniformity in Diagnostic Magnetic Resonance Images
- . NEMA MS 8-2008 Characterization of the Specific Absorption Rate for Magnetic Resonance Imaging Systems
- . NEMA MS 2-2008 Determination of Two-Dimensional Geometric Distortion in Diagnostic Magnetic Resonance Images
### Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the Magnetic Resonance Imaging System. The device complies with the IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC.
### Animal and clinical study
The subject of this premarket submission, Magnetic Resonance Imaging System, does not require animal or clinical studies to support substantial equivalence.
#### VIII. CONCLUSIONS
The non-clinical data support the safety of the device should perform as intended in the specified use conditions. Jiangsu Magspin Instrument Co., Ltd. considers the Magnetic Resonance Imaging System substantially equivalent to the predicate device and does not raise any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.