← Product Code [NUH](/productcode/NUH) · K151034

# PainKARE (K151034)

_Triowave Technologies · NUH · Nov 21, 2015 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151034

## Device Facts

- **Applicant:** Triowave Technologies
- **Product Code:** [NUH](/productcode/NUH.md)
- **Decision Date:** Nov 21, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.5890
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

The PainKARE is indicatied for Temporary relief of pain of sore and aching muscles, joints, and tissues due to strain from exercise or normal household and work activities.

## Device Story

PainKARE is an OTC transcutaneous electrical nerve stimulator (TENS) for temporary pain relief. The device consists of two stimulator pads connected by a USB cable. It operates via a microprocessor-controlled system that delivers biphasic electrical waveforms to the body. The device is powered by a rechargeable battery. Users operate the device to provide electrotherapy, with adjustable treatment times ranging from 30 minutes to 4 hours. The device includes safety features such as automatic overload protection, automatic no-load alarm, and automatic shut-off. It is intended for use by individuals to manage pain from muscle or joint strain. The healthcare provider or user monitors the device status via an operational indicator. The device provides a non-pharmacological method for pain management.

## Clinical Evidence

Bench testing only. The device was tested for electrical safety and electromagnetic compatibility per AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-6, AAMI HA60601-1-11, and IEC 60601-2-10. No clinical data was provided.

## Technological Characteristics

Microprocessor-controlled TENS device; ABS plastic enclosure; rechargeable battery power source; biphasic waveform output; 50μA regulated current; 0.3Hz frequency; Type BF patient leakage current classification; compliant with AAMI ES 60601-1, IEC 60601-1-2, IEC 60601-1-6, AAMI HA60601-1-11, and IEC 60601-2-10.

## Regulatory Identification

A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.

## Predicate Devices

- Painmaster MCT Patch ([K090042](/device/K090042.md))
- Painmaster MCT Patch ([K130114](/device/K130114.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized human face in profile, with three overlapping heads. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 21, 2015

TrioWave Technologies Attn: Ms. Parul Chansoria Regulatory Consultant for TrioWave Technologies 3984 Washington Blvd. #166 Fremont, CA 94538

> Re: K151034 Trade Name: PainKARE Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NUH Dated: October 20, 2015 Received: October 23, 2015

Dear Ms. Chansoria:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code 27f Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Carlos L. Pena -S

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K151034

Device Name PainKARE

Indications for Use (Describe)

The PainKARE is indicatied for Temporary relief of pain of sore and aching muscles, joints, and tissues due to strain from exercise or normal household and work activities.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 5. 510 (K) Summary

This 510(k) Summary has been created per the requirements of the Safe Medical Device Act (SMDA) of 1990, and the content is provided in conformance with 21 CFR Part 807.92.

#### 5.1 Submitter's Information

|                                 | Table 1-Submitter's Information                                |
|---------------------------------|----------------------------------------------------------------|
| Submitter's Name:               | Huiyou Zhu, Founder                                            |
| Company:                        | TrioWave Technologies.                                         |
| Address:                        | 3984 Washington Blvd., #166<br>Fremont, CA 94538               |
| Contact Person:                 | Parul Chansoria,<br>Elexes; Elexes Medical Consulting Pvt Ltd. |
| Phone:                          | 650-528-2445                                                   |
| Fax:                            | --------                                                       |
| Email:                          | parul@elexes.com                                               |
| Date of Summary<br>Preparation: | October 20, 2015                                               |

### 5.2 Device Information

| Table 2-Device Information |                                                     |
|----------------------------|-----------------------------------------------------|
| Trade Name:                | PainKARE                                            |
| Common Name:               | Transcutaneous electrical nerve<br>Stimulator       |
| Classification Name:       | Stimulator, nerve, Transcutaneous, Over-the-counter |
| Classification Number:     | Class II per 21 CFR 882.5890(b)                     |
| Product Code:              | NUH                                                 |
| Classification Panel:      | Neurology                                           |

# 5.3 Predicate Device Information

Painmaster MCT Patch (K090042; K130114)

{4}------------------------------------------------

#### 5.4 Device Description

PainKARE is an OTC medical device intended to temporarily relieve pain due to strain from exercise or normal household work or activities via electro-therapy. The Device consists two Stimulator pads connected by a USB cable for Electrotherapy.

PainKARE's Stimulator pads operate in a programmable micro-current mode and deliver Biphasic waveforms which in turn provide electrical stimulation to the body that aids in relieving pain.

#### 5.5 Indications for Use

The PainKARE system is indicated for "temporary relief of pain of sore and aching muscles, joints, and tissues due to strain from exercise or normal household and work activities".

#### 5.6 Technological Characteristics

The PainKARE (Subject Device) makes use of two Predicate Devices; Painmaster MCT Patch (K090042; K130114) for the Stimulator pads (for electrotherapy).

| Table 3-PainKARE (Stimulator pads) vs. Painmaster MCT Patch (K090042; K130114) |                                                                                                                                                                              |                                                                                                                                                                                                                   |                                                                                                                                                                           |                                                               |
|--------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------|
| Parameter                                                                      | Painmaster MCT<br>Patch<br>(Predicate Device)<br>K090042                                                                                                                     | Painmaster MCT Patch<br>(Predicate Device)<br>K130114                                                                                                                                                             | Proposed Device<br>(Subject Device)<br>PainKARE<br>Stimulator                                                                                                             |                                                               |
| Manufacturer                                                                   | Newmark Inc.                                                                                                                                                                 | Newmark Inc.                                                                                                                                                                                                      | TrioWave<br>Technologies                                                                                                                                                  |                                                               |
| Indications for use                                                            | Temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the lower<br>back due to strain<br>from exercise or<br>normal household<br>and work activities. | Temporary relief of<br>pain associated with<br>sore and aching<br>muscles in the upper<br>and lower extremities<br>(arm and/or leg) due<br>to strain from exercise<br>or normal household<br>and work activities. | Temporary relief of<br>pain of sore and<br>aching muscles,<br>joints, and tissues<br>due to due to strain<br>from exercise or<br>normal household<br>and work activities. |                                                               |
| Operation Mode                                                                 | Waveform: pulsed<br>square wave.<br>Amplitude: 50μA<br>Frequency: 0.5Hz<br>Polarity: Mono-phasic                                                                             | Waveform: pulsed<br>square wave.<br>Amplitude: 50μA<br>Frequency: 0.5Hz<br>Polarity: Mono-phasic                                                                                                                  | Waveform: pulsed<br>square wave.<br>Amplitude: 50μA<br>Frequency: 0.3Hz<br>Polarity: Biphasic<br>(Alternating Mono-<br>phasic)                                            |                                                               |
| Power Source(s)†                                                               | Primary battery only.<br>One time use                                                                                                                                        | Primary battery only.<br>One time use                                                                                                                                                                             | Primary Battery only.<br>Rechargeable                                                                                                                                     |                                                               |
| Micro-processor<br>Control                                                     | Yes                                                                                                                                                                          | Yes                                                                                                                                                                                                               | Yes                                                                                                                                                                       |                                                               |
| Table 3-PainKARE (Stimulator pads) vs. Painmaster MCT Patch (K090042; K130114) |                                                                                                                                                                              |                                                                                                                                                                                                                   |                                                                                                                                                                           |                                                               |
| Parameter                                                                      | Painmaster MCT<br>Patch<br>(Predicate Device)<br>K090042                                                                                                                     | Painmaster MCT Patch<br>(Predicate Device)<br>K130114                                                                                                                                                             | Proposed Device<br>(Subject Device)<br>PainKARE<br>Stimulator                                                                                                             |                                                               |
| Automatic<br>Overload<br>Protection                                            | Yes                                                                                                                                                                          | Yes                                                                                                                                                                                                               | Yes                                                                                                                                                                       |                                                               |
| Automatic No Load<br>Alarm                                                     | Yes                                                                                                                                                                          | Yes                                                                                                                                                                                                               | Yes                                                                                                                                                                       |                                                               |
| User Override<br>Control                                                       | No                                                                                                                                                                           | No                                                                                                                                                                                                                | Yes                                                                                                                                                                       |                                                               |
| Operational<br>Indicator                                                       | Yes                                                                                                                                                                          | Yes                                                                                                                                                                                                               | Yes                                                                                                                                                                       |                                                               |
| Time Range                                                                     | Not adjustable. Up to<br>250 hours                                                                                                                                           | Not adjustable. Up to<br>250 hours                                                                                                                                                                                | 30 minutes to 4<br>hours, depending on<br>pain                                                                                                                            |                                                               |
| Housing Materials<br>and Construction                                          | Plastic (ABS)<br>enclosure                                                                                                                                                   | Plastic (ABS) enclosure                                                                                                                                                                                           | Plastic (ABS)<br>enclosure                                                                                                                                                |                                                               |
| Body contact<br>material                                                       | Electrode, CE certified                                                                                                                                                      | Electrode, CE certified                                                                                                                                                                                           | Electrodes, FDA<br>approved                                                                                                                                               |                                                               |
| Maximum Average<br>current                                                     | 50 μA                                                                                                                                                                        | 50 μA                                                                                                                                                                                                             | 50 μA                                                                                                                                                                     |                                                               |
| OTC use                                                                        | Yes                                                                                                                                                                          | Yes                                                                                                                                                                                                               | Yes                                                                                                                                                                       |                                                               |
| Dimensions (in.) [W<br>x H x D]                                                | 1.55 x 0.29x1                                                                                                                                                                | 1.55 x 0.29x1                                                                                                                                                                                                     | 1.2" x0.6" x1.6"                                                                                                                                                          |                                                               |
| Weight                                                                         | 0.03 lb.                                                                                                                                                                     | 0.03 lb.                                                                                                                                                                                                          | 0.1 lb.                                                                                                                                                                   |                                                               |
| Operating<br>temperature and<br>humidity                                       | 10-45°C, 20%-90%                                                                                                                                                             | 10-45°C, 20%-90%                                                                                                                                                                                                  | 10-45°C, 20%-90%                                                                                                                                                          |                                                               |
| Method of Line<br>-<br>Current Isolation                                       | N/A                                                                                                                                                                          | N/A                                                                                                                                                                                                               | N/A                                                                                                                                                                       |                                                               |
| Patient Leakage<br>-<br>Current++                                              | Type CF                                                                                                                                                                      | Type CF                                                                                                                                                                                                           | Type BF                                                                                                                                                                   |                                                               |
| - Normal<br>Condition (μΑ)                                                     | Unspecified                                                                                                                                                                  | Unspecified                                                                                                                                                                                                       | 0                                                                                                                                                                         |                                                               |
| - Single Fault<br>Condition (μΑ)                                               | Unspecified                                                                                                                                                                  | Unspecified                                                                                                                                                                                                       | 50μA                                                                                                                                                                      |                                                               |
| Average DC current<br>through electrodes<br>when device is on                  | 0                                                                                                                                                                            | 0                                                                                                                                                                                                                 | N/A (Pulse always<br>applied when on)                                                                                                                                     |                                                               |
| Table 3-PainKARE (Stimulator pads) vs. Painmaster MCT Patch (K090042; K130114) |                                                                                                                                                                              |                                                                                                                                                                                                                   |                                                                                                                                                                           |                                                               |
| Parameter                                                                      |                                                                                                                                                                              | Painmaster MCT<br>Patch<br>(Predicate Device)<br>K090042                                                                                                                                                          | Painmaster MCT Patch<br>(Predicate Device)<br>K130114                                                                                                                     | Proposed Device<br>(Subject Device)<br>PainKARE<br>Stimulator |
| but no pulses are<br>being applied (µA)                                        |                                                                                                                                                                              |                                                                                                                                                                                                                   |                                                                                                                                                                           |                                                               |
| Number of output<br>modes                                                      |                                                                                                                                                                              | 1                                                                                                                                                                                                                 | 1                                                                                                                                                                         | 2                                                             |
| Number<br>of Output<br>Channels<br>††††:                                       | Synchro<br>nous or<br>Alternati<br>ng?                                                                                                                                       | 1 channel,<br>monophasic                                                                                                                                                                                          | 1 channel, monophasic                                                                                                                                                     | 1 channel, biphasic<br>(alternating<br>monophasic)            |
|                                                                                | Method<br>of<br>Channel<br>Isolation                                                                                                                                         | NA                                                                                                                                                                                                                | NA                                                                                                                                                                        | NA                                                            |
| Regulated Current<br>or Regulated<br>Voltage?                                  |                                                                                                                                                                              | Regulated Current                                                                                                                                                                                                 | Regulated Current                                                                                                                                                         | Regulated Current                                             |
| Software/Firmware<br>/Microprocessor<br>Control?                               |                                                                                                                                                                              | Yes                                                                                                                                                                                                               | Yes                                                                                                                                                                       | Yes                                                           |
| Automatic<br>Overload<br>Trip?                                                 |                                                                                                                                                                              | No                                                                                                                                                                                                                | No                                                                                                                                                                        | Yes                                                           |
| Automatic No-Load<br>Trip?                                                     |                                                                                                                                                                              | Yes                                                                                                                                                                                                               | Yes                                                                                                                                                                       | Yes                                                           |
| Automatic Shut<br>Off?                                                         |                                                                                                                                                                              | Yes                                                                                                                                                                                                               | Yes                                                                                                                                                                       | Yes                                                           |
| Indicator<br>Display:                                                          | On/Off<br>Status?                                                                                                                                                            | Yes                                                                                                                                                                                                               | Yes                                                                                                                                                                       | Yes                                                           |
|                                                                                | Low<br>Battery?                                                                                                                                                              | No                                                                                                                                                                                                                | No                                                                                                                                                                        | Yes                                                           |
|                                                                                | Voltage/<br>Current<br>Level?                                                                                                                                                | No                                                                                                                                                                                                                | No                                                                                                                                                                        | No                                                            |
| Table 3-PainKARE (Stimulator pads) vs. Painmaster MCT Patch (K090042; K130114) |                                                                                                                                                                              |                                                                                                                                                                                                                   |                                                                                                                                                                           |                                                               |
| Parameter                                                                      | Painmaster MCT<br>Patch<br>(Predicate Device)<br>K090042                                                                                                                     | Painmaster MCT Patch<br>(Predicate Device)<br>K130114                                                                                                                                                             | Proposed Device<br>(Subject Device)<br>PainKARE<br>Stimulator                                                                                                             |                                                               |
| Compliance with<br>Voluntary<br>Standards?                                     | Yes, UL260,<br>EN60601-2 Part1, A1<br>& A2<br>EN 60601-2-10, Part<br>2-10<br>EN 60601-1-2 EMI                                                                                | Yes, UL260,<br>EN60601-2 Part1, A1 & A2<br>EN 60601-2-10, Part 2-<br>10<br>EN 60601-1-2 EMI                                                                                                                       | Yes, AAMI ES 60601-<br>1,<br>AAMI HA60601-1-<br>11,<br>IEC 60601-1-2<br>IEC 60601-1-6, IEC<br>60601-2-10,                                                                 |                                                               |
| Compliance with 21<br>CFR 8987?                                                | N/A                                                                                                                                                                          | N/A                                                                                                                                                                                                               | N/A                                                                                                                                                                       |                                                               |

#### 5.6.1 PainKARE (Stimulator pads) vs. Painmaster MCT Patch (K090042; K130114)

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

#### Similarities:

- . The Subject and the Predicate devices have the same intended use and indications for use.
- Both the Subject and Predicate devices have similar technological characteristics, namely, number of output channel and output current (50μA), and have the same operating temperature and humidity ranges.
- Both devices are microprocessor controlled, and utilize certified/cleared electrode pads.
- The Subject and the Predicate Devices also have same features like the automatic overload trip protection, automatic no load trip, automatic shut off, operational Indicator and no current display.
- . The subject device has rechargeable battery whereas the Predicates do not have a rechargeable battery.

#### Differences:

- . The Subject device operates in biphasic waveforms/modes, whereas the Predicates only operates in monophasic mode/waveform.
- There are some minor technological differences namely user override control etc.

#### 5.7 Performance Data

The following Performance testing has been performed on the Subject Device in accordance with appropriate FDA guidance documents and relevant standards, to support the determination of substantial equivalence. The test data showed that the Subject device is safe and effective for its intended use as compared to the Predicate.

- . The Subject Device was tested to conform to the electrical and safety requirements established in AAMI ES 60601-1 and the electromagnetic compatibility requirements in IEC 60601-1-2.
- The Subject Device was also tested to conform to IEC 60601-1-6, and AAMI HA60601-1-11 and IEC 60601-2-10.

{8}------------------------------------------------

## 5.8 Conclusion

Based on the comparison of intended use and key technological characteristics, the PainKARE (Subject Device) is substantially equivalent to the Predicate Device.

---

**Source:** [https://fda-staging.innolitics.com/device/K151034](https://fda-staging.innolitics.com/device/K151034)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
