K151025 · Ihear Medical, Inc. · EWO · Dec 29, 2015 · Ear, Nose, Throat
Device Facts
Record ID
K151025
Device Name
iHearTest
Applicant
Ihear Medical, Inc.
Product Code
EWO · Ear, Nose, Throat
Decision Date
Dec 29, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1050
Device Class
Class 2
Indications for Use
The iHearTest is an air-conduction over the counter hearing screener for profiling your hearing ability based on guidelines by the World Health Organization (WHO). The iHearTest is intended for home use by adults of at least 18 years of age. Hearing test results are displayed on the computer screen and stored in our HIPAA-compliant remote server for review anytime using secure online access.
Device Story
iHearTest is an OTC home hearing screener for adults. System includes iHearUSB device, factory-calibrated in-ear earphones, and browser-based software. User connects iHearUSB to PC; software controls sound output (500-4000 Hz, 25-85 dB HL) and prompts user response. Device includes microphone to monitor background noise levels. Results are displayed on PC and uploaded to HIPAA-compliant remote server. Software computes hearing ability per WHO guidelines and provides potential hearing aid recommendations. Benefits include convenient, self-administered hearing profiling. Healthcare providers do not operate the device; users review results via secure online access.
Clinical Evidence
Prospective IRB-approved study of 96 subjects (192 ears, ages 22-90). Primary endpoint: agreement with WHO method for identifying disabling hearing impairment. Results: 96.4% agreement with WHO method; 95% sensitivity; 97% specificity. Ordinal assessment showed 85.9% level-by-level agreement and 100% agreement within ±1 level. Variability (14.1%) was consistent with standard practice audiometry (13.5%). No adverse effects reported.
Technological Characteristics
Air-conduction audiometer; USB-connected hardware with integrated microphone for ambient noise monitoring. In-ear earphones factory-calibrated. Standards: ANSI S3.6-2010 (Type 4), EN/IEC 60601-1, EN/IEC 60601-1-2. Software-controlled sound output (25-85 dB HL at 500, 1000, 2000, 4000 Hz). Hardware-based sound output limiter (<90 dB HL).
Indications for Use
Indicated for adults aged 18+ for home-based air-conduction hearing screening. Contraindicated for individuals requiring physician-signed medical clearance if specific contraindications exist.
Regulatory Classification
Identification
An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird in flight, composed of three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 29, 2015
iHear Medical, Inc. Ms. Angela Foreman Director, Quality Assurance & Regulatory Affairs 15250 Hesperian Blvd. Ste. #102 San Leandro, CA 94578
Re: K151025
Trade/Device Name: iHearTest Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO Dated: November 17, 2015 Received: December 2, 2015
Dear Ms. Foreman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet
address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go
to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet
address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Srinivas Nandkumar -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151025
Device Name iHearTest
Indications for Use (Describe)
The iHearTest is an air-conduction over the counter hearing screener for profiling your hearing ability based on guidelines by the World Health Organization (WHO). The iHearTest is intended for home use by adults of at least 18 years of age. Hearing test results are displayed on the computer screen and stored in our HIPAA-compliant remote server for review anytime using secure online access.
Type of Use (Select one or both, as applicable)
| <input type="checkbox"/> Production Use (Part 21 CFR 601 Subpart E) |
|------------------------------------------------------------------------------|
| <input checked="" type="checkbox"/> Compassionate Use (21 CFR 601 Subpart E) |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
| Date Prepared: | December 21, 2015 |
|----------------|-------------------|
|----------------|-------------------|
#### Submitter Information:
| Company Name: | iHear Medical Inc. |
|------------------|--------------------------------------------|
| Company Address: | 15250 Hesperian Blvd. Suite #102 |
| | San Leandro, CA 94578 |
| Contact Person: | Angela Lee Foreman, Ph.D., Director, QA/RA |
| | iHear Medical, Inc. |
| | Tel: (510) 276-4437 |
| | Fax: (510) 686-8560 |
| | e-mail: angela@ihearmedical.com |
#### De vice Information:
| Trade Name: | iHearTest |
|----------------------|------------------------------|
| Common Name: | Hearing Screener, Audiometer |
| Classification Name: | Audiometer |
| Device Class/Code: | Class II Exempt, EWO |
| Regulation Number: | 21 CFR §874.1050 |
| Type of 510(k): | Traditional |
#### Predicate Devices:
- 1. Maico MA1 Handheld Audiometer, (Establishment # 2113281) Maico Diagnostics, Exempt under 21 CFR §874.1050
- 2. Earscan 3 Screening Audiometer (K 982878) Micro Audiometric Corp.; approved for marketing on November 3, 1998
- 3. InSound HandFit (K 021867) InSound Medical, part of the InSound XT Hearing Aid and HandFit system, approved for marketing on November 19, 2002
{4}------------------------------------------------
#### K151025 iHearTest
Device Description: The iHearTest is an over the counter (OTC) hearing screener for profiling hearing ability based on guidelines by the World Health Organization (WHO). The iHearTest is intended for home use by adults of at least 18 years of age.
## The iHearTest Software System consists of:
- The iHearUSB device for connection to a personal computer via USB port. ●
- Factory calibrated in-ear earphones, marked for right and left ears, with 3 tip sizes. .
- iHearTest software application.
- . Instructions and warranty information provided in the User Guide and online.
- A quick Install guide to assist in the initial software installation, registering the USB ● device, and initiating the iHearTest.
The iHearTest Software System requires a standard personal computer (PC) with USB port running Microsoft Windows or OS X. The iHearTest software application is a browser-based software application executed from the consumer's personal computer that provides the user interface for a hearing screening test on the PC's monitor, controls the output of test sound signals from an iHearUSB device connected to the PC's USB port, displays hearing test results and sends the results to a remote HIPAA compliant server for review anytime by the consumer using secure online access, and conducts periodic checks of earphone calibration via the calibration tool incorporated into the iHearUSB device.
The User Guide and the online instructions instruct the individual in operations and functions. The individual follows the instructions to initiate and proceed with the hearing screening procedure.
The iHearTest kit is designed to produce air-conducted sounds via the in-ear earphone within the supra-threshold range of 25-85 dB HL at test frequencies of 500, 1000, 2000 and 4000 Hz. The iHearTest software application prompts the user to respond to test sounds and determines the individual's hearing level at each test frequency. The user's overall hearing ability is then computed and displayed for both the right and left ear with a recommendation for potential hearing aid use as per WHO guidelines. The user is provided, both online and in the User Guide, with disclaimers and safety information pertaining to the iHearTest. If a contraindication exists, the consumer must submit a physician-signed Medical Clearance prior to proceeding with the hearing screening.
The iHearUSB device incorporates a microphone for sensing the background noise during the hearing screening process, and ensuring that the background noise is within an acceptable level range.
Indications for Use: The iHearTest is an air conduction over the counter hearing screener for profiling your hearing ability based on guidelines by the World Health Organization (WHO). The iHearTest is intended for home use by adults of at least 18 years of age. Hearing test results
{5}------------------------------------------------
#### K151025 iHearTest
are displayed on the computer screen and stored in our HIPAA-compliant remote server for review anytime using secure online access.
Intended Use: The online iHearTest is intended for the adult consumer to provide basic hearing screening in the convenience of home or office.
Comparison to Predicate Device: Device features and parameters are similar for the iHearTest, which is the subject of this 510(k) Notification, and the aforementioned predicate devices. The intended use (including patient population) for the iHearTest is mostly consistent with the cited predicate devices in terms of technological characteristics (i.e., design, material, and sound emissions). The iHearTest is designed for self-administration by the adult consumer; it is not a diagnostic tool for the professional health care provider.
## Performance testing:
Verification testing was conducted on the iHearTest software and the iHearUSB firmware to confirm the correct functioning of the Software System as described above, including correct output of sounds at test frequencies of 500, 1000, 2000 and 4000 Hz. Even when other Windows programs/apps were running in parallel. Also manipulation of Windows audio output settings were verified to not affect the intended sound output, consistent with the design of the iHear system.
Earphones were confirmed to be factory-calbrated using industry standard callbration tools, and are confirmed to maintain calibration after simulated aging and weathering tests.
The iHearUSB device's built-in calibration tool is confirmed to have identical calibration test results as industry standard calibration tools both at the time of manufacturing and after simulated aging and weathering tests.
Engineering validation was performed to demonstrate performance in accordance with recognized standards for this type of device, namely ANSI S3.6-2010 Type 4 air-conduction audiometer and the World Health Organization (WHO) guidelines for determination of hearing impairment.
## Clinical Study Summary:
An Institutional Review Board (IRB)-approved clinical study was conducted to assess the safety and efficacy of the iHearTest as a home air-conduction hearing screener. The study included 96 subjects (192 ears): 54 males and 42 females, ages 22-90 years with median age of 52. The iHearTest scoring results were compared to World Health Organization (WHO) grading of hearing impairment using standard practice audiometry. The study targeted adult subjects with known or suspected hearing loss. The primary objective of the iHearTest clinical study was to determine identification of disabling hearing impairment (corresponding to WHO levels 2-4), and the extent to which the iHearTest agreed with the WHO method for determining hearing impairment. The clinical study results showed 96.4% agreement with the WHO method using
{6}------------------------------------------------
#### K151025 iHearTest
standard practice audiometry for the assessment of disabling hearing impairment, with sensitivity and specificity of 95% and 97%, respectively. For the ordinal assessment of hearing ability, the iHearTest showed 85.9% level-by-level (WHO grades of 0-4) agreement and 100% agreement within ±1 level with respect to WHO grading using standard practice instruments. The 14.1% variability in ordinal assessment for the iHearTest with respect to the standard practice method was consistent with the 13.5% variability across two standard practice audiometry methods evaluated in the study.
In addition to determining agreement with the WHO method using standard practice audiometry, the usability of the iHearTest system was also evaluated to obtain comments on the subjective experience with the iHearTest device, its packaging, software installation and setup. The iHearTest was found to be safe and effective as an over-the-counter hearing screening device.
In summary, subjects tested included both male and female adult participants, of the appropriate age range, and varied hearing impairment levels (i.e., subjects with normal hearing and mild to severe impairment). No adverse effects or complications were observed. The performance data collected supports a substantial equivalence determination.
Safety: The design of the iHearTest device provides safety to the individual user. To prevent the unlikely possibility of excessive exposure to high-level sound, hardware and software were implemented to limit maximum sound output to below 90 dB HL. The iHearTest device has been designed to conform to the following performance and safety standards:
| ANSI S3.6-2010 | Specifications for Audiometers, American National<br>Standards Institute |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EN/IEC 60601-1 | Medical electrical equipment: Part 1: General<br>requirements for safety |
| EN/IEC 60601-1-2 | Medical electrical equipment: Part 1: General<br>requirements for safety; Section 1-2 Collateral standard:<br>Electromagnetic compatibility - Requirements and tests |
Conclusion: The information and data provided in this 510(k) Notification establish that the iHearTest device with the online test performed by the consumer using his/her home computer is substantially equivalent to the aforementioned legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.