← Product Code [PGY](/productcode/PGY) · K151007

# 21.3 inch (54 cm) Color Digital Mammography LCD Monitor CCL550i2(CL21550) (K151007)

_Jvckenwood Corporation · PGY · May 8, 2015 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K151007

## Device Facts

- **Applicant:** Jvckenwood Corporation
- **Product Code:** [PGY](/productcode/PGY.md)
- **Decision Date:** May 8, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

21.3 inch (54 cm) inch Color Digital Mammography 5M pixel LCD Monitor CCL550i2 (CL21550) is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. It is to be used in digital mammography PACS and modalities including FFDM.

## Device Story

The CCL550i2 is a 21.3-inch color LCD monitor designed for displaying high-resolution medical images, specifically for digital mammography and FFDM. It receives image data via DVI or DisplayPort from a PACS or imaging modality. The device features an LED-backlit panel and internal 16-bit LUTs to ensure compliance with the DICOM GSDF for grayscale rendering. It includes a front sensor and ambient light sensor to automate luminance stabilization and grayscale calibration, reducing the need for manual intervention. The monitor is used in clinical settings by radiologists and medical practitioners to view diagnostic images. The output is a high-fidelity visual representation of medical data, which supports clinical decision-making by providing consistent, calibrated image quality. The device benefits patients by ensuring accurate, reliable visualization of mammographic images for diagnostic assessment.

## Clinical Evidence

Bench testing only. Performance was validated through system and integration testing, including luminance response, uniformity, geometrical distortion, veiling glare, chromaticity, and temporal response. Calibration accuracy using the integrated front sensor was verified against external sensors. Compliance with AAPM-TG18 guidelines and international safety/EMC standards was confirmed.

## Technological Characteristics

21.3-inch AMLCD panel; LED backlight; 2048x2560 resolution; 10-bit per pixel intrinsic depth; 16-bit LUT; DVI and DisplayPort connectivity; integrated front sensor and ambient light sensor for calibration; power supply 100-240V AC. Complies with ANSI/AAMI ES60601-1 and CAN/CSA-C22.2 No. 601-1.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- MS55i2 (ML21055) ([K131137](/device/K131137.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three abstract human profiles in a row, with a triple-strand helix wrapping around them. The symbol is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 8, 2015

JVC KENWOOD Corporation % Mr. Tsukasa Tashiro Engineering Specialist 3-12, Moriya-cho, Kanagawa-ku Yokohama-shi, Kanagawa 221-0022 JAPAN

Re: K151007

Trade/Device Name: 21.3 inch (54 cm) Color Digital Mammography LCD Monitor CCL550i2 (CL21550) Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: PGY Dated: April 14, 2015 Received: April 15, 2015

Dear Mr. Tashiro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

### Indications for Use

510(k) Number (if known)

K151007

Device Name CCL550i2 (CL21550)

Indications for Use (Describe)

21.3 inch (54 cm) inch Color Digital Mammography 5M pixel LCD Monitor CCL550i2 (CL21550) is intended to be used in displaying and viewing medical images for diagnosis by trained medical practitioners. It is to be used in digital mammography PACS and modalities including FFDM.

| Type of Use (Select one or both, as applicable)                                                                                         |                                             |
|-----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <div> <span> <span style="text-decoration: underline;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)             </span> </div> | Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services  

Food and Drug Administration  

Office of Chief Information Officer  

Paperwork Reduction Act (PRA) Staff  

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

PSC Publishing Services (301) 443-6740 EF

Form Approved: OMB No. 0910-0120

Expiration Date: January 31, 2017

See PRA Statement below.

{3}------------------------------------------------

# TOTOKU

## 510(k) SUMMARY

| Submitted Information:   | JVC KENWOOD CORPORATION<br>3-12, MORIYA-CHO, KANAGAWA-KU,<br>YOKOHAMA-SHI, KANAGAWA, 221-0022 JAPAN                                                                                                                                                                                              |
|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:          | Tsukasa Tashiro, Engineering Specialist<br>Email: tashiro.tsukasa@jvckenwood.com<br>Tel: +81.258.24.6611<br>Fax: +81.258.24.6617                                                                                                                                                                 |
| Date Prepared:           | April 14, 2015                                                                                                                                                                                                                                                                                   |
| Device Name:             | 21.3 inch (54 cm) inch Color Digital Mammography LCD Monitor<br>CCL550i2 (CL21550)                                                                                                                                                                                                               |
| Common Name:             | CCL550i2, CL21550                                                                                                                                                                                                                                                                                |
| Classification Name:     | Class II<br>(Part 892 Radiology Devices<br>Sec. 892.2050 Picture Archiving and Communication System)                                                                                                                                                                                             |
| Predicate Device:        | 21.3 inch (54 cm) inch Monochrome Digital Mammography LCD<br>Monitor MS55i2 (ML21055) (K131137)                                                                                                                                                                                                  |
| Device Description:      | CCL550i2 (CL21550) is a 21.3 inch (54 cm) color LCD<br>monitor whose display resolution is 2048 x 2560 (landscape), 2560 x<br>2048 (portrait) supporting DVI (digital visual interface) and<br>DisplayPort.                                                                                      |
| Intended Use:            | 21.3 inch (54 cm) inch Color Digital Mammography 5M pixel LCD<br>Monitor CCL550i2 (CL21550) is intended to be used in displaying<br>and viewing medical images for diagnosis by trained medical<br>practitioners. It is to be used in digital mammography PACS and<br>modalities including FFDM. |
| Substantial Equivalence: | CCL550i2 (CL21550) shares the same characteristics with our<br>predicate device MS55i2 (ML21055) (K131137) except for the LCD<br>panel and standard software.                                                                                                                                    |

### JVC KENWOOD Corporation

Professional & Healthcare Division 3-12, Moriya-cho, Kanagawa-ku, Yokohama-shi, Kanagawa, 221-0022 Japan

{4}------------------------------------------------

| Display System Description                                                                                                                                                                                                                                |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. Active-Matrix Liquid-Crystal Displays (AMLCD) panel manufacturer, technology, Screen size<br>and pixel pitch:                                                                                                                                          |
| Screen size: 422.4(H)mm x 337.92(V) mm at landscape display, 337.92(H)mm x 422.4(V)mm<br>at portrait display, Pixel pitch: 0.165mm(H) x 0.165mm(V) at landscape display, 0.165mm(H)<br>x 0.165(V) at portrait display                                     |
| • Communicating ports:                                                                                                                                                                                                                                    |
| Serial communication: USB (upstream x1, downstream x2)                                                                                                                                                                                                    |
| 2. Graphics card and software                                                                                                                                                                                                                             |
| • Digital-to-Analog converters: speed and precision:                                                                                                                                                                                                      |
| Only Digital operation, Digital to Analog converters are not provided.                                                                                                                                                                                    |
| • Software included: calibration                                                                                                                                                                                                                          |
| JVCKENWOOD's calibration software called "FCAL" optimizes the display's performance                                                                                                                                                                       |
| using DICOM GSDF gamma and luminance. Calibration is recommended quarterly.                                                                                                                                                                               |
| • On-screen GUI: On-screen GUI is not provided.                                                                                                                                                                                                           |
| • Panel user controls: power switch                                                                                                                                                                                                                       |
| Technical Specification                                                                                                                                                                                                                                   |
| 1. Measuring Equipment<br>Refer to List.                                                                                                                                                                                                                  |
| 2. Declared and actual (measured with test pattern in the screen) array size:<br>Declared array size (spec): 422.4(H)mm x 337.92(V) mm<br>Actual array size (measured): 422.7(H)mm x 338.0(V)mm                                                           |
| 3. Luminance response                                                                                                                                                                                                                                     |
| • max and min achievable luminance:<br>[SPEC] Luminance Max (Lmax)=1000cd/m²,<br>Luminance Min (typ) = 0.77cd/m² at Lmax=1000cd/m²                                                                                                                        |
| • max and min recommended (operational) luminance:<br>[SPEC] Luminance Max recommend: Lmax(r)=500cd/m², Lmin(r)=0.8cd/m²                                                                                                                                  |
| • intrinsic bit-depth of the panel:<br>[SPEC] 10 bit per each pixel                                                                                                                                                                                       |
| • true output bit-depth by performing visual test with gradient test pattern:<br>[SPEC] 10.0bit grayscale (1024 steps) achieved by 16bit LUT (Look-up Table) in<br>the Display hardware.                                                                  |
| • intrinsic luminance response at 256 digital values:<br>[SPEC] dJND per dP<3.000, dJNDs/dP max error<2.000,<br>JNDs/P RMS error<1.000<br>Above spec is based on AAPM-TG18 Advanced Luminance Response, 4.3.5<br>Refer to actual luminance response data. |
| 4. Angular Dependency                                                                                                                                                                                                                                     |
| • angular dependency of luminance:<br>[SPEC] LR'≥175, Κδ>30%                                                                                                                                                                                              |
| 5. Luminance uniformity<br>[SPEC] Less than 30% based on AAPM-TG18 4.4. Refer to actual Luminance<br>uniformity data.                                                                                                                                     |
| 6. Geometrical distortion<br>[SPEC] Less than 2.0% based on AAPM-TG18. Refer to actual Geometrical<br>Distortion data.                                                                                                                                    |

-

{5}------------------------------------------------

- 7. Display reflectance
	- Bi-directional reflection distribution function [SPEC] Refer to Max allowable ambient luminance in Tables 4 and 5 on AAPM-TG18 4.2.4.

Refer to actual Reflectance Data.

- 8. Noise
	- · Pixel fill factor [SPEC] 30% Min. Refer to Data of Pixel fill factor.
	- · Noise power spectrum (weiner spectrum) [SPEC] Refer to actual data.

### 9. Veiling glare

[SPEC] Glare Ratio (GR) ≥ 400 GR=(Lb-Ln)/(L-Ln) based on AAPM-TG18 4.7.4 Refer to actual veiling glare ratio data.

- 10. Chromaticity
[SPEC] Delta (u', v') ≤ 0.01 measured at 80% Lmax based on AAPM-TG18 4.8.4 Refer to Chromaticity actual data.

### 11. Artifacts

- pixel dropouts including spatial distribution [SPEC] Class (pixel) II. Refer to table 3 on 3.4.13 ISO13406-2.
- phase/clock issues flicker
- · miscellaneous including ringing, ghosting, image sticking [SPEC] By visible check, no ringing, ghosting image sticking
- 12. Spatial resolution, spatial MTF [DATA] Refer to actual MTF data.

### 13. Temporal response

- rise and fall time constants for 5-95% and 40-60% transitions [SPEC] Rise Time Tr, Fall Time Tf, Tr, Tf<55mS. Refer to actual data.
- 14. Stability (possibly determined via temperature or time stress tests)
	- · of luminance response, of temporal response, of described artifacts [SPEC] Ko within +/- 10% deviation to DICOM GSDF, based on AAPM-TG18 Luminance Response 4.3.4 via temperature stress 0 degC, 20 degC, 25 degC, 30 deqC, 40 deqC Refer to actual measured data.

[fefxos = for example, following ... or similar]

{6}------------------------------------------------

|                                                     |                                                                                                                                                                                | 5M Color Digital Mammograpl                                                           |
|-----------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
|                                                     |                                                                                                                                                                                | CD Monitor CCL550i2 (CL2155                                                           |
| 0(k) Numb                                           |                                                                                                                                                                                |                                                                                       |
| isplay Are put Signa                                |                                                                                                                                                                                | lorizontal: 337.92mm, Vertical: 422.41                                                |
|                                                     |                                                                                                                                                                                | DisplayPort                                                                           |
| aximum Displa                                       |                                                                                                                                                                                | 048 x 2560 dots                                                                       |
| canning Frequer                                     |                                                                                                                                                                                | Landscap<br>Portrait<br>izontal:129.1KHz Vertical:50Hz izontal:103.5KHz Vertical:50Hz |
| aximum Image Cloc                                   |                                                                                                                                                                                | 85 MHz                                                                                |
| aximum Luminancı  alibrated Luminanc  วntrast Ratio |                                                                                                                                                                                | Min.700cd/m² Typ. 1000cd/m² (as LCD component) (500cd/m² calibrated lluminance)       |
|                                                     |                                                                                                                                                                                |                                                                                       |
|                                                     |                                                                                                                                                                                | Min.1000 : 1      Typ. 1300 : 1                                                       |
|                                                     | Min.900 : 1    Typ.1200 : 1 CR>50 Typ.176 horizontal/vertical Software (Standard): Medivisor Software (Standard): Medivisor Calibration Sensor (Optional): Chroma5 (X-Rit      | R>50 Typ.176 horizontal/vertic:                                                       |
| ewing Angle uminance Calibratio                     |                                                                                                                                                                                | tware (Standard): FCAL                                                                |
| Optional)                                           |                                                                                                                                                                                | ibration Sensor (Optional): Chroma5 (X-Rite                                           |
| erial Communicati                                   |                                                                                                                                                                                | B: Downstream port x 2, Upstream port                                                 |
| ayscale                                             | Uest Downstream Boot in 2. Uperream Port x 1 10.3 bit it CoSA C22.2 No. 601-1. FOOD, V. FOOD, V. FOOD, V. FOOD, V. FOOD, V. POOD, V. POOD, V. POOD, V. POOLES NGC 3. BOOKWAY N | 0.0 bit (1024 gradation                                                               |
| afety Standa                                        |                                                                                                                                                                                | 60601-1,<br>NSI/AAMI ES60601-1, CAN/CSA-C22.2 No.                                     |
|                                                     |                                                                                                                                                                                | B, VCCI-B, MDD/CE                                                                     |
| eight & Dimensi                                     |                                                                                                                                                                                | Net                                                                                   |
|                                                     |                                                                                                                                                                                |                                                                                       |
|                                                     |                                                                                                                                                                                |                                                                                       |
|                                                     |                                                                                                                                                                                |                                                                                       |
|                                                     |                                                                                                                                                                                |                                                                                       |
| ower Suppl                                          | AC100-240V, 1.5-0.6A, 50/60Hz                                                                                                                                                  | C100-240V, 1.5-0.6A, 50/60Hz                                                          |

# bstantial Equivalence Compariso

Similarities

Similarities:
CCL55012 (CL21550) employs the same driver board, tilt stand, power supply, etc. except of those of LCD Panel and standard software

Differences:
CCL550i2 (CL21550) employs different LCD panel and standard softwar

{7}------------------------------------------------

CCL550i2 (CL21550) can be considered to have equivalent display performances to those of the predicate device MS55i2 (ML21055) due to the following reasons:

- a. The maximum display sizes (2048*2560) and the active area sizes (337.92mm (H) x 422.4mm (V)) used for the both devices are the same.
- b. The DICOM calibrated luminance value of the both devices are the same (500 cd/m²) though the typical maximum luminance value (typ. 1000cd/m²) is lower than that of the predicate device (typ. 1200cd/m²). And Panels of the both devices are used LED backlight. The high luminance to be maintained constantly was realized by the employment of LED backlight deteriorating more slowly than conventional CCFL backlights. We have not recognized any adverse effects of the LED backlight on the quality of displayed images. Refer to "Technical Data" where several image quality characteristics of the proposed device are compared with those of the predicate device.
- c. The both devices display images in accordance with DICOM GSDF by default utilizing the factory calibrated display mode stored in lookup tables inside of them.
- d. Both devices support Digital Visual Interface (DVI) and DisplayPort.
- e. As for the maintenance, the same QC software is used for both devices have Front Sensor to stabilize the luminance.
- f. As for built-in sensors, both devices have 2 (two) kinds of common sensors, Front Sensor and Ambient Light Sensor. Front Sensor is related to the maintenance or calibration and Ambient Light Sensor is used to measure the ambient light by lx. Front sensor enables automatic grayscale calibration by measuring the luminance at the screen surface. Without Front sensor, the grayscale calibration process requires human intervention and the use of and external sensor. The accuracy data of the calibration with external sensors and that with Front Sensor is included in section 8 "Validation & Verification: Integration Test Report" and section 9 "Validation & Verification: System Test Report" in "Application".
- g. The overall design of the CCL550i2 (CL21550) was validated in accordance with internationally recognized Safety and EMC standards by third-party certifiers. Besides, JVC KENWOOD Corporation performed a range of system and performance tests to ensure that the CCL550i2 (CL21550) performs in accordance with its specifications. None of the tests revealed behaviors inconsistent with the expected performance.

### Conclusion

The 5M pixel Color LCD Monitor, CCL550i2 (CL21550) is substantially equivalent to the predicate device with respect to technical characteristics, application and intended use. The specifications of the primary components employed by the proposed device are the same to those of the predicate device except for the LCD panel, and the differences have been independently validated. Any differences between the devices do not affect safety or effectiveness.

---

**Source:** [https://fda-staging.innolitics.com/device/K151007](https://fda-staging.innolitics.com/device/K151007)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
