K150976 · Anatomage, Inc. · LLZ · Jun 4, 2015 · Radiology
Device Facts
Record ID
K150976
Device Name
Collage
Applicant
Anatomage, Inc.
Product Code
LLZ · Radiology
Decision Date
Jun 4, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Collage is a software application intended for viewing of 3D medical image files from scanning devices, such as CT, MRI, or 3D Ultrasound as well as 2D patient images, such as patient photographs, intraoral photographs, and dental x-rays. Images and data can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. It is intended for use by doctors, clinicians, and other qualified individuals utilizing standard PC hardware. This device is not indicated for mammography use.
Device Story
Collage is an interactive imaging software application for visualization, storage, and management of 2D and 3D medical image data. Input data includes CT, MRI, 3D Ultrasound, patient photographs, intraoral photographs, and dental x-rays. The software runs on standard off-the-shelf x86 Windows-based PCs. It allows clinicians to retrieve, process, render, review, store, and print medical images. The device acts as a platform to bridge different patient data sets for comprehensive studies. It does not perform automated diagnosis; it serves as a tool for qualified healthcare providers to interpret images and make clinical decisions. The device is intended for use in clinical settings by doctors and dental clinicians. It provides a centralized system for managing diverse patient imaging data, potentially improving workflow efficiency and diagnostic review.
Clinical Evidence
No clinical data. Bench testing only. Performance was evaluated by comparing images rendered by Collage against the predicate software. An expert in radiology evaluated the rendering of 2D and 3D images to confirm stability, functionality, and equivalence to the predicate.
Technological Characteristics
Software-only medical device for digital image processing. Operates on standard off-the-shelf x86 processor-based computers running the Windows operating system. Does not contact the patient. No life-sustaining functions. Connectivity via computer networks. No specific hardware requirements beyond standard PC specifications.
Indications for Use
Indicated for doctors, clinicians, and qualified individuals to view, store, communicate, process, and display 3D medical images (CT, MRI, 3D Ultrasound) and 2D images (photographs, dental x-rays). Not indicated for mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 4, 2015
Anatomage Inc. % Mr. Sean Xu Program Director 111 N. Market Street. #500 SAN JOSE CA 95113
Re: K150976 Trade/Device Name: Collage Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 14, 2015 Received: April 15, 2015
Dear Mr. Xu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150976
Device Name Collage
#### Indications for Use (Describe)
Collage is a software application intended for viewing of 3D medical image files from scanning devices, such as CT, MRI, or 3D Ultrasound as well as 2D patient images, such as patient photographs, intraoral photographs, and dental x-rays. Images and data can be stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. It is intended for use by doctors, clinicians, and other qualified individuals utilizing standard PC hardware. This device is not indicated for mammography use.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;"><b> </b></span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510K Summary
| Submitter | Anatomage Inc.<br>111 N. Market Street #500<br>San Jose, CA 95113<br>Phone: (408) 885-1474<br>Fax: (408) 295-9786 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Regulatory Contact Person | Sean Xu, Program Director<br>Phone Number: (408) 885-1474 ext 108 |
| Device Name | Collage |
| 510(k) Preparation Date | 04/08/2015 |
| Common Name | System, Image Processing, Radiological |
| Classification | Class II |
| Classification Name | Imaging Processing System, LLZ, 21 CFR 892.2050 |
| Product Code | LLZ, 21 CFR 892.2050 |
| Device Description | Collage is an interactive imaging software used for the<br>visualization, storage, and management of 3D medical image files<br>from scanning devices, such as CT, MRI, or 3D Ultrasound as<br>well as 2D patient images, such as patient photographs, intraoral<br>photographs, and dental x-rays. Doctors, dental clinicians, and<br>other qualified individuals can retrieve, process, render, review,<br>store and print images, utilizing standard PC hardware. The<br>software runs in Windows operating systems and visualizes<br>medical imaging data on the computer screen. The Collage<br>software is intended as a platform bridging different sets of patient<br>data for comprehensive studies. With Collage software, doctors<br>can manage all of their patient images, including both 2D and 3D<br>image data, in a single software. |
| Intended Use | Collage is a software application intended for viewing of 3D<br>medical image files from scanning devices, such as CT, MRI, or<br>3D Ultrasound as well as 2D patient images, such as patient<br>photographs, intraoral photographs, and dental x-rays. Images and<br>data can be stored, communicated, processed and displayed within<br>the system and or across computer networks at distributed<br>locations. It is intended for use by doctors, clinicians, and other |
| Equivalent Devices | K101342, OsiriX MD |
| Technological Characteristic | Collage is a software that handles digital medical images. This<br>device does not contact the patient, nor does it control any life<br>sustaining devices. Diagnosis is not performed by this software<br>but by doctors, clinicians and other qualified individuals. A<br>physician, providing ample opportunity for competent human<br>intervention, interprets images and information being displayed<br>and printed.<br>Collage is installed on standard off-the-shelf x86 processor based<br>computers running the Windows operating system. |
| Non-Clinical Test Results | The following quality assurance measures were applied to the<br>development of the system:<br>· Risk Analysis<br>· Requirements Reviews<br>· Design Reviews<br>· Performance testing<br>· Usability testing<br>· Final acceptance testing<br>· Bench testing to compare with predicate software<br><br>Testing confirmed that the software is stable and operating as<br>designed. Testing also confirmed that the software has been<br>evaluated for hazards and that risk has been reduced to acceptable<br>levels.<br><br>Bench testing of the software with predicate software was<br>performed by evaluation of images rendered by Collage and<br>predicate software. This testing and evaluation included testing of<br>rendering both 2D and 3D images in both predicate and subject<br>software and was evaluated by an expert in the field of radiology.<br>This testing confirms that Collage is as effective as its predicate in<br>its ability to perform its essential functions of rendering and<br>managing medical images. |
| Summary | Based on the intended use, product, performance, and testing<br>information provided in this notification, the subject device has<br>been shown to be substantially equivalent in the area of technical<br>characteristics, general functionality, and indicated use to the<br>currently marketed predicate device and does not introduce any<br>new potential risks. |
{4}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.