The TRU Implant System can be used in dental implant applications for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible. Implant retained restorations may consist of single crowns or bridges as well as complete or partial dentures. The TRU Implant System is intended for delayed loading. It is also indicated for immediate loading with good primary stability and appropriate occlusal loading. The TRU Implant System is only intended for use with straight abutments. The TRU implant body is intended to be placed such no angle correction is necessary.
Device Story
TRU Dental Implant System consists of self-tapping, double-thread, root-form screw implants; manufactured from Grade 4 titanium with acid-etched surface. Available in various diameters (3.5-6.0mm) and lengths. Used by dentists/oral surgeons in clinical settings for oral rehabilitation. Implants placed in maxilla or mandible to support prosthetic restorations (crowns, bridges, dentures). System includes healing, straight, and UCLA abutments. Special driver facilitates insertion without implant mount. Device provides stable foundation for dental prosthetics; benefits patients by restoring chewing function and aesthetics. No angle correction permitted; intended for straight abutment use only.
Clinical Evidence
Bench testing only. Performance testing conducted per FDA guidance for root-form endosseous dental implants. Sterility, shelf life, and biocompatibility data leveraged from previously cleared Sterngold devices. Fatigue testing deemed unnecessary due to similarity to predicates and restriction to straight abutments without angle correction.
Indicated for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible requiring single crowns, bridges, or complete/partial dentures. Suitable for delayed or immediate loading given good primary stability.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
Nobel Active Internal Connection Implant (K071370)
Sterngold Acid Etch Dental Implant System (K023580)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a stylized symbol. The symbol consists of three abstract human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 24,2015
Sterngold Dental, LLC Maria Rao Director of QA & Regulatory Affairs 23 Frank Mossberg Drive Attleboro, Massachusetts 02703
Re: K150968
Trade/Device Name: TRU Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE, NHA Dated: June 26, 2015 Received: June 30, 2015
#### Dear Ms. Rao:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Susan Runno, DDS, MA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known): _K150968
Device Name: TRU Dental Implant System
## Indications for Use:
The TRU Implant System can be used in dental implant applications for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible. Implant retained restorations may consist of single crowns or bridges as well as complete or partial dentures. The TRU Implant System is intended for delayed loading. It is also indicated for immediate loading with good primary stability and appropriate occlusal loading.
The TRU Implant System is only intended for use with straight abutments. The TRU implant body is intended to be placed such no angle correction is necessary.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-the -Counter Use (21 CFR 807 Subpart D)
{3}------------------------------------------------
#### K150968 510(k) Summary
| Trade Name: | TRU Dental Implant System |
|----------------------|-------------------------------------------------------------------------|
| Sponsor: | Sterngold Dental, LLC<br>23 Frank Mossberg Drive<br>Attleboro, MA 02703 |
| | Contact: Maria Rao, QA/RA Director<br>Ph: 508-226-5660 ext 1206 |
| Date: | July 23, 2015 |
| Device Generic Name: | Endosseous Dental Implant |
Classification: According to Section 513 of the Federal Food, Drug, and Cosmetic Act, the device classification is Class II
Product Code: DZE (21CFR 872.3640) NHA (21CFR 872.3630)
#### Predicate Devices:
The TRU Dental Implant System is substantially equivalent to other currently marketed dental implant systems that have been cleared by FDA through the 510(k) Premarket Notification process, including the Sterngold Acid Etch Dental Implant System, and Nobel Active Internal Connection Implant System.
## Product Description:
The TRU is a self-tapping, double thread screw implant with a micro groove section, manufactured from pure grade 4 titanium. The implant surface is acid etched. The implants are manufactured with four body diameters 3.5 mm, 4.3 mm, 5.0 mm, and 6.0 mm. The 3.5 mm implants have a Narrow Platform (NP) prosthetic head. The 4.3 mm, 5.0 mm, and the 6.0 mm implants have a Regular Platform (RP) prosthetic head. They provide for non-rotational single and multiple tooth restorations in both the maxilla and mandible. They are packaged sterile in a double blister, which contains the implant in a titanium tube. A special driver is available for the implants which simplant insertion by eliminating the need for an implant mount.
The proposed TRU Dental Implant will be available in a range of lengths and diameters.
| Thread Major Diameter (mm) | Overall Implant Length (mm) |
|----------------------------|-----------------------------|
| 3.5mm | 8mm 10mm, 12mm, 14mm |
| 4.3mm | 8mm 10mm, 12mm, 14mm |
| 5.0mm | 8mm 10mm, 12mm |
| 6.0mm | 8mm, 10mm |
The prosthetic components for the TRU implant System include Healing abutments, Straight Abutments and UCLA Abutments.
{4}------------------------------------------------
Healing Abutments are intended for placement onto the implant. As the tissue heals, it conforms to the contours of the abutment. As the gingiva heals, the tissue adapts to the form of the healing abutment, creating permanent oral access to the implant and prepares the gingiva for the restorative phase.
Straight Abutments attach directly to the implant and provide the transitional link between the head of the implant and the restorative components. They are designed to be used for screw retained multiimplant prostheses. They are available in a narrow platform with 1.5mm and a 3.0mm cuff size and in a regular platform with 1.5mm and a 3.0mm cuff size.
UCLA Abutments attach directly to the implant and provide a pattern for the creation of a screw retained veneered crown. They are available with two apical (bottom) ends; one with a hex to engage the hex of the implant and one without a hex that does not engage the implant. The engaging/hexed abutment is used on restoration of a single tooth to prevent rotation of the completed crown, and the non-engaging/non-hexed is used on multiple tooth restorations.
## Indications for Use:
The TRU Implant System can be used in dental implant applications for oral rehabilitation of edentulous and partially dentate patients in the maxilla and mandible. Implant retained restorations may consist of single crowns or bridges as well as complete or partial dentures. The TRU Implant System is intended for delayed loading. It is also indicated for immediate loading with good primary stability and appropriate occlusal loading.
The TRU Implant System is only intended for use with straight abutments. The TRU implant body is intended to be placed such no angle correction is necessary.
## Substantial Equivalence:
The proposed TRU Dental implants are substantially equivalent to other currently marketed dental implant systems that have been cleared by FDA through the 510(k) Premarket Notification process. The Indications for Use, basic design, fundamental operating principles are the same or similar as the predicate devices.
- . Nobel Active Internal Connection Implant (K071370) - Primary Predicate
- Sterngold Acid Etch Dental Implant System (K023580) Reference Predicate ●
Compatibility and substantial equivalency was determined by comparing the design features including diameters, lengths, cuff sizes, materials, implant-to-abutment connection platform, implant surface texture, abutment fixation method, and intended use of proposed device to predicate devices.
Any differences between the proposed devices and predicate devices do not render the device NSE.
{5}------------------------------------------------
## Technological Characteristics:
The TRU dental Implant provides the same or similar functions as well as design and technological characteristics as the predicate devices. In addition the surface treatment is identical to the cleared Sterngold Acid Etched Implant (K023580).
The materials, technology and processes used to produce the TRU dental implants and abutments are the same as other Sterngold Dental devices previously cleared by the FDA.
See Substantial Equivalence Comparison table below.
| Property | TRU Dental<br>Implant System<br>(Proposed Device) | Nobel Active Internal Connection Implant<br>(K071370)<br>Primary Predicate | Sterngold Acid Etch Dental<br>Implant System<br>(K023580)<br>Reference Predicate |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Implant<br>Design | Self-tapping, threaded, Root-form<br>implant | Self-tapping, threaded, Root-form implant | Self-tapping, threaded, Root-<br>form implant |
| Implant Sizes<br>diameter x<br>length | 3.5x8mm,10mm,12mm,14mm<br>4.3x8mm,10mm,12mm,14mm<br>5.0mmx8mm,10mm,12mm<br>6.0mmx8mm,10mm,12mm | 3.5x8.5mm,10mm,11.5mm,13mm,15mm<br>4.3x8.5mm,10mm,11.5mm,13mm,15mm | 3.3x10mm,11.5mm,13mm<br>4.0x8.5mm,10mm,11.5mm,13mm,15mm<br>5.0x8.5mm,10mm |
| Abutment<br>Compatibility | Internal Connection | Internal Connection | External Connection |
| Implant Stage | Stage 1 and Stage 2 | Stage 1 and Stage 2 | Stage 1 and Stage 2 |
| Placement<br>Accessories | Implant drills, countersinks, bone taps | Implant drills, countersinks, bone taps | Implant drills, countersinks,<br>bone taps |
| Supplied<br>Sterile | Yes | Yes | Yes |
| Intended Use | Implant applications for oral<br>rehabilitation of edentulous and<br>partially dentate patients in the maxilla<br>and mandible. Implant retained<br>restorations may consist of single<br>crowns or bridges as well as complete<br>or partial dentures. | Long term surgical implantation in the bone of<br>the patient's upper or lower arch to provide<br>immediate load or delayed load of prosthetic<br>systems, such as artificial teeth, in order to<br>restore the patient's chewing function. Also<br>indicated for for immediate loading with good<br>primary stability and appropriate occlusal<br>loading. | Implantation into any area of<br>the partially and/or fully<br>edentulous maxilla and<br>mandible for the support of a<br>removable or fixed dental<br>prosthesis. Intended for single<br>tooth or multiple unit<br>prosthesis. |
| Implant<br>Material | Wrought Titanium 6AL-4V ELI | CP Titanium titanium grade 4 | CP Titanium titanium grade 4 |
| Implant<br>Surface | Blasted with aluminum oxide particles<br>and acid etched | Blasted with aluminum oxide particles and acid<br>etched | Blasted with aluminum oxide<br>particles and acid etched |
{6}------------------------------------------------
### Performance Testing:
Non clinical Testing was performed following "Guidance for Industry and FDA Staff -Class 11 Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments." Performance testing demonstrated that the device performs appropriately for the proposed indications for use.
Fatigue testing is not required because design features and technological features are similar to predicate devices; the implant body is not intended to be placed with any angle correction, and this submission does not contain any abutments for angle correction.
Proposed devices have the same sterilization process and radiation dose, same shelf life and biocompatibility as previous cleared Sterngold devices, therefore sterility, shelf life and bio-compatibility testing performed on previous cleared Sterngold devices is applicable to the proposed new devices. As a result, the TRU Dental Implant System including abutments is substantial equivalent to the predicates.
#### Conclusion:
Based on the above analysis, technological characteristics and performance testing, the subject device is substantially equivalent in intended use, material, design and performance to its predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.