The Deployable Oxygen Generator System - Small (DOGS-S) is intended for the administration of supplemental oxygen. This device is not intended for life support nor does it provide any patient monitoring capabilities. The system will be operated by trained personnel.
Device Story
DOGS-S is a portable, battery-powered oxygen concentrator for military and clinical use. It draws ambient air through a molecular sieve (Pressure Swing Adsorption) to separate oxygen from nitrogen; produces continuous flow of 93% (±3%) USP oxygen. Operated by trained medical technicians/respiratory therapists in deployed medical assemblages (EMEDS, ERPSS) or clinical settings. Output is delivered via nasal cannula. Device includes detachable lithium-ion battery, internal charger, and AC power capability. Provides supplemental oxygen to patients with respiratory insufficiency; does not monitor patients or provide life support. Healthcare providers use output to manage patient oxygenation levels; benefits include reliable, portable oxygen supply in austere or clinical environments.
Clinical Evidence
Bench testing only. No clinical data. Performance verified against ISO 80601-2-69, AAMI ES60601-1, and IEC 60601-1-2. Biological evaluation (ISO 10993-1) included cytotoxicity, sensitization, and irritation testing. Additional testing included particulate matter, VOC, ozone, and military standards (MIL-STD-810f/g, MIL-STD-461e). Results confirm 93% ±3% oxygen purity and flow rate accuracy.
Technological Characteristics
Portable oxygen concentrator using 5-bed Pressure Swing Adsorption (PSA) with molecular sieve. Materials: G/10 F/R fiberglass tube, PC/ABS. Flow: 0.5-15 LPM continuous. Purity: 93% ±3% (USP). Power: AC 100/240V or 674 Whr Li-ion battery. Connectivity: Standalone. Standards: ASTM F1464-93, ISO 80601-2-69, AAMI ES60601-1, IEC 60601-1-2.
Indications for Use
Indicated for patients suffering from discomfort due to ailments affecting lung efficiency in transferring oxygen to the bloodstream. Intended for use in home, hospital, or medical facility settings, including military deployed scenarios (e.g., wartime, humanitarian, contingency operations). Not for life support.
Regulatory Classification
Identification
A portable oxygen generator is a device that is intended to release oxygen for respiratory therapy by means of either a chemical reaction or physical means (e.g., a molecular sieve).
Predicate Devices
OMNI 2 (SAROS) from SeQual Technologies / A Chart Industries Company (K083163)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 18, 2015
AirSep Corporation Mr. Ted Vlahopoulos Regulatory Specialist 260 Creekside Drive Buffalo, NY 14228
Re: K150930
Trade/Device Name: Deployable Oxygen Generator System - Small (DOGS-S) Regulation Number: 21 CFR 868.5440 Regulation Name: Portable Oxygen Generator Regulatory Class: II Product Code: CAW Dated: November 12, 2015 Received: November 16, 2015
Dear Mr. Vlahopoulos:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150930
Device Name
Deployable Oxygen Generator System - Small (DOGS-S)
Indications for Use (Describe)
The Deployable Oxygen Generator System - Small (DOGS-S) is intended for the administration of supplemental oxygen. This device is not intended for life support nor does it provide any patient monitoring capabilities. The system will be operated by trained personnel.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### Submitter:
Airsep Corporation 260 Creekside Drive Buffalo, NY 14228
Phone: (716) 691-0202
Fax: (716) 691-1255
Registration Number: 1319044
#### Contact person:
Peter Weisenborn Quality & Regulatory Manager Email: peter.weisenborn@chartindustries.com
#### Preparation Date: 03/31/15
#### Device Trade Name:
DOGS-S Deployable Oxygen Generation System-Small
| Classification Name: | Portable Oxygen Generator |
|----------------------|---------------------------|
|----------------------|---------------------------|
| Regulation Number: | 868.5440 |
|--------------------|----------|
|--------------------|----------|
Product Code: CAW
Device Class: Class II
Classification Panel: Anesthesiology and Respiratory Therapy Devices
#### Predicate Devices:
OMNI 2 (SAROS) from SeQual Technologies / A Chart Industries Company Product Code: CAW 510(k) number: K083163
{4}------------------------------------------------
# Device Description:
The DOGS-S system is a portable battery powered oxygen concentrator that will be operationally used for ground-based medical assemblages, such as the Expeditionary Medical Support (EMEDS), and En-Route Patient Staging System (ERPSS) in deployed scenarios, including wartime operations, deterrence, humanitarian and contingency operations.
# Oxygen Concentrator Process
Oxygen concentrators were introduced in the mid-1970s and have become the most convenient, reliable source of supplemental oxygen available today.
The air we breathe contains approximately 21% oxygen, 78% nitrogen, and 1% other gases. In the DOGS-S unit, room air passes through a regenerative adsorbent material called molecular sieve. This material separates the oxygen from the nitrogen and other gases. The result is a constant supply of high concentration supplemental oxygen that is delivered to the patient.
# User/Operator Profile:
AirSep's Concentrators are intended to supply supplemental Oxygen to patients suffering from discomfort due to ailments which affect the efficiency of ones lungs to transfer the oxygen in air to their bloodstream. Patients can benefit from supplemental oxygen therapy for respiratory care at home, in the hospital, or at a medical facility. Oxygen Concentrator use requires a physician's prescription, and is not intended for life support use.
DOGS-S was designed for Military Medical Service use in deployed scenarios, including wartime operations, deterrence, humanitarian and contingency operations.
The DOGS-S will be operated by a trained medical technician with respiratory therapy background. In addition to the normal operation of the DOGS-S, the USER/OPERATOR shall have a working knowledge of oxygen concentrators.
# Technological Characteristics:
DOGS-S will generate 15 litres per minute (LPM) gaseous supply pressure oxygen with a regulated flow adjustment ranging from 0.5 LPM to 15 LPM (ASTM F1464-93) in 0.5 increments will concentrate oxygen from the atmosphere to a purity of 93% (ASTM F1464-93) United States Pharmacopeia (USP). DOGS-S supplies continuous flow of product through a standard patient distribution port. An operator supplied nasal cannula connects to the distribution port to deliver supplement oxygen to the patient. The device includes a detachable battery and internally contained battery charger. The battery charger is capable of charqing the battery while the DOGS-S is off or being used while plugged into AC power source. DOGS-S is a portable unit using handle accessory and/or transit case with handles. DOGS-S will be provided with a non-reusable box for initial shipment and a reusable carrying case.
{5}------------------------------------------------
Below is the comparison table.
| Substantial Equivalence<br>Comparison: | DOGS-S | OMNI 2 (SAROS) |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use: | The Deployable Oxygen<br>Generator System – Small<br>(DOGS-S)<br><br>is intended for the<br>administration of<br>supplemental oxygen. The<br>device is not intended for<br>life support nor does it<br>provide any patient<br>monitoring capabilities. The<br>system will be operated by<br>trained personnel. | The Omni 2 Oxygen System<br><br>is intended for the<br>administration of<br>supplemental oxygen. The<br>device is not intended for<br>life support nor does it<br>provide any patient<br>monitoring capabilities. |
| Technological<br>Characteristics: | | |
| Major Components: | | |
| Compressor: | Brushless DC motor Type - Rotary Swing | Brushless DC motor Type - Scroll |
| PSA Valve type: | (PSA) Pressure Swing Adsorption Adsorption - Molecular sieve 5 Bed design Type - Rotary | (PSA) Pressure Swing Adsorption Adsorption - Molecular sieve 5 Bed design Type Rotary |
| Battery: | Lithium Ion | Lithium Ion |
| Other Characteristics<br>Comparisons: | | |
| Electrical<br>requirements: | AC 100/240V 50/60Hz Battery 674 Whr peak | AC (100VAC, 50/60 Hz) Power Cartridge (Battery) 88 Whr |
| Purity: | 93% +/-3% USP | 93% +/-3% USP |
| Flow Rate: | 0.5 to 15LPM continuous flow control panel readout | 1 to 3LPM continuous flow control panel readout |
| Filtration: | • Dust<br>Compressed Inlet<br>HEPA | • Dust<br>Compressed Inlet<br>HEPA |
| Output pressure: | • 10.0 psig nominal | • 5.0 psig nominal |
| Power consumption: | • 550 Watts @ 15LPM | • 128 Watts @ 3LPM |
| Weight: | • 37 lbs. w/o battery | • 10 lbs. w/o battery |
| Diameter: | • 10" | • 4.4" |
| Height: | • 33" | • 26" |
| Sound Level: | • <70dBA | • <59dBA |
| Oxygen Purity<br>Warning: | • <85% | • 70-85% |
| Oxygen Purity Low: | • <85% | • <70% |
| Alarm indicators: | • Low Oxygen purity | • Low Oxygen purity |
| | • O2 Flow High or Low | • O2 Flow High or Low |
| | • Low power indicator<br>(Battery) | • Low power Cartridge<br>(Battery) |
| | • Unit malfunction | • Unit malfunction |
| Enclosure: | • MATERIAL: G/10 F/R<br>FIBERGLASS TUBE. | • MATERIAL: G/10 F/R<br>FIBERGLASS TUBE. |
| | • PC \ ABS | • PC / ABS |
{6}------------------------------------------------
# SUMMARY:
This summary demonstrates that the Device has no significant technological differences from the predicate device that would adversely affect product safety and effectiveness, i.e. all technological characteristics are covered.
# Accessories:
The following components are part of the DOGS-S submission; a battery pack / lithium ion, power supply, power cord, handle, and manuals. There are no accessories that have received prior FDA 510k clearance.
# Intended Use:
The Deployable Oxygen Generator System – Small (DOGS-S) is intended for the administration of supplemental oxygen. This device is not intended for life support nor does it
{7}------------------------------------------------
provide any patient monitoring capabilities. The system will be operated by trained personnel. The differences in language in the intended use for the SAROS and DOGS-S are minor.
#### Testing:
The DOGS-S has been tested and verified in various phases, internal testing, verification and validation as well as external testing and validation. The design was verified throughout the design process. Risk analysis was done, appropriate measures were implemented and their effectiveness verified. External test house was used to confirm compliance to EMC requirements and standards for electrical safety.
The testing confirms the Deployable Oxygen Generator System - Small (DOGS-S) meets the ISO 80601-2-69 standards for Oxygen Concentrator devices. Testing demonstrates that the product is in compliance to: AAMI ES60601-1 3° Edition Medical electrical equipment—Part:1 General Requirements for Basic Safety and Essential Performance and to IEC 60601-1-2 3ª Edition Medical Electrical Equipment, Part 1: General Requirements for Safety-Collateral Standard: Electrical Compatibility - Requirements and Tests.
#### Performance Data:
The oxygen flow rate and purity at the operating pressure was tested and verified in the Performance Bench Testing section of this submission, the instruction manual and in accordance with
- ISO 80601-2-69 . Medical Electrical Equipment, Part 2-69
- IEC 60601-1 Safety Requirements (Medical Electrical Equipment Part 1: General Requirements)
- . IEC 60601-1-2 EMC (Electro Magnetic Compatibility Testing)
- ISO 62304 Medical Device Software-Software Life-Cycle Processes
- Biological Evaluation of Medical Devices . ISO 10993-1
- Cytotoxicity
- Sensitization
- Irritation
- . Compliance with USP 93% +/-3%
- . Particulate Matter Testing
- Volatile Organic Compound Testing .
{8}------------------------------------------------
- . Ozone Testing
- MIL-STD-810f
- MIL-STD-810g
- . MIL-STD-461e
The test platform ensures compliance to recognized consensus standards and therefore does not raise new questions of safety and effectiveness.
#### Conclusion:
The DOGS-S performance tests demonstrate substantial equivalence to the predicate device. The performance tests results also confirm ability to provide 93% +/- 3% USP oxygen for supplemental oxygen use only.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.