The BR128 Breast Pump is an electrically powered suction device intended to express and collect milk from a lactating woman's breasts. This is for a single user.
Device Story
BR128 is an electrically powered breast pump for single-user home use. Device uses a reciprocating vacuum pump to generate negative pressure (0-170mmHg) to express milk. A single-chip microcomputer controls the suction cycle, imitating a baby's suckling action. User interfaces with the device via a massage cushion; motor unit connects to a collection bottle via silicone tubing. Suction strength is adjustable via a button; cycle rates are fixed by software. Powered by batteries or mains adapter. Healthcare provider does not operate the device; patient self-operates. Output is collected breast milk. Device benefits patient by facilitating milk expression for breastfeeding support.
Clinical Evidence
No clinical data. Bench testing included biocompatibility (ISO 10993-5, -10), electrical safety (AAMI/ANSI ES60601-1), electromagnetic compatibility (IEC 60601-1-2), food contact material safety (21 CFR 177.2600/177.1520), software validation, use life evaluation, and performance testing (pressure, cycle speed, sound, low voltage alert).
Technological Characteristics
Materials: Silicone, PP, ABS (ISO 10993 compliant). Principle: Reciprocating vacuum pump. Energy: AC adapter or 2 AA batteries. Connectivity: None. Software: Single-chip microcomputer for pressure modulation. Sterilization: Not specified (cleaning via boiling water).
Indications for Use
Indicated for lactating women to express and collect breast milk. For single-user use.
Regulatory Classification
Identification
A powered breast pump in an electrically powered suction device used to express milk from the breast.
Predicate Devices
Evenflo Advanced Double Electric Breast Pump (K131153)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, composed of three stylized human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 5, 2015
Guangzhou Jinxinbao Electronic Co.,ltd % Helen Nan General Manager Wenzhou Cytech Information Service Co., Ltd. Room302, No.21 Building, Kaiyu Garden, Xishan South Road Wenzhou. Zhejiang, 325000 China
Re: K150923 Trade/Device Name: BR128 Breast Pump Regulation Number: 21 CFR 884.5160 Regulation Name: Powered Breast Pump Regulatory Class: II Product Code: HGX Dated: December 20, 2014 Received: September 8, 2015
Dear Helen Nan.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA mav publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Herbert P. Lerner -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K150923
Device Name BR128 Breast Pump
Indications for Use (Describe)
The BR128 Breast Pump is an electrically powered suction device intended to express and collect milk from a lactating woman's breasts. This is for a single user.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo with the letters JXB in a stylized font. The letters are arranged horizontally, with the 'J' and 'X' in black and the 'B' in blue. The 'B' has a pixelated effect on the upper right corner, giving it a digital or tech-related feel. The logo appears to be for a company or organization with the initials JXB.
510(k) Summary
## (As required by 21 CFR 807.92(a))
## 1.0 Submitter Information
- · Company: Guangzhou Jinxinbao Electronic Co.,Ltd
- · Address: No.38 Huanzhen Xi Road, Dagang Town, Panyu, Guangzhou,
Guangdong, 511470, CHINA
- ・ Phone: 086-20-34938449
- ・ Fax: 086-20-22865301
- · Contact: Jiacheng Guo, General Manager
- · Date: November 5, 2015
## 2.0 Device Information
- · Trade/Device Name: BR128 Breast Pump
Image /page/3/Picture/14 description: The image shows the word "Trademark" preceded by a bullet point. The text is written in a clear, serif font. The bullet point is slightly offset to the left of the word "Trademark".
Image /page/3/Picture/15 description: The image features the word "Rycom" in a stylized logo. The "R" is in blue, while the "ycom" portion is in green. A blue line underlines the entire word, and a registered trademark symbol appears to the right of the word.
- · Common Name: Powered Breast Pump
- · Classification: Device: Pump, Breast, Powered
Regulation Description: Powered breast pump
Review Panel: Obstetrics/Gynecology
Product Code: HGX
Submission Type: 510(K)
Regulation Number: 21 CFR 884.5160
Device Class: II
# 3.0 Predicate Device Information
Evenflo Advanced Double Electric Breast Pump
submitted by Evenflo Feeding, Inc.
510K Number: K131153
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo with the letters JXB. The letters are stylized and connected. The J and X are in black, while the B is in blue. There are some pixelated graphics to the right of the B.
#### 4.0 Device Description
The BR128 is an electrically powered breast pump that uses negative pressure to express milk from the breasts of lactating women. It adopts a Single-Chip Microcomputer to imitate a baby's suckling action and a reciprocating vacuum pump to generate the suction to extract the milk at vacuum levels between 0 and 170mmHg. It is intended for single pumping.
It is designed to mechanically interface with a mother's breast via a massage cushion and withdraw then collect the breast milk. The motor unit of the device is connected to the bottle via a silicone tube. Its suction strength is adjustable via the suction adjustment button on the device, but its cycles per minute are controlled by software and cannot be adjusted. The device can be powered by batteries or the provided mains adapter.
## 5.0 Indications for Use
The BR128 Breast Pump is an electrically powered suction device intended to express and collect milk from a lactating woman's breasts. This is for a single user.
# 6.0 Comparison with the Predicate Device
Comparison Table of Technological Characteristic with the Predicate Device
| Device & Predicate Device(s): | Subject Device:<br>BR128 Breast Pump | Predicate Device:<br>Evenflo Advanced<br>Double Electric<br>Breast Pump |
|-------------------------------|--------------------------------------|-------------------------------------------------------------------------|
| | (K150923) | (K131153) |
| Intended Use | | |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters J, X, B, and D. The letters J and X are in black, while the letters B and D are in blue. The letter D has a pixelated design on the upper right side.
| Intended Use | An electrically powered<br>suction device intended<br>to express and collect<br>milk from a lactating<br>woman's breasts. This<br>is for a single user. | An electrically powered<br>suction device intended<br>to express and collect<br>milk from a lactating<br>woman's breasts. This<br>is for a single user. |
|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------|
| Patient Population | Lactating women | Lactating women |
| Environment of Use | Home | Home |
| Prescription | OTC | OTC |
| Designation | | |
| Single User | Yes | Yes |
| General Device Characteristics | | |
| Pump Style | Reciprocating Vacuum<br>Pump | Reciprocating Vacuum<br>Pump |
| Single/Double | Single | Double/Single |
| Adjustable Suction<br>Levels | Yes | Yes |
| Adjustable Cycle Rates | No | Yes |
| Stimulating Velocity | 100-150 cycles/min | 30-80 cycles/min |
| Stimulating Intensity | 0-170mmHg | 50-270mmHg (single) |
| Sucking Velocity<br>(Cycle Rate) | 13-18 cycles/min | 30-80 cycles/min |
| Sucking Intensity | 0-170mmHg | 50-270mmHg (single) |
| Overflow Protection | No | Yes |
| Pressure Modulation<br>Control | Microprocessor | Microprocessor |
| Software | Yes | Yes |
| Power Source | AC adaptor or 2 AA<br>Batteries | AC adaptor or 6 AA<br>batteries |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with the letters JXB. The letters are in different colors, with J and X in black and B in blue. There are some pixelated designs on the top right of the letter B. The logo is simple and modern.
| Cleaning Method for<br>Accessories | Boiling water | Boiling water |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------|---------------|
| Patient Contact<br>Material | Silicone, PP, ABS<br>(have been verified by<br>relevant ISO 10993<br>standards) | Unknown |
| Food Contact Material | PP & Silicone<br>(have been evaluated<br>by relevant food criteria<br>as set forth in 21 Code<br>of Federal Regulations) | Unknown |
The subject device has the same intended use as the predicate device.
The subject and predicate device have the following technological
characteristics in common:
- Both use reciprocating vacuum pump
- Both have adjustable suction levels ●
- Both are controlled by software .
- . Both have the same power source
However, the subject and predicate device have the following differences
in technological characteristics.
- Pumping Options (single vs. sing/double) ●
- Suction Pressure ●
- Cycle Speed ●
- Overflow Protection .
- Patient Contacting Material
These differences do not raise different questions of safety or effectiveness.
# 7.0 Discussion of Tests Performed
{7}------------------------------------------------
# 7.1 Clinical Tests
Clinical testing was not performed for BR128 Breast Pump as part of the submission.
# 7.2 Non-Clinical Tests
The subject device was tested/analyzed according to the following standards in order to ensure its safety:
Biocompatibility according to AAMI / ANSI / ISO 10993-5:2009/(R) 2014, Biological Evaluation of Medical Devices -- Part 5: Tests for In Vitro Cytotoxicity (L929 Assay) and AAMI / ANSI / ISO 10993-10:2010, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization. (Vaginal Irritation and Guinea Pig Maximization Sensitization)
Electrical Safety according to AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and Essential Performance (IEC 60601-1:2005, Mod). (General II (ES/EMC))
# Electromagnetic Compatibility according to AAMI / ANSI / IEC
60601-1-2:2007/(R)2012, Medical Electrical Equipment - Part 1-2: General Requirements for Basic Safety and Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements and Tests (Edition 3). (General II (ES/EMC))
Safety of Food Contact Materials according to FDA 21 CFR 177.2600 and FDA 21 CFR 177.1520
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains a logo with the letters "JXB" in a stylized font. The "J" and "X" are in black, while the "B" is in blue. There is a pixelated design element extending from the top right of the "B", also in blue. The logo appears to be for a company or organization with the initials "JXB".
Software Validation per the FDA Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices," dated May 11, 2005.
Use Life based on evaluation of device specifications following simulated use testing.
Performance Testing including minimum and maximum pressure testing, cycle speed, sound volume, and low voltage alert testing.
# 8.0 Conclusion:
The subject device - BR128 Breast Pump is substantially equivalent to the predicate device – Evenflo Advanced Double Electric Breast Pump (K131153).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.