K150921 · Ypsomed AG · FPA · Apr 12, 2016 · General Hospital
Device Facts
Record ID
K150921
Device Name
Orbit Infusion Set
Applicant
Ypsomed AG
Product Code
FPA · General Hospital
Decision Date
Apr 12, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Orbit®micro Infusion Sets are intended for the subcutaneous delivery of fluids and medication, such as insulin, from an external infusion pump.
Device Story
Orbit®micro Infusion Set is a subcutaneous infusion device for external insulin pumps. It consists of tubing with a luer lock connection to a medication reservoir and an adhesive base that anchors a stainless steel indwelling cannula into subcutaneous tissue. A patented design allows 360° tubing rotation at the adhesive site and disconnection of the tubing from the infusion base. The device is used by patients in home or clinical settings to deliver medication. It replaces the soft Teflon cannula of the predicate device with a stainless steel cannula. The healthcare provider or patient attaches the device to the skin; the cannula delivers medication from the pump. The device facilitates continuous subcutaneous insulin infusion, potentially improving glycemic control for patients requiring insulin therapy.
Clinical Evidence
No clinical data was required or provided. Substantial equivalence was established through bench testing, including material strength, pressure leak, penetration force, needle/catheter retention, bond strength, and tape adhesion. Biocompatibility testing included cytotoxicity, hemolysis, acute systemic injection, sensitization, intracutaneous reactivity, and bacterial endotoxin testing.
Technological Characteristics
Subcutaneous infusion set with stainless steel cannula (31G or 27G). Components: PVC/LDPE tubing, MABS caps, polyisoprene septum, nonwoven medical tape, PC base. Sterilization: Ethylene Oxide. Connectivity: Mechanical luer lock. Standards: ISO 9626 (cannula strength), ISO 10555-1 (needle retention), ISO 8536-8 (catheter retention), ISO 10993-1 (biocompatibility).
Indications for Use
Indicated for subcutaneous delivery of fluids and medication (e.g., insulin) from an external infusion pump in patients requiring such therapy.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem depicting a stylized human figure with three faces in profile, layered on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 12, 2016
Ypsomed AG % Lee Leichter President P/L Biomedical 10882 Stonington Avenue Fort Myers, Florida 33913
Re: K150921
Trade/Device Name: Orbit®micro Infusion Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: FPA Dated: March 8, 2016 Received: March 11, 2016
Dear Mr. Lee Leichter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Erin I. Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K150921
Device Name Orbit®micro Infusion Set
Indications for Use (Describe)
Orbit®micro Infusion Sets are intended for the subcutaneous delivery of fluids and medication, such as insulin, from an external infusion pump.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) SUMMARY for K150921
| Submitted By/ Contact Person: | Stephan Affolter<br>Head of Quality System & Regulatory Affairs<br>Ypsomed AG<br>Brunnmattstrasse 6<br>3401 Burgdorf<br>Switzerland<br>Tel. 0041-34 424 3382<br>Fax 0041-34 424 4122<br>E-mail: <a href="mailto:stephan.affolter@ypsomed.com">stephan.affolter@ypsomed.com</a> |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Alternative Contact: | Lee Leichter<br>President<br>P/L Biomedical 10882 Stonington Avenue<br>Fort Myers, FL 33913<br>USA<br>Tel. (239) 244-1448<br>Fax. (815) 550-0162<br>E-mail: <a href="mailto:leichter@plbiomedical.com">leichter@plbiomedical.com</a> |
| Date Prepared: | March 08, 2016 |
| 1. Trade/Proprietary Name: | Orbit <sup>®</sup> micro Infusion Set |
| 2. Common/Usual Name: | Subcutaneous Infusion Set |
| 3. Classification Name: | Intravascular Administration Set |
| Classification: | Class: II<br>Panel: 80<br>Product Code: FPA<br>Regulation: 21 CFR 880.5440 |
- 4. Predicate Device: K130468 Orbit Infusion Set
- 5. Purpose of Submission:
To market a product line of subject devices, designed with stainless steel cannulas as an alternative to the predicate devices designed with soft catheters and stainless steel insertion needles.
{4}------------------------------------------------
- 6. Device Description
All Orbit micro Infusion Sets are used for the subcutaneous delivery of fluids and medication from an external infusion pump. All Orbit micro Infusion sets are comprised of tubing that is connected on one end to the medication reservoir of the infusion pump using a luer lock connection and on the other end to the patient, attached to the skin by an adhesive base, anchoring a catheter that is inserted into the subcutaneous tissue. All Orbit micro Infusion sets have a patented design feature which allows the tubing to freely rotate 360° at the adhesive attachment and to disconnect the tubing set from the infusion base. The subject devices use a stainless steel indwelling cannula instead of the Teflon cannula in the predicate devices.
- 7. Intended Use:
The intended use of the subject devices remains the same as the predicate devices (K130468):
Orbit®micro Infusion Sets are intended for the subcutaneous delivery of fluids and medication, such as insulin, from an external infusion pump.
- 8. Technological Characteristics:
The technological characteristics of the subject devices are the same as for the predicate devices (K130468). There are some changes in components, materials and dimensions. All significant differences are described below.
Otherwise the Subject devices have the same indications for use, intended use and the same design including connectors, tubing lengths and diameters, materials, dimensions and operating principles as the currently marketed predicate devices.
| Characteristic | Predicate device | Subject device | Comments |
|------------------------|--------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|-------------------------|
| Dimensions | | | |
| Cannula Diameter | Soft Cannula:<br>ID 0.44mm<br>OD 0.76mm<br><br>Introducer Needle:<br>ID 0.20 mm<br>OD 0.41 (27G) | Stainless steel cannula (31G):<br>ID 0.11 mm<br>OD 0.25 mm<br><br>Stainless steel cannula (27G):<br>ID 0.24 mm<br>OD 0.41 mm | Different<br>dimensions |
| Cannula Length | 6mm, 9mm | 5.5mm, 8mm, 8.5mm | Different<br>dimensions |
| Tubing Length | 30cm (12 inch) - 105cm (42 inch) | 30cm (12 inch) - 105cm (42 inch) | Identical |
| Tubing Diameter | 0.020 x 0.062 inch (0.51 x 1.57mm ) | 0.020 x 0.062 inch (0.51 x 1.57mm ) | Identical |
| Material of components | | | |
| Female Luer | PVC | PVC | Identical |
| Tubing | PVC - outer layer<br>PE Polyethylene (LDPE) - inner<br>layer<br>co-extruded | PVC - outer layer<br>LDPE - inner layer co-extruded | Identical |
| Cap, Tubing | MABS | MABS | Identical |
| Septum | Polyisoprene | Polyisoprene | Identical |
| Adhesive Tape | Nonwoven Medical Tape | Nonwoven Medical Tape | Identical |
| Base | PC | PC | Identical |
| Cap, Needle | MABS | MABS | Identical |
| Lubrication | Silicon | Silicon | Identical |
| Needle Protector | HDPE/LDPE | HDPE/LDPE | Identical |
| Tape | Cohesive Tape | Cohesive Tape | Identical |
{5}------------------------------------------------
| Metal anchor<br>(Eyelet) | Stainless Steel | N/A | N/A |
|--------------------------|-------------------------------|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|
| Introducer Needle | Stainless Steel | N/A | N/A |
| Blue needle cover | HDPE/LDPE | N/A | N/A |
| Characteristic | Predicate device | Subject device | Comments |
| Catheter / Cannula | FEP catheter | Stainless Steel cannula | Different material<br>for the cannula,<br>but the same<br>stainless steel as<br>for the introducer<br>needle of the<br>subject device |
| UV Glue bond | Adhesive | Adhesive | Same type of<br>adhesive (acrylic<br>urethane adhesive,<br>UV-cured),<br>different viscosity |
| Packaging Material | | | |
| Primary Packaging | Blister sealed with Tyvek lid | Blister sealed with Tyvek lid | Identical |
| Sterilization | | | |
| Sterilization Method | Ethylene Oxide | Ethylene Oxide | Identical |
#### 9. Performance Data
Y psomed has performed the relevant assessments specified in the following international and internal standards and protocols and confirmed compliance of the modified devices and equivalence to the predicate devices. The Orbit®micro Infusion Sets have met the requirements of the relevant sections of the following standards:
| Test | Specification | Results |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|---------|
| Material strength<br>of steel cannula | Material strength per ISO 9626 | Pass |
| Activated pressure leak | No leak when subjected to pumping pressures up to 20psi<br>under normal delivery conditions and occluded fluid path<br>conditions | Pass |
| Penetration force | Needle and catheter shall penetrate a 0.025 inch thick<br>membrane with a speed of 50mm/min. and a force of less<br>than 0.8N | Pass |
| Needle retention | No separation of the needle from the cap when subjected to<br>a minimum force of 10N (ISO 10555- 1, Annex B) | Pass |
| Catheter retention | No separation of the catheter from the base when<br>subjected to a minimum force of 3N (ISO 8536-8) | Pass |
| Bond strength of<br>tubing/fittings | No separation of the tubing assembly when subjected to a<br>static tensile force of 15N for 15 sec. | Pass |
| Bond strength of<br>tape/base | No separation of the tape from the base when subjected to<br>a minimum force of 18N | Pass |
| Engagement force<br>tubing cap/base | The cap locks on the base with a force less than 13N | Pass |
{6}------------------------------------------------
| Disengagement force<br>tubing cap/base | The force to remove the tubing cap from the base is more<br>than 13N | Pass |
|----------------------------------------|------------------------------------------------------------------------------------------------------------|------|
| Occlusion test | No occlusion of the device when tested with a water flow<br>at a hydrostatic pressure of 0.1 bar | Pass |
| Tape adhesion | Removal of adhesive from a stainless steel plate with a 90<br>degree peel force of minimum 2.5N (0.56lbs) | Pass |
The following biocompatibility tests were completed on the subject device in accordance to ISO 10993-1 2009 with acceptable results:
- · Cytotoxicity ISO MEM Elution Assay
- Hemolysis ASTM Assay Extract Method
- · Acute Systemic Injection
- Guinea Pig Maximization Sensitization
- Intracutaneous reactivity study ●
- . Bacterial Endotoxin
The Subject infusion sets were tested, and met all acceptance criteria, for all the requirements as provided above. Based on the results Ypsomed concluded that the device performance is acceptable for the product.
#### 10. Clinical Data
Clinical Data per 807.92(b)(2) was not required to establish substantial equivalence for these devices.
#### 11. Conclusion
Y psomed AG concludes based on the information presented that the subject infusion sets are substantially equivalent to the predicate device legally marketed in the US.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.