K150915 · Imedicom Co., Ltd. · GEI · Mar 18, 2016 · General, Plastic Surgery
Device Facts
Record ID
K150915
Device Name
SPINAUT-P
Applicant
Imedicom Co., Ltd.
Product Code
GEI · General, Plastic Surgery
Decision Date
Mar 18, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SPINUAT-P is indicated for ablation, coagulation, and decompression of disc material to treat symptomatic patients with contained herniated discs.
Device Story
SPINAUT-P is a minimally invasive RF electrode for percutaneous disc decompression. Device consists of two stainless steel shafts with straight electrode configuration, housed in a polyamide coating. Used with standard RF generators (2-pin, 22mm outlet, 1kVp max) to deliver radiofrequency energy to the epidural space or posterolateral/posterocentral disc area. Principle of operation involves coblation: conductive media creates a plasma layer to ablate, remove, or coagulate disc tissue. Operated by physicians in a clinical setting. Output is the physical ablation/coagulation of disc material, enabling decompression. Benefits include minimally invasive treatment of symptomatic contained herniated discs.
Clinical Evidence
No clinical data. Evidence consists of bench testing, including sterilization validation (ISO 11137), shelf-life testing (ASTM F1980), biocompatibility (ISO 10993, USP 37), electrical safety (IEC 60601-2-2, ANSI/AAMI HF18), and a comparative performance study using porcine spines to evaluate tissue removal, durability, and navigation against the L'DISQ predicate.
Technological Characteristics
RF electrode; bipolar; stainless steel 304 shafts; polyamide coating; 496mm length; 1.0mm thickness. Compatible with standard RF generators. Single-use; gamma radiation sterilized. Complies with IEC 60601-2-2, ANSI/AAMI HF18, ISO 10993, ISO 11137, ASTM F1980, ASTM D882, ASTM F88, and ASTM F1929.
Indications for Use
Indicated for ablation, coagulation, and decompression of disc material in symptomatic patients with contained herniated discs.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
ArthroCare® Coblator IQ™ Perc-D®, SpineWand® (Coblator IQ DLR and DLG SpineWands) (K132099)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, with flowing lines extending from their necks. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" is arranged in a circular pattern around the design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 18, 2016
IMEDICOM Co., Ltd. c/o Ms. Priscilla Chung LK Consulting Group USA, Inc. 2651 E. Chapman Ave., Ste. 110 Fullerton, CA 92831
Re: K150915
Trade/Device Name: SPINAUT-P Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessoriesv Regulatory Class: Class II Product Code: GEI Dated: February 24, 2016 Received: February 26, 2016
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
## Jennifer R. Stevenson -A
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150915
Device Name
SPINAUT-P
Indications for Use (Describe)
The SPINUAT-P is indicated for ablation, and decompression of disc material to treat symptomatic patients with contained herniated discs.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> |
|-----------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
(K150915)
This summary of 510(K) - safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: Mar 17, 2016
### 1. Applicant / Submitter
IMEDICOM Co., Ltd. #612, 172, LS-ro, Gunpo-si, Gyeonggi-do, 435-824 Republic of Korea Tel:+82-31-479-1156 Fax:+82-31-479-1157
#### 2. Submission Correspondent
LK Consulting Group USA, Inc. 2651 E Chapman Ave. Ste 110, Fullerton CA 92831 Priscilla Chung Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
#### 3. Device
- I Trade Name: SPINAUT-P
- י Common Name: Electrosurgical accessory
- " Classification Name: Electrosurgical cutting and coagulation device and accessories
- י Product Code: GEI
- י Classification regulation: 21 CFR 878.4400
#### 4. Predicate Device:
- ArthroCare® Coblator IQ™ Perc-D®, SpineWand® (Coblator IQ DLR and DLG י SpineWands) by Arthrocare Corp. (K132099)
- L'DISQ by U&I Corporation (K132797) י
#### 5. Description:
The SPINAUT-P is a minimally invasive RF electrode device offered in 496mm length for percutaneous disc decompression surgery using radiofrequency energy and designed
{4}------------------------------------------------
for use with radiofrequency generators in RF procedure. SPINAUT-P is intended to be placed into the epidural space or posterolateral / posterocentral area in the disc to remove, ablate or coagulate disc material in the area. The electrode consists of two stainless steel shafts with straight electrode configuration design. The shaft is housed in a coating composed of Polyamide. The device is single-use gamma radiation sterilized and is compatible with most standard electrosurgical generators that provide a cable with two pin outlet (2-pin, 22mm) with 1kVp maximum electrical capacity.
#### 6. Indication for use:
The SPINUAT-P is indicated for ablation, coagulation, and decompression of disc material to treat symptomatic patients with contained herniated discs.
#### 7. Basis for Substantial Equivalence
The subject device is the same as the predicate device in the indications for use, materials, energy source, performance specifications, electrode thickness, mode of operation (bipolar) and sterilization method. The major difference is the electrode length but this difference does not raise a question of safety and performance since the electrode is coated with the polyamid to protect the patient and the user from electrical leakage and burn. In this regard, the longer electrode does not raise a question of safety issue. The electrode length would not be a performance affecting factor either since the thickness of the tip would matter the most which is the only part exposed to the patient. The performance test results submitted in this 510K submission also supports that the subject device is substantially equivalent to the predicate devices despite this difference.
| Trade name | SPINAUT-P | ArthroCare®<br>Coblator IQ™ Perc-<br>D®<br>SpineWand®<br>(Coblator IQ DLR<br>and DLG<br>SpineWands) | L'DISQ | |
|------------------|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacture | IMEDICOM Co., Ltd. | Arthrocare Corp. | U&I Corporation | |
| 510(k) number | - | K132099 | K132797 | |
| Product Code | GEI | GEI | GEI | |
| | Indications for Use | The SPINUAT-P is indicated for ablation, coagulation, and decompression of disc material to treat symptomatic patients with contained herniated discs. | The ArthroCare Coblator IQ Perc-D Spine Wands (Coblator IQ DLR and Coblator IQ DLG SpineWands) are indicated for ablation, coagulation, and decompression of disc material to treat symptomatic patients with contained herniated lumbar and lumbosacral discs. The Wands are designed to be used exclusively with the ArthroCarc Coblator IQ Controller. | The L'DISQ is indicated for coagulation and ablation of disc material to treat symptomatic patients with contained herniated discs. |
| | Principle of Operation | Electrode is activated using the generator which starts the coblation process to ablate disc tissue. The coblation process involves use of a conductive media to create a plasma layer to ablate tissue. | Electrode is activated using the generator which starts the coblation process to ablate disc tissue. The coblation process involves use of a conductive media to create a plasma layer to ablate tissue. | Electrode is activated using the generator which starts the coblation process to ablate disc tissue. The coblation process involves use of a conductive media to create a plasma layer to ablate tissue. |
| Material | Electrode | Stainless Steel 304 | Stainless Steel 304 | Stainless Steel 304 |
| | Handle | Polyamide | Polyamide | Polyamide |
| Perfor-<br>mance | RF leakage current | ≤7.2 dfL mA | ≤7.2dfL mA | ≤7.2dfL mA |
| Size | Electrode length | 496.0 mm | 300.0~350.0mm | 210.0mm |
| | Electrode thickness | 1.0 mm | 0.8~1.0mm | 0.7mm<br>1.0mm |
| | Compatibility w/other devices | Standard RF generator | with the ArthroCare Coblator IQ Controller | Standard RF Generator |
| | Chemical Safety | Biocompatible | Biocompatible | Biocompatible |
{5}------------------------------------------------
{6}------------------------------------------------
* dfL
d : the cable's smallest external dimension(unit : mm)
f : test frequency(unit : MHz)
L : length of cable submerged in saline solution
### 8. Non-Clinical Testing
- · Sterilization validating testing has been performed in accordance with ISO 11137-1, 2, 3 and ISO 11737-1, 2, 3, and the test results met the pre-set criteria.
- · The tests to validate the shelf life of the device through the proposed shelf life were conducted using the accelerated aging method in accordance to ASTM F1980-07 and the test results validated 3 year shelf life.
| No. | Test | Test Description / Referenced Standard |
|-----|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Appearance | Visual Inspection |
| 2 | Electrical Conductivity Test | Connect an electric circuit meter to the<br>device and check electrical conductivity<br>by electric circuit tester. |
| 3 | Tensile Strength Test of Electrode<br>Cord | Test in accordance with ANSI/AAMI<br>HF18:2001 Electrosurgical devices<br>Clause 4.2.5.5 |
| 4 | Package Tensile Test | Test in accordance with ASTM D882-12,<br>standard test methods for tensile<br>properties of thin plastic sheeting.<br>(Ophthalmic) |
| 5 | Package Peeling Test | Test in accordance with ASTM<br>F88/F88M-09, standard test method for<br>seal strength of flexible barrier materials.<br>(Sterility) |
| б | Package Dye Penetration Test | Test in accordance with ASTM F1929-12,<br>standard test method for detecting seal<br>leaks in porous medical packaging by dye<br>penetration. (Sterility) |
| 7 | Sterility Test | Test in accordance with ISO 11737-2<br>Second edition 2009-11-15, sterilization<br>of medical devices - microbiological<br>methods - part 2: tests of sterility<br>performed in the definition, validation and<br>maintenance of a sterilization process.<br>(Sterility) |
{7}------------------------------------------------
- · The following biocompatibility tests were performed in accordance with ISO 10993-4, 5, 10, 11 and USP 37 <151>, and the test results supported that the subject device is biocompatible.
| No. | Test | Test Description / Referenced Standards |
|-----|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Extraction Test | |
| 2 | Cytotoxicity | ISO 10993-5:2009/(R) 2014, biological<br>evaluation of medical devices -- part 5:<br>tests for in vitro cytotoxicity.<br>(Biocompatibility) |
| 3 | Skin Sensitization | ISO 10993-10 Third Edition 2010-08-01,<br>biological evaluation of medical devices -<br>part 10: tests for irritation and skin<br>sensitization. (Biocompatibility) |
| 4 | Intracutaneous reactivity | ISO 10993-10 Third Edition 2010-08-01,<br>biological evaluation of medical devices -<br>part 10: tests for irritation and skin<br>sensitization. (Biocompatibility) |
| 5 | Acute Systemic Toxicity Test | ISO 10993-11 Second edition 2006-08-15,<br>biological evaluation of medical devices<br>part 11: tests for systemic toxicity.<br>(Biocompatibility) |
| 6 | Pyrogen | USP 37-NF32:2014, <151> pyrogen test<br>(usp rabbit test). (Sterility) |
| 7 | Haemolysis | ISO 10993-4:2002 Biological evaluation of<br>medical devices -- Part 4: Selection of tests<br>for interactions with blood |
- The following electrical safety and bench tests were performed to evaluate the performance and the safety of the subject device and the test results met the pre-set criteria.
| No. | Test | Test Description / Referenced Standards |
|-----|--------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Electrical Conductivity Test | Evaluate if the electrode is able to pass<br>electricity. |
| 2 | Cable Dielectric Strength Test | IEC 60601-2-2 Edition 5.0 2009-02,<br>medical electrical equipment - part 2-2:<br>particular requirements for the basic safety<br>and essential performance of high<br>frequency surgical equipment and high<br>frequency surgical accessories [including:<br>technical corrigendum 1 (2014)]. (General<br>Plastic Surgery/General Hospital) |
{8}------------------------------------------------
| | | • Clause 201.8.8.3.104 Line voltage<br>withstanding capability<br>• Clause 201.8.8.3.103 High frequency<br>voltage withstanding capability |
|---|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| 3 | High Frequency Leakage Test of<br>Electrode | Test in accordance with ANSI/AAMI HF<br>18:2001 Electrosurgical Devices<br>• Clause 4.2.5.2 |
| 4 | Appearance | Visual Inspection |
| 5 | Size Measurement | Measure the size of the test article by<br>vernier calipers. |
| 6 | Tensile Strength Test of<br>Electrode Cord | Test in accordance with ANSI/AAMI HF<br>18:2001 Electrosurgical Devices<br>• Clause 4.2.5.5 |
| 7 | Sterility test | Korean Pharmacopeia (10th Edition) :<br>General Requirements for Tests and<br>Assays / Sterility Test |
- 〈We compared the tissue removal, durability, and ability to navigation in the lesion (performance to be introduced, directed to and extracted from the target lesion) of our RF Electrode SPINAUT-P and the predicate device, L'disQby U and I (K132797) using porcine spines. The test results demonstrated the device being as safe and as effective as the predicate device.
| No. | Test | Test Description |
|-----|-------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Performance Comparison Test with the<br>predicate device using porcine spines | • Tissue Removal Performance<br>• Durability of Electrode<br>• Ability to be introduced, directed<br>to and extracted from the surgical<br>site(target area) |
All the test results supported that the subject device is substantially equivalent to the predicate devices.
#### 9. Conclusion
Based on the similarities, we conclude that the SPINAUT-P is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.