XM-01 Automatic Electronic Blood Pressure Monitor is a home use digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 13.5 cm to 19.5 cm. The device is not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated.
Device Story
Non-invasive blood pressure monitor; uses oscillometric method to measure systolic/diastolic pressure and pulse rate. Hardware consists of main unit (microprocessor, pressure sensor, pump, deflation valve, 3.7V lithium battery) and wrist cuff (nylon/cotton/PVC). Device lacks onboard display; communicates via Bluetooth to mobile application (iOS/Android). App displays results, stores data locally, and transmits to remote server. Intended for home use by adult patients. Healthcare providers review results via mobile device or server data to assist in monitoring blood pressure trends; device does not provide diagnostic conclusions.
Clinical Evidence
Bench testing only. Device performance verified against ISO 81060-2, ISO 80601-2-30, IEC 60601-1, IEC 60601-1-2, FCC Part 15, and ISO 10993-1. No clinical trial data presented.
Technological Characteristics
Oscillometric measurement; electronic pressure sensor; Bluetooth connectivity. Materials: nylon/cotton/PVC cuff. Power: 3.7V/400mAH rechargeable lithium battery. Standards: ISO 14971, ISO 81060-2, ISO 80601-2-30, IEC 60601-1, IEC 60601-1-2, ISO 10993-1.
Indications for Use
Indicated for adult patients with wrist circumference 13.5-19.5 cm for home measurement of blood pressure and pulse rate. Not for diagnostic use.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{0}------------------------------------------------
Image /page/0/Picture/12 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with a flowing, ribbon-like element beneath them. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 16, 2015
Sky Innovation Technology (Shanghai) Limited % Charles Shen Official Correspondent Manton Business And Technology Services 37 Winding Ridge Oakland. New Jersey 07436
Re: K150908
Trade/Device Name: XM-01 Electronic Automatic Blood Pressure Monitor Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN Dated: October 28, 2015 Received: October 30, 2015
Dear Charles Shen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mitchell Stein
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150908
Device Name
XM-01 Automatic Electronic Blood Pressure Monitor
Indications for Use (Describe)
XM-01 Automatic Electronic Blood Pressure Monitor is a home use digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 13.5 cm.
The device is not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
|-----------------------------------------------------------------------|---------------------------------------------------------------------------------|
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5: 510(k) Summary:
This summary of 510k safety and effectiveness information is being submitted In accordance with the requirements of 21CFR 807.92
#### Submitter & Foreign Manufacture Identification 5.1
Sky Innovation Technology (Shanghai) Limited 9th Floor Hechuan Building, 2016 Yishan Rd, Minhang District, Shanghai Postal code: 201103, China Tel: (086) 400-9999-523 Submitter's FDA Registration Number: N/A
#### Contact Person 5.2
Charles Shen Manton Business and Technology Services 37 Winding Rdg, Oakland, NJ 07436 Tel: 608-217-9358 Email: cyshen@aol.com
#### Date of Summary: April 1, 2015 5.3
| 5.4 Device Name: | |
|------------------------|---------------------------------------------------|
| Proprietary Name: | XM-01 Automatic Electronic Blood Pressure Monitor |
| Common Name: | Blood Pressure Monitor |
| Classification Name: | System, Measurement, Blood-Pressure, Non-invasive |
| Device Classification: | II |
| Regulation Number: | 21 CFR 870.1130 |
| Panel: General | Cardiovascular |
| Product Code: | DXN |
#### ર્ડ Predicate Device Information:
- K123669, "Transtek Wrist Blood Pressure Monitor", manufactured by (1) "Zhongshan Transtek Electronics Co., Ltd."
#### 5.6 Device Description:
XM-01 Automatic Electronic Blood Pressure Monitor is designed to measure the systolic and diastolic blood pressure and pulse rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the wrist. This is a home use device.
{4}------------------------------------------------
Measurement method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury nmometer. With a sensor, the microprocessor converts tiny alterations in cuff pressure to electrical signals, and analyze those signals to define the systolic and diastolic blood pressure and calculating heartbeat rate, which is a well-known technique in the market called the "Oscillometric method". The sensor also detect pulse rate at the same time.
XM-01 Automatic Electronic Blood Pressure Monitor achieves its function by integrating a single-mounted device hardware to a mobile application software, which is downloadable to a mobile device such as Apple or Android phone. As the hardware itself does not include a LCD or other display components, it is necessary for the device hardware to communicate with mobile devices containing the supporting software to constitute a complete blood pressure measurement system. This communication is realized through an embedded Bluetooth module.
The single-mounted device hardware is composed of the main unit and cuff unit. The cuff unit, which is applicable to wrist circumference approximately between 13.5 and 19.5 cm, includes the inflatable bladder and nylon/cotton/PVC shell. The main unit consists of the microprocessor, pressure sensor, pump, the electromagnetic deflation control valve and the battery. The subject device is powered by a 3.7V/400mAH lithium battery. The battery can be re-charged by either a DC5V 500mA charger or a USB cable.
The mobile application, downloadable to mobile devices such as Apple IOS or Android phone, receives measurement results from the single mounted device. The mobile App allows the test results to be displayed on the mobile devices (such as cell phone) for user review. It also transmits the data to the database in remote server for storage and future retrieval.
#### 5.7 Indications for Use:
XM-01 Automatic Electronic Blood Pressure Monitor is a home use digital monitor intended for use in measuring blood pressure and pulse rate in adult patient population with wrist circumference ranging from 13.5 cm to 19.5 cm.
The device is not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated.
#### 5.8 Design Control and Performance Testing Summary:
Design control activities for this modification were performed and bench tests have been done. Those performance tests, risk management, and design verification tests provide demonstration that the device does not raise any new questions of safety and effectiveness.
XM-01 Automatic Electronic Blood Pressure Monitor conforms to the following standards:
{5}------------------------------------------------
- ISO 14971, Risk management to medical device
- o ISO 81060-2. Non-invasive sphygmomanometers -- Part 2: Clinical investigation of automated measurement type
- o ISO 80601-2-30, Medical electrical equipment -- Part 2-30: Particular requirements for basic safety and essential performance of automated noninvasive sphygmomanometers
- IEC60601-1, Electrical safety;
- o IEC6060 1-1-2, Electromagnetic compatibility
- FCC Part 15, EMI tests of FCC Radiation & RF rules and regulations o
- ISO 10993-1, Biological evaluation of medical devices -- Part 1: Evaluation and ● testing within a risk management process
XM-01 Automatic Electronic Blood Pressure Monitor conforms to the following guidance document:
- FDA guidance document Non-Invasive Blood Pressure (NIBP) Monitor Guidance (March, 1997).
#### 5.9 Technological Comparison with Predicate Device
The following table shows similarities and differences of use, mechanism, design, and labeling between our device and the predicate devices.
| Table 5.1: Comparison of Intended Use, Mechanism, Labeling, and Design | | |
|------------------------------------------------------------------------|--|--|
| | | |
| Description | Subject Device | Predicate Device (K123669) | Comparison |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Indication<br>for Use | XM-01 Automatic Electronic Blood<br>Pressure Monitor is a home use<br>digital monitor intended for use in<br>measuring blood pressure and pulse<br>rate in adult patient population with<br>wrist circumference ranging from<br>13.5 cm to 19.5 cm.<br><br>The device is not intended to be a<br>diagnostic device. Contact your<br>physician if hypertensive values are<br>indicated. | Transtek Wrist Blood Pressure Monitor<br>TMB-1014-BT is a digital monitor<br>intended for use in measuring blood<br>pressure and heartbeat rate in adult patient<br>population with wrist circumference<br>ranging from 13.5 cm to 21.5 cm (about 5<br>1/4-S8 1/2 inches).<br><br>This device detects the appearance of<br>irregular heartbeats during measurement<br>and gives a warning signal with readings.<br><br>The Wrist Blood Pressure Monitor<br>compares average blood pressure results to<br>pre-established AHA (American Heart<br>Association) hypertension guideline of<br>135/85 mm Hg.<br><br>Transtek Wrist Blood Pressure Monitor,<br>TMB-1014-BT is not intended to be a | Similar |
{6}------------------------------------------------
| Component | Main Unit, Cuff, Battery | Main Unit, Cuff, Battery<br>diagnostic device. Contact your physician<br>if hypertensive values are indicated. | Similar |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Measurement Method | Oscillographic | Oscillographic | Same |
| User Control | User controls measurement from<br>mobile devices through mobile<br>application software | User controls measurement from LCD<br>screen on hardware | Minor<br>difference |
| Labelling | Company Name and Address,<br>Specifications, Product Descriptions,<br>Indications for Use, Contraindication<br>for Use, Precautions, Warnings,<br>Safety Terms and Conditions, Safety<br>Alert Description, Safety and<br>Performance Standards, etc. | Company Name and Address,<br>Specifications, Product Descriptions,<br>Indications for Use, Contraindication for<br>Use, Precautions, Warnings, Safety Terms<br>and Conditions, Safety Alert Description,<br>Safety and Performance Standards, etc. | Same |
| Power Source | Rechargeable Lithium Battery | Battery | Similar |
| Cuff | Wrist warp around | Wrist warp around | Same |
| Display | Remote display on mobile devices | Local liquid crystal digital display, and<br>remote display on mobile devices | Similar |
| Wireless Mode | Bluetooth | Bluetooth | Same |
| Data Storage | On mobile devices and on server | Mobile devices | Similar |
Our device is essentially identical to the predicate device in terms of indications for use, design, mechanism, and labeling between subject device and the predicate device. The several minor differences do not affect the safety and effectiveness of the device
#### 5.10 Substantial Equivalence Conclusion
It has been shown in this 510(k) submission that XM-01 Automatic Electronic Blood Pressure Monitor and its predicate devices have the identical indications for use, design, mechanism, labeling, and similar performance.
The difference between the XM-01 Automatic Electronic Blood Pressure Monitor and their predicate device do not raise any question regarding its safety and effectiveness.
XM-01 Automatic Electronic Blood Pressure Monitor, as designed and manufactured, are as safe and effective as its predicate device, and therefore is substantially equivalent as its predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.