The NovoPen Echo® is a re-useable pen injector designed for single patient use by diabetics for the self-injection of a desired dose of insulin. The pen injector uses PenFill® 3 mL cartridge of NovoLog®, 100 units/mL (U-100) finsulin aspart (rDNA origin) injection] and a single use detachable and disposable pen needle (supplied separately). The pen injector allows the user to dial the desired dose from 0.5 to 30 units in 0.5 unit increments.
Device Story
Reusable mechanical pen injector for insulin delivery; patient self-use. Input: manual rotation of dose button (0.5 to 30 units, 0.5 unit increments). Operation: mechanical coupling drives piston in insulin cartridge to expel dose. Memory function: sensor in coded cylinder detects rotation; LCD displays last dose quantity and elapsed time since last injection. Memory function operates independently of mechanical injection mechanism. Benefits: precise dosing and tracking of last injection for diabetes management.
Clinical Evidence
Bench testing only. Verification testing confirmed compliance with ISO 11608-1:2012. Software verification and validation (static, unit, integration, and system regression testing) passed.
Technological Characteristics
Reusable mechanical pen injector; 0.5-30 unit dosing range; 0.5 unit increments. Complies with ISO 11608-1:2012. Features integrated memory function with LCD display. Powered by internal battery for memory/display; mechanical injection. Materials updated for specific components; manufacturing site transfer to Tianjin, China.
Indications for Use
Indicated for diabetics requiring self-injection of insulin using PenFill® 3 mL cartridges (NovoLog® 100 units/mL).
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three human profiles facing right, connected by a flowing ribbon-like design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 31, 2015
Novo Nordisk Inc. Ms. Poonam Tuliani Associate Director, Regulatory Affairs P.O. Box 846 Plainsboro, New Jersey 08536
Re: K150874
Trade/Device Name: NovoPen Echo® Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: FMF Dated: June 30, 2015 Received: July 1, 2015
Dear Ms. Tuliani:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
in filməsinə filmləri filmləri filmləri filmləri filmləri filmləri filmləri filmləri filmləri filmləri filmləri filmləri filmləri filmləri filml Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150874
Device Name NovoPen Echo®
#### Indications for Use (Describe)
The NovoPen Echo® is a re-useable pen injector designed for single patient use by diabetics for the self-injection of a desired dose of insulin. The pen injector uses PenFill® 3 mL cartridge of NovoLog®, 100 units/mL (U-100) finsulin aspart (rDNA origin) injection] and a single use detachable and disposable pen needle (supplied separately). The pen injector allows the user to dial the desired dose from 0.5 to 30 units in 0.5 unit increments.
Type of Use (Select one or both, as applicable)
| <span>☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|-------------------------------------------------------------|
| <span>☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Date: Version: Status: Page:
#### 510(k) Summary K150874
# 21 CFR 807.87(h)
As required by 21 CFR 807.92(a)
### Date Prepared: July 29, 2015
This 510(k) Summary is being submitted in accordance with 21 CFR 807.92.
### 1. Submitter's Name:
Novo Nordisk Inc. P.O. Box 846 Plainsboro, NJ 08536
## Contact Person:
Poonam Tuliani Associate Director, Regulatory Affairs Tel: 609-786-5210 Fax: 609-987-3916
### 2. Device Name:
Proprietary Name: NovoPen Echo® Common Name: Dial-A-Dose Insulin Delivery Device (Pen Injector) Classification Name: Syringe/Piston (FMF) Regulation: 21 CFR 880.5860 Class: Class II
### 3. Substantial Equivalence:
The manufacturing site for NovoPen Echo® is to be transferred from Site Device Manufacturing Development (DMD), Denmark to Site Tianjin, China (STJ). Hereinafter, the original cleared device is referred to as NovoPen Echo® (DMD) and the modified device is referred to as NovoPen Echo® (STJ).
{4}------------------------------------------------
CONFIDENTIAL
Date: Version: Status: Page:
In order to improve manufacturability, production automation, and to align manufacturing with other devices within the NovoPen® family, the following changes have been implemented to NovoPen Echo®.
- Update of software to support manufacturing automation as well as including other minor changes
- . Update of selected components and component material changes for a small number of components
- Labelling changes
The above changes do not affect intended use, indications for use, technological characteristics, functionality of components, principle of operation, shelf life, biocompatibility and device user interface. Therefore, NovoPen Echo® (manufactured in Novo Nordisk (NN), Site Tianjin, China) is substantially equivalent to Novo Nordisk's NovoPen Echo® (manufactured in NN, Site DMD, Denmark), cleared under 510(k) K123766 in August, 2013.
The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence as found in the Federal Food, Drug, and Cosmetic Act, as amended and as supplied under 21 CFR 807, Subpart E under which a device can be marketed without pre-market approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statement related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent laws or their application by the court.
# 4. Device Description:
The NovoPen Echo® (STJ) is a reusable mechanical pen-injector capable of injecting a dose of up to 30 units of insulin, in 0.5 unit increments, from a PenFill® 3 mL cartridge of NovoLog®, 100 units/mL (U-100) [Insulin aspart Injection (rDNA origin)] . The memory function enables the user of the device to see the quantity of last given dose and hours that have passed since last dose. The operation of the pen-injector for the injection of insulin is entirely independent of the operation of the memory function. The intended dose is mechanically set by rotating a dose button. The insulin is injected by depressing the dose button which via mechanical coupling causes the piston in the insulin cartridge to move forward thereby expelling the intended dose. The device has a memory function and liquid crystal display that allows the user to review the units of the last dose and the number of hours that have elapsed since the last dose was taken. Rotation of the dose button during dose setting causes a sensor to rotate within a coded cylinder. The movement detected by this sensor is stored for later display on the liquid crystal display in the dosage selector module as the
{5}------------------------------------------------
| NovoPen Echo<br>Section 5. Special 510(k) Summary<br>novoDOCs id:002057078 | CONFIDENTIAL | Date: | 31 July 2015 | Novo Nordisk |
|----------------------------------------------------------------------------|--------------|----------|--------------|--------------|
| | | Version: | 4.0 | |
| | | Status: | Final | |
| | | Page: | 4 of 5 | |
number of units injected. The pen-injector is intended for use with PenFill® 3 mL cartridge of NovoLog", 100 units/mL (U-100) [Insulin aspart Injection (rDNA origin)] and a single-use, detachable and disposable pen needle (supplied separately by Novo Nordisk ). NovoPen Echo® (STJ) is substantial equivalent to the primary predicate device NovoPen Echo® (DMD)
5. Intended Use including Indication for Use:The Intended Use including Indication for Use for the modified device are as follows:
The Novo Pen Echo® is a re-usable pen injector designed for single-patient use by diabetics for the self-injection of a desired dose of insulin. The pen injector uses PenFill® 3 mL cartridge of NovoLog", 100 units/mL (U-100) [ [Insulin aspart Injection (rDNA origin)] and a single use detachable and disposable pen needle (supplied separately). The pen injector allows the user to dial the desired dose from 0.5 to 30 units in 0.5- unit increments.
# 6. Technological Characteristics:
The NovoPen Echo® (STJ) is considered substantially equivalent to the predicate device in intended use, principle of operation, materials, energy source and performance.
The injection technique used to administer insulin is the same as predicate device, NovoPen Echo® (DMD). Both devices require a needle to penetrate the subcutaneous layer of tissue, an insulin source and a plunger to push the insulin from the source, through the needle and into the subcutaneous layer of tissue. Both devices are capable of multiple injections from a single insulin source (cartridge).
Both devices have the same 0.5 increment dosing capability and comply with requirements specified in ISO11608-1:2012 Needle-based injection systems for medical use - Requirements and test methods - Part 1: Needle-based injection systems.
# As required by 21 CFR 807.92(b)
# Testing
# Performance Data:
Device Verification Tests were run to verify all product requirements are still met and all changes work properly and complies with the requirements specified in ISO11608-1:2012 Needle-based injection systems for medical use - Requirements and test methods - Part 1: Needle-based injection systems.
{6}------------------------------------------------
| NovoPen Echo |
|-----------------------------------|
| Section 5. Special 510(k) Summary |
| novoDOCs id:002057078 |
CONFIDENTIAL
Software verification and validation regression testing all static, unit, integration and system, tests as well as additional tests for the changes and related risk control measures was conducted and all tests passed.
# Shelf Life:
The NovoPen Echo® has an in-use lifetime of 5 years provided that the pen is taken into use within two years of the production date. The NovoPen Echo® informs the user via the memory display when the pen has reached its end of life.
# Conclusions:
In conclusion, the results of the testing demonstrate that NovoPen Echo® (STJ) is as safe and effective and performs as well as the predicate device.
The results of the testing to standards provide additional evidence that NovoPen Echo® (STJ) is substantially equivalent to the predicate device, NovoPen Echo® (DMD).
The differences between NovoPen Echo® (STJ) and the predicate device, NovoPen Echo® (DMD), do not raise any new issues of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.